WEEK ENDING OCTOBER 2
PRODUCT 8870 software application card Version AAA 02, BBB 04, BBC 02, and BBD 01, which
is used in conjunction with the Model 8840 N'Vision Clinician Program.
CODE Software Version AAA 02, BBB 04, BBC 02, and BBD 01.
RECALLING FIRM/MANUFACTURER Medtronic Inc, Neurological & Spinal Division, Columbia Heights,
MN.
REASON Users may mistakenly enter a periodic bolus interval into the minutes field,
rather than the hours field, resulting in drug overdoses. This issue is limited
to programming the SynchroMed and SynchroMed EL pumps.
VOLUME OF PRODUCT IN COMMERCE 4,466 cards.
DISTRIBUTION Nationwide and Internationally.
PRODUCT The LTV Series ventilators provides continuous or intermittent ventilatory support
for the care of individuals who require mechanical ventilation. This is a prescription
medical device intended for use by personnel under the direction of a physician.
The ventilator is suitable for use in institutional, home and transport settings.
CODE Ventilators manufactured prior to September 2003. Affected LTV Model/ Serial's:
LTV 1000- S# A06500, LTV 950- S# C07000, LTV 900- S# B03500, LTV 800- S# D01400.
RECALLING FIRM/MANUFACTURER Pulmonetic Systems, Inc, Minneapolis, MN
REASON LTV Series of Ventilators (Models 1000, 950, 900 and 800). The ventilator has
not consistently switched to internal battery operation when the external DC
power source has become inadequate to supply stable power, resulting in loss
of ventilation to the patient.
VOLUME OF PRODUCT IN COMMERCE 10,299 units.
DISTRIBUTION Nationwide and Internationally.
PRODUCT
a) ACOM.M angiographic x-ray system. Model number 4371741.
b) ACOM.M angiographic x-ray system. Model number 5215707.
CODE
a) Serial numbers: 1024 through 1639;
b) Serial numbers 1087, 1088,1089,1093, 1094, 1493, 1174, 1176, and 1212.
RECALLING FIRM/MANUFACTURER Siemens Medical Solutions USA, Inc., Malvern,
PA.
REASON Possible problem reviewing images on the ACOM.M.
VOLUME OF PRODUCT IN COMMERCE 237 units.
DISTRIBUTION Nationwide.
PRODUCT Hill-Rom brand VersaCare Bed; Model P3200.
CODE All units.
RECALLING FIRM/MANUFACTURER Hill-Rom, Inc., Batesville, IN.
REASON Pivot bolts on the sleep deck may become loose and fall out, allowing the head
section to shift or to drop to one side, and a grounding warning label is missing.
VOLUME OF PRODUCT IN COMMERCE 5,048 units.
DISTRIBUTION Nationwide and Internationally.
PRODUCT Trio Monitor. Cardiac Monitor (including cardiotachometer and rate alarm).
CODE Serial Numbers: MC01475-D3; MC02325-K3; MC01476-D3; MC02120-I3; MC02017-I3;
MC02004-I3; MC02011-I3; MC02018-I3; MC02047-I3; MC02039-I3; MC02196-I3; MC02279-K3;
MC02295-K3; MC02298-K3; MC01478-D3; MC01491-D3; MC01496-D3; MC01865-G3; MC01455-D3;
MC02281-K3; MC02237-I3; MC05179-C4; MC05180-C4; MC05191-C4; MC05194-C4; MC05206-C4;
MC05218-C4; MC05223-C4; MC01497-D3; MC01452-D3; MC01473-D3; MC01490-D3; MC01493-D3;
MC01861-G3; MC02354-K3; MC02293-K3; MC02303-K3; MC02316-K3; MC02134-I3; MC02138-I3;
MC02148-I3; MC02321-K3; MC02292-K3; MC01487-D3; MC02285-K3; MC02309-K3; MC02313-K3;
MC05082-C4; MC05107-C4; MC02306-K3; MC01862-G3; MC01918-H3; MC01479-D3; MC05183-C4;
MC05189-C4; MC01474-D3; MC01477-D3; MC01450-D3; MC05072-C4; MC05075-C4; MC05083-C4;
MC05089-C4; MC05090-C4; MC05105-C4; MC05178-C4; MC05195-C4; MC05209-C4; MC05210-C4;
MC05212-C4; MC05213-C4; MC05214-C4; MC05215-C4; MC05217-C4; MC05224-C4; MC01448-D3;
MC01449-D3; MC01464-D3; MC01481-D3; MC01495-D3; MC01859-G3; MC01867-G3; MC01890-H3;
MC02348-K3; MC02359-K3; MC02363-K3; MC01453-D3; MC01860-G3; MC02199-I3; MC01460-D3;
MC01806-G3; MC01954-H3; MC02135-I3; MC02136-I3; MC02146-I3; MC02168-I3; MC02171-I3;
MC02278-K3; MC02280-K3; MC05172-C4; MC05173-C4; MC05174-C4; MC05181-C4; MC05184-C4;
MC05188-C4; MC05190-C4; MC05201-C4; MC05203-C4; MC05208-C4; MC05219-C4; MC05222-C4;
MC05225-C4; MC05227-C4; MC02002-I3; MC02005-I3; MC02006-I3; MC02008-I3; MC02020-I3;
MC02023-I3; MC02025-I3; MC02027-I3; MC02030-I3; MC02034-I3; MC02043-I3; MC02044-I3;
MC02048-I3; MC02052-I3; MC05182-C4; MC05185-C4; MC05192-C4; MC05199-C4; MC01869-G3;
MC02326-K3; MC02344-K3; MC02357-K3; MC02365-K3; MC01866-G3; MC01868-G3; MC01929-H3;
MC01949-H3; MC02187-I3; MC02277-K3; MC01482-D3; MC01488-D3; MC02177-I3; MC02290-K3;
MC02317-K3.
RECALLING FIRM/MANUFACTURER Datascope Corp., Mahwah, NJ.
REASON The Trio Monitor has latex feet. It is not labeled per 21 CFR801.437 to indicate
that it contains latex.
VOLUME OF PRODUCT IN COMMERCE 146 units.
DISTRIBUTION Nationwide.
PRODUCT CryoValve, Conduit and Pulmonary Valve and Conduit. Donor #66949. Model #SGPV00.
CODE Serial #7988716.
RECALLING FIRM/MANUFACTURER CryoLife, Inc., Kennesaw, GA.
REASON CryoLife received information regarding a positive culture for Group A Streptococcus
infection in a recipient of tissue procured from the attached donor.
VOLUME OF PRODUCT IN COMMERCE 1 tissue.
DISTRIBUTION NM.
PRODUCT
a) Gluteraldehyde Concentration Indicators "Browne GA Indicator for CIDEX PLUS Products.
b) Gluteraldehyde Concentration Indicators "Browne GA Indicator for CIDEX Products.
c) Orthophthaladehyde Concentration Indicator Browne CIDEX OPA Indicator.
CODE
a) CIDEX PLUS code 2924 lots: 9467, 9514, 9517, 9533 CIDEX PLUS code 2926, lots: 9504, 9518, 9534, 9624, 9655;
b) CIDEX Activated code 2927, lots: 9520, 9656, 9695, 9804, 9896, 10001 CIDEX Activated Code 2920, lots: 9447, 9522,
9621, 9622, 9661, 9662, 96666, 9764, 9897;
c) CIDEX OPA code 20392, lots: 9272, 9273, 9433, 9434, 9436, 9461, 9512, 9513, 9537, 9538, 9578, 9593, 9664, 9699, 9743,
9744, 9745, 9746, 9814, 9815, 9816, 9854, 9935, 9984, 9985, 10019, 10028, 10036, 10037, 8101, 8144, 8218, 8233, 8321,
8482, 8483, 8510, 8511, 8512, 8513, 8514, 8618, 8630, 8631, 8632, 8636, 8637, 8638, 8792, 8793, 8794, 8795, 8796, 8797,
8798, 8978, 8980, 8981, 9130, 9131, 9133, 9270, 9271, 9272 CIDEX OPA code 20393, lots: 8079, 8102, 8219, 8223, 8421,
8456, 8515, 8550, 8633, 8639, 8970, 8982, 9075, 9260, 9274, 9275, 9536, 9564, 9663, 9796, 9907, 9908, 10000.
RECALLING FIRM/MANUFACTURER Advanced Sterilization Products, Irvine,
CA.,
REASON Performance failure complaints, moisture ingress into the bottles was causing
failure or variability in results.
VOLUME OF PRODUCT IN COMMERCE 122,914 cases.
DISTRIBUTION Nationwide and Internationally
.Pulmonetic Systems LTV Series of Ventilators
Pulmonetic Systems, Inc. notified healthcare professionals of a recall of the LTV series of ventilators, models 1000, 950, 900 and 800, designed to automatically switch to internal battery operation, allowing uninterrupted ventilation, when an external power source is removed or is no longer adequate to power the ventilator. The ventilators malfunction when switching to the internal battery, causing failure of the ventilator to breathe for the patient.
WEEK ENDING OCTOBER 9
PRODUCT
Alveolus Tracheobronchial Stent Technology System (TB-STS).
CODE
Size 12x40, Part #90129-20, Lot #AC0406; Size 14x40, Part #90129-21, Lot #AC0409; Size 16x40, Part #90129-22, Lot #AA0411; Size 16x60, Part #90129-23, Lot #A0413; Size 12x20, Part #90129-24, Lot #AA0404; Size 14x40, Part #90129-25, Lot 3AA0409; Size 14x40, Part #90129-26, Lot #AC0409; Size 16x60, Part #90129-27, Lot #AA0413.
RECALLING FIRM/MANUFACTURER
Alveolus, Inc., Charlotte, NC.
REASON
The tracheal stent may collapse resulting in blockage of the air passage because the stents were not properly coated with the polyurethane material to the very end of the
stent.
VOLUME OF PRODUCT IN COMMERCE
47 units.
DISTRIBUTION
FL, SC, and TX.
PRODUCT
NT Intera Magnetic Resonance Image System.
CODE
Part #s 9896 030 09981, 9896 030 10251 Site #s 504531, 104343, 59488, 41391, 37996, 37979, 85119, 26269, 59435, X1281, 504409, 102353, 13606, 103166, 37929, 104392, 103149, 84418, 6947, 63046, 86293, 102978, 103904, 105315, 504866, X0038, 103582, 100745, 101946, 103333, 504412, 13056, 100786, 59428, 103818, 105314, 50430, 104808, 85139, 45016, X0716, X0942, X1272, X1086, X0214, 103324, 101420, 103579, 103847, 6808, 104467, 38652, 103347, 14117, 86335, 68196, 104032, 103417, 103243, 103207, 104102, 52408, 6841, 47029, 41346, 45065, X0164, X1082, 104912, X0083, 104054, 83325, 103911, 101477, 250156, 25958, 104193, 102547, 85172, 104897, 104070, 506190, 103946, 504608, 104340, 100462, 505007, 101595, 104095, 13932, 87527, 38966, 87604, 50336, 86085, 103291, 103592, 103098, 38257, 104532, 101162, 504897, 87018, 104336, X1693, 59637, 103143, 102933, 505021, 103631, 103552, 82678, X1635, X0162, 14065, 103530, 505568, 500034, 250498, 103099, 38438, 76359, 104138, 101561, 101560, 47043, 17394, 105271, X1215, 13057, 250475, 102569, 103770, 86313, X1179, X1523, 504632, 504527, X1412, 104942, 38317, 103101, 26119, 103509, 103544, 85433, 103547, 250122, 102896, 85158, 41342, 86012, 103907, 103909, 101408, 38420.
RECALLING FIRM/MANUFACTURER
Philips Medical Systems Sales & Service Region No. America, Bothell, WA.
REASON
Potential for patient burn.
VOLUME OF PRODUCT IN COMMERCE
156 units.
DISTRIBUTION
Nationwide.
PRODUCT
Medtronic FLP VAD Venous Cannula for Ventricular Assist, Models 95036 and CB95036.
CODE
Lot numbers 2003080199, 2003090147, 2003100101, 2004010371, and 2004030070.
RECALLING FIRM/MANUFACTURER
Medtronic, Inc., Minneapolis, MN.
REASON
Spiral tip from DLP VAD cannula may be loose and potentially detach during use.
VOLUME OF PRODUCT IN COMMERCE
753 units.
DISTRIBUTION
Nationwide and Internationally.
PRODUCT
a) Impervious Stockinette, Large, 12” X 48” or 12” X 54”, Ref
#9806-54, #9816-54, and #9978-12; X-Large, 16” X 48”, Ref
#9978-16.
b) Stockinette, Single Ply Econo Weave, 8 X 72", Ref #9999-72.
c) Stockinette, Single Ply, Standard Weave, 8 X 72” and 10 X 48”, Ref #9508-72, #9510-48, #9510-48, #9510-60.
d) Stockinette, Double Ply, Standard Weave, 6 X 60", 6 X 72", 8 X 48", 8 X 60", 8 X 72", 10 X 48".
e) Stockinette, Double Ply, Econo Weave, 6 X 54", 6 X 60", 6 X 72", 8 X 48", 8 X 72", Ref #9984-54, #9984-54, #9985-60, #9986-72, #9987-48, #9988-60, #9988-72.
f) Stockinette, Double Ply, Econo Weave, EZ Roll, 6 X 60”, 8 X
60”, Ref#9991-60, #9993-60.
CODE
a) Lot # 1007298 1007335 1007360 1030228 1030244 1045526 1049113
1081519 1107441 1141972 1161316 1161332 1200979 1225252
1306538 1310369 1315021 1330597 1330600 1342811 1375681
1375808 1375816 1490540 1504008 1504032 1521713 1543630
1571438 1650988 1736028 1759511 1802072 1911543 1911615
2076658 2076738 2137810 2137828 2138062 2187067 2228121
2376204 2701345 765821 765855 804316 805431 805440 825580
825598 825635 825643 850507 900691 932991 933071 933097
933118 933151 950663 950680 950743 974585 974593 974665
b) Lot: 1503929
c) Lots: 1030279 1600315 929506 929514
d) Lot: 1007212 1030172 1030181 1142043 1225324 1225332 1225341 1225359 1306298 1330669 1330677 1341190 1375832 1375841 1375867 1375875 1407072 1431304 1521772 1543672 1543681 1627016 1634152 1678542 1678606 1736079 1736087 1736095 1736108 1736116 1800384 1800608 1876426 1911375 1911421 2055689 2075217 2136454 2136471 2268684 2375877 728683 783421 805360 805378 805482 825555 900762 900797 900800 974681 974729
e) Lots: 9984-54 9985-60 9991-60 9986-72 9987-48 9988-72
f) Lot: 932924 1330482 1407048.
RECALLING FIRM/MANUFACTURER
Deroyal Surgical-Rose Hill, Rose Hill, VA.
REASON
Surgical drapes and covers were packaged in containers with incomplete seals which compromise product sterility.
VOLUME OF PRODUCT IN COMMERCE
3,098 cases.
DISTRIBUTION
Nationwide, Ireland, Peru and Israel.
PRODUCT
Architect Anti-HBs Reagent Kit; list 7C18-20, 4 x 100 tests, list 7C18-25, 100 tests and list 7C18-30, 4 x 500
tests.
CODE
List 7C18-20, lots 10014M200, 08672M200, 07780M300 and 17048M100; list 7C18-25, lots 08672M201, 10014M201, 15692M100 and 17048M101; list 7C18-30, lots 09140M300, 09238M200, 08458M200, 02348M300, 10015M200 and 13251M100. list 7C18-20, lots 10014M200, 08672M200, 07780M300 and 17048M100; list 7C18-25, lots 08672M201, 10014M201, 15692M100 and 17048M101; list 7C18-30, lots 09140M300, 09238M200, 08458M200, 02348M300, 10015M200 and 13251M100.
RECALLING FIRM/MANUFACTURER
Abbott Laboratories HPD/ADD/GPRD, Abbott Park, IL.
REASON
There is a potential for significant differences in quantitative results for certain specimens when using the affected reagent lots.
VOLUME OF PRODUCT IN COMMERCE
6,691 kits.
DISTRIBUTION
Canada, Japan, Australia, New Zealand, Germany and England.
PRODUCT
Neonatal Total Galactose Test Kits. Reagents for 960 (4800) assays, Item #NG-4000.
CODE
Lot #244468.
RECALLING FIRM/MANUFACTURER
PerkinElmer LAS Inc., Norton, OH.
REASON
The test kits are producing lower values than expected which may result in an increased number of false positive results increasing the risk of missing a baby with
galactosemia.
VOLUME OF PRODUCT IN COMMERCE
49 kits.
DISTRIBUTION
IN, MO, TX, WI, and MN.
PRODUCT Crosseal, fibrin sealant (human), One Applicator, Two Vial cups, Sterile, Disposable, Single Use, Rx Only.
CODE
Lot 0404 exp. Oct. 2008; Lot 0417 exp. Jan. 2009; Lot 0419 exp. Jan. 2009.
RECALLING FIRM/MANUFACTURER
Omrix Biopharmaceuticals, Ltd., Plasma Fract. Inst., Magen David Adom Blood Services Ctr., Ramat
Gan, Israel.
REASON
5 mL Crosseal fibrin sealant (human) is difficult to expel from the device resulting in product leakage from the vial insertion port.
VOLUME OF PRODUCT IN COMMERCE
4,277 units.
DISTRIBUTION
Nationwide.
WEEK ENDING OCTOBER 16
PRODUCT
Aquasorb Æ Border, Hydrogel Wound Dressing with Polyurethane Film Border, 25
individual packets per case, Catalog # 46-511, 2.5' x 2.5', labeled in part
***Sterile***.
CODE
Lot 360025.
RECALLING FIRM/MANUFACTURER
DeRoyal Wound Care, Rose Hill, VA.
REASON
Wound dressing failed to meet sterility specifications.
VOLUME OF PRODUCT IN COMMERCE
4 cases.
DISTRIBUTION
CA, OH and PA.
PRODUCT
a) Automated Differential Cell Counter.
b) Automated Differential Cell Counter.
CODE
a) Part Numbers: Pentra 80 P8000000NUA10;
b) Part Numbers: Pentra 80XL P80XL000NUA10.
RECALLING FIRM/MANUFACTURER
ABX Diagnostics, Inc., Irvine, CA.
REASON
Device can generate overestimated platelet results.
VOLUME OF PRODUCT IN COMMERCE
94 units.
DISTRIBUTION
Nationwide.
PRODUCT
Ameriwater Portable RO+, model numbers MR01 and MR02. The device is reverse
osmosis system for dialysis.
CODE
Serial Numbers effected: 21013, 21014, 21026, 21031, 21042, 21052, 21053, 21071,
21072, 21073, 21079, 21080, 21090, 21091, 21092, 21100, 21101, 21110, 21111,
21121, 21122, 21125, 21127, 21193, 21194, 21200, 21201, 21203, 21204, 21205,
21208, 21210, 21215, 21222, 21223, 21236, 21237, 21240, 21258, 21259, 21263,
21270, 21328, 21333, 21336, 21338, 21352, 21371, 21379, 21384, 21387, 21388,
21015, 21016, 21017, 21021, 21023, 21024, 21032, 21033, 21034, 21035, 21036,
21043, 21049, 21050, 21051, 21054, 21066, 21089, 21093, 21094, 21095, 21097,
21098, 21108, 21109, 21123, 21126, 21128, 21142, 21143, 21156, 21157, 21160,
21162, 21163, 21164, 21165, 21166, 21167, 21168, 21169, 21176, 21178, 21179,
21180, 21183, 21185, 21186, 21191, 21192, 21212, 21216, 21217, 21224, 21225,
21228, 21229, 21244, 21249, 21260, 21268, 21269, 21271, 21272, 21274, 21275,
21276, 21282, 21283, 21298, 21299, 21300, 21302, 21307, 21311, 21316, 21329,
21332, 21339, 21342, 21343, and 21346.
RECALLING FIRM/MANUFACTURER
Ameriwater Inc., Dayton, OH.
REASON
The device has a potential risk of failure caused by faulty wiring within the
electrical component. The wires can overheat and eventually break in two.
VOLUME OF PRODUCT IN COMMERCE
148 units.
DISTRIBUTION
Nationwide.
PRODUCT
AxSYM Total B-hCG Reagent, Product list number: 7A59-22 and 7A59-21.
CODE
Lot #: 18357Q100, 18357Q101 (France), 18357Q102, 18357Q103 and 18357104(France).
Exp. date 11/11/04.
RECALLING FIRM/MANUFACTURER
Abbott Health Products, Inc., Barceloneta, PR.
REASON
Assay may give elevated patients results.
VOLUME OF PRODUCT IN COMMERCE
7,830 kits.
DISTRIBUTION
Nationwide and Internationally.
PRODUCT
Protege GPS Biliary Stent System (6Fr 9mm x 30mm).
CODE
SERB65-09-30-120 (.035) (Biliary) LOT #: 257441.
RECALLING FIRM/MANUFACTURER
Ev3, Plymouth, MN.
REASON
A customer complaint received indicated that a ProtÈgÈ stent, PN SERB65-09-30-120
was found in a pouch that was not properly sealed. The ProtÈgÈ GPS Stent System
is 'double barrier' packaged and consists of a sealed inner tray in a sealed
pouch. The pouch seal ensures sterility of the external surface of the inner
tray. The inner tray seal containing the device was not affected.
VOLUME OF PRODUCT IN COMMERCE
5 systems.
DISTRIBUTION
SC, NC, and TX.
PRODUCT
Applied Blunt Tip Trocar Models C0717 and C0718.
CODE
Trocars are assembled into procedure kits. Trocars have thier own serial number
but the kit lots are relevant and include: 45036301001 to 45036301007 containing
1 each of lot 126803 = 71 each; 45208001001 to 452084001008 containing 1 each
of lot 127322=8 each; 452451005001 to 452451005003 containing 1 each of lot
127322=3 each; 453029001001 to 453029001017 containing 1 each of lot 127322
= 17 each; 453715004002 containing 1 each of lot 126803 = 1 each; 454116005001
to 454116005002 containing 1 each of lot 127322 = 2; 454608003001 to 454608003006
containing 1 each of lot 172528 = 6 each.
RECALLING FIRM/MANUFACTURER
Applied Medical Resources, Corp., Rancho Santa Margarita, CA.
REASON
Inadequate trocar tip assembly that might yield parts to patients.
VOLUME OF PRODUCT IN COMMERCE
1,025 Systems.
DISTRIBUTION
Nationwide and Ireland.
PRODUCT
Lactosorb pin, 2.0 x 20 mm, sterile; ref. 948205.
CODE
Lot 439010 Exp. 2006-11.
RECALLING FIRM/MANUFACTURER Biomet, Inc., Warsaw, IN.
REASON
Mislabeled as to size; 1.55 mm pins are labeled as 2mm.
VOLUME OF PRODUCT IN COMMERCE
37 pins.
DISTRIBUTION
CO, MS, MO, MT, NY, OH, PA, TX, and WI.
PRODUCT
Mystic M-Style Mushroom Cup Mityvac Vacuum Assisted Delivery System Product
No: 10047.
CODE
Lot Number: 30115.
RECALLING FIRM/MANUFACTURER CooperSurgical, Inc., Trumbull, CT.
REASON
Product is mislabeled as a Mushroom Cap Vacuum but actually contains a Bell
Style Assisted Delivery System.
VOLUME OF PRODUCT IN COMMERCE
495 units.
DISTRIBUTION
Nationwide and Italy.
PRODUCT
BC Thrombin Reagent. Thrombin Time Test.
CODE
Lot 540624, Lot 540625, and Lot 540626.
RECALLING FIRM/MANUFACTURER
Dade Behring, Inc., Newark, DE.
REASON
The product does not demonstrate the expected prolonged thrombin times for specimens
obtained from patients receiving heparin.
VOLUME OF PRODUCT IN COMMERCE
3,942 kits.
DISTRIBUTION
CT, FL, IA, IL, NJ, PA, WI, and Internationally.
PRODUCT
a) Medtronic MiniMed Paradigm Model 712.
b) Medtronic MiniMed Paradigm Model 512.
CODE
a) Serial numbers that end in ì-Dxx3î. Where ìxî can be any
number or letter;
b) Serial numbers that end in ì-Axx4î. Where ìxî can be any
number or letter.
RECALLING FIRM/MANUFACTURER
Medtronic MiniMed, Northridge, CA.
REASON
Several complaints of pumps exhibiting A47 alarms were received. Investigation
revealed the alarms occurred when the pump attempted to display text that exceeded
the maximum display length of the pump screen. The investigation revealed this
only occurs when the pump is programmed to display information in Spanish.
VOLUME OF PRODUCT IN COMMERCE
10,600 units.
DISTRIBUTION
Nationwide and Spain.
WEEK ENDING OCTOBER 23
PRODUCT
a) Mouse anti-TTF-1 concentrate used in a panel designed to
differentiate between different types of cancer when used to
test a patient.
b) Mouse anti-TTF-1 2nd Gen Predilute used in a panel designed to
differentiate between different types of cancer when used to
test a patient.
CODE
a) Product Number 18-0221, no 510(k) number, Product Code MXZ,
lot code 40280262;
b) Product Number 18-1221, no 510(k) number, Product Code MXZ,
lot code 40180205.
RECALLING FIRM/MANUFACTURER
Zymed Laboratories Inc., South San Francisco, CA.
REASON
Mouse anti-TTF-1 had become contaminated by immunoglobulins from mouse serum
that was co-eluted during purification leading to false readings.
VOLUME OF PRODUCT IN COMMERCE
24 vials of concentrate, and 4 vials of 2nd Gen Predilute.
DISTRIBUTION
MD, GA, TX, OH, IL, DC, NY, CA, France, Portugal, China, Italy, Israel, Philippines,
and Germany.
PRODUCT
Aeroset Analyzer Systems Operation Manual.
CODE
All Systems Operations Manuals.
RECALLING FIRM/MANUFACTURER
Abbott Laboratories, Inc., Irving, TX.
REASON
Changes to the printer default settings made by users/operators cause printed
data to be truncated and misidentified.
VOLUME OF PRODUCT IN COMMERCE
736 units.
DISTRIBUTION
Nationwide and Internationally.
PRODUCT
Fresenius Peritoneal Dialysis Premier Luer Lock Transfer Set 8", Sterile
with Female Safe Lock Connector Luer Lock and Slide Clamp Catalog Number: 050-30034.
CODE
Lot Numbers: 3LR240, 3NR190, 3SR074, 4AR119, 4BR187, 4CR290.
RECALLING FIRM/MANUFACTURER
Fresenius Medical Care North America, Lexington, MA.
REASON
Peritoneal Dialysis Transfer Set may have occluded pathway.
VOLUME OF PRODUCT IN COMMERCE
3,510 units.
DISTRIBUTION
Nationwide and Canada.
PRODUCT
Trio Monitor. Cardiac Monitor (including cardiotachometer and rate alarm).
CODE
Serial Number: MC02017-I3; MC02134-I3; MC02138-I3; MC02148-I3; MC02321-K3; MC02002-I3;
MC02005-I3; MC02006-I3; MC02008-I3; MC02020-I3; MC02023-I3; MC02025-I3; MC02027-I3;
MC02030-I3; MC02034-I3; MC02043-I3; MC02044-I3; MC02048-I3; MC02052-I3.
RECALLING FIRM/MANUFACTURER
Datascope Corp., Mahwah, NJ.
REASON
Software anomaly where the variable heart rates may be displayed inaccurately
or intermittent "dashes" may be displayed when the patient's heart
rate is derived from ECG.
VOLUME OF PRODUCT IN COMMERCE
19 units.
DISTRIBUTION
CA, IN and SC.
PRODUCT
HealthPax (HP-1). Cranial Electrotherapy Stimulator GZJ.
CODE
HP-1.
RECALLING FIRM/MANUFACTURER
Health Directions, Inc., Morrisville, PA.
REASON
Implied claims (extremes of compulsive or addictive behavior) are outside the
scope of clearance for the device.
VOLUME OF PRODUCT IN COMMERCE
3,475 units.
DISTRIBUTION
CA, MA, NJ, NY, PA, TX and WA.
WEEK ENDING OCTOBER 30
PRODUCT
Emission Computed Tomography System. Including Model numbers: 2200967, ASM001080,
ASM001083, ASM001092, ASM001099, ASM000085, 10036800109, and Varicam.
CODE
All serial numbers are being recalled.
RECALLING FIRM/MANUFACTURER
General Electric Med. Systems LLC, Waukesha, WI.
REASON
GE Healthcare received reports of intermittent low ejection fraction (EF) results
for MUGA (multi-gated acquisitions) studies on Millennium VG systems. This issue
can occasionally lead to distortions in the volume curve in some MUGA studies,
and result in calculated EF values lower (never higher) than actual values.
A low EF value could lead to a physician decision to stop needed treatment.
VOLUME OF PRODUCT IN COMMERCE
853 units.
DISTRIBUTION
Nationwide and Internationally.
PRODUCT
a) Stryker Navigation System - Hip Module Patient Tracker,
green; Model 6007-005-000.
b) Stryker Navigation System - Hip Module Patient Tracker, blue;
Model 6007-010-000.
CODE
a) Serial numbers PG04-0001 through PG04-0017;
b) Serial numbers PG04-0006 through PG04-0017.
RECALLING FIRM/MANUFACTURER
Stryker Instruments, Instruments Div., Kalamazoo, MI.
REASON
The tracker is not locking adequately to the anchoring pin, resulting in unacceptable
movement, which will affect the patient tracker system.
VOLUME OF PRODUCT IN COMMERCE
17.
DISTRIBUTION
Nationwide.
PRODUCT
iBOT 3000 Mobility System, Independence iBOT 3000 Mobility System (Stair-climbing
wheelchair). Catalog No.'s IT000101-IT000148 (inclusive).
CODE
Vehicle Identification Numbers ("VIN"s): 020404-000347, 020104-000309,
020504-000357, 021203-000308, 020204-000331, 020104-000314, 020104-000318, 021203-000301,
020304-000335, 020104-000310, 020404-000341, 021203-000299, 020204-000330, 020104-000322,
020204-000332, 020404-000350, 020404-000348, 020304-000340, 020404-000346, 020404-000342,
021203-000307, 020204-000333, 020104-000319, 021203-000304, 020404-000352, 020304-000337,
020304-000336, 020504-000356, 020404-000349, 021203-000305, 020604-000360, 020104-000317,
021203-000306, 020504-000358, 020404-000345, 020504-000354, 020304-000334, 020304-000338,
020404-000344, 020404-000351, 020504-000355, 020104-000323, 020104-000324, 021103-000291,
021203-000300, 020304-000339, 020404-000343, 020504-000359, 020604-000363, 020604-000364,
020604-000365, 020604-000366, 020604-000367, 020604-000368, 020604-000362, 021103-000290,
021103-000295, 020104-000315, 020104-000316, 020104-000320, 020104-000311, 020104-000313,
020204-000329, 020504-000353, 020104-000312, 020604-000361, 021003-000280, 020204-000328,
020104-000321, 020204-000326, 021203-000297, 021203-000298, 021203-000303, 020204-000327,
021103-000294, 021103-000286, 021103-000287, 021003-000273, 021103-000282, 021103-000292,
021103-000293, 021003-000276, 021003-000277, 021103-000285, 021103-000289, 020204-000325,
021103-000296, 021003-000278, 021003-000281, 021003-000272, 021003-000279, 021003-000274,
021003-000275, 021103-000283, 021103-000288, 021103-000284, 021203-000302.
RECALLING FIRM/MANUFACTURER
Independence Technology LLC, Warren, NJ.
REASON
Rocks and debris can enter into the wheel cluster housing through a damaged
wheel cap causing wheel malfunction.
VOLUME OF PRODUCT IN COMMERCE
81 chairs.
DISTRIBUTION
Nationwide, UK, Ireland, and New Zealand.
PRODUCT
Stryker brand Rugged EZ-Pro R4 Ambulance Cot, model 6092.
CODE
Serial numbers 040639739, 040639849, 040639737, 040639738, 040639675, 040639674,
040639670, 040639671, 040639672, 040639716, 040639847, 040639848, 040639935,
040639936 and 040639937.
RECALLING FIRM/MANUFACTURER
Stryker Medical, Kalamazoo, MI.
REASON
Locking pins may be installed backward, and cot may drop from highest to lowest
position when the users are not prepared for the drop.
VOLUME OF PRODUCT IN COMMERCE
15 units.
DISTRIBUTION
AL, CA, FL, KS, MT, OK, TX and VA.
PRODUCT
Difco (TM) BBL (TM) Gram Crystal Violet, 250 mL bottles in packs of four, catalog
# 212525 and 3.8 L bottles, catalog # 212526, staining reagent contained in
Becton Dickinson's BD Gram Stain Kits and Reagents, catalog #212539.
CODE
Lots 4223220, Exp 5/28/05 4226886, Exp 5/28/05 4233156, Exp 5/28/05 4237134,
Exp 6/30/05 4226882, Exp 5/28/05 4237144, Exp 6/30/05.
RECALLING FIRM/MANUFACTURER
Becton Dickinson and Co, Sparks, MD.
REASON
Reagent for microbiological testing may cause inconsistent staining characteristics
and subsequent misidentification of bacteria in patient samples.
VOLUME OF PRODUCT IN COMMERCE
8,685 bottles.
DISTRIBUTION
Nationwide and Internationally.
PRODUCT CryoValve, Pulmonary Valve & Conduit.
CODE
Donor Number: 68668, Serial Number: 8037696, Model Number: PV00.
RECALLING FIRM/MANUFACTURER Cryolife, Inc., Kennesaw, GA.
REASON
A pre-implant swab culture of the aortic valve, which was procured from the
same donor, was positive for Methicillin resistant Staphyococcus aureus.
VOLUME OF PRODUCT IN COMMERCE
1 tissue.
DISTRIBUTION
CA.
PRODUCT
SOLO-care® Plus with Aqualube, Multi-purpose solution*** AquaSoft (Private
label), Multi-purpose solution***Sterile***For soft lenses, Made in Canada,
Assembled in Canada, The product is distributed in units of single or twin packs
containing 2 to 12 ounces of product.
CODE
All sizes, configurations and all third part branded products.
RECALLING FIRM/MANUFACTURER
Ciba Vision Corporation, Duluth, GA.
REASON
Stability tests at 24 months revealed failing results for disinfection efficacy.
VOLUME OF PRODUCT IN COMMERCE
5,913,120 units.
DISTRIBUTION
Nationwide and Internationally.
PRODUCT
EP Healing Abutment -- An abutment device is a premanufactured prosthetic component
directly connected to the endosseous dental implant. Product Code: THA54.
CODE
Lot No. 249229.
RECALLING FIRM/MANUFACTURER
Implant Innovations, Inc., Palm Beach Gardens, FL.
REASON
The nylon bag holding the product was not completely sealed before sterilization.
Patients could be infected if the abutment is implanted.
VOLUME OF PRODUCT IN COMMERCE
216 units.
DISTRIBUTION
Nationwide and Internationally.
PRODUCT
GE CIC Central Station Monitoring Product, Versions 3.x, 4.0.5, and 4.0.6.
CODE
All serial numbers for versions 3.x, 4.0.5, and 4.0.6.
RECALLING FIRM/MANUFACTURER
General Electric Medical Systems Information Technology, Milwaukee, WI.
REASON
The following two conditions may occur with the printing of full disclosure
strips and reports, printing of the alarm history events stored on a GE Unity
Network Patient Data Server (PDS), and printing of caliper reports in the GE
CIC Pro central station monitoring product, 1) a delay or cessation in updating
portions of th display screen: or, 2) System reset/restart.
VOLUME OF PRODUCT IN COMMERCE
11,778 units.
DISTRIBUTION
Nationwide and Internationally.
PRODUCT
Remel Chocolate Agar, catalog # 01301, containing 15 plates/pkg, for in-vitro
diagnostic use.
CODE
Lot #435902, Exp. 2004-11-04.
RECALLING FIRM/MANUFACTURER Remel, Inc., Lenexa, KS.
REASON
The product does not perform as intended with some quality control organisms (Haemophilus
influenzae).
VOLUME OF PRODUCT IN COMMERCE
308/15-plate packages.
DISTRIBUTION
IA, KS, MO, FL, NY, IL, AR, TX, and OK.
Gillette Announces Recall of Oral-B CrossAction Power and PowerMAX Toothbrushes and Refills
Reports have been received concerning instances where the brush head of an Oral-B CrossAction Power or PowerMAX toothbrush has become loose in the mouth during brushing.
It is therefore recommend that consumers stop using the CrossAction Power or PowerMAX toothbrushes and brush refills to assist in brushing the teeth of people with special needs. No other Oral-B product is affected.
If you are currently using a CrossAction Power or PowerMAX to assist in brushing the teeth of a person with special needs, Oral-B will replace it -- at no charge -- with our Oral-B Advance Power 400, which is better suited for use in assisted-brushing environments because it has a brush head without a latch.
Please call 1-800-496-6557 between the hours of 9:00 AM and 5:00 PM (EST) on Monday through Friday and Oral-B will send you a prepaid mailing envelope to return your CrossAction Power or PowerMAX toothbrush to us. Once received, we will send you an Oral-B Advance Power 400 battery toothbrush.
Also, caregivers can find helpful information about brushing the teeth of people with special needs on the website of the American Dental Association:
http://www.ada.org/public/topics/special_needs.asp
General Information
Gillette is redesigning the brush heads for these toothbrushes to eliminate
issues related to the unlatching of the brush head. The new brush head refills,
when available, will fit into the current CrossAction Power and PowerMAX toothbrushes.
The new Oral-B CrossAction Power and CrossAction PowerMAX toothbrushes and brush
refills are scheduled to be available in stores in January 2005.
You can call 1-800-496-6557 as noted above if you have any questions regarding the use of these products or this notice.