MARCH 2003
WEEK ENDING MARCH 1
***CORRECTION***
Enforcement Report for February 19, 2003, (03-08), the recall listing for
Baxter, Recall # Z-0504-3/0523-3, list Baxter Healthcare, Corp., Deerfield,
IL., as the manufacturer, the correct manufacturer is Baxter Healthcare, Corp.,
Cleveland, MS., and Baxter Healthcare, Corp., Albonito, PR.
PRODUCT
Revolution
XR/d Wall Stand, Model Number 2291655, for Revolution XR/d Digital Radiographic
Imaging Systems .
CODE All serial numbers in distribution at the time of the recall.
RECALLING FIRM/MANUFACTURER General Electric Med Systems, Waukesha, WI,
REASON
The devices
do not meet requirements in 21 CFR 1010.2 in that they lack certification
labels.
VOLUME OF PRODUCT IN COMMERCE 78 devices.
DISTRIBUTION Nationwide.
PRODUCT Pharmaseal Thoracentesis Tray with Catheter, catalog number
4341B; a sterile, single use, Rx, latex free procedure tray containing a
7" long Thoracentesis Catheter with a 14 gauge x 2" needle, catheter
guard, 60 ml syringe, needles, 5 ml syringe, drainage tube with needle, fluid
collection bag, three prelabeled specimen vials with caps, 2 swab sticks, three
gauze pads, fenestrated drape with adhesive strips, towel, 5 ml -1% lidocaine
hydrochloride, hospital wrap and bandage.
CODE L1J035X, L1J088, L1K042, L1K058, L1K070, L1L057X, L1N018, L1N031,
L1N039X, L1N094, L1P017, L1P045, L1S015, L2A012, L2A021, L2A048, L2A075,
L2B039, L2B047 ,L2B073, L2C012, L2C038, L2C083, L2D041, L2D067, L2E036
RECALLING FIRM/MANUFACTURER Recalling Firm: Allegiance Healthcare,
Corp., McGaw Park, IL,
Manufacturer: Allegiance Healthcare, Corp., Mannford, OK
REASON Raw material used in the manufacture of catheters may contain a
contaminant predisposing the catheter to become brittle, possibly breaking
during use.
VOLUME OF PRODUCT IN COMMERCE 123,050 trays.
DISTRIBUTION Nationwide.
PRODUCT a) Y-Type Blood/Solution Set, catalog #1C8029; Standard Blood
Filter (170 to 260 micron), Drip Chamber Pump, Injection Site, Male Luer Lock Adapter,
111'' long, 10 drops per mL.
b)
Y-Type Blood/Solution Set, catalog #1C8117; Standard Blood Filter (170 to 260 micron), Drip Chamber Pump, Injection
Site, Male Luer Lock Adapter, 88'' long,
10 drops per Ml
c)
Y-Type Blood/Solution Set, catalog #1C8372; Standard Blood Filter (170 to 260
micron), Drip Chamber Pump, 2 Injection
Sites, Male Luer Lock Adapter, 111'' long, 10 drops per mL.
d)
Y-Type Blood/Solution Set, catalog #1C8459; Standard Blood Filter (170 to 260
micron), Drip Chamber Pump, 3 Injection Sites, Male Luer Lock Adapter,
111" long, 10 drops per mL.
e)
Y-Type Blood/Solution Set with Pressure Pump, catalog #1C8658; Standard Blood
Filter (170 to 260 micron), Drip Chamber Pump, Injection Site, Male Luer Lock
Adapter, 109" long, 10 drops per mL.
CODE
a) Lots GR182881, GR183558, GR184200, GR185900;
b) Lot GR183715;
c) Lots GR183152, GR183483, GR184622, GR185652;
d) Lots GR182873, GR185678;
e) GR184184, GR186551.
RECALLING FIRM/MANUFACTURER Recalling Firm: Baxter Healthcare Corp.,
Round Lake, IL,
Manufacturer: Baxter Healthcare Corp.,
Cleveland, MS.
REASON Leakage may occur at the inlet port.
VOLUME OF PRODUCT IN COMMERCE
a) 21,134 sets;
b) 3,792 sets;
c) 42,330 sets;
d) 4,320 sets;
e) 1,392 sets.
DISTRIBUTION Nationwide.
PRODUCT
Stryker
brand Neptune waste management system, model 711-27-1
CODE
All systems
with a rover top cap assembly 711-27-001 manufactured from 1/1/01 to 12/31/01.
RECALLING FIRM/MANUFACTURER Stryker Instruments, Instruments Div.,
Kalamazoo, MI,
REASON
Cap may be cracked
and fail, exposing O.R. staff to patient blood, and cause an unexpected loud
noise in the O.R.
VOLUME OF PRODUCT IN COMMERCE 370.
DISTRIBUTION Nationwide.
\
PRODUCT a) Pursuit ureteral dilation balloon catheters, 6 mm x 4 cm. Order # PAB-050080-6-4.0
b)
Pursuit ureteral dilation balloon catheters, 6 mm x 8 cm. Order #
PAB-050080-6-8.0.
CODE
a) Lot 1049345;
b) Lot 1050160.
RECALLING FIRM/MANUFACTURER Cook Urological, Inc., Spencer, IN,
REASON
Product is
labeled sterile, but was not sterilized.
VOLUME
OF PRODUCT IN COMMERCE 24.
DISTRIBUTION Ak, MN, OH, TN, Hawaii, and Ireland.
PRODUCT Steris Quick Connect
(tethered) C1406, QLC1406, C1473, and QLC1473 Quick Connect for Olympus
30/40/160/200/240 Series Bronchoscopes in the C1140 Universal Processing Trays. Catalog numbers: C11473 and
QLC1473.
CODE Lot nos: 3508686 or smaller (C1406 and QLC1406);
Lot nos: 3147303 or smaller
(C1473 and QLC1473).
RECALLING FIRM/MANUFACTURER Steris Corp., Mentor, OH,
REASON The Steris device may not
attach properly to retrofitted Olympus
ronchoscopes, and become loose during the sterile processing cycle.
VOLUME OF PRODUCT IN
COMMERCE 1,140
units.
DISTRIBUTION Nationwide and
Internationally.
PRODUCT Spine-Tech Ancillary
Discectomy Instrument (Box Curette), Catalog Number 07.00254.001, in Spine-Tech
Ancillary Discectomy Instrument Set, Catalog Number 07.00316.001.
CODE Lot numbers P010900 and P010957.
RECALLING FIRM/MANUFACTURER Sulzer Spine-Tech,
Minneapolis, MN,.
REASON The product may have a
twisted or broken tip.
VOLUME OF PRODUCT IN
COMMERCE 14
instruments.
DISTRIBUTION CA and MS.
PRODUCT ChemMate (brand) Keratin
Primary Antibody, Clones AE1, AE3, CAM 5.2 and 35betaH11. Catalog number
PAB111.
CODE Lot numbers: R20061A, R20429A, R21002A, and R21002002B.
RECALLING FIRM/MANUFACTURER Ventana Medical Systems,
Inc., Tucson, AZ,
REASON Inappropriate staining of
tissue specimen caused by decreased sensitivity with certain component of
antibody cocktail.
VOLUME OF PRODUCT IN
COMMERCE 60.
DISTRIBUTION Nationwide.
PRODUCT Boston Medical Products
Otocell Ear Wick 15 mm, sterile Ref: EW-0915 .
CODE Lot Number: 56003 Exp. Date
: 7/2005.
RECALLING FIRM/MANUFACTURER Boston Medical Products,
Inc., Westborough, MA, by
REASON Product sterility may be
compromised.
VOLUME OF PRODUCT IN
COMMERCE 267
units.
DISTRIBUTION Nationwide.
PRODUCT Prism 3000S and 3000XP Gamma
Camera Systems.
CODE All serial numbers of Models N210060 and N210335.
RECALLING FIRM/MANUFACTURER Philips Medical Systems,
Inc., Cleveland, OH,
REASON The radius motion may exceed
the hardware limits, resulting in the detector head to drop towards the
patient.
VOLUME OF PRODUCT IN
COMMERCE 311
units.
DISTRIBUTION Nationwide and
Internationally.
PRODUCT
a) Hip Prosthesis as a
component of Primaloc
Cementless/Collared Hip System.
b) Hip Prosthesis as a
component of Primaloc
Cementless/Collarless Hip System.
c) Hip Prosthesis as a
component of Primaloc Cemented Hip
System.
d) Hip Prosthesis as a
component of TriFix J Cemented Hip.
(Manufactured for distribution in Japan only.)
CODE a) Catalog # 112-0915, Lot
# OD-18031, W003152, WO#003152. Catalog
# 112-0917, Lot # OD-19103-1, OD-20567,W0#003109. Catalog # 112-1015, Lot # OD-19038, W0#003155. Catalog # 112-1017, Lot # OD-17403. Catalog # 112-1115, Lot # WO#003045. Catalog # 112-1117, Lot # OD-23210,
OD-23211. Catalog # 112-1119, Lot #
OD-23232. Catalog # 112-1217, Lot #
OD-23206. Catalog # 112-1219, Lot # OD-23233. Catalog # 112-1221, Lot # OD-23252 Catalog # 112-1223, Lot # OD-17321, W003725, WO#003350. Catalog
# 112-1319, Lot # OD-20572, OD-20867-1, OD-23234, OD-23234-1, OD-23246. Catalog # 112-1321, Lot # OD-23253. Catalog
# 112-1421, Lot # OD-23248, OD-23248-1, OD-23251, OD-23251-1. Catalog # 112-1425, Lot # OD-21874. Catalog # 112-1623, Lot # OD-21147-1. Catalog # 112-1625, Lot # W0#003130,
WO#003139.
b) Catalog # 111-0917, Lot #
OD-22372; Catalog # 111-1015, Lot #
OD-21589, OD-21589-1; Catalog #
111-1017, Lot # OD-23209; Catalog #
111-1019, Lot # OD-23235; Catalog #
111-1115, Lot # OD-22050; Catalog # 111-1117, Lot # OD-23205, OD-23205-1,
OD-23298; Catalog # 111-1119, Lot # OD-22846, W002952, OD-23222; Catalog #
111-1217, Lot # OD-23207, OD-23208, OD-23299; Catalog # 111-1219, Lot #
OD-22847, OD-20859-1, OD-23230, OD-22847-1; Catalog # 111-1221, Lot # OD-23254,
OD-23249; Catalog # 111-1223, Lot # OD-18165, W002524;
Catalog # 111-1319, Lot #
OD-23231, OD-23231-1;Catalog # 111-1321, Lot # OD-23247; Catalog # 111-1421,
Lot # OD-23250, OD-23255;
Catalog # 111-1423, Lot #
OD-23256; Catalog # 111-1425, Lot # OD-20837-1, W002683; Catalog # 111-1625,
Lot # OD-21873, OD-19626-1,
OD-22537.
c) Catalog # 114-0917, Lot #
OD-18030-5, W003600, W003601, W003602; Catalog # 114-1017, Lot # OD-18563,
OD-18563-1,
OD-23190, OD-23191,
OD-23221; Catalog # 114-1219, Lot # OD-23192, OD-23192-1, OD-23193, OD-23194;
Catalog # 114-1421, Lot # OD-17306, OD-17412; Catalog # 114-1623, Lot #
W0#003264, W003614;
d) Catalog # 115-1015, Lot #
W016134, W017080; Catalog # 115-1319, Lot # W016434.
RECALLING FIRM/MANUFACTURER Ortho Development
Corporation, Draper, UT,
REASON Cleaning agent residue from
packaging material was adhering to implantable hip stems.
VOLUME OF PRODUCT IN
COMMERCE a)
78; b) 105; c) 41; d) 14.
DISTRIBUTION TX and Japan.
PRODUCT
a) Ear Loop Procedure Masks,
for one time use only, Glass-Free,
Reorder No. 2201, packed in a cardboard Box (unit box). Made in China. Product is labeled with a Latex-free symbol that states
"This product does not contain any natural rubber latex".
b) Ear Loop Procedure Masks
with plastic shield, for one time use only, Glass-Free, Reorder No. 2202,
packed in a cardboard box (unit box).
Made in China. Product is
labeled with a Latex-free symbol that states "This product does not
contain any natural rubber latex".
CODE
a) Lot Number 14929, Barcode
# 16784 22012.
b) Lot Number 14929, Barcode
# 16784 22023.
RECALLING FIRM/MANUFACTURER Dynarex Corporation,
Orangeburg, NY
REASON Product labeled as
"latex free" but recent tests indicate that it contains minute
amounts of latex.
VOLUME OF PRODUCT IN
COMMERCE
a) 1937 cases;
b) 396 cases.
DISTRIBUTION Nationwide, Puerto Rico, and
the Bahamas.
PRODUCT Nobel Biocare Replace Select
Radiographic Template, part number 30159.
CODE No codes apply. All
physicians need to ensure the template is replaced.
RECALLING FIRM/MANUFACTURER Nobel Biocare USA, Inc., CA,
by recall letter and web.
REASON Misprinted radiographic
template plates. Misprinted labeling.
VOLUME OF PRODUCT IN
COMMERCE Unspecified.
DISTRIBUTION Nationwide and
Internationally.
PRODUCT Convertors three quarter
sheet (Drape Sheet), catalog #29350.
Individually packaged drapes, 20 packages per case.
CODE Lot number 02LFRW95.
RECALLING FIRM/MANUFACTURER Recalling Firm: Allegiance Healthcare, Corp., McGaw Park,
IL,
Manufacturer: Allegiance Healthcare, Corp., El Paso, TX
REASON The product labeled as
sterile had not been through its sterilzation cycle.
VOLUME OF PRODUCT IN
COMMERCE 320
drapes.
DISTRIBUTION NJ, GA, WA, and FL.
PRODUCT Surgisis peripheral vascular
patch; catalog number C-VSLH-$S-lx10.
CODE Lots SB102671, SB102908, LB203536, SB10189, SB101148,
SB101402,SB102006 and SB102670.
RECALLING FIRM/MANUFACTURER Cook Biotech, Inc., West
Lafayette, IN
REASON Adverse events-suspected
pseudoaneurysm.
VOLUME OF PRODUCT IN
COMMERCE 153.
DISTRIBUTION IN.
PRODUCT Life Vest (WCD 3000)
Electrode Belt. Wearable Cardioverter
Defibrillator. Product label 20A0051-FI
CODE Product No 10A0889-A01.
RECALLING FIRM/MANUFACTURER Lifecor, Inc., Pittsburgh,
PA,
REASON Wires may break or pull out,
causing device to malfunction.
VOLUME OF PRODUCT IN
COMMERCE 174
units.
DISTRIBUTION Nationwide and
Internationally.
PRODUCT
a) Gamma Drill
5.5x300mm. Recall # Z-0579-03;
b) Gamma AO Drill
5.5x300mm. Recall # Z-0580-03;
c) Gamma Drill Dia.
4.2x300mm. Recall # Z-0581-03;
d) T2 3.5x130mm AO Drill
Bit. Recall # Z-0582-03;
e) T2 3.5x130mm Tri Flat
Drill Bit. Recall # Z-0583-03;
f) T2 4.2mmx130mm AO Drill
Bit. Recall # Z-0584-03;
g) T2 4.2mmx130mm Tri Flat
Bit. Recall # Z-0585-03;
CODE
a) 1214-5300S -- Gamma Drill
5.5x300mm -- K400056; K412682; K423453; K451275; K467231; K469197; K494341;
K494342;K508650; K516457; K526734; K533284; K542636; K548132; K553788; K553789;
K562650; K562654; K573121; K573122; K579936; K579937; K579938; K619518;
K622108; K622109; K637404; K641097; K653378; K665203; K665204; K665205;
K678519; K679127; K713867; K713868; K714918; K714919;
K714920; K714921; K714922;
K714923; K714924; K714925; K714926; K717459; K717462; K721328; K724574;
K732012; K732013; K733474; K748814; K748817; K753097; K753098; K774660;
K775180; K784772; K784774; K784775; K821925; K829527; K829529; K834976;
K957002.
b) 1212-5300S -- Gamma AO
Drill 5.5x300mm -- K626400;K626401; K626402; K821823; K821824.
c) 1314-3042S Gamma Drill
Dia. 4.2x300mm -- K602221; K624126; K67480; K740593; K810155; K929678; K966326.
d) 1806-3550S T2 3.5x130mm
AO Drill Bit -- K561203; K569685; K569686; K591212; K591213; K653373; K713876;
K725975; K743641; K743642.
e) 1806-3555S T2 3.5x130mm
Tri Flat Drill Bit -- K569689; K591094; K591215.
f) 1806-4280S T2 4.2mmx130mm
AO Drill Bit -- K410511; K419243; K419247; K462304; K462306; K579531; K591907;
K591914; K591927; K591937; K637448; K713209; K713210; K723426; K737377;
K743644; K770908; K776715; K776716;
K776717; K776720; K776721; K796725; K796726; K823858; K823859; K823860; K823861; K834973.
g) 1806-4285S T2 4.2mmx130mm
Tri Flat Bit -- K727627;K727628; K727630; K823224; K823225; K824691;
K824692; K824693.
RECALLING FIRM/MANUFACTURER Stryker Howmedica Osteonics,
Mahwah, NJ,
REASON Certain sterile packed
drills with center tip may penetrate the foam and blister resulting in loss of
sterility.
VOLUME OF PRODUCT IN COMMERCE
a) 5062;
b) 53;
c) 12;
d) 290;
e) 27;
f) 2344;
g) 1.
DISTRIBUTION Nationwide.
PRODUCT Carton Label: "TDM Performance Verifier I"; 6x2
mL vials****Catalog Number 896 2540.
CODE T3606, Exp 31 Mar 2004.
RECALLING FIRM/MANUFACTURER Ortho-Clinical Diagnostics,
Rochester, NY,
REASON Cartons labeled "TDM
Performance Verifier I" may contain TDM Performance Verified II.
VOLUME OF PRODUCT IN
COMMERCE 1,338
cartons.
DISTRIBUTION Nationwide.
PRODUCT Vitros Immunodiagnostic
Products***Universal Wash Reagent***for in vitro diagnostic use only. The product is sold in cartons containing 2
bottles (5 liters each). CAT 838
9793.
CODE Lot 5003, EXP. 12 September
2003.
RECALLING FIRM/MANUFACTURER Ortho-Clinical Diagnostics,
Rochester, NY,
REASON Label printing
irregularities on cartons and/or bottles.
VOLUME OF PRODUCT IN
COMMERCE 284
cartons/568 bottles.
DISTRIBUTION Nationwide.
PRODUCT Boston Scientific
Microvasive Radial Jaw 3 Single-Use Maximum Capacity Biopsy Forceps Needle, 240
cm length, 3.3 mm Jaw Outside Diameter.
Each device is packaged in a Tyvek laminate pouch. The pouch assemblies are packaged in a
five-pack product box. CODE The code for the individual pouches is
Catalog 1589 and Lot # 0551847, expiration date of 2005-07;The code for the
five pack boxes is Catalog No. 1589 and lot number 0551847, expiration date of
2005-07.
RECALLING FIRM/MANUFACTURER Symbiosis, Corporation,
Miami, FL,
REASON Product does not have the
needle component as part of the jaw assembly.
VOLUME OF PRODUCT IN
COMMERCE 155
5-packs.
DISTRIBUTION Nationwide.
PRODUCT
a) HomeChoice Automated Peritoneal Dialysis Systems; Made in U.S.A.; catalog numbers 5C4471, 5C4471R, 5C4474, 5C4474R, T5C4441, T5C4441R
b) HomeChoice PRO Automated Peritoneal
Dialysis Systems; Catalog numbers 5C8310, 5C8310R, R5C8320, R5C8320R, T5C8300, T5C8300R.
CODE
a) All HomeChoice Systems
with software versions 8.51 and 8.52;
b) All HomeChoice PRO
Systems with software versions 8.51 and 8.52.
RECALLING FIRM/MANUFACTURER Recalling Firm: Baxter Healthcare, Corp., Round Lake, IL,
Manufacture: Baxter Healthcare, Corp., Largo, FL.
REASON Potential for low
ultrafiltrate volumes when using the Low Fill Mode during continuous cycling
peritoneal dialysis (CCPD) therapy.
VOLUME OF PRODUCT IN
COMMERCE 12,667
units.
DISTRIBUTION Nationwide and
Internationally.
PRODUCT Act 5 diff Cap Pierce
Hematology Analyzer 864.5220 Automated differential cell counter. .
CODE Not applicable.
RECALLING FIRM/MANUFACTURER Beckman Coulter, Inc., Brea,
CA,
REASON Failure to cycle properly
among samples.
VOLUME OF PRODUCT IN
COMMERCE 351.
DISTRIBUTION Nationwide and Canada.
PRODUCT Coulter LH 700 series and
GEN S Hematology Analyzers 864.5220 Automated differential cell counter.
CODE In the LH 700 series, all
serial numbers up to AE1088. In the
GENS series, all instruments.
RECALLING FIRM/MANUFACTURER Beckman Coulter, Inc., Brea,
CA
REASON Device design allows for
cross contamination to occur, possibly affecting assays and results reports.
VOLUME OF PRODUCT IN
COMMERCE 447.
DISTRIBUTION Nationwide and Canada.
PRODUCT X-Cel MB-700 A/C Podiatry X-Ray System.
CODE Model MB-700 A/C, serial numbers 622686, 622687,622688, 622689,
320367, and 325120.
RECALLING FIRM/MANUFACTURER X-Cel X-Ray Corp., Crystal Lake, IL,
REASON The X-Ray system contained a
defective X-Ray tube shield.
VOLUME OF PRODUCT IN COMMERCE 3 units.
DISTRIBUTION IL.
PRODUCT
a) COOK Stiffened Micropuncture(r) Introducer Set, 4 FR. Product Reorder
Number: MPIS-401-10.0-SC.
b) COOK Stiffened Micropuncture(r) Introducer Set, 4 FR. Product Reorder
Number: MPIS-401-10.0-SC-NT.
c) COOK Stiffened Micropuncture(r) Introducer Set, 4 FR. Product Reorder
Number: MPIS-401-10.0-SC-NT-U.
d) COOK Stiffened Micropuncture(r) Introducer Set, 4 FR. Product Reorder
Number: MPIS-401-15.0-SC-NT.
e) COOK Stiffened Micropuncture(r) Introducer Set, 4 FR. Product Reorder
Number: MPIS-401-20.0-SC-NT-UA-082900.
f) COOK Stiffened Micropuncture(r) Introducer Set, 5 FR. Product Reorder
Number: MPIS-501-10.0-SC.
g) COOK Stiffened Micropuncture(r) Introducer Set, 5 FR. Product Reorder
Number: MPIS-501-10.0-SC-LONDON-062593.
h) COOK Stiffened Micropuncture(r) Introducer Set, 5 FR. Product Reorder
Number: MPIS-501-10.0-SC-NT.
i) COOK Stiffened Micropuncture(r) Introducer Set, 5 FR. Product Reorder
Number: MPIS-501-10.0-SC-NT-U.
j) COOK Stiffened Micropuncture(r) Introducer Set, 5 FR. Product Reorder
Number: MPIS-501-10.0-SC-U.
k) COOK Stiffened Micropuncture(r) Introducer Set, 5 FR. Product Reorder
Number: MPIS-501-15.0-SC.
l) COOK Stiffened Micropuncture(r) Introducer Set, 5 FR.Product Reorder Number:
MPIS-501-15.0-SC-NT.
m) COOK Stiffened Micropuncture(r) Introducer Set, 5 FR. Product Reorder
Number: MPIS-501-15.0-SC-NT-MKDT-090600.
CODE All lots on the market as of November 25, 2002.
RECALLING FIRM/MANUFACTURER Cook, Inc., Bloomington, IN, by letters on
November 26, 2002.
REASON The tip of the stiffened
cannula might break during use and remain in the patient.
VOLUME OF PRODUCT IN COMMERCE 39,906.
DISTRIBUTION Nationwide, Brazil, Canada, Dominican Republic and Israel.
PRODUCT
a) Megadyne Mega Tip, E-Z Clean Non Stick Cautery Tip,Single Use Only, STERILE.
b) Megadyne Mega Tip, Modified E-Z Clean Non Stick Cautery Tip, Single Use
Only, STERILE.
CODE a) Catalog Number 0600. Lots 00664, 00845, 10038, 10190, 10541,
10669, 10703, 10766, 10951, 11110, 11356, 20046, 20219, 20520, 20592, 20757,
20777, 20856, 20997,21075;
b) Catalog Number 0600M. Lots 00994, 01880, 01181, 01112,10156, 10670, 11008,
11388, 20260, 20814.
RECALLING FIRM/MANUFACTURER Megadyne Medical Products, Inc., Draper, UT,
REASON The metal substrate cautery
tip may break during use.
VOLUME OF PRODUCT IN COMMERCEa) 19464;
b) 4,855.
DISTRIBUTION Nationwide and Internationally.
PRODUCT ThermaCool TC 1.0 cm2 Treatment Tip, Single Patient Use Only.
CODE M00023, M00024, M00029, M00031, M00033R, M00034, M00035.
RECALLING FIRM/MANUFACTURER Thermage, Hayward, CA,
REASON The product tip has
potential to liquid leakage into the internal circuitry causing malfunction and
may result in epidermal injury.
VOLUME OF PRODUCT IN COMMERCE 1,326 units.
DISTRIBUTION Nationwide.
PRODUCT MultiDiagnost 4 - x-ray system, multi functional/universal
system for general use.
CODE Serial numbers include (but are not limited to): 4584520996,
4594911196, 4595090397, 4595290697, 4594871196, 4594841196, 4595130197,
4615641197, 4595240797, 459500197, 4615511097, 4615581097, 4584570996,
4615320897, 459516497, 4595070197, 4584320996, 4584721096, 4615661197,
4594980197 Part numbers: 9896 000 58511, 9896 000 58512, 9896 000 58521, 9896
000 58522.
RECALLING FIRM/MANUFACTURER Philips Medical Systems, Bothell, WA,
REASON Internal short circuit - May
allow for stand and table to move spontaneously in upward or downward movement.
VOLUME OF PRODUCT IN COMMERCE 22 units.
DISTRIBUTION Nationwide.
PRODUCT
a) Modular femoral and proximal tibial replacement system consisting of femoral
head/neck, body, stem and condyle components. Also proximal tibial body
rotating hinge-type knee and stem components.
b) Modular Replacement System with Extra-Cortical Porous Coating, Femoral
Module Stem and Tibial Module Stem.
CODE Catalog No. Lot No. Description
6485-3-000 D6MR MRS SML FEM STEM 10 X 102; 6485-3-011 D6CEB MRS 11MM X 127MM
FEMORAL STEM; 6485-3-013 D6CFB MRS 13MM X 127MM FEMORAL STEM; 6485-3-013 D6CFG
MRS 13MM X 127MM FEMORAL STEM; 6485-3-013 D6CFE MRS 13MM X 127MM FEMORAL STEM;
6485-3-013 D6GZE MRS 13MM X 127MM FEMORAL STEM; 6485-3-013 D6GZC MRS 13MM X
127MM FEMORAL STEM; 6485-3-013 D6HAB MRS 13MM X 127MM FEMORAL STEM; 6485-3-013
D6HAC MRS 13MM X 127MM FEMORAL STEM; 6485-3-013 D6HAD MRS 13MM X 127MM FEMORAL
STEM; 6485-3-015 D6CHD MRS 15MM X 127MM FEMORAL STEM;
6485-3-017 D6BZC MRS 17MM X 127MM FEMORAL STEM; 6485-3-017 D6GPA MRS 17MM X
127MM FEMORAL STEM; 6485-3-017 D6GPD MRS 17MM X 127MM FEMORAL STEM; 6485-3-311
D6CAB MRS 11MM X 203MM FEMORAL STEM; 6485-3-311 D6CAA MRS 11MM X 203MM FEMORAL
STEM; 6485-3-311 D6GTA MRS 11MM X 203M FEMORAL STEM; 6485-3-311 D6GTB MRS 11MM
X 203MM FEMORAL STEM; 6485-3-313 D6CBC MRS 13MM X 203MM FEMORAL STEM;
6485-3-313 D6CBD MRS 13MM X 203MM FEMORAL STEM; 6485-3-313 D6CBB MRS 13MM X
203MM FEMORAL STEM; 6485-3-313 D6GWB MRS 13MM X 203MM FEMORAL STEM;
6485-3-313 D6GWC MRS 13MM X 203MM FEMORAL STEM; 6485-3-317 D6GM MRS 17MM X
203MM FEMORAL STEM; 6485-6-409 D6HMA MRS 9MM TIBIAL STEM; 6485-6-411 D5XWB MRS
11MM TIBIAL STEM; 6485-6-411 D6HNA MRS 11MM TIBIAL STEM;
b) Catalog No. Lot No. Description 6485-3-008 D6JC SMALL FEMORAL STEM SECTION
8MM X 102MM; 6485-3-011 D6CDA FEMORAL MODULE STEM SECTION 11MM X 127MM;
6485-3-011 D6CDE FEMORAL MODULE STEM SECTION 11MM X 127MM; 6485-3-011 D6CDF
FEMORAL MODULE STEM SECTION 11MM X 127MM; 6485-3-011 D6CEA FEMORAL MODULE STEM
SECTION 11MM X
127MM; 6485-3-013 D6ADA FEMORAL MODULE STEM SECTION 13MM X 127MM; 6485-3-013
D6ADB FEMORAL MODULE STEM SECTION 13MM X 127MM; 6485-3-013 D6ADC FEMORAL MODULE
STEM SECTION 13MM X 127MM; 6485-3-013 D6CFA FEMORAL MODULE STEM SECTION 13MM X
127MM; 6485-3-013 D6CFD FEMORAL MODULE STEM SECTION 13MM X 127MM; 6485-3-013
D6CFF FEMORAL MODULE STEM SECTION 13MM X 127MM; 6485-3-013 D6CFH FEMORAL MODULE
STEM SECTION 13MM X 127MM; 6485-3-013 D6CFI FEMORAL MODULE STEM SECTION 13MM X
127MM; 6485-3-013 D6CFJ FEMORAL MODULE STEM SECTION 13MM X 127MM; 6485-3-013
D6CFK FEMORAL MODULE STEM SECTION 13MM X 127MM; 6485-3-013 D6CGA FEMORAL MODULE
STEM SECTION 13MM X 127MM; 6485-3-013 D6CGF FEMORAL MODULE STEM SECTION 13MM X
127MM; 6485-3-013 D6NJA FEMORAL MODULE STEM SECTION 13MM X 127MM; 6485-3-013
D6NKB FEMORAL MODULE STEM SECTION 13MM X
127MM; 6485-3-013 D6NKC FEMORAL MODULE STEM SECTION 13MM X 127MM; 6485-3-013
D6NKD FEMORAL MODULE STEM SECTION 13MM X 127MM; 6485-3-015 D6CHA FEMORAL MODULE
STEM SECTION 15MM X 127MM; 6485-3-015 D6CHC FEMORAL MODULE STEM SECTION 15MM X
127MM; 6485-3-015 D6CHE FEMORAL MODULE STEM SECTION 15MMX 127MM; 6485-3-015
D6CHF FEMORAL MODULE STEM SECTION 15MM D6GRB FEMORAL MODULE STEM SECTION 15MM X
127MM; 6485-3-015 D6GRC FEMORAL MODULE STEM SECTION 15MM X 127MM; 6485-3-015
D6GRD FEMORAL MODULE STEM SECTION 15MM X 127MM;
6485-3-015 D6GRE FEMORAL MODULE STEM SECTION 15MM X 127MM; 6485-3-017 D5YEF
FEMORAL MODULE STEM SECTION 17MM X 127MM; 6485-3-311 D5YKAA FEMORAL MODULE 11MM
X 203MM CURVED LONGSTEM; 6485-3-311 D5YKAB FEMORAL MODULE 11MM X 203MM CURVED
LONGSTEM; 6485-3-311 D5YKAC FEMORAL MODULE 11MM X 203MM CURVED LONGSTEM;
6485-3-313 D6CBA FEMORAL MODULE 13MM X 203MM CURVED LONGSTEM; 6485-6-409 D5GBE
TIBIAL MODULE STEM SECTION 9MM X 127MM; 6485-6-411 D5XWA TIBIAL MODULE STEM
SECTION 11MM X 127MM.
RECALLING FIRM/MANUFACTURER Stryker Howmedica Osteonics, Mahwah, NJ,
REASON Some components have
disassociated post-operatively resulting in additional surgery to the patient.
VOLUME OF PRODUCT IN COMMERCE a) 136; b) 311.
DISTRIBUTION Nationwide and Internationally.
PRODUCT JOGRAPHY Cardiovascular Angiographic Catheter, 5F *** pigtail
*** tapered with perfusion ports *** Single use, sterile, disposable
intravascular catheters with an traumatic tip.
CODE Lot Number 41340.
RECALLING FIRM/MANUFACTURER Jomed Catheters, Inc., Queensbury, NY
REASON 6 Fr. Catheters packaged and
labeled as 5 Fr.
VOLUME OF PRODUCT IN COMMERCE 50 catheters.
DISTRIBUTION Germany, Hungary and Italy.
PRODUCT
a) Viridia/Agilent Information Center Model: M3150A#C01.
b) Viridia/Agilent Information Center Model: M3151A#C01.
c) Intellivue Information Center Model: M3150B#C01.
d) Intellivue Information Center Model: M3150BU#C01.
CODE
a) Serial 4102A02856 4102A10464 4102A10465 4102A10466 4102A10467 4102A10468
4102A10470 4102A10473 4102A10483 4102A10488 4102A10491 4102A10492 4102A10498
4102A10499 4102A10506 4102A10509 4102A10520 4102A10523 4102A10524 4102A10525
4102A10526 4102A10527 4102A10531 4102A10534 4102A10535 4102A10536 4102A10537
4102A10539 4102A10540 4102A10546 4102A10549 4102A10550 4102A10554 4102A10555
4102A10556 4102A10557 4102A10558 4102A10567 4102A10568 4102A10573 4102A10574
4102A10576 4102A10578 4102A10587 4102A10589 4102A10591 4102A10597 4102A10599
4102A10624 4102A10625 4102A10628 4102A10633 4102A10634 4102A10637 4102A10653
4102A10656 4102A10659 4102A10660 4102A10662 4102A10663 4102A10664 4102A10679
4102A10688 4102A10689 4102A10692 4102A10695 4102A10707 4102A10709 4102A10710
4102A10712 4102A10721 4102A10723 4102A10726 4102A10727 4102A10728 4102A10729
4102A10730 4102A10731 4102A10732 4102A10738 4102A10743 4102A10744 4102A10746
4102A10749 4102A10752 4102A10753 4102A10756 4102A10757 4102A10758 4102A10759
4102A10760 4102A10766 4102A10768 4102A10769 4102A10771 4102A10774 4102A10775
4102A10777 4102A10780 4102A10781 4102A10782 4102A10793 4102A10818 4102A10821
4102A10825 4102A10827 4102A10828 4102A10854 4102A10858 4102A10862 4102A10863
4102A10864 4102A10874 4102A10875 4102A10876 4102A10890 4102A10893 4102A10900
4102A10901 4102A10908 4102A10909 4102A10910 4102A10911 4102A10926 4102A10928
4102A10931 4102A10933 4102A10934 4102A10936 4102A10942 4102A10943 4102A10954
4102A10957 4102A10960 4102A10966 4102A10967 4102A10968 4102A10970 4102A10972
4102A10974 4102A10981 4102A10982 4102A11000 4102A11003 4102A11004 4102A11008
4102A11009 4102A11010 4102A11014 4102A11018 4102A11019 4102A11023 4102A11030
4102A11031 4102A11039 4102A11042 4102A11043 4102A11053 4102A11054 4102A11058
4102A11059 4102A11060 4102A11061 4102A11063 4102A11067 4102A11072 4102A11073
4102A11079 4102A11081 4102A11082 4102A11083 4102A11084 4102A11086 4102A11103
4102A11106 4102A11108 4102A11122 4102A11125 4102A11126 4102A11127 4102A11146
4102A11147 4102A11148 4102A11154 4102A11158 4102A11172 4102A11173 4102A11174
4102A11175 4102A11176 4102A11182 4102A11183 4102A11184 4102A11186 4102A11189
4102A11194 4102A11195 4102A11196 4102A11208 4102A11214 4102A11215 4102A11216
4102A11217 4102A11227 4102A11233 4102A11236 4102A11237 4102A11247 4102A11248
4102A11249 4102A11253 4102A11256 4102A11257 4102A11258 4102A11259 4102A11260
4102A11262 4102A11263 4102A11264 4102A11265 4102A11266 4102A11268 4102A11269
4102A11270 4102A11271 4102A11274 4102A11278 4102A11282 4102A11283 4102A11288
4102A11302 4102A11303 4102A11305 4102A11306 4102A11307 4102A11308 4102A11314
4102A11315 4102A11327 4102A11330 4102A11334 4102A11336 4102A11337 4102A11338
4102A11348 4102A11349 4102A11350 4102A11351 4102A11352 4102A11353 4102A11356
4102A11357 4102A11359 4102A11363 4102A11371 4102A11382 4102A11386 4102A11387
4102A11388 4102A11397 4102A11398 4102A11399 4102A11400 4102A11404 4102A11405
4102A11411 4102A11412 4102A11413 4102A11416 4102A11422 4102A11424 4102A11429
4102A11430 4102A11433 4102A11436 4102A11451 4102A11453 4102A11454 4102A11456
4102A11458 4102A11463 4102A11467 4102A11471 4102A11472 4102A11473 4102A11481
4102A11482 4102A11493 4102A11495 4102A11496 4102A11500 4102A11501 4102A11507
4102A11511 4102A11516 4102A11517 4102A11518 4102A11520 4102A11530 4102A11531
4102A11532 4102A11536 4102A11537 4102A11543 4102A11544 4102A11545 4102A11549
4102A11552 4102A11561 4102A11562 4102A11564 4102A11565 4102A11566 4102A11567
4102A11568 4102A11569 4102A11576 4102A11577 4102A11578 4102A11579 4102A11588
4102A11590 4102A11591 4102A11592 4102A11593 4102A11595 4102A11598 4102A11600
4102A11606 4102A11607 4102A11608 4102A11613 4102A11614 4102A11618 4102A11619
4102A11631 4102A11632 4126A11603 4126A11604 4126A11624 4126A11625 4126A11629
4126A11633 4126A11639 4126A11642 4126A11645 4126A11663 4126A11664 4126A11666
4126A11667 4126A11668 4126A11670 4126A11676 4126A11677 4126A11680 4126A11681
4126A11682 4126A11693 4126A11694 4126A11700 4126A11701 4126A11703 4126A11719
4126A11721 4126A11727 4126A11728 4126A11729 4126A11731 4126A11733 4126A11735
4126A11740 4126A11741 4126A11742 4126A11757 4126A11758 4126A11759 4126A11760
4126A11765 4126A11771 4126A11772 4126A11773 4126A11774 4126A11775 4126A11776
4126A11777 4126A11778 4126A11784 4126A11798 4126A11800 4136A11801 4136A11802
4136A11804 4136A11805 4136A11806 4136A11807 4136A11808 4136A11809 4136A11810
4136A11814 4136A11815 4136A11817 4136A11818 4136A11821 4136A11823 4136A11824
4136A11825 4136A11826 4136A11827 4136A11851 4136A11854 4136A11855 4136A11856
4136A11863 4136A11864 4136A11865 4136A11866 4136A11867 4136A11868 4136A11869
4136A11871 4136A11872 4136A11873 4136A11874 4136A11879 4136A11880 4136A11881
4136A11885 4136A11886 4136A11887 4136A11888 4136A11889 4136A11890 4136A11891
4136A11893 4136A11896 4136A11902 4136A11908 4136A11909 4136A11919 4136A11932
4136A11938 4136A11939 4136A11941 4136A11942 4136A11943 4136A11944 4136A11947 4136A11948
4136A11949 4136A11958 4136A11959 4136A11960 4136A11962 4136A11963 4136A11964
4136A11970 4136A11971 4136A11973 4136A11974 4136A11975 4136A11980 4136A11989
4136A11990 4136A11991 4136A11992 4136A11993 4136A11996 4136A12005 4136A12007
4136A12008 4136A12009 4136A12015 4136A12016 4136A12017 4136A12023 4136A12024
4136A12026 4136A12027 4136A12028 4136A12029 4136A12034 4136A12035 4136A12036
4136A12037 4136A12044 4136A12057 4136A12065 4136A12066 4136A12067 4136A12068
4136A12069 4136A12070 4136A12090 4136A12096 4136A12097 4136A12099 4136A12100
4136A12101 4136A12103 4136A12104 4136A12105 4136A12107 4136A12109 4136A12110
4136A12111 4136A12112 4136A12124 4136A12125 4136A12126 4136A12128 4136A12133
4136A12134 4136A12139 4136A12140 4136A12143 4143A03606 4143A11919 4143A12153
4143A12158 4143A12159 4143A12164 4143A12167 4143A12169 4143A12190 4143A12191
4143A12192 4143A12193 4143A12195 4143A12197 4143A12200 4143A12204 4143A12218
4143A12219 4143A12223 4143A12230 4143A12231 4143A12246 4143A12249 4143A12254
4143A12258 4143A12266 4143A12267 4143A12271 4143A12272 4143A12274 4143A12275
4143A12276 4143A12277 4143A12281 4143A12287 4143A12288 4143A12293 4143A12294
4143A12316 4143A12323 4143A12332 4143A12342 4143A12351 4143A12355 4143A12356
4143A12366 4143A12377 4143A12379 4143A12383 4143A12384 4143A12385 4143A12388
4143A12389 4143A12390 4143A12392 4143A12393 4143A12394 4143A12395 4143A12396
4143A12397 4143A12398 4143A12402 4143A12421 4143A12422 4143A12427 4143A12428
4143A12436 4143A12439 4143A12446 4143A12447 4143A12460 4143A12469 4143A12472
4143A12473 4143A12491 4143A12492 4143A12493 4143A12498 4143A12501 4143A12504
4143A12515 4143A12516 4143A12532 4143A12537 4143A12549 4143A12550 4143A12554
4143A12555 4143A12559 4143A12560 4143A12561 4143A12562 4143A12563 4143A12571
4143A12573 4143A12576 4143A12579 4143A12580 4143A12583 4143A12584 4143A12593
4143A12594 4143A12595 4143A12596 4143A12597 4143A12598 4143A12602 4143A12609
4143A12614 4143A12626 4143A12628 4143A12633 4143A12637 4143A12638 4143A12646
4143A12647 4143A12648 4143A12649 4143A12650 4143A12651 4143A12652 4143A12653
4143A12654 4143A12657 4143A12658 4143A12659 4143A12660 4143A12676 4143A12682
4143A12683 4143A12684 4143A12685 4143A12694 4143A12695 4143A12699 4143A12701
4143A12702 4143A12704 4143A12708 4143A12709 4143A12711 4143A12712 4143A12722
4143A12728 4143A12730 4143A12731 4143A12732 4143A12735 4143A12736 4143A12738
4143A12739 4143A12740 4143A12743 4143A12746 4143A12749 4143A12750 4143A12751
4143A12752 4143A12756 4143A12758 4143A12761 4143A12762 4143A12763 4143A12764
4143A12765 4143A12768 4143A12769 4143A12772 4143A12773 4143A12774 4143A12775
4143A12776 4143A12778 4143A12780 4143A12781 4143A12783 4143A12784 4143A12792
4143A12802 4143A12803 4143A12806 4143A12807 4143A12808 4143A12813 4143A12815
4143A12816 4143A12824 4143A12827 4143A12839 4143A12843 4143A12857 4143A12858
4143A12864 4143A12867 4143A12868 4143A12871 4143A12876 4143A12877 4143A12885
4143A12886 4143A12896 4143A12897 4143A12898 4143A12907 4143A12913 4143A12922
4143A12930 4212A12809 4212A12810 4212A12811 4212A12916 4212A12917 4212A12918
4212A12919 4212A12923 4212A12931 4212A12935 4212A12937 4213A01418 4213A12875
4213A12948 4213A12949 4213A12951 4213A12954 4213A12956 4213A12961 4213A12964
4213A12965 4213A12968 4213A12973 4213A12975 4213A12976 4213A12991 4213A12992
4213A12994 4213A12995 4213A12996 4213A12997 4213A13000 4213A13001 4213A13002
4213A13008 4213A13015 4213A13023 4213A13046 4213A13048 4213A13049 4213A13051
4213A13052 4213A13057 4213A13063 4213A13064 4213A13065 4213A13080 4213A13081
4213A13086 4213A13087 4213A13097 4213A13098 4213A13099 4213A13104 4213A13105
4213A13106 4213A13107 4213A13109 4213A13110 4213A13111 4213A13112 4213A13123
4213A13130 4213A13131 4213A13132 4213A13136 4213A13137 4213A13145 4213A13148
4213A13153 4213A13154 4213A13155 4213A13168 4213A13169 4213A13171 4213A13180
4213A13182 4213A13185 4213A13186 4213A13187 4213A13188 4213A13191 4213A13198
4213A13215 4213A13216 4213A13222 4213A13225 4213A13226 4213A13229 4213A13230
4213A13231 4213A13232 4213A13233 4213A13234 4213A13235 4213A13236 4213A13237
4213A13238 4213A13242 4213A13243 4213A13250 4213A13252 4213A13267 4213A13268
4213A13269 4213A13273 4213A13274 4213A13276 4213A13278 4213A13282 4213A13284
4213A13295 4213A13296 4213A13297 4213A13298 4213A13306 4213A13313 4213A13317
4213A13318 4213A13320 4213A13325 4213A13326 4213A13327 4213A13328 4213A13340
4213A13343 4213A13345 4213A13350 4213A13351 4213A13352 4213A13353 4213A13354
4213A13358 4213A13359 4213A13360 4213A13361 4213A13364 4213A13379 4213A13382
4213A13386 4213A13387 4213A13389 4213A13390 4213A13397 4213A13402 4213A13407
4213A13409 4213A13410 4213A13416 4213A13422 4213A13423 4213A13431 4213A13434
4213A13436 4213A13437 4213A13439 4213A13446 4213A13449 4213A13450 4213A13451
4213A13453 4213A13455 4213A13456 4213A13457 4213A13460 4213A13461 4213A13463
4213A13474 4213A13479 4213A13480 4213A13481 4213A13482 4213A13483 4213A13489
4213A13490 4213A13493 4213A13494 4213A13500 4213A13501 4213A13502 4213A13504
4213A13505 4213A13507 4213A13521 4213A13522 4213A13523 4213A13528 4213A13529
4213A13530 4213A13531 4213A13532 4213A13534 4213A13535 4213A13536 4213A13544
4213A13567 4213A13568 4213A13570 4213A13583 4213A13586 4213A13587 4213A13588
4213A13589 4213A13590 4213A13591 4213A13592 4213A13593 4213A13594 4213A13595
4213A13597 4213A13598 4213A13599 4213A13601 4213A13602 4213A13604 4213A13608
4213A13609 4213A13610 4213A13611 4213A13617 4213A13619 4213A13621 4213A13622
4213A13626 4213A13628 4213A13629 4213A13630 4213A13639 4213A13640 4213A13641
4213A13648 4213A13648 4213A13660 4213A13661 4213A13662 4213A13663 4213A13670 4213A13673
4213A13676 4213A13678 4213A13679 4213A13685 4213A13686 4213A13687 4213A13697
4213A13698 4213A13699 4213A13700 4213A13711 4213A13712 4213A13714;
b) Serial# 4102A00874
4102A00889 4102A00890 4102A00894 4102A00905 4102A00920 4102A00945 4102A00957
4102A00971 4102A00972 4102A00984 4102A01011 4102A01029 4102A01030 4102A01031
4102A01032 4136A01100 4136A01101 4136A01102 4136A01103 4136A01104 4136A01105
4136A01106 4136A01116 4136A01117 4143A01190 4143A01191 4143A01192 4143A01213
4143A01227 4143A01244 4143A01289 4143A01290 4143A01291 4143A01292 4143A01311
4143A01312 4213A01338 4213A01345 4213A01346 4213A01383 4213A01386 4213A01398
4213A01399 4213A01400 4213A01401 4213A01402 4213A01403 4213A01404 4213A01405
4213A01406 4213A01407 4213A01412 4213A01416 4213A01420 4213A01421 4213A01425
4213A01442 4213A01452 4213A01468 4213A01472 4213A01479 4213A01481 4213A01482
4213A01483 4213A01484 4213A01485 4213A01486 4213A01490 4213A01491 4143A01175
4143A01249 4143A01250 4143A01251 4143A01252 4143A01253 4213A01347 4143A01178 4213A01469
4102A00903 4102A00904 4102A00892;
c) Serial 4225A00106
4225A00107 4225A00108 4225A00111 4225A00112 4225A00113 4225A00114 4225A00116
4225A00117 4225A00118 4225A00119 4225A00127 4225A00145 4225A00149 4225A00150
4225A00151 4225A00152 4225A00164 4225A00165 4225A00166 4225A00167 4225A00169
4225A00188 4225A00189 4213A13698 4213A13700 4225A00122 4225A00136 4225A00138
4225A00137 4225A00139 4225A00140 4225A00142 4225A00143 4225A00144 4225A00101
4225A00102 4225A00103 4225A00104;
d) 4225A00169 4225A00171
4225A00181 4225A00182 4213A13456 4131A13457.
RECALLING FIRM/MANUFACTURER Philips Medical Systems, Inc. Cardiac &
Monitoring Systems, Andover, MA,
REASON Central Station Monitoring
Information Centers may experience screen freeze, i.e. loss of display,
keyboard or mouse lock up.
VOLUME OF PRODUCT IN COMMERCE
a) 972 units;
b) 82 units;
c) 39 units;
d) 6 units.
DISTRIBUTION Nationwide and Canada, Europe, Asia, Latin America.
PRODUCT
a) Bard Stinger Ablation Catheter: Item Numbers 210001 218501 L210001 210002
218502 L210002 210003 218503 L210003 210004 218504 L210004 210005 218505
L210005 210006 218506 L210006 210007 218507 L210007 210015 218516 L210008
210016 218518 L210009 210017 218519
L210010 210018 218520 L210011 210019 218521 L210012 210020 218523 L210013
210023 218524 L210014 210024 218525 L210015 210025 218526 L210016 210026 218527
L210017 210027 218528 L210018 210028 218529 210029 210030.
b) Bard Stinger M Ablation Catheters: Item Numbers 210001M 218501M 210002M
218502M 210003M 218503M 210004M 218504M
210005M 218505M 210006M 218506M 210007M 218507M 210015M 218516M 210016M 218518M
210017M 218519M 210018M 218520M
210019M 218521M 210020M 218523M 210023M 218524M 210024M 218525M 210025M 218526M
210026M 218527M 210027M 218528M
210028M 218529M 210029M 210030M.
c) Bard Stinger S Ablation
Catheter: Item Numbers 210001S 218501S 6210015S 210002S 218502S 6210016S
210003S 218503S 6210017S 210004S 218504S 6210018S 210005S 218505S 6210019S
210006S 218506S 6210020S 210007S 218507S 210015S 218516S 210016S 218518S
210017S 18519S
210018S 218520S 210019S 218521S 210020S 218523S 210023S 218524S 210024S 218525S
210025S 218526S 210026S 218527S 210027S 218528S 210028S 218529S 210029S
210030S.
d) Bard Stinger SM Ablation Catheter Item Numbers 210001SM 218501SM 210002SM
218502SM 210003SM 218503SM 210004SM
218504SM 210005SM 218505SM 210006SM 218506SM 210007SM 218507SM 210015SM
218516SM 210016SM 218518SM 210017SM
218519SM 210018SM 218520SM 210019SM 218521SM 210020SM 218523SM 210023SM
218524SM 210024SM 218525SM 210025SM
218526SM 210026SM 218527SM 210027SM 218528SM 210028SM 218529SM 210029SM
210030SM.
e) Bard Orbiter ST Artrial
Mapping Steerable Catheter Item
Numbers 320001 7FAM00001 7FAM00010 7FAM00019 7FAM0008B 7FAM0017B 320002
7FAM00002 7FAM00011 7FAM00020 7FAM0009B
7FAM0018B 320003 7FAM00003 7FAM00012 7FAM0001B 7FAM0010B 7FAM0019B 320004
7FAM00004 7FAM00013 7FAM0002B 7FAM0011B 7FAM0020B 320005 7FAM00005 7FAM00014
7FAM0003B 7FAM0012B 320006 7FAM00006 7FAM00015 7FAM0004B 7FAM0013B 320007
7FAM00007 7FAM00016 7FAM0005B 7FAM0014B 320008 7FAM00008 7FAM00017 7FAM0006B
7FAM0015B 320009 7FAM00009 7FAM00018 7FAM0007B 7FAM0016B.
f) Bard Orbiter PV Steerable
Catheter Item #: 320100 (Marketed in Europe Only).
CODE All lot numbers where the 4th digit contains the letter A through
M. Ex. 07A"A"1234.
RECALLING FIRM/MANUFACTURERC. R. Bard, Inc. Electrophysiology Division,
Billerica, MA,
REASON Sterile barrier maybe perforated compromising
sterility of the device.
VOLUME OF PRODUCT IN COMMERCE 25,502 units;
DISTRIBUTION Nationwide.
PRODUCT External Drainage Sets labeled NMT NeuroSciences Implants or
Integra NeuroSciences Implants, S.A.
CODE The External Drainage Sets lot numbers: 60300370, 60306608,
60306609, 60321387, 60300373, and 60376191. These lots of EDS may be included
in the following set configurations: Catalog #: 910-109 Lot#:
0118080; 0118369; 0118572; 0118981; 0119845; 0117525.
Catalog# 910-110A Lot #: 60300370; 60300373; 60306608;
60300370-1; 60321387; 60306608-1; 60306609; 60321387-1; 60306609-1; 60321387-2;
60376191; 60321387-3.
Catalog # 910-112A Lot #: 0117569; 0117609; 0117834; 0117478;
0117833; 0117957; 0118062; 0118175; 0118420; 0118421; 0118573; 0118693.
Catalog #910-118A Lot #: 0117570; 0117479; 0118023; 0118489;
0119150; 0119505. Catalog #910-120A Lot #: 0117536; 0117371;
0117453;
0117537; 0117902; 0117948; 0118047; 0118161; 0118422; 0118543; 0118616;
0118744; 0118808; 0118912; 0118955; 0119020; 0119081; 0119348;
0119500; 0119582.
Catalog #910-123A Lot #: 0117805; 0118576; 0118845; 0118957.
RECALLING FIRM/MANUFACTURER Integra LifeSciences Corp., Plainsboro, NJ,
REASON The check valve in the
patient line may stick in the closed position.
VOLUME OF PRODUCT IN COMMERCE 8952.
DISTRIBUTION Nationwide and Internationally.
PRODUCT
a) Compat Enteral Delivery Pump Set with Piercing Spike.
b) Compat 500 mL Baggle Top Fill Enteral Feeding Container with Preattached
Enteral Delivery Pump Set.
c) Compat DualFlo Enteral Delivery Pump Set with 1000-mL Twist Cap Formula
Vinyl Bag and 1000-mL Twist Cap Water Bag.
CODE
a) Item number 19930700, lot 3092D;
b) Item number 19931200, lots 3152A and 3182A;
c) Item number 19944700, lot 3232E.
RECALLING FIRM/MANUFACTURER Novartis Nutrition Corp., St Louis Park, MN,
REASON Some of the sets may leak or
disconnect near the pump adapter during use. The patient could ingest air and
vomit causing a safety concern.
VOLUME OF PRODUCT IN COMMERCE 930 administration sets.
DISTRIBUTION Nationwide and
Canada.
PRODUCT
a) Datex-Ohmeda S/5 Oxygen Saturation Module (M-OSAT).
b) Datex-Ohmeda Cardiocap/5 Anesthesia Monitor with N-XOSAT SpO2 measurement
option.
CODE
a) All serial numbers of the units distributed prior to the recall. September
16, 2002.
b) All units with N-XOSAT SpO2 measurement option distributed prior to the
recall. September 16, 2002.
RECALLING FIRM/MANUFACTURER Datex-Ohmeda, Madison, WI,
REASON Falsely high blood oxygen saturation readings in some
circumstances, which could result in no alarm in a critical life-threatening
situation.
VOLUME OF PRODUCT IN COMMERCEa) 1019;
b) 85.
DISTRIBUTION Nationwide and
Internationally.
PRODUCT Scimed ChoICE PT "J" Tip 182 cm Guide Wire, a single
use, sterile (EO), device packaged in a carrier tube, which is held in a coil
by plastic clips. The carrier tube assembly of the guide wire is sealed into a
pouch. The pouch assemblies are packaged in a five-pack product box.
CODE Pouch codes are: Catalog No. 12160-01J, UPN H7491216001J1, batch
4536936, expiration date of 2004-01. The five pack box codes are: Catalog No.
12160-01J, UPN H7491216001J2, batch 4562262, expiration date 2004-01.
RECALLING FIRM/MANUFACTURER Symbiosis Corp., Miami, FL,
REASON Labeled as J tip but may be straight tip.
VOLUME OF PRODUCT IN COMMERCE 24 units.
DISTRIBUTION FL, KS, NH, IA, PA, MI and CA.
PRODUCT
a) Scimed ChoICE PT Extra Support PTCA Guide Wire with ICE Hydrophilic Coating,
300 cm length. Recall Z-0590-03.
b) Scimed PT Graphix Intermediate PTCA Guide Wire with ICE Hydrophilic Coating,
300 cm length. Recall Z-0591-03.
CODE
a) Individual Devices in Pouches with Catalog No. 12155-01, UPN H74912155011,
Lot Number 4773987, exp. Date 2004-5.
These pouches were packaged in two lots of five-pack boxes with the following
codes: Catalog 12155-01, lot 4791304 Catalog 12155-011, lot 4798294.
b) Individual devices in pouches as follows: Catalog No. 14914-01 UPN
H74914914011 Lot Number Expiration Date 4772607 2004-0 4772610 " 4762379
" 4762384 " 4772618 " 4766619 " 4758393 " 4758432
" 4753775 " 4758389 " 4758427 " 4756521 " 4762390
" 4756485 " 4753781 "
4758397 " 4758423 " 4753783 " 4753785 " 4740328 2004-4
4753779 2004-5 4753759 " 4753769 " 4756534 " 4756536 "
4756531 " 4724690 2004-4 4740329 " 4736552 "
The above PT Graphix(tm) guide wire pouch lots were packaged in 18 lots of
five-pack boxes with the following codes: Catalog No. 1215T UPN H74914914012
Lot Number Expiration Date 4789558 2004-05 4791600 " 4789558 "
4791126 " 4785009 " 4787988 " 4786427 "
4783177 " 4780051 " 4780186 " 4783316 " 4780190 " 4782079
" 4780939 " 4771889 " 4777410 " 4775761 " 4776272
" 4766792 "
RECALLING FIRM/MANUFACTURER Symbiosis, Corp., Miami, FL, by letters on
August 22, 2002.
REASON Product labeled as Scimed
Choice PT Extra Support PTCA Guide Wire 300 cm were actually Scimed PT Graphix
Intermediate PTCA Guide Wire 300 cm.
VOLUME OF PRODUCT IN COMMERCE a) 27/5 pack boxes. b) 228/5 pack boxes.
DISTRIBUTION Nationwide and Internationally.
PRODUCT Baxter Meridian Hemodialysis Instrument. The instrument is part of
a high permeability hemodialysis system which consists of a controlled
dialysate delivery system that incorporates an ultra filtration controller to
prevent excessive loss of water from the patient's blood, an extra corporeal
blood set, and a high permeability dialyzer. The standard features of the
Meridian instrument include a high blood flow rate capacity (for shortened
hemodialysis treatment time), automatic ultra filtration control, standard and
variable bicarbonate and sodium capabilities and automated chemical
disinfection. The Meridian will operate in either the bicarbonate or acetate
mode for concentrates.
CODE Product codes 5M5576, 5M5576R, serial numbers 200000 through
202864.
RECALLING FIRM/MANUFACTURER Baxter Healthcare, Corp., Round Lake, IL,
REASON Meridian hemodialysis
instrument may reset when in shutdown mode causing the system to reset with
potential for patient increased bioburden exposure.
VOLUME OF PRODUCT IN COMMERCE 2,864 units.
DISTRIBUTION Nationwide, Mexico, China, South Korea, Hong Kong.
PRODUCT
SKBM
Microkeratome cutting tool.
CODE All codes are affected by Safety Alert: 98-3002-005;
98-3002-01-017; 98-3002-02-001; 98-3002-02-002; 98-3002-02-007-03;
98-3002-03-001; 98-3002-03-003; 98-3002-03-004; 98-3002-03-006; 98-3002-03-007;
98-3002-03-010; 98-3002-03-013; 98-3002-03 018; 092K14; 98-3002-04-001;
98-3002-04-002; 98-3002-04-003; 98-3002-04-004; 98 3002-04-005; 98-3002-04-006;
98-3002-04-007; 98-3002-04-008; 98-3002-04-009; 98-3002-04-010; 98-3002-04-011;
98-3002-04-012; 98-3002-04-013; 98-3002-04-014; 98-3002-04-015; 98-3002-04-016;
98-3002-04-017; 98-3002-04-018; 98-3002-04-019; 98-3002-04-020; 98-3002-04-021;
98-3002-04-022; 98-3002-04-023; 98-3002-04-024; 98-3002-04-025; 98-3002-04-026;
98-3002-04-027; 98-3002-04-028; 98-3002-04-029; 98-3002-04-030; 99-0204-001-IR;
99-0204-002-IR; 99-0304-003-IR; 99-0304-004-IR; 99-0304-005-IR; 99-0304-006-IR;
99-0304-007-IR; 99-0304-008-IR; 99-0304-009-IR; 99-0304-010-IR; 99-0504-011
-IR; 99-0404-001-GE; 99-0404-002-GE; 99-0404-003-GE; 99-0404-004-GE;
99-0404-005-GE; 99-0404-006-GE; 99-0404-007-GE; 99-0404-011-IR; 99-0404-012-IR;
99-0404-013-IR; 99-0404-014-IR; 99-0404-015-IR; 99-0404-016-IR; 99-0404-017-IR;
99-0404-018-IR; 99-0504-001-GE; 99-0504-002-GE; 99-0504-003-GE; 99-0504-004-GE;
99-0504-005-GE; 99-0504-006-GE; 99-0504-007 GE; 99-0504-008-GE; 99-0504-009-
GE; 99-0504-010-GE; 99-0504-019-IR; 99-0504-020-IR; 99-3002-04-001;
99-3002-04-002; 99-3002-04-003; 99-3002-04-004; 99-3002-04-005; 99-3002-04-006;
99-3002-04-007; 99-3002-04-008; 99-3002-04-009; 99-3002-04-010; 99-3002-04-011;
99-3002-04-012; 99-3002-04-013; 99-3002-04-014; MK31001; MK31002; MK31003;
MK31004; MK31005; MK31006; MK31007; MK31008; MK31009; MK31010; MK31011;
MK31012; MK31013; MK31014; MK31015; MK31016; MK31017; MK31018; MK31019;
MK31020; MK31021; MK31022; MK31023; MK31024; MK31025; MK31026; MK31027;
MK31028; MK31029; MK31030; MK31031; MK31032; MK31033; MK31034; MK31035;
MK31036; MK31037; MK31038; MK31039; MK31040; MK31041; MK31042; MK31043;
MK31044; MK31045; MK31046; MK31047; MK31048; MK31049; MK31050; MK31051;
MK31052; MK31053; MK31054; MK31055; MK31056; MK31057; MK31058; MK31059;
MK31060; MK31061; MK31062; MK31063; MK31064; MK31065; MK31066; MK31067;
MK31068; MK31069; MK31070; MK31071; MK31072; MK31073; MK31074; MK31075;
MK31076; MK31077; MK31078; MK6021/G; MK6022/G; MK6023/G; MK6024/G; MK6025/G;
MK6026/G; MK6027/G; MK6028/G; MK6029/G; MK6030/G; MK6031/G; MK6032/G; MK6033/G;
MK6034/G; MK6035/G; MK6036/G; MK6037/G; MK6038; MK6039; MK6040; MK6041/J;
MK6042; MK6043/G; MK6044; MK6045; MK6046; MK6047; MK6048; MK6060E; MK7000;
MK7001; MK7002; MK7003; MK7004; MK7005; MK7006; MK7007; MK7008; MK7009; MK7010;
MK7011; MK7012; MK7013; MK7014; MK7015; MK7016; MK7017; MK7018; MK7019; MK7020;
MK7021; MK7022; MK7023; MK7024; MK7025; MK7026; MK7027; MK7028; MK7029; MK7030;
MK7031; MK7032; MK7033; MK7034; MK7035; MK7036; MK7037; MK7038; MK7039; MK7040;
MK7041; MK7042; MK7043; MK7044; MK7045; MK7046; MK7047; MK7048; MK7049; MK7050;
MK7051; MK7052; MK7053; MK7054; MK7055; MK7056; MK7057; MK7058; MK7059; MK7060;
MK7061; MK7062; MK7063; MK7064; MK7065; MK7066; MK7067; MK7068; MK7069; MK7070;
MK7071; MK7072; MK7073; MK7074; MK7075; MK7076; MK7077; MK7078; MK7079; MK7080;
MK7081; MK7082; MK7083; MK7084; MK7085; MK7086; MK7087; MK7088; MK7089; MK7090;
MK7091; MK7092; MK7093; MK7094; MK7095; MK7096; MK7097; MK7098; MK7099; MK7100;
MK7101; MK7102; MK7103; MK7104; MK7105; MK7106; MK7107; MK7108; MK7109; MK7110;
MK7111; MK7112; MK7113; MK7114; MK7115; MK7116; MK7117; MK7118; MK7119; MK7120;
MK7121; MK7122; MK7123; MK7124; MK7125; MK7126; MK7127; MK7128; MK7129; MK7130;
MK7131; MK7132; MK7133; MK7134; MK7135; MK7136; MK7137; MK7138; MK7139; MK7140;
MK7141; MK7142; MK7143; MK7144; MK7145; MK7146; MK7147; MK7148; MK7149; MK7150;
MK7151; MK7152; MK7153; MK7154; MK7155; MK7156; MK7157; MK7158; MK7159; MK7160;
MK7161; MK7162; MK7163; MK7164; MK7165; MK7166; MK7167; MK7168; MK7169; MK7170;
MK7171; MK7172; MK7173; MK7174; MK7175; MK7176; MK7177; MK7178; MK7179; MK7180;
MK7181; MK7182; MK7183; MK7184; MK7185; MK7186; MK7187; MK7188; MK7189; MK7190;
MK7191; MK7192; MK7193; MK7194; MK7195; MK7196; MK7197; MK7198; MK7199; MK7200;
MK7201; MK7202; MK7203; MK7204; MK7205; MK7206; MK7207; MK7208; MK7209; MK7210;
MK7211; MK7212; MK7213; MK7214; MK7215; MK7216; MK7217; MK7218; MK7219; MK7220;
MK7221; MK7222; MK7223; MK7224; MK7225; MK7226; MK7227; MK7228; MK7229; MK7230;
MK7231; MK7232; MK7233; MK7234; MK7235; MK7236; MK7237; MK7238; MK7239; MK7240;
MK7241; MK7242; MK7243; MK7244; MK7245; MK7246; MK7247; MK7248; MK7249; MK7250;
MK7251; MK7252; MK7253; MK7254; MK7255; MK7256; MK7257; MK7258; MK7259; MK7260;
MK7261; MK7262; MK7263; MK7264; MK7265; MK7266; MK7267; MK7268; MK7269; MK7270;
MK7271; MK7272; MK7273; MK7274; MK7275; MK7276; MK7277; MK7278; MK7279; MK7280;
MK7281; MK7282; MK7283; MK7284; MK7285; MK7286; MK7287; MK7288; MK7289; MK7290;
MK7291; MK7292; MK7293; MK7294; MK7295; MK7296; MK7297; MK7298; MK7299; MK7300;
MK7301; MK7302; MK7303; MK7304 MK7305; MK7306; MK7307; MK7308; MK7309; MK7310;
MK7311; MK7312; MK7313; MK7314; MK7315; MK7316; MK7317; MK7318; MK7319; MK7320;
MK7321; MK7322; MK7323; MK7324; MK7325; MK7326; MK7327; MK7328; MK7329; MK7330;
MK7331; MK7332; MK7333; MK7334; MK7335; MK7336; MK7337; MK7338; MK7339; MK7340;
MK7341; MK7342; MK7343; MK7344; MK7345; MK7346; MK7347; MK7348; MK7349; MK7350;
MK7351; MK7352; MK7353; MK7354; MK7355; MK7356; MK7357; MK7358; MK7359; MK7360;
MK7361; MK7362; MK7363; MK7364; MK7365; MK7366; MK7367; MK7368; MK8000; MK8001;
MK8002; MK8003; MK8004; MK8005; MK8006; MK8007; MK8008; MK8074; MK8075; MK8076;
MK8077; MK8078; MK8079; MK8080; MK8081; MK8082; MK8083; MK8084; MK8085; MK8086;
MK8087; MK8088; MK8089; MK8090; MK8091; MK8092; MK8093; MK8094; MK8095; MK8096;
MK8097; MK8098; MK8099; MK8100; MK8101; MK8102; MK8103; MK8104; MK8105; MK8106;
MK8107; MK8108; MK8109; MK8110; MK8111; MK8112; MK8113; MK8114; MK8115; MK8116;
MK8117; MK8118; MK8119; MK8120; MK8121; MK8122; MK8123; MK8124; MK8125; MK8127;
MK9001; MK9002; MK9003; MK9004; MK9005; MK9006; MK9007; MK9008; MK9009; MK9010;
MK9012; MK8009; MK8010; MK8011; MK8012; MK8013; MK8014; MK8015; MK8016; MK8017;
MK8018; MK8019; MK8020; MK8021; MK8022; MK8023; MK8024; MK8025; MK8026; MK8027;
MK8028; MK8029; MK8030; MK8031; MK8032; MK8033; MK8034; MK8035; MK8036; MK8037;
MK8038; MK8039; MK8040; MK8041; MK8042; MK8043; MK8044; MK8045; MK8046; MK8047;
MK8048; MK8049; MK8050; MK8051; MK8052; MK8053; MK8054; MK8055; MK8056; MK8057;
MK8058; MK8059; MK8060; MK8061; MK8062; MK8063; MK8064; MK8065; MK8066; MK8067;
MK8068; MK8069; MK8070; MK8071; MK8072; MK8073.
RECALLING FIRM/MANUFACTURER Alcon Laboratories, Inc., Fort Worth, TX
REASON Degradation of glue causes glass movement resulting in
misalignment that can cause deeper cuts than anticipated.
VOLUME OF PRODUCT IN COMMERCE 567 units.
DISTRIBUTION Nationwide and Internationally.
PRODUCT Misys Laboratory System.
CODE Versions 5.2, 5.23 and 5.3.
RECALLING FIRM/MANUFACTURER Misys Healthcare Systems, Tucson, AZ.
REASON Software anomaly. Rejecting the selected Autofiling status may
result in the change being accepted rather than rejected.
VOLUME OF PRODUCT IN COMMERCE 542.
DISTRIBUTION Nationwide and Internationally.
PRODUCT
a) Chlamydia OIA. For in vitro diagnostic use.
b) Chlamydia OIA. For in vitro diagnostic use.
CODE
a) Catalog No. CHL30, Lot No. 132117, 132117A, 133101;
b) Catalog No. CHLINT, Lot No. 133354, 133174. Foreign distribution.
RECALLING FIRM/MANUFACTURER Thermo Biostar, Inc., Louisville, CO.
REASON Chlamydia IVD test kits may contain diluted conjugate reagent.
Use may produce a false negative result with a low positive patient sample.
VOLUME OF PRODUCT IN COMMERCE 342 units.
DISTRIBUTION Nationwide, France, New Zealand and Spain.
PRODUCT Integrated Orbiter Single Detector General Purpose Gamma Camera
System.
CODE
Model
35-26-932, serial numbers 102, 1001/1217, 1169A; Model 35-44-637, serial
numbers 0101/0142, 120A.
RECALLING FIRM/MANUFACTURER Siemens Medical Solutions USA, Inc., Hoffman
Estates, IL.
REASON Unintended yoke rotational motion.
VOLUME OF PRODUCT IN COMMERCE 256 units.
DISTRIBUTION Nationwide and Internationally.
PRODUCT Atakr II RF Power Generator, Models 4802 and 4803.
CODE All units.
RECALLING FIRM/MANUFACTURER Medtronic, Inc., Rice Creek Facility,
Fridley, MN.
REASON
An RF pulse
occurs each time the temperature set point increase button is activated while
ablating and this can result in heart rhythm disturbances.
VOLUME OF PRODUCT IN COMMERCE 775 units.
DISTRIBUTION Nationwide.
PRODUCT Access and Access 2 Immunoassay Systems Discrete photometric
chemistry analyzer for clinical use.
CODE Not applicable.
RECALLING FIRM/MANUFACTURER Beckman Coulter, Inc., Brea, CA.
REASON Spring failure, design defect in component.
VOLUME OF PRODUCT IN COMMERCE 2418.
DISTRIBUTION Nationwide.
PRODUCT
a) Custom Angiographic Kit.
b) Custom Angiographic Kit.
c) Custom Angiographic Kit. .
CODE
a) Model Number: L04-00278/A. Lot Numbers: A2253131;
b) Model Number: K08-MP5130. Lot Numbers: A146066. A188604;
c) Mo