JUNE 2002
WEEK ENDING JUNE 1
PRODUCT EBI XFIX DFS OptiROM Elbow Fixator. The product is
sold non-sterile and is packaged either in an inner foam liner aluminum metal
casing or in plastic bag.
CODE DC09575 or 09575 will be etched on the unit and the following
lot numbers: 40994, 44782, 206883, 204881, 204882, 271159, 288313,
316281,
355170, 393638, 422625, 439685, 456129, 493368, 500176, 504015, 522720, 516609.
RECALLING FIRM/MANUFACTURER EBI, l.p., Parsippany, NJ
REASON Pin migration at the connection of OptiROM central
body and fixator.
VOLUME OF PRODUCT IN COMMERCE 339.
DISTRIBUTION Nationwide and Internationally.
PRODUCT Seprapack-Unilateral Application, sterile.
(Bioresorbable nasal packing and sinus stent).
CODE Product Code: 7089-0807 Lot Number: 51579V.
RECALLING FIRM/MANUFACTURER Genzyme Biosurgery, a
division of Genzyme Corp., Fall River, MA
REASON Product packaging may have voids in the seal
compromising sterility of device.
VOLUME OF PRODUCT IN COMMERCE 348 units.
DISTRIBUTION TN.
PRODUCT Bayer Immuno 1(R) Troponin I Method. Common Name: Immuno 1(R) Troponin I Method.
CODE Product No. T01-3887-51, Lot V60275.
RECALLING FIRM/MANUFACTURER Recalling Firm: Bayer Corp., Business Group Diagnostics,
Tarrytown, NY
Manufacturer: Fisher Diagnostics Middletown, VA.
REASON Complaints received of low recoveries on Immuno 1
Troponin I Reagent Lot V60275.
VOLUME OF PRODUCT IN COMMERCE 6488 units.
DISTRIBUTION Nationwide and Puerto Rico.
PRODUCT Puritan Bennett-840 Ventilator System with optional
compressor.
CODE Ventilators with Rev. E BD CPU boards with Serial
numbers 3510000007 through 3510012304.
RECALLING FIRM/MANUFACTURER Recalling Firm: TYCO Healthcare, Nellcor Puritan Pennett,
Pelasanton, CA
Manufacturer: Nellcor Puritan Bennett Ireland, ltd.
Mervue, Galway, Ireland
REASON Failure of optional compressor on ventilator.
VOLUME OF PRODUCT IN COMMERCE 2,599 ventilators systems
and 299 field replacement units with Rev. E BD CPU boards
DISTRIBUTION Nationwide and Internationally.
PRODUCT a) Lightcycler microchemistry analyzer, catalog
number 2011468, printers; b) Lightcycler microchemistry analyzer, catalog
number 2043912, printers; c) Lumi Imager microchemistry analyzer, catalog
number 2012847, printers.
CODE All Hewlett Packard Deskjet 800 and 900 series and
PhotoSmart 1000, 1100, 1200 and 1300 series printers shipped between April 2001
and January 2002 with a Longwell "LS-7C" gray two-prong power cord.
RECALLING FIRM/MANUFACTURER Recalling Firm: Roche Diagnostics Corp., Indianapolis, IN
Manufacturer: Hewlett Packard Company, Loveland, CO.
REASON Printer power cord poses a shock hazard.
VOLUME OF PRODUCT IN COMMERCE 220.
DISTRIBUTION Nationwide.
PRODUCT Linvatec brand Double-Armed, Straight, Trocar Point
13” (33cm) Suture Needle Polyester, Green Braided, 36” (91.4cm) 2-0 U.S.P 3.0.
metric.
CODE , 32188, 430382, 432189, 432190, 434760, 456391,
465144, 469384, 466961, 473349, 80470, 486004.
RECALLING FIRM/MANUFACTURER ReGen Biologics, Inc.,
Redwood City, CA
REASON Sterility may be compromised.
VOLUME OF PRODUCT IN COMMERCE 10,760 units.
DISTRIBUTION Nationwide.
PRODUCT Synapse Image and Information Management System.
CODE Synapse Image and Information Management System with
software versions 2.0.2 and 2.1.1.
RECALLING FIRM/MANUFACTURER Fujifilm Medical Systems U.S.A., Inc., Stamford, CT
REASON Rotated/flipped Images not viewed on the screen may
not rotate but left / right markers will be flipped.
VOLUME OF PRODUCT IN COMMERCE 70.
DISTRIBUTION Nationwide.
PRODUCT Liebel-Flarsheim Angiomat Illumena Media Power
Injectors, P/N 900001, using Version 8.0 software with new console.
CODE Serial Numbers: 0202-5000 to 0402-5126.
RECALLING FIRM/MANUFACTURER Recalling Firm: Mallinckrodt, Inc. / Liebel-Flarsheim
Business, Cincinnati, OH
Manufacturer: Liebel-Flarsheim Co., Cincinnati, OH
REASON Pushing the scoll button repeatedly / rapidly
results in erroneous protocol ID.
VOLUME OF PRODUCT IN COMMERCE 62.
DISTRIBUTION Nationwide and Canada, Taiwan, China, So.
Africa and Australia.
PRODUCT Stryker Secure II hospital beds with an optional
110V outlet box assembly.
CODE All beds manufactured between June 1, 2001 and
February 19, 2002 that has the optional 110V outlet assembly box.
RECALLING FIRM/MANUFACTURER Stryker Corp., Kalamazoo, MI
REASON Shock hazard.
VOLUME OF PRODUCT IN COMMERCE 1,4460 beds.
DISTRIBUTION United States and Canada.
PRODUCT Duoflo, Applicator Dispenser Kit, with Tisseal VH,
Fibrin Sealant.
CODE Product Code 921023; lot number 03130001 in USA.
Lot
05290106 in foreign distribution.
RECALLING FIRM/MANUFACTURER
Recalling
Firm: Baxter Healthcare Corporation Glendale, CA
Manufacturer:
Hemaedics, Inc., Brentwood, CA
REASON Tip breakage into surgical site.
VOLUME OF PRODUCT IN COMMERCE 11,624.
DISTRIBUTION Nationwide and Internationally.
PRODUCT a) 3T vent valves for perfusion circuits sterile; b)
3T vent valves for perfusion circuits; non-sterile.
CODE
a) CATALOG # 030 (STERILE)- all lots bearing a sterilization
date prior to April 5, 2002.
b)
CATALOG #030NS (NON-STERILE)- all lots sold to manufacturers prior to
initiation of the recall.
RECALLING FIRM/MANUFACTURER Recalling Firm: 3T Medical Systems, LLC, Dearborn, MI
Manufacturer: MPC/Geneva Medical Products, Prairie Du
Chien, WI
REASON Inadequate bond allows blood leakage or separation
during use.
VOLUME OF PRODUCT IN COMMERCE 7,200.
DISTRIBUTION Nationwide.
PRODUCT EB-250S and EB-450S bronchoscope.
CODE Model/Catalog No. EB-250S: Serial Numbers: 2B027C002; 2B027C051, 2B027C061, 2B027C003, 2B027C067,
2B027C005, 2B027C001, 2B027C004, 2B027C017, 1B027C016, and 1B027C017
Model/Catalog No. EB-450S: Serial
Numbers: 1B028C048, 1B028C011, 1B028B040, 1B028C004, 1B028C013, 2B028C007,
2B028C008, 2B028C070, 2B028C076, 2B028C023, 2B028C001, 2B028C013, 2B028C015,
2B028C048,
2B028C002,
2B028C009, 2B028C010, 2B028C003, 2B028C033, 2B028C017, and HB028A038.
RECALLING FIRM/MANUFACTURER Recalling Firm: Fujinon, Inc., Wayne, NJ
Manufacturer: Fuji Photo Optical Co. Ltd., Omiya, Saitama,
300-0036, Japan
REASON During brochoscopy, suction valve stuck in active
position.
VOLUME OF PRODUCT IN COMMERCE 32.
DISTRIBUTION Nationwide and Canada.
PRODUCT Padgett Instruments, Inc. roller cylinder for the
Mesh Skin Graft Expander, Model P-170, for use in producing a honeycomb pattern
on the graft so it will expand.
CODE Not coded.
RECALLING FIRM/MANUFACTURER Recalling Firm: Padgett
Instruments, Inc., Kansas City, MO
Manufacturer:
Dadson Mfg. Corp., Grain Valley, MO.
REASON The roller assembly might be defective.
VOLUME OF PRODUCT IN COMMERCE 39 P-170 units as of
4/24/02.
DISTRIBUTION FL, WV, NY and TX and Brazil, Portugal,
Chile, Finland, England, Spain and Ireland.
WEEK
ENDING JUNE 8
PRODUCT LTV Series Ventilators. Models: a) LTV 1000 (cat.
no. 10130); b) Model LTV 950 (cat. no. 10950); c) Model LTV 900 (cat. no.
10638); d) Model LTV 800 (cat. no. 11800)
CODE Forthcoming.
RECALLING FIRM/MANUFACTURER Pulmonetic Systems, Inc.,
Colton, CA
REASON Shorts to inoperative condition.
VOLUME OF PRODUCT IN COMMERCE 1,890 finished devices and
82 turbine assemblies.
DISTRIBUTION Nationwide and Internationally.
PRODUCT Ti Cann Troch Fixation Nail. ; Trochanteric Fixation
Nail.
CODE Part number 456.318S, lot numbers 4333900 exp.
11/10, 4333901 exp. 12/10, and 4340289 exp. 12/10.
Part
number 456.322S, lot numbers 4333904 exp. 11/10, 4340293 exp. 12/10, 4340295
exp. 12/10.
Part
number 456.325S, lot number 4320398 exp. 12/10.
Part
number 456.328S, lot number 4320395 exp. 12/10.
Part
number 456.414S, lot number 4349527 exp. 12/10.
Part
number 456.416S, lot numbers 4333921 exp. 11/10, 4333922 exp. 11/10, 4340324
exp. 12/10, and 4348124 exp. 12/10.
Part
number 456.418S, lot number 4340328 exp. 12/10.
Part
number 456.421S, lot number 4340334 exp. 12/10.
Part
number 456.422S, lot number 4333929 exp. 11/10.
Part
number 456.425S, lot number 4346091 exp. 12/10.
Part
number 456.511S, lot number 4348136 exp. 12/10.
Part
number 456.322S, lot number 4333903 exp. 12/10.
RECALLING FIRM/MANUFACTURER Recalling Firm: Synthes
(USA), Paoli, PA
Manufacturer: Synthes (USA), Monument, CO.
REASON Hole not properly centered.
VOLUME OF PRODUCT IN COMMERCE 71 units.
DISTRIBUTION Nationwide and Switzerland.
PRODUCT Hill-Rom brand Horizon P100705 monitor slide
CODE All units built prior to May 2001.
RECALLING FIRM/MANUFACTURER Hill-Rom Co. Inc.,
Batesville, IN
REASON Monitor may fall.
VOLUME OF PRODUCT IN COMMERCE 381 units.
DISTRIBUTION Nationwide and Egypt.
PRODUCT a) Ballard Express Pre-Assembled Biliary Stent
Systems.
b) Ballard
Standard Biliary Stent Sets with and without guide wire.
c)
Ballard Holland Biliary Stent Sets with and without guide wire.
d)
Ballard Pancreatic Stent Sets with and without guide wire.
All
sets are packaged in sterile 1-each package.
CODE a) Catalog Number: 63000 Lot numbers: 101296,
120483, 122702, 125804, 126787, 128723, 129804, 132833, 142812, 151458, 156020,
91425, 91426, 91427
Catalog
Number: 63001 Lot Numbers: 101297, 102446, 103780, 107268, 110876,117880,
1125805, 128187, 128724, 130780, 133712, 137831, 139088,
156997,
160503, 165093.
Catalog
Number: 63002 Lot Number: 101298,
102454, 103781, 109316, 116585, 120484, 121332, 122703, 127468, 128725, 128993,
137023, 139993,
144126,
146988, 148556, 150150, 152326, 157776, 165094, 99726.
Catalog
Number: 63003 Lot Number: 101299, 102455, 107269, 119047, 120485, 121333,
122704, 126788, 128726, 129805, 131552, 135976, 144190, 158318.
Catalog
Number: 63004 Lot Number: 101300, 120486, 121334, 125554, 125806, 128727,
128994, 132045, 135977.
Catalog
Number: 63005 Lot Number: 101301, 1119749, 120487, 121335, 125555,
125807,128188, 128728, 130781, 133713, 134524, 135394, 140688,
142813,
144128, 144191, 165095.
Catalog
Number: 63006 Lot Number: 101302, 125556, 128729, 128995, 129806, 132834,
135395, 135979, 139089, 142814, 143480, 148557.
Catalog
Number: 63007 Lot Number: 101303, 125557, 139995, 142107, 143481,
144129,146989, 149414, 150151, 151459, 152327, 156022.
Catalog
Number: 63008 Lot Number: 101304, 125559, 132504, 135396, 135980, 138023,140689,
142815, 144192, 153100, 156999, 158320, 160505, 165096.
Catalog
Number: 63009 Lot Number: 121336, 125560, 128189, 132505, 135981,
139090,141320.
Catalog
Number: 63130 Lot Number: 137415, 139095, 141325, 145264.
Catalog
Number: 63131 Lot Number: 137416, 140000, 144743, 145265, 146110.
Catalog
Number: 63132 Lot Number: 137417, 144744, 144972.
Catalog
Number: 63133 Lot Number: 137418, 145267, 147015.
Catalog
Number: 63134 Lot Number: 137419
b)
Catalog Number: 63040 Lot Number: 101161, 125561, 138028.
Catalog
Number: 63041 Lot Number: 101162, 125562, 146095.
Catalog
Number: 63042 Lot Number: 101163, 125563.
Catalog
Number: 63043 Lot Number: 101164, 125564.
Catalog
Number: 63044 Lot Number: 101165, 125565, 150153.
Catalog
Number: 63045 Lot Number: 101166
Catalog
Number: 63046 Lot Number: 101167, 125567.
Catalog
Number: 63047 Lot Number: 101168, 125568, 140691, 143483, 154735,
157309,160462, 160506.
Catalog
Number: 63048 Lot Number: 101169, 119753, 125569, 128720, 129813,
132837,135397, 135982, 138029, 139091.
Catalog
Number: 63049 Lot Number: 101170, 125570, 130782, 133714, 135398, 135983.
Catalog
Number: 63050 Lot Number: 102456, 125571.
Catalog
Number: 63051 Lot Number: 125572.
Catalog
Number: 63052 Lot Number: 125573, 130783, 133715.
Catalog
Number: 63053 Lot Number: 125574, 128722.
Catalog
Number: 63054 Lot Number: 125575.
Catalog
Number: 63055 Lot Number: 125576.
Catalog
Number: 63056 Lot Number: 101183, 109317, 127855, 128767,129814, 138030.
Catalog
Number: 63057 Lot Number: 125577.
Catalog
Number: 63058 Lot Number: 125578.
Catalog
Number: 63061 Lot Number: 112612, 125581.
Catalog
Number: 63062 Lot Number: 101187, 125582, 142426.
Catalog
Number: 63063 Lot Number: 101188, 125583, 141321.
Catalog
Number: 63084 Lot Number: 101206, 125590.
Catalog
Number: 63085 Lot Number: 102495, 125591.
Catalog
Number: 63086 Lot Number: 101207, 125592.
Catalog
Number: 63087 Lot Number: 101208, 125593.
Catalog
Number: 63088 Lot Number: 125594.
Catalog
Number: 63089 Lot Number: 101210, 125595, 128762, 130784, 133717.
Catalog
Number: 63090 Lot Number: 103784, 125596, 131553, 134526, 135474, 137832,
138025, 144130, 145262, 149422.
Catalog
Number: 63091 Lot Number: 101211, 125597, 131554, 131690, 134527,
137833,138026, 139996, 143489, 146102, 147000, 150157.
Catalog
Number: 63092 Lot Number: 101212, 125598, 128763, 132838, 135984, 138027,
142110.
Catalog 63093 Lot Number: 101213, 125599.
Catalog Number: 63094 Lot Number: 101214,
125600.
Catalog Number: 63095 Lot Number: 102498,
125601.
Catalog Number: 63096 Lot Number: 125602
Catalog Number: 63097 Lot Number: 125603
Catalog Number: 63150 Lot Number: 103791
Catalog Number: 63165 Lot Number: 103792,
105512.
Catalog Number: 63166 Lot Number: 101049,
103893, 108831, 144139.
Catalog Number: 63167 Lot Number: 103794,
112611.
Catalog Number: 63168 Lot Number: 101050.
Catalog Number: 63169 Lot Number: 122712,
130786.
Catalog Number: 63171 Lot Number: 101053.
Catalog Number: 63172 Lot Number: 101054,
128766.
Catalog Number: 63173 Lot Number: 101055
Catalog Number: 63174 Lot Number: 101056.
c)
Catalog Number: 63010 Lot Number: 91422, 91424.
Catalog Number: 63012 Lot Number: 101147,
119750.
Catalog Number: 63013 Lot Number: 105952,
120489.
Catalog Number: 63015 Lot Number: 119751.
Catalog Number: 63016 Lot Number: 101151,
128773, 132835.
Catalog Number: 63017 Lot Number: 101152,
128774, 132836, 155223, 156555.
Catalog Number: 63018 Lot Number: 101153,
129811, 146991, 157308.
Catalog Number: 63020 Lot Number: 105953,
120492.
Catalog Number: 63023 Lot Number: 101159
Catalog Number: 63070 Lot Number: 101193,
118685, 137027.
Catalog Number: 63071 Lot Number: 101194,
128770.
Catalog Number: 63072 Lot Number: 101195,
127469.
Catalog Number: 63073 Lot Number: 101196,
118686, 148560.
Catralog Number: 63074 Lot Number: 101197,
121305, 127470, 143485.
Catalog Number: 63075 Lot Number: 101198,
128772, 161901.
Catalog Number: 63076 Lot Number: 101199,
118687, 138024, 143486, 148957.
Catalog Number: 63077 Lot Number: 101200,
122706, 143487, 144871.
Catalog Number: 63078 Lot Number: 101201,
119754.
Catalog Number: 63079 Lot Number: 101202,
120496, 137028.
Catalog Number: 63080 Lot Number: 101203,
118688, 146101.
Catalog Number: 63081 Lot Number: 101204,
103782, 107270, 119755, 120497, 121306, 137029, 141322, 143488, 145261, 148561.
Catalog Number: 63082 Lot Number: 103783,
112062, 127856, 137030, 150939.
Catalog Number: 63083 Lot Number: 101205,
119756.
d) Catalog
Number: 63175 Lot Number: 146116, 155441, 91410.
Catalog Number: 63176 Lot Number: 101057,
119763, 125618, 128918, 134531, 158321.
Catalog: 63177 Lot Number: 101058, 119764,
125619, 128194.
Catalog Number: 63178 Lot Number: 101059,
119765, 125620.
Catalog Number: 63179 Lot Number: 101060.
Catalog Number: 63181 Lot Number: 125623,
126790, 128783, 129821, 132056.
Catalog Number: 63182 Lot Number: 101063,
125624, 128195.
Catalog Number: 63183 Lot Number: 125625
Catalog Number: 63184 Lot Number: 101065,
125626.
Catalog Number: 63185 Lot Number: 101066,
91416, 91417, 91418.
Catalog Number: 63186 Lot Number: 101067,
105513, 126791, 128735, 132843.
Catalog Number: 63187 Lot Number: 101068,
107272, 125627, 132844.
Catalog Number: 63188 Lot Number: 101069,
125628, 132845.
Catalog Number: 63189 Lot Number: 101070,
125629, 128196, 132846.
Catalog Number: 63190 Lot Number: 101071,
125630, 127859, 130789, 132847.
Catalog Number: 63191 Lot Number: 101072,
105514, 126793, 130790, 132848.
Catalog Number: 63192 Lot Number: 101073,
109325, 111068, 132058.
Catalog Number: 63193 Lot Number: 101074,
119766, 125631, 132849, 135994, 140003.
Catalog Number: 63194 Lot Number: 101075,
125632, 130791, 132850.
RECALLING FIRM/MANUFACTURER Recalling Firm: Ballard
Medical Products Draper, UT
Manufacturer:
Ballard Medical Products, Pocatello, ID.
REASON Stent anchor wings may not project, thereby failing
to anchor stent.
VOLUME OF PRODUCT IN COMMERCE 6,312 units.
DISTRIBUTION Nationwide and Internationally.
PRODUCT Portex Tracheostomy Tube, Laryngectomy tube, and
Nasopharyngeal Airway products as follows:
Flex
D.I.C.(TM) Trach Tubes, Product Number, 504070,504080, 504090, 505070, 505080,
505090, 505100.
D.I.C.(TM)
Trach Tubes, Product Number 502060, 502080, 502090, 503060, 503070, 503080,
503090, 503100, 512060, 512080, 512090, 513070, 513080.
Laryngectomy
Tube Product Number, 562090.
Nasopharyngeal
Airway Product Number, 340060, 340070, 340080, 340090.
Pediatric
Trach Tubes Product Number, 555030, 555035, 555040, 555050, 555055.
Per
Fit (R) Percutaneous Trach Tubes Product Number, 522070, 522080, 522090.
Lo
Profile Trach Tube Product Number, 592070.
Blue
Line(R) Trach Tubes Product Number, 508070, 518070, 518090, 521070, 521080,
521090, 530060, 530070, 530080, 530090, 531080, 531100, 550060, 550070,
550080, 550090.
CODE Portex Tracheostomy Tube, Laryngectomy Tube, and
Nasopharyngeal Airway Products with the Lot Numbers listed below:
Flex
D.I.C.(TM) Trach Tubes
Product
Number Lot Number
504070 K108407
504080 110562, K108408
504090 108359, K108409
505070 109606, 110184, 110625,K101267, K101268, K112021
505080 108657
505090 108679, 110219, K101160, K112025
505100 108728, K105823
D.I.C.(TM)
Trach Tubes
Product
Number Lot Number
502060 109233
502080 K109382
502090 108614, K111020
503060 111010, 111072, K107619, K109385
503070 110075, 111482
503080 109057, 110632, 111069, 111070, K103444,
K107019, K107021, K107620,
K112685, K112686
503090 108356, 108466
503100 K112023
512060 K105824, K109398, K112345
512080 K107023
512090 108616
513060 K106753
513070 111484, 111485
513080 K102747, K112026
Laryngectomy
Tube
Product
Number Lot Number
562090 108468
Nasopharyngeal
Airway
Product
Number Lot Number
340060 111187
340070 111073, K100232, K108108
340080 111582
340090 111388, K109963
Pediatric
Trach Tubes
Product
Number Lot Number
555030 K103059
555035 K107624
555040 111488, K102749
555050 K108417
555055 K106090
Per
Fit (R) Percutaneous Trach Tubes
Product
Number Lot Number
522070 108683, 111337
522080 111096
522090 111097, 111120, 111577
Lo
Profile Trach Tube
Product
Number Lot Number
592070 K101755
Blue
Line(R) Trach Tubes
Product
Number Lot Number
508070 109132
518070 108127, 109134
518090 109135, K108443
521070 111006
521080 110626, K112122
521090 110633, K108444
530060 110628, K105829
530070 110113, 111270
530080 109511, K101751, K101752, K101753, K112125
530090 K108445
531080 K105831
531100 108681
550060 K103057
550070 K105832
550080 K103058
550090 K108415.
RECALLING FIRM/MANUFACTURER Recalling Firm: Sims Portex,
Inc., Keene, NH
Manufacturer:
Portex, Inc., Keene, NH.
REASON Product package may have defective seals and
compromise sterility of device.
VOLUME OF PRODUCT IN COMMERCE 22,794 units.
DISTRIBUTION Nationwide and Bahamas, Canada, Puerto Rico,
Mexico and the United Kingdom.
PRODUCT AXIS/IRIX Gamma Camera System. A Prism XPVT System,
Tomography, Computed, Emission, 90KPS System.
System Catalog # Gantry S/N affected
AXIS 210714 094-193
AXIS 210880 101-291
AXIS 211037 181-613
AXIS 211039 140-613
IRIX 210857 096-177,435
IRIX 210881 102-167
IRIX 211038 190-613
IRIX 211040 096-613.
CODE Serial Numbers: 094 through 613.
RECALLING FIRM/MANUFACTURER Recalling Firm: Philips
Medical Systems, fka: Picker International, Inc., Highland Heights, OH
REASON If the patient table is lowered onto a nearby
object, such as a gurney, it may cause it to tip over.
VOLUME OF PRODUCT IN COMMERCE 500.
DISTRIBUTION Nationwide and Internationally.
PRODUCT INRAD brand Express core biopsy device, single use,
disposable, sterile; 18 gauge x 15 cm.
Catalog # 901518.
CODE Lot 6020-1003, exp. 1/2005.
RECALLING FIRM/MANUFACTURER Recalling Firm: INRAD, Inc.,
Kentwood, MI
Manufacturer:
Arlanda Medicinska Instrument Co. AB, Sollentuna,
REASON Extended expiration date on outer carton.
VOLUME OF PRODUCT IN COMMERCE 105.
DISTRIBUTION CA, SC, PA, TN and WA.
WEEK ENDING JUNE 15
PRODUCT Hematocrit and creatinine tests, as part of the
i-STAT in-vitro diagnostic test system.
Field Correction affected cartridges:
i-STAT
6+, E3+, EC4+, EC6+, EC8+.Recall cartridges: Crea.
CODE
Lot Cartridge Number Total
CREA R01331 9150
CREA R01318 7800
CREA R01327 7450
CREA R01305 5725
CREA R01349B 4500
CREA R01349 2575
CREA R01310 1750
CREA R01349A 1700
CREA R01355 1475
CREA R01310A 1075.
RECALLING FIRM/MANUFACTURER I-Stat Corp., East Windsor,
NJ
REASON Erroneously high hematocrit results generated with
cartridges.
VOLUME OF PRODUCT IN COMMERCE 43,200 crea cartridges.
DISTRIBUTION Nationwide.
PRODUCT Clearview HCG II Pregnancy Test
CODE Lot No.
4357, exp. 8/14/2002 4359, exp.
8/11/2002.
RECALLING FIRM/MANUFACTURER Recalling Firm: Medpointe, Inc. (Wampole Laboratories
Division), Cranbury, NJ
Manufacturer: Unipath Ltd., Bedford, the United Kingdom.
REASON Sensitivity deterioration of pouched PRODUCT.
VOLUME OF PRODUCT IN COMMERCE Lot 4357 -- 448 kits; Lot
4359 -- 1,672 kits.
DISTRIBUTION Nationwide.
PRODUCT Mitek Clearfix Meniscal Screw, 2 mm x 10 mm.
CODE Catalog Number: 253101 Lot Numbers: 011097, 0111229,
0111337, 0112145, 0112291.
RECALLING FIRM/MANUFACTURER Mitek products, Norwood, MA
REASON Meniscal screw may be out of specification and not
deploy properly.
VOLUME OF PRODUCT IN COMMERCE 698.
DISTRIBUTION Nationwide and Internationally.
PRODUCT Brand name: Uropump Tube, Uropump Damping
Chamber.The tube and chamber are sold stand-alone and in four kits, the Uropump
Starter Kit, the Cavernosometry Starter Kit, and two different configurations
of the Instrument Set Up Kit. When sold stand alone, the UPT is sold in a box
of 50, and the UPDCC in a box of 25.
CODE UPDCC Lot numbers:
2835603 through 2836509; 2900101 through 2936509; 2A00101 through
2A36509; 2B00101 through 2B36103.
UPT
Lot numbers: 2A27002 through 2A36502; 2B00101 through 2B34401.
RECALLING FIRM/MANUFACTURER Life-Tech Intl., Inc.,
Stafford, TX
REASON Device may not be sterile.
VOLUME OF PRODUCT IN COMMERCE 11,500 boxes.
DISTRIBUTION Nationwide and Internationally.
PRODUCT Hill-Rom brand Affinity Three birthing bed.
CODE All models distributed prior to March 1, 2002.
RECALLING FIRM/MANUFACTURER Hill-Rom, Inc., Batesville,
IN
REASON Bed may change position on its own.
VOLUME OF PRODUCT IN COMMERCE 1,927.
DISTRIBUTION Nationwide and Internationally.
PRODUCT HEARTPORT Brand Arterial Cannulas described as
follows:
Endo
Return, 21 Fr.
Endo
Return, 23 Fr.
Direct
Flow, 24 Fr.
StraightShot,
Angled, 23 Fr.
Arterial,
19 Fr.
The
above products are distributed separately as well as in the following Kit Model
Numbers:
EAVR23A1
ECPB212
ECPB233
ECPBA214
ED241
EAVR-23-A
EAVR-23-S
ED-KIT
EAVR23A2
ECPB213
ECPB234
ECPBA231
ED242
SS-23-A
SS-23-S
SC-24
EAVR23S1
ECPB214
ECPBA211
ECPBA232
ED243
ECPB-AR21
ECPB-AR23
ED-24
EAVR23S2
ECPB231
ECPBA212
ECPBA233
ED244
ECPB-A21
ECPB-A23
ECPB211
ECPB232
ECPBA213
ECPBA234
SC24
ECPB-KIT
ECPB-AR-19.
CODE
ER21
(07043) for Endo Return, 21 Fr.
ER23
(07044) for Endo Return, 23 Fr.
DFK24
(06723) for Direct Flow, 24 Fr.
SS23A
(06725) for StraightShot, Angled, 23 Fr.
AC19
(04000) for Arterial, 19 Fr.
SCK
(06724) for Soft Clamp, 24 Fr.
SS23S
(06726) for StraightShot, Straight, 23 Fr.
RECALLING FIRM/MANUFACTURER Heartport, Inc., Redwood
City, CA
REASON A lap joint in certain arterial cannulas can
separate.
VOLUME OF PRODUCT IN COMMERCE 1,800 units.
DISTRIBUTION Nationwide and Internationally.
PRODUCT a) Monotube Triax Carbon Tube 15x200mm, Yellow, used
for the stabilization of open and/or unstable fractures.
b) Monotube Triax Carbon Tube 15x150mm,
Yellow.
CODE a) Catalog No. 5150-2-381, Lot J326229; b) Catalog
No. 5150-2-371, Lot J326228.
RECALLING FIRM/MANUFACTURER Recalling Firm: Howmedica Osteonics, Corp., Rutherford, NJ
Manufacturer: Stryker Trauma SA, Geneva, Switzerland.
REASON 200mm monotube length package actually contains a
150mm length monotube.
VOLUME OF PRODUCT IN COMMERCE a) 4 ea; b) 2 ea.
DISTRIBUTION NJ.
PRODUCT VITROS Immunodiagnostics.Free T3 Reagent Packs
For
in vitro diagnostic use. 100 test units per pack.
CAT
131 5589 Responsible firm on the label: Ortho-Clinical Diagnostics, Amersham,
the United Kingdom.
CODE Lot numbers:
1)
500, Exp. 10 April 2002
2)
511, Exp. 22 May 2002
3)
512, Exp. 9 August 2002.
RECALLING FIRM/MANUFACTURER Recalling Firm:
Ortho-Clinical Diagnostics, Inc., Rochester, NY
Manufacturer:
Ortho-Clinical Diagnostics, Inc., Cardiff, Wales, the United Kingdom.
REASON Failure to include Supplemental Info. Sheet that
contains updated values for Free Thyroid Control lots 60, 70 & 80.
VOLUME OF PRODUCT IN COMMERCE 262 domestic; and 1,981
international.
DISTRIBUTION Nationwide.
PRODUCT STERRAD 100S Sterilization System
CODE Version 3 or lower is subject to upgrade. Newer versions contain software
"fix". Current version is
Version 6.
RECALLING FIRM/MANUFACTURER Advanced Sterilization
Products, Irvine, CA
REASON Software anomaly allows half cycle run when full
cycle displayed.
VOLUME OF PRODUCT IN COMMERCE 1,405
DISTRIBUTION Nationwide.
PRODUCT Interplak power plaque remover toothbrush, battery
operated toothbrush used to brush teeth and remove plaque.
CODE Model number 12209W.
RECALLING FIRM/MANUFACTURER Recalling Firm: Conair Corp., East Windsor, NJ
Manufacturer: Gimelli Laboratories Co. Ltd., Shenzen
Bao'An, Guangdong.
REASON Batteries leaking hydrogen gas during charging
popping off the end cap.
VOLUME OF PRODUCT IN COMMERCE 15,758 units.
DISTRIBUTION Nationwide.
PRODUCT Invacare IRC 2121 Select Flow, CGS 870, 0-6L medical
oxygen regulators.
CODE Serial numbers 0203 00322 through 0203 00496.
RECALLING FIRM/MANUFACTURER Essex Industries, Inc., St.
Louis, MO
REASON Mislabeled as 0-6 LPM rather than 0-8 LPM.
VOLUME OF PRODUCT IN COMMERCE 175 regulators.
DISTRIBUTION IN.
The
FDA has shut down certain operations of Air Liquide America Corp. (ALAC), a medical
gas facility in Coatesville, Penna. The company had failed to assure that its
industrial-grade trailers (also used for storing and hauling medical gases) are
clean, purged, and completely free of contaminants prior to having them filled
with medical gas, thereby creating a potential hazard for patients who use the
product.
ALAC
had agreed to comply with FDA product safety regulations called Current Good
Manufacturing Practices (CGMP) in a consent decree (a court-sanctioned
agreement between two parties) signed in November, 1997, but an FDA inspection
on April 2, 3, and 10, 2002, found that the Coatesville facility had failed to
meet the agreed upon standards. As a result, the inspected plant must
immediately stop having their trailers filled with medical gas and must stop
distributing medical gases.
Of
greatest concern is that the use of trailers that have not been determined to
be free of industrial gas could result in a contaminated medical gas. The
consequences of inhaling contaminated medical gas may include aspiration
pneumonia. ALAC must also determine whether any other facilities of the company
are in violation of CGMPs, provide the FDA with a list of such facilities, and
provide to FDA a certification that the violative practices have ceased.
ALAC may resume having its trailers filled with medical gas and distributing it to its medical customers once it has an expert consultant certify to FDA that its operations are in compliance with CGMP.
WEEK
ENDING JUNE 29
PRODUCT NEMIO, SSA550A, Diagnostic
Ultrasound System
CODE Serial Numbers: B1552349,
C1572413, C1572414, C1572415, C1572417, C1572418, C1572420, C1572423, C1572428,
C1572429, C1572430
C1572431, C1572432, C1572433, C1572421.
RECALLING FIRM/MANUFACTURER Toshiba America Medical
Systems, Inc., Tustin, CA
REASON Software anomally caused
error in calculating VM_P.
VOLUME OF PRODUCT IN COMMERCE 15.
DISTRIBUTION CA, ID, LA, MA, MD, MS, MT,
OR and TX.
PRODUCT Puritan Bennett-840 Ventilator
System with optional compressor
CODE Ventilators with Rev. E BD
CPU boards with Serial numbers 3510000007 through 3510012304.
RECALLING FIRM/MANUFACTURER Recalling Firm: Tyco Healthcare, Nellcor Puritan Bennett,
Pelasanton, CA
Manufacturer:
Nellcor Puritan Bennett, Ireland, Ltd. Mervue, Galway, Ireland.
REASON Failure of optional
compressor on ventilator.
VOLUME OF PRODUCT IN COMMERCE 2,599 ventilators systems
and 299 field replacement units with Rev. E BD CPU boards.
DISTRIBUTION Nationwide and
Internationally.