FEBRUARY
2009
WEEK ENDING FEBRUARY 7
PRODUCT
a) Medtronic SynchroMed EL Programmable Pump, Model 8626-10. 10 mL
Reservoir. The contents [STERILE] of the inner package have been sterilized by
ethylene oxide gas. The implantable Medtronic SynchroMed EL programmable Pump is
part of the SynchroMed EL Infusion System designed to contain and administer
parenteral drugs to a specific site. The implantable components of the
SynchroMed EL Infusion System include the pump with or without a side catheter
access port, catheters, and catheter accessories. The SynchroMed EL lnfusion
System is indicated for use when patient therapy requires the chronic infusion
of drugs or fluids. The EL Infusion System with a 10 reservoir volume is
indicated for use in the smaller patient who has insufficient body mass to
support the larger SynchroMed EL pump or in patients who choose a smaller, lower
profile pump when presented with the choice, when patient therapy requires the
chronic infusion of drugs or fluids.
b) Medtronic SynchroMed EL Programmable Pump, Model 8626L-10. 10 mL Reservoir. The contents [STERILE] of the inner package have been sterilized by ethylene oxide gas. The implantable Medtronic SynchroMed EL programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site. The implantable components of the SynchroMed EL Infusion System include the pump with or without a side catheter access port, catheters, and catheter accessories. The SynchroMed EL lnfusion System is indicated for use when patient therapy requires the chronic infusion of drugs or fluids. The EL Infusion System with a 10 reservoir volume is indicated for use in the smaller patient who has insufficient body mass to support the larger SynchroMed EL pump or in patients who choose a smaller, lower profile pump when presented with the choice, when patient therapy requires the chronic infusion of drugs or fluids.
c) Medtronic SynchroMed EL Programmable Pump, Model 8626-18. 18 mL Reservoir. The contents [STERILE] of the inner package have been sterilized by ethylene oxide gas. The implantable Medtronic SynchroMed EL programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site. The implantable components of the SynchroMed EL Infusion System include the pump with or without a side catheter access port, catheters, and catheter accessories. The SynchroMed EL lnfusion System is indicated for use when patient therapy requires the chronic infusion of drugs or fluids. The EL Infusion System with a 10 reservoir volume is indicated for use in the smaller patient who has insufficient body mass to support the larger SynchroMed EL pump or in patients who choose a smaller, lower profile pump when presented with the choice, when patient therapy requires the chronic infusion of drugs or fluids.
d) Medtronic SynchroMed EL Programmable Pump, Model 8626L-18. 18 mL Reservoir. The contents [STERILE] of the inner package have been sterilized by ethylene oxide gas. The implantable Medtronic SynchroMed EL programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site. The implantable components of the SynchroMed EL Infusion System include the pump with or without a side catheter access port, catheters, and catheter accessories. The SynchroMed EL lnfusion System is indicated for use when patient therapy requires the chronic infusion of drugs or fluids. The EL Infusion System with a 10 reservoir volume is indicated for use in the smaller patient who has insufficient body mass to support the larger SynchroMed EL pump or in patients who choose a smaller, lower profile pump when presented with the choice, when patient therapy requires the chronic infusion of drugs or fluids.
e) Medtronic SynchroMed EL Programmable Pump, Model 8627-10. 10 mL Reservoir. The contents [STERILE] of the inner package have been sterilized by ethylene oxide gas. The implantable Medtronic SynchroMed EL programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site. The implantable components of the SynchroMed EL Infusion System include the pump with or without a side catheter access port, catheters, and catheter accessories. The SynchroMed EL lnfusion System is indicated for use when patient therapy requires the chronic infusion of drugs or fluids. The EL Infusion System with a 10 reservoir volume is indicated for use in the smaller patient who has insufficient body mass to support the larger SynchroMed EL pump or in patients who choose a smaller, lower profile pump when presented with the choice, when patient therapy requires the chronic infusion of drugs or fluids.
f) Medtronic SynchroMed EL Programmable Pump, Model 8627L-10. 10 mL Reservoir. The contents [STERILE] of the inner package have been sterilized by ethylene oxide gas. The implantable Medtronic SynchroMed EL programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site. The implantable components of the SynchroMed EL Infusion System include the pump with or without a side catheter access port, catheters, and catheter accessories. The SynchroMed EL lnfusion System is indicated for use when patient therapy requires the chronic infusion of drugs or fluids. The EL Infusion System with a 10 reservoir volume is indicated for use in the smaller patient who has insufficient body mass to support the larger SynchroMed EL pump or in patients who choose a smaller, lower profile pump when presented with the choice, when patient therapy requires the chronic infusion of drugs or fluids.
g) Medtronic SynchroMed EL Programmable Pump, Model 8627-18. 18 mL Reservoir. The contents [STERILE] of the inner package have been sterilized by ethylene oxide gas. The implantable Medtronic SynchroMed EL programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site. The implantable components of the SynchroMed EL Infusion System include the pump with or without a side catheter access port, catheters, and catheter accessories. The SynchroMed EL lnfusion System is indicated for use when patient therapy requires the chronic infusion of drugs or fluids. The EL Infusion System with a 10 reservoir volume is indicated for use in the smaller patient who has insufficient body mass to support the larger SynchroMed EL pump or in patients who choose a smaller, lower profile pump when presented with the choice, when patient therapy requires the chronic infusion of drugs or fluids.
h) Medtronic SynchroMed EL Programmable Pump, Model 8627L-18. 18 mL Reservoir. The contents [STERILE] of the inner package have been sterilized by ethylene oxide gas. The implantable Medtronic SynchroMed EL programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site. The implantable components of the SynchroMed EL Infusion System include the pump with or without a side catheter access port, catheters, and catheter accessories. The SynchroMed EL lnfusion System is indicated for use when patient therapy requires the chronic infusion of drugs or fluids. The EL Infusion System with a 10 reservoir volume is indicated for use in the smaller patient who has insufficient body mass to support the larger SynchroMed EL pump or in patients who choose a smaller, lower profile pump when presented with the choice, when patient therapy requires the chronic infusion of drugs or fluids.
i) Medtronic SynchroMed II Programmable Pump, Model 8637-20. 20 mL Reservoir. The contents [STERILE] of the inner package have been sterilized by ethylene oxide gas. The implantable Medtronic SynchroMed II programmable Pump is part of the SynchroMed II Infusion System designed to contain and administer prescribed drugs to a specific site. The implantable components of the SynchroMed II Infusion System include the pump, catheter, and catheter accessories. The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the follosing drugs or fluids: -chronic intraspinal (epidural and intrathecal) infusion of preservative-free morphine sulfate sterile solution in the treatment of chronic intractable pain. Preservative-free sterile saline, USP, can be used to achieve the physician-prescribed concentration of preservative-free morphine sulfate sterile solution. - chronic intrathecal infusion of preservative-free ziconotide sterile solution for the management of serve chronic pain. Preservative-free sterile saline, USP, can be used to achieve the physician-prescribed concentration of preservative-free ziconotide sterile solution after the initial fill of the pump with this drug. -chronic intrathecal infusion of Lioresal Intrathecal (baclofen injection) in the management of severe spasticity. Preservative-free sterile saline, USP, can be used to achieve the physician-prescribed concentration of Lioresal Intrathecal (baclofen Injection). -Chronic intravascular infusion of floxuridine, (FUDR) or methotrexate for the treatment of primary or metastatic cancer. Bacteriostatic water or preservative-free sterile saline, USP, can be used to achieve the physician-prescribed concentration of chemotherapy drugs or to flush the pump reservoir. Saline or heparinized saline (unless contraindicated) may be used during an interruption in chemotherapy to maintain pump and catheter patency.
j) Medtronic SynchroMed II Programmable Pump, Model 8637-40. 40 mL Reservoir.
The contents [STERILE] of the inner package have been sterilized by ethylene
oxide gas. The implantable Medtronic SynchroMed II programmable Pump is part of
the SynchroMed II Infusion System designed to contain and administer prescribed
drugs to a specific site. The implantable components of the SynchroMed II
Infusion System include the pump, catheter, and catheter accessories. The
SynchroMed II Programmable Pump is indicated when patient therapy requires the
chronic infusion of the follosing drugs or fluids: -chronic intraspinal
(epidural and intrathecal) infusion of preservative-free morphine sulfate
sterile solution in the treatment of chronic intractable pain. Preservative-free
sterile saline, USP, can be used to achieve the physician-prescribed
concentration of preservative-free morphine sulfate sterile solution. - Chronic
intrathecal infusion of preservative-free ziconotide sterile solution for the
management of serve chronic pain. Preservative-free sterile saline, USP, can be
used to achieve the physician-prescribed concentration of preservative-free
ziconotide sterile solution after the initial fill of the pump with this drug. –
Chronic intrathecal infusion of Lioresal Intrathecal (baclofen injection) in the
management of severe spasticity. Preservative-free sterile saline, USP, can be
used to achieve the physician-prescribed concentration of Lioresal Intrathecal (baclofen Injection). - Chronic intravascular infusion of
floxuridine, (FUDR) or
methotrexate for the treatment of primary or metastatic cancer. Bacteriostatic
water or preservative-free sterile saline, USP, can be used to achieve the
physician-prescribed concentration of chemotherapy drugs or to flush the pump
reservoir. Saline or heparinized saline (unless contraindicated) may be used
during an interruption in chemotherapy to maintain pump and catheter patency.
CODE
All SynchroMed EL and
SynchroMed II pumps
RECALLING FIRM/MANUFACTURER Medtronic
Neuromodulation, Minneapolis, MN,
REASON Medtronic issued a
letter to healthcare professionals providing important safety information
regarding MRI (magnetic resonance imaging) effects on SynchroMed EL and
SynchroMed II implantable infusion pumps. As stated in product labeling, the
magnetic field of an MRI will temporarily stop the rotor of the pump motor and
suspend drug infusion for the duration of MRI exposure for all SynchroMed pumps.
The pump should resume normal operation when removed from the MRI magnetic
field, however new information has been identified related to the following MRI
effects: 1) There is the potential for a delay in the return of proper drug
infusion after an MRI (this affects all SynchroMed pumps) and 2) There is the
potential for a delay in the logging of motor stall events after an MRI (this
only affects SynchroMed II pumps).
VOLUME OF PRODUCT IN
COMMERCE Approx
76,253
DISTRIBUTION Nationwide and countries of Algeria,
Aruba, Austria, Belarus, Belgium, Croatia, Cyprus Czech Republic, Denmark,
Egypt, Finland, France, French Polynesia, Germany, Greece, Hungary, Iceland,
India, Iran, Iraq, Ireland, Israel, Italy , Jordan Kuwait, Lebanon, Luxembourg,
Malta, Netherlands, Netherlands Antilles, Norway, Pakistan, Poland, Portugal,
Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia and
Montenegro, Slovakia, Slovenia, Sough Africa, Spain, Sweden, Switzerland,
Tunisia, Turkey, United Arab Emirates, United Kingdom, Vatican City State, New
Caledonia, Brazil, Dominican Republic, Mexico, Hong Kong, Singapore, Canada,
Australia and
Japan
PRODUCT Liko AB Uno Lifts, Powered (non-AC) Patient Lift: a) Uno 100
EM/EE;
b) Uno 101. Uno 100 and 102 are battery powered mobile lifts, designed to be
used in most of the common lifting situations, for example transfers between bed
and wheelchair, to and from the toilet and bathtub or for lifting to and from
the floor.
CODE
a)
Serial numbers: 7090001 through 7096200;
b) Serial
numbers: 10001 through 1000. Production Dates: Manufactured prior to June 26,
2008
RECALLING FIRM/MANUFACTURER
Recalling Firm:
Liko North America Inc., Franklin, MA,
Manufacturer: Liko AB,
Lulea, Sweden.
REASON Actuator of the lift has
potential to separate when the upper arm is fully extended.
VOLUME OF
PRODUCT IN COMMERCE 24,070
units
DISTRIBUTION Nationwide and
Canada
PRODUCT DLP(R)
Pericardial/Intracardiac Sump Cannulae, 20 Fr., P/N: 12112.
This product is intended for draining the pericardial sac or the cardiac
chambers.
CODE Lot number:
2006100305
RECALLING FIRM/MANUFACTURER
Recalling Firm:
Medtronic Cardiovascular Revascularization & Surgical Therap,
Brooklyn
Park, MN,
Manufacturer: Atek Medical
Manufacturing, Grand Rapids, MI.
REASON Separates: Six complaints were received
from two locations. In all cases the sump tip assembly separated from the tubing
during device removal and had to be retrieved from the patient. There were no
adverse events affecting these patients.
VOLUME OF PRODUCT IN
COMMERCE 560 units
DISTRIBUTION NC, PA, IA,
WI, IN, VA, IL, KY, and TX, and countries of Australia, France, Hong Kong,
Netherlands, and
Poland.
PRODUCT
a) CAD SCIENCES LLC 3 TP TheraMapTM Software, for oncology
treatment planning, assessment and monitoring of the patient response.
b) CAD SCIENCES LLC 3 TP PrecisionPointTM Software, for biopsy guidance.
CODE
All software
versions
RECALLING FIRM/MANUFACTURER iCAD, Inc.,
Nashua, NH,
REASON Software modules not
approved for this indication.
VOLUME OF PRODUCT IN
COMMERCE 35
units
DISTRIBUTION Nationwide
PRODUCT
a) Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm
Hook Probe. Product Code EPS01.
b) Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Spatula Probe. Product Code EPS02.
c) Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Right Angle Probe. Product Code EPS03.
d) Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Curved Dissector Probe. Product Code EPS04.
e) Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Hook Probe. Product Code EPS05.
f) Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Spatula Probe. Product Code EPS06.
g) Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Right Angle Probe. Product Code EPS07.
h) Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Curved Dissector Probe. Product Code EPS08.
i) Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Needle
Probe. Product Code EPS09.
CODE
a)
Lot # E4K594, Exp Date 2012-12; Lot # E4K67R, Exp Date 2012-12; Lot # E4K93J,
Exp Date 2012-12; Lot # E4KC5J, Exp Date 2013-01; Lot # E4KF21, Exp Date
2013-01; Lot # E4KK1M, Exp Date 2013-02; Lot # E4KK7Z, Exp Date 2013-02; Lot #
E4KL9T, Exp Date 2013-02; Lot # E4KM4A, Exp Date 2013-02; Lot # E4KP38, Exp Date
2013-02; Lot # E4KT81, Exp Date 2013-03; Lot # E4KU2X, Exp Date 2013-03; Lot #
E4KX2D, Exp Date 2013-03; Lot # E4L01C, Exp Date 2013-04; Lot # E4L02M, Exp Date
2013-04; Lot # E4L21T, Exp Date 2013-04; Lot # E4L66N, Exp Date 2013-04; Lot #
E4L738, Exp Date 2013-05; Lot # E4L82C, Exp Date 2013-05; Lot # E4LA6P, Exp Date
2013-05; Lot # E4LC88, Exp Date 2013-05; Lot # E4LE62, Exp Date 2013-05; Lot #
E4LG4C, Exp Date 2013-06; Lot # E4LJ8W, Exp Date 2013-06; Lot # E4LK99, Exp Date
2013-06; Lot # E4LM57, Exp Date 2013-06; Lot # E4LP5G, Exp Date 2013-07; Lot #
E4LU30, Exp Date 2013-07; Lot # E4LV48, Exp Date 2013-07; Lot # E4LX91, Exp Date
2013-07; Lot # E4M068, Exp Date 2013-08; Lot # E4M12K, Exp Date 2013-08; Lot #
E4M23T, Exp Date 2013-08; Lot # E4M33W, Exp Date 2013-08; Lot # E4M387, Exp Date
2013-08; Lot # E4M43R, Exp Date 2013-08; Lot # E4M54H, Exp Date 2013-08; Lot #
E4M75K, Exp Date 2013-09; Lot # E4M973, Exp Date 2013-09; and Lot # E4MA8R, Exp
Date 2013-09;
b) Lot # E4K181, Exp. Date 2012-12; Lot # E4K58C, Exp. Date 2012-12; Lot # E4KC5X, Exp. Date 2013-01; Lot # E4KD1P, Exp. Date 2013-01; Lot # E4KF2N, Exp. Date 2013-01; Lot # E4KG70, Exp. Date 2013-01; Lot # E4KK22, Exp. Date 2013-02; Lot # E4KP2F, Exp. Date 2013-02; Lot # E4KV17, Exp. Date 2013-03; Lot # E4KY1A, Exp. Date 2013-03; Lot # E4KY70, Exp. Date 2013-03; Lot # E4L03H, Exp. Date 2013-04; Lot # E4L519, Exp. Date 2013-04; Lot # E4L73Z, Exp. Date 2013-05; Lot # E4L916, Exp. Date 2013-05; Lot # E4LC3K, Exp. Date 2013-05; Lot # E4LD8L, Exp. Date 2013-05; Lot # E4LF60, Exp. Date 2013-05; Lot # E4LG4D, Exp. Date 2013-06; Lot # E4LL7P, Exp. Date 2013-06; Lot # E4LT6E, Exp. Date 2013-07; Lot # E4LX4H, Exp. Date 2013-07; Lot # E4M045, Exp. Date 2013-08; Lot # E4M12L, Exp. Date 2013-08; Lot # E4M27N, Exp. Date 2013-08; Lot # E4M466, Exp. Date 2013-08; Lot # E4M535, Exp. Date 2013-08; Lot # E4M54F, Exp. Date 2013-08; Lot # E4M69H, Exp. Date 2013-09; Lot # E4M838, Exp. Date 2013-09; Lot # E4MA4M, Exp. Date 2013-09; and Lot # E4MC3A, Exp. Date 2013-09;
c) Lot # E4K182, Exp. Date 2012-12; Lot # E4K59R, Exp. Date 2012-12; Lot # E4K67T, Exp. Date 2012-12; Lot # E4K81D, Exp. Date 2012-12; Lot # E4K938, Exp. Date 2012-12; Lot # D4JL20, Exp. Date 2012-09; Lot # D4JL21, Exp. Date 2012-09; Lot # D4JL22, Exp. Date 2012-09; Lot # E4KC1M, Exp. Date 2013-01; Lot # E4KF20, Exp. Date 2013-01; Lot # E4KK23, Exp. Date 2013-02; Lot # E4KK6V, Exp. Date 2013-02; Lot # E4KL0N, Exp. Date 2013-02; Lot # E4KR6Y, Exp. Date 2013-03; Lot # E4KV1X, Exp. Date 2013-03; Lot # E4KX2Z, Exp. Date 2013-03; Lot # E4KY5H, Exp. Date 2013-03; Lot # E4L18A, Exp. Date 2013-04; Lot # E4L404, Exp. Date 2013-04; Lot # E4L66M, Exp. Date 2013-04; Lot # E4L69A, Exp. Date 2013-05; Lot # E4L871, Exp. Date 2013-05; Lot # E4LA7Z, Exp. Date 2013-05; Lot # E4LD0T, Exp. Date 2013-05; Lot # E4LF5F, Exp. Date 2013-05; Lot # E4LJ46, Exp. Date 2013-06; Lot # E4LJ4F, Exp. Date 2013-06; Lot # E4LK9C, Exp. Date 2013-06; Lot # E4LN16, Exp. Date 2013-06; Lot # E4LP57, Exp. Date 2013-07; Lot # E4LR66, Exp. Date 2013-07; Lot # E4LT3Y, Exp. Date 2013-07; Lot # E4LU2Z, Exp. Date 2013-07; Lot # E4LV56, Exp. Date 2013-07; Lot # E4M00W, Exp. Date 2013-08; Lot # E4M01X, Exp. Date 2013-08; Lot # E4M120, Exp. Date 2013-08; Lot # E4M20L, Exp. Date 2013-08; Lot # E4M23R, Exp. Date 2013-08; Lot # E4M31A, Exp. Date 2013-08; Lot # E4M376, Exp. Date 2013-08; Lot # E4M39C, Exp. Date 2013-08; Lot # E4M572, Exp. Date 2013-08; Lot # E4M57U, Exp. Date 2013-08; Lot # E4M62M, Exp. Date 2013-08; Lot # E4M914, Exp. Date 2013-09; Lot # E4MA2R, Exp. Date 2013-09; and Lot # E4MD45, Exp. Date 2013-09;
d) Lot # E4KC2J, Exp. Date 2013-01; Lot # E4KG8M, Exp. Date 2013-01; Lot # E4KP8H, Exp. Date 2013-02; Lot # E4KV0T, Exp. Date 2013-03; Lot # E4L030, Exp. Date 2013-04; Lot # E4L892, Exp. Date 2013-05; Lot # E4LD0R, Exp. Date 2013-05; Lot # E4LK7V, Exp. Date 2013-06; Lot # E4LX1D, Exp. Date 2013-07; Lot # E4M072, Exp. Date 2013-08; Lot # E4M28F, Exp. Date 2013-08; Lot # E4M91D, Exp. Date 2013-09; and Lot # E4MA2P, Exp. Date 2013-09;
e) Lot # E4K72Z, Exp. Date 2012-12; Lot # E4K92E, Exp. Date 2012-12; Lot # E4KE29, Exp. Date 2013-01; Lot # E4KH2L, Exp. Date 2013-01; Lot # E4KP2E, Exp. Date 2013-02; Lot # E4KU5K, Exp. Date 2013-03; Lot # E4KX1Y, Exp. Date 2013-03; Lot # E4L03G, Exp. Date 2013-04; Lot # E4L69Y, Exp. Date 2013-05; Lot # E4LA5X, Exp. Date 2013-05; Lot # E4LE64, Exp. Date 2013-05; Lot # E4LF6H, Exp. Date 2013-05; Lot # E4LN1K, Exp. Date 2013-06; Lot # E4LR3W, Exp. Date 2013-07; Lot # E4LU50, Exp. Date 2013-07; Lot # E4M00X, Exp. Date 2013-08; Lot # E4M24N, Exp. Date 2013-08; Lot # E4M506, Exp. Date 2013-08; Lot # E4M62L, Exp. Date 2013-08; Lot # E4M79W, Exp. Date 2013-09; and Lot # E4MA4P, Exp. Date 2013-09;
f) Lot # E4K730, Exp. Date 2012-12; Lot # E4K937, Exp. Date 2012-12; Lot # E4KE28, Exp. Date 2013-01; Lot # E4KF4E, Exp. Date 2013-01; Lot # E4KK3D, Exp. Date 2013-02; Lot # E4KK9Z, Exp. Date 2013-02; Lot # E4KT18, Exp. Date 2013-03; Lot # E4KU5A, Exp. Date 2013-03; Lot # E4KW7W, Exp. Date 2013-03; Lot # E4KY71, Exp. Date 2013-03; Lot # E4KZ5L, Exp. Date 2013-03; Lot # E4KZ9G, Exp. Date 2013-04; Lot # E4L189, Exp. Date 2013-04; Lot # E4L41U, Exp. Date 2013-04; Lot # E4L53G, Exp. Date 2013-04; Lot # E4L54M, Exp. Date 2013-04; Lot # E4L891, Exp. Date 2013-05; Lot # E4LA6F, Exp. Date 2013-05; Lot # E4LD1N, Exp. Date 2013-05; Lot # E4LF86, Exp. Date 2013-05; Lot # E4LJ4C, Exp. Date 2013-06; Lot # E4LN1J, Exp. Date 2013-06; Lot # E4LT6C, Exp. Date 2013-07; Lot # E4LU3P, Exp. Date 2013-07; Lot # E4M20H, Exp. Date 2013-08; Lot # E4M338, Exp. Date 2013-08; Lot # E4M457, Exp. Date 2013-08; Lot # E4M54G, Exp. Date 2013-08; Lot # E4M634, Exp. Date 2013-08; and Lot # E4MA4N, Exp. Date 2013-09;
g) Lot # E4KC1V, Exp Date 2013-01; Lot # E4KJ1J, Exp Date 2013-02; Lot # E4KM09, Exp Date 2013-02; Lot # E4KV1N, Exp Date 2013-03; Lot # E4KZ5M, Exp Date 2013-03; Lot # E4L464, Exp Date 2013-04; Lot # E4L664, Exp Date 2013-04; Lot # E4LA0W, Exp Date 2013-05; Lot # E4LC3M, Exp Date 2013-05; Lot # E4LE63, Exp Date 2013-05; Lot # E4LJ89, Exp Date 2013-06; Lot # E4LK9K, Exp Date 2013-06; Lot # E4LR3Y, Exp Date 2013-07; Lot # E4LX68, Exp Date 2013-07; Lot # E4LX88, Exp Date 2013-07; Lot # E4M071, Exp Date 2013-08; Lot # E4M43P, Exp Date 2013-08; Lot # E4M83K, Exp Date 2013-09; and Lot # E4M98Y, Exp Date 2013-09;
h) Lot # E4KF22, Exp. Date 2013-01; Lot # E4L088, Exp. Date 2013-04; Lot # E4LR3X, Exp. Date 2013-07; and Lot # E4MD46, Exp. Date 2013-09;
i) Lot # E4KC2H, Exp. Date 2013-01; Lot # E4KJ1H, Exp. Date 2013-02; Lot #
E4KR73, Exp. Date 2013-03; Lot # E4KW8R, Exp. Date 2013-03; Lot # E4L465, Exp.
Date 2013-04; Lot # E4LA5T, Exp. Date 2013-05; Lot # E4LD9H, Exp. Date 2013-05;
Lot # E4LF6M, Exp. Date 2013-05; Lot # E4LK7U, Exp. Date 2013-06; Lot # E4LU3N,
Exp. Date 2013-07; Lot # E4M33D, Exp. Date 2013-08; Lot # E4M69J, Exp. Date
2013-09; and Lot # E4M83N, Exp. Date 2013-09
RECALLING
FIRM/MANUFACTURER
Recalling Firm: Ethicon Endo-Surgery Inc.,
Cincinnati, OH,
Manufacturer:
Ethicon Endo-Surgery, Inc. S.A. de C.V. Planta II, Ciudad Juarez,
Chihuahua, Mexico.
REASON Nickel exposure: The stainless steel tips on the
affected product were degrading, causing a small amount of nickel to be
released, which may lead to health risks in some patients, particularly those
with a known or suspected nickel allergy or sensitivity. Ethicon Endo-Surgery
determined the presence of free nickel in the affected product could elicit a
reaction in those patients with nickel sensitivity or could sensitize certain
patients to nickel exposure in the future.
VOLUME OF PRODUCT IN
COMMERCE 109,374
instruments
DISTRIBUTION Nationwide and countries of
Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia,
Dominican Republic, Egypt, England, Equator, Germany, Greece, India, Indonesia,
Israel, Italy, Japan, Korea, Malaysia, Mexico, Pakistan, Panama, Poland,
Republic OF Yemen, Russia, Singapore, South Africa, Spain, Switzerland, Taiwan,
Thailand, Trinidad Turkey United Arab Emirates, Uruguay, and
Venezuela
PRODUCT Stelkast Dual Taper Wedge Pressed Fit Femoral Component. Hip
Prosthesis, Part number: SC1439-1000.
CODE Lot number: 17877-071707 exp.
7/2017
RECALLING FIRM/MANUFACTURER Stelkast Co.,
McMurray, PA,
REASON Sterility may be
compromised due to breach in packaging.
VOLUME OF PRODUCT IN
COMMERCE 6 units
DISTRIBUTION TX,
WV
PRODUCT
a) Medtronic Maximo II DR D28DRG, for ventricular pacing and
defibrillation.
b) Medtronic Maximo II VR D284VRC, for ventricular pacing and defibrillation.
c) Maximo II CRT D284TRK, for ventricular pacing and defibrillation.
CODE
a) Lot Serial Number: PZM600131S
PZM600054S PZM600616S PZM600617S PZM600620S PZM600615S PZM600523S PZM600111S
PZM600213S PZM600502S PZM600595S PZM600600S PZM600277S PZM600078S PZM600581S
PZM600582S PZM600597S PZM600602S PZM600530S PZM600108S PZM600495S PZM600497S
PZM600498S PZM600500S PZM600501S PZM600151S PZM600152S PZM600153S PZM600446S
PZM600447S PZM600448S PZM600264S PZM600278S PZM600469S PZM600094S PZM600401S
PZM600052S PZM600053S PZM600517S PZM600621S PZM600622S PZM600233S PZM600539S
PZM600129S PZM600130S PZM600132S PZM600160S PZM600162S PZM600207S PZM600208S
PZM600138S PZM600140S PZM600145S PZM600146S PZM600148S PZM600156S PZM600198S
PZM600056S PZM600092S PZM600104S PZM600105S PZM600147S PZM600168S PZM600252S
PZM600051S PZM600199S PZM600057S PZM600273S PZM600075S PZM600232S PZM600472S
PZM600049S PZM600203S PZM600276S PZM600071S PZM600143S PZM600098S PZM600110S
PZM600167S PZM600101S PZM600109S PZM600126S PZM600150S PZM600491S PZM600492S
PZM600080S PZM600081S PZM600281S PZM600282S PZM600283S PZM600284S PZM600285S
PZM600286S PZM600287S PZM600288S PZM600289S PZM600290S PZM600184S PZM600192S
PZM600245S PZM600180S PZM600262S PZM600526S PZM600643S PZM600473S PZM600186S
PZM600189S PZM600243S PZM600266S PZM600247S PZM600254S PZM600556S PZM600557S
PZM600558S PZM600559S PZM600560S PZM600561S PZM600562S PZM600563S PZM600584S
PZM600586S PZM600112S PZM600544S PZM600599S PZM600486S PZM600405S PZM600406S
PZM600407S PZM600408S PZM600409S PZM600410S PZM600411S PZM600412S PZM600413S
PZM600414S PZM600415S PZM600416S PZM600417S PZM600418S PZM600419S PZM600420S
PZM600421S PZM600422S PZM600423S PZM600424S PZM600425S PZM600426S PZM600427S
PZM600428S PZM600429S PZM600430S PZM600431S PZM600432S PZM600433S PZM600434S
PZM600435S PZM600436S PZM600437S PZM600438S PZM600439S PZM600440S PZM600441S
PZM600442S PZM600443S PZM600444S PZM600445S PZM600059S PZM600060S PZM600061S
PZM600062S PZM600063S PZM600064S PZM600065S PZM600066S PZM600135S PZM600136S
PZM600137S PZM600116S PZM600117S PZM600118S PZM600119S PZM600120S PZM600121S
PZM600122S PZM600123S PZM600124S PZM600125S PZM600079S PZM600206S PZM600578S
PZM600215S PZM600216S PZM600217S PZM600218S PZM600219S PZM600220S PZM600221S
PZM600222S PZM600223S PZM600224S PZM600265S PZM600083S PZM600470S PZM600528S
PZM600533S PZM600640S PZM600577S PZM600076S PZM600594S PZM600241S PZM600242S
PZM600272S PZM600274S PZM600573S PZM600574S PZM600576S PZM600161S PZM600163S
PZM600234S PZM600133S PZM600134S PZM600193S PZM600194S PZM600090S PZM600509S
PZM600072S PZM600142S PZM600173S PZM600623S PZM600624S PZM600482S PZM600196S
PZM600197S PZM600535S PZM600536S PZM600067S PZM600068S PZM600077S PZM600490S
PZM600190S PZM600611S PZM600612S PZM600084S PZM600087S PZM600088S PZM600089S
PZM600231S PZM600201S PZM600202S PZM600141S PZM600471S PZM600546S PZM600235S
PZM600236S PZM600048S PZM600128S PZM600550S PZM600551S PZM600073S PZM600085S
PZM600055S PZM600158S PZM600251S PZM600165S PZM600280S PZM600070S PZM600170S
PZM600204S PZM600097S PZM600545S PZM600058S PZM600166S PZM600587S PZM600512S
PZM600513S PZM600291S PZM600292S PZM600293S PZM600107S PZM600510S PZM600050S
PZM600149S PZM600102S PZM600238S PZM600240S PZM600598S PZM600601S PZM600086S
PZM600091S PZM600596S PZM600099S PZM600100S PZM600093S PZM600155S PZM600157S
PZM600159S PZM600253S PZM600171S PZM600590S PZM600493S PZM600074S PZM600209S
PZM600580S PZM600494S PZM600069S PZM600200S PZM600144S PZM600154S PZM600047S
PZM600095S PZM600275S PZM600547S PZM600225S PZM600113S PZM600613S PZM600614S
PZM600542S PZM600237S PZM600402S PZM600403S PZM600404S PZM600514S PZM600515S
PZM600258S PZM600259S PZM600487S PZM600248S PZM600279S PZM600508S PZM600256S
PZM600257S PZM600608S PZM600609S PZM600488S PZM600489S PZM600261S PZM600263S
PZM600641S PZM600642S PZM600270S PZM600516S PZM600589S PZM600592S PZM600255S
PZM600249S PZM600169S PZM600172S PZM600174S PZM600175S PZM600176S PZM600177S
PZM600178S PZM600179S PZM600181S PZM600182S PZM600183S PZM600187S PZM600191S
PZM600246S PZM600474S PZM600644S PZM600507S PZM600543S PZM600250S PZM600244S
PZM600534S PZM600548S PZM600549S PZM600537S PZM600538S PZM600552S PZM600553S
PZM600185S PZM600188S PZM600521S PZM600522S PZM600271S PZM600527S PZM600607S
PZM600627S PZM600555S PZM600646S PZM600294S PZM600295S PZM600511S PZM600713S
PZM600714S PZM600260S PZM600524S PZM600525S PZM600554S PZM600268S PZM600269S
PZM600475S PZM600476S PZM600477S PZM600214S PZM600541S PZM600210S PZM600212S
PZM600520S PZM600540S PZM600114S PZM600115S PZM600226S PZM600229S PZM600230S
PZM600483S PZM600484S PZM600485S PZM600529S PZM600205S PZM600572S PZM600575S
PZM600267S PZM600106S PZM600164S PZM600103S and PZM600645S;
b) Lot Serial No PZN600430S PZN600431S PZN600436S PZN600428S PZN600422S PZN600385S PZN600317S PZN600383S PZN600308S PZN600418S PZN600388S PZN600409S PZN600410S PZN600250S PZN600391S PZN600392S PZN600374S PZN600375S PZN600116S PZN600333S PZN600334S PZN600335S PZN600336S PZN600337S PZN600145S PZN600146S PZN600147S PZN600148S PZN600348S PZN600349S PZN600057S PZN600058S PZN600351S PZN600352S PZN600247S PZN600248S PZN600328S PZN600462S PZN600419S PZN600199S PZN600200S PZN600129S PZN600130S PZN600131S PZN600132S PZN600134S PZN600151S PZN600152S PZN600153S PZN600154S PZN600167S PZN600181S PZN600182S PZN600201S PZN600208S PZN600209S PZN600210S PZN600211S PZN600212S PZN600213S PZN600214S PZN600249S PZN600325S PZN600326S PZN600327S PZN600329S PZN600330S PZN600362S PZN600381S PZN600463S PZN600464S PZN600082S PZN600083S PZN600084S PZN600085S PZN600086S PZN600071S PZN600088S PZN600149S PZN600138S PZN600079S PZN600429S PZN600442S PZN600101S PZN600135S PZN600140S PZN600183S PZN600111S PZN600067S PZN600361S PZN600074S PZN600055S PZN600107S PZN600143S PZN600170S PZN600171S PZN600266S PZN600169S PZN600341S PZN600254S PZN600255S PZN600256S PZN600257S PZN600258S PZN600259S PZN600260S PZN600261S PZN600262S PZN600263S PZN600197S PZN600237S PZN600239S PZN600219S PZN600246S PZN600236S PZN600203S PZN600227S PZN600252S PZN600164S PZN600202S PZN600396S PZN600397S PZN600398S PZN600399S PZN600400S PZN600401S PZN600402S PZN600403S PZN600404S PZN600405S PZN600425S PZN600426S PZN600427S PZN600196S PZN600318S PZN600195S PZN600354S PZN600316S PZN600127S PZN600123S PZN600313S PZN600312S PZN600268S PZN600269S PZN600270S PZN600271S PZN600272S PZN600273S PZN600274S PZN600275S PZN600276S PZN600277S PZN600278S PZN600060S PZN600061S PZN600062S PZN600063S PZN600064S PZN600574S PZN600136S PZN600137S PZN600359S PZN600547S PZN600251S PZN600370S PZN600372S PZN600114S PZN600117S PZN600184S PZN600185S PZN600186S PZN600187S PZN600188S PZN600189S PZN600190S PZN600191S PZN600307S PZN600103S PZN600379S PZN600380S PZN600087S PZN600407S PZN600408S PZN600180S PZN600363S PZN600364S PZN600365S PZN600366S PZN600414S PZN600415S PZN600465S PZN600466S PZN600467S PZN600133S PZN600166S PZN600100S PZN600075S PZN600340S PZN600115S PZN600434S PZN600435S PZN600320S PZN600168S PZN600068S PZN600069S PZN600324S PZN600078S PZN600423S PZN600094S PZN600420S PZN600097S PZN600098S PZN600099S PZN600122S PZN600090S PZN600091S PZN600175S PZN600176S PZN600141S PZN600303S PZN600304S PZN600389S PZN600206S PZN600207S PZN600049S PZN600050S PZN600053S PZN600054S PZN600128S PZN600424S PZN600073S PZN600095S PZN600076S PZN600077S PZN600059S PZN600080S PZN600139S PZN600155S PZN600302S PZN600350S PZN600105S PZN600106S PZN600215S PZN600216S PZN600546S PZN600144S PZN600118S PZN600119S PZN600120S PZN600052S PZN600437S PZN600438S PZN600439S PZN600112S PZN600113S PZN600440S PZN600441S PZN600096S PZN600241S PZN600109S PZN600110S PZN600102S PZN600065S PZN600072S PZN600331S PZN600066S PZN600081S PZN600413S PZN600421S PZN600332S PZN600070S PZN600172S PZN600173S PZN600142S PZN600433S PZN600048S PZN600108S PZN600104S PZN600056S PZN600125S PZN600386S PZN600051S PZN600265S PZN600217S PZN600267S PZN600345S PZN600346S PZN600231S PZN600232S PZN600321S PZN600223S PZN600339S PZN600229S PZN600230S PZN600305S PZN600306S PZN600322S PZN600323S PZN600235S PZN600369S PZN600245S PZN600347S PZN600406S PZN600228S PZN600432S PZN600156S PZN600157S PZN600158S PZN600159S PZN600161S PZN600163S PZN600165S PZN600218S PZN600220S PZN600221S PZN600222S PZN600198S PZN600360S PZN600238S PZN600338S PZN600240S PZN600224S PZN600226S PZN600367S PZN600390S PZN600377S PZN600378S PZN600393S PZN600394S PZN600160S PZN600162S PZN600355S PZN600356S PZN600357S PZN600358S PZN600204S PZN600205S PZN600371S PZN600395S PZN600469S PZN600387S PZN600264S PZN600342S PZN600628S PZN600629S PZN600233S PZN600368S PZN600243S PZN600244S PZN600353S PZN600382S PZN600314S PZN600315S PZN600319S PZN600384S PZN600416S PZN600192S PZN600193S PZN600311S PZN600417S PZN600443S PZN600194S PZN600309S PZN600310S PZN600124S PZN600126S PZN600289S PZN600344S PZN600343S PZN600373S PZN600178S PZN600179S PZN600242S PZN600376S PZN600089S and PZN600449S;
c) Lot Serial No PZP600096S PZP600610S PZP600608S PZP600546S PZP600149S
PZP600569S PZP600587S PZP600589S PZP600612S PZP600450S PZP600557S PZP600193S
PZP600558S PZP600559S PZP600531S PZP600533S PZP600534S PZP600535S PZP600536S
PZP600199S PZP600215S PZP600473S PZP600474S PZP600475S PZP600476S PZP600477S
PZP600478S PZP600479S PZP600480S PZP600607S PZP600513S PZP600514S PZP600515S
PZP600516S PZP600530S PZP600092S PZP600093S PZP600562S PZP600571S PZP600609S
PZP600596S PZP600166S PZP600467S PZP600529S PZP600638S PZP600134S PZP600135S
PZP600172S PZP600133S PZP600509S PZP600128S PZP600196S PZP600103S PZP600111S
PZP600140S PZP600142S PZP600169S PZP600527S PZP600611S PZP600636S PZP600656S
PZP600614S PZP600090S PZP600141S PZP600195S PZP600137S PZP600089S PZP600112S
PZP600178S PZP600127S PZP600173S PZP600211S PZP600213S PZP600511S PZP600524S
PZP600577S PZP600542S PZP600091S PZP600154S PZP600174S PZP600466S PZP600177S
PZP600454S PZP600455S PZP600456S PZP600457S PZP600458S PZP600459S PZP600462S
PZP600463S PZP600464S PZP600465S PZP600543S PZP600552S PZP600434S PZP600510S
PZP600438S PZP600439S PZP600584S PZP600221S PZP600521S PZP600180S PZP600544S
PZP600604S PZP600605S PZP600606S PZP600436S PZP600517S PZP600588S PZP600113S
PZP600114S PZP600115S PZP600116S PZP600117S PZP600118S PZP600119S PZP600120S
PZP600121S PZP600122S PZP600097S PZP600098S PZP600099S PZP600100S PZP600101S
PZP600163S PZP600164S PZP600165S PZP600448S PZP600197S PZP600200S PZP600202S
PZP600203S PZP600204S PZP600205S PZP600206S PZP600207S PZP600208S PZP600209S
PZP600210S PZP600136S PZP600550S PZP600637S PZP600568S PZP600109S PZP600586S
PZP600159S PZP600198S PZP600160S PZP600176S PZP600161S PZP600581S PZP600582S
PZP600132S PZP600108S PZP600168S PZP600170S PZP600214S PZP600512S PZP600526S
PZP600545S PZP600191S PZP600192S PZP600566S PZP600567S PZP600110S PZP600528S
PZP600634S PZP600155S PZP600126S PZP600129S PZP600130S PZP600131S PZP600148S
PZP600123S PZP600124S PZP600171S PZP600508S PZP600158S PZP600105S PZP600153S
PZP600095S PZP600094S PZP600167S PZP600212S PZP600547S PZP600538S PZP600216S
PZP600217S PZP600435S PZP600162S PZP600641S PZP600146S PZP600532S PZP600592S
PZP600593S PZP600594S PZP600595S PZP600143S PZP600144S PZP600555S PZP600556S
PZP600618S PZP600620S PZP600506S PZP600507S PZP600657S PZP600560S PZP600561S
PZP600646S PZP600649S PZP600654S PZP600107S PZP600104S PZP600147S PZP600615S
PZP600617S PZP600537S PZP600598S PZP600102S PZP600194S PZP600583S PZP600157S
PZP600175S PZP600087S PZP600088S PZP600444S PZP600575S PZP600576S PZP600151S
PZP600106S PZP600525S PZP600471S PZP600472S PZP600599S PZP600231S PZP600232S
PZP600520S PZP600226S PZP600453S PZP600541S PZP600229S PZP600230S PZP600522S
PZP600523S PZP600432S PZP600433S PZP600623S PZP600624S PZP600223S PZP600585S
PZP600228S PZP600653S PZP600225S PZP600179S PZP600182S PZP600183S PZP600184S
PZP600185S PZP600186S PZP600188S PZP600190S PZP600218S PZP600219S PZP600222S
PZP600452S PZP600629S PZP600632S PZP600437S PZP600539S PZP600224S PZP600573S
PZP600574S PZP600564S PZP600565S PZP600578S PZP600579S PZP600481S PZP600482S
PZP600483S PZP600484S PZP600442S PZP600621S PZP600622S PZP600580S PZP600551S
PZP600644S PZP600645S PZP600468S PZP600469S PZP600470S PZP600540S PZP600655S
PZP600572S PZP600625S PZP600553S PZP600554S PZP600440S PZP600626S PZP600628S
PZP600630S PZP600631S PZP600150S PZP600152S PZP600548S PZP600549S PZP600570S
PZP600633S PZP600518S PZP600597S PZP600145S PZP600563S PZP600139S and
PZP600643S
RECALLING FIRM/MANUFACTURER
Recalling Firm:
Medtronic Inc. Cardiac Rhythm Management, Mounds View, MN,
Manufacturer: Medtronic Puerto Rico
Operations Company, Juncos, PR.
REASON Maximo II VR
ICD, DR ICD and CRT-D
inadvertently include the Capture Management algorithm (Atrial/Right
Ventricular/Left Ventricular Capture management). The Capture management
algorithm monitors daily pacing thresholds and may increase or decrease pacing
amplitudes in response to patients' needs. Although the algorithm is operating
correctly, it is not accessible to the clinician for reprogramming.
VOLUME OF PRODUCT IN COMMERCE 1,108
units
DISTRIBUTION Nationwide and countries of Austria,
Denmark, Finland, Germany, Italy, Spain, Sweden, UK, Greece, Hong Kong, India,
Jordan and Saudi Arabia
PRODUCT Stratus CS Acute Care Troponin I
TestPak, Catalog number: CCTNI, in
vitro diagnostic,
CODE Lot numbers:
238070002 exp 2008-09-06, 238077002 exp 2008-09-13, 238084002 exp 2008-09-20,
238091002 exp 2008-09-27, 238105002 exp 2008-10-11, 238112002 exp 2008-10-18,
238119002 exp 2008-10-25, 238126002 exp 2008-11-01, 238133002 exp 2008-11-08,
and 238140002 exp 2008-11-15
RECALLING
FIRM/MANUFACTURER Dade Behring, Inc., Newark, DE,
REASON
Incorrect result: Potential for
false elevation of CTnl results.
VOLUME OF PRODUCT IN
COMMERCE 5,868
cartons
DISTRIBUTION Nationwide and countries of
Singapore, Saudi Arabia, Japan, Germany, Canada, Bahrain, Australia, Korea, and
India
PRODUCT Hardy Diagnostic McFarland Standard 0.5; in vitro diagnostic, glass
vial, one vial per box. Model number: Hardy = MS05.
CODE
Lot number: 08182; Expiration date:
10/28/08
RECALLING FIRM/MANUFACTURER
Recalling Firm: Trek
Diagnostic Systems, Cleveland, OH,
Manufacturer: Hardy Diagnostics, Santa Maria, CA.
REASON
Incorrect result:
The absorbance has changed as the standards have aged; the standards’ turbidity
has become denser. The recalled lot is now reading 0.15 ABS @ 625 nm. TREK’S
internal specification and the manufacturer’s (Hardy Diagnostics) specification
is 0.08 – 0.10 ABS @ 625 nm. CLSI specifies 0.08 – 0.13 ABS @ 625
nm.
VOLUME OF PRODUCT IN COMMERCE 4
vials
DISTRIBUTION OH,
IL
PRODUCT GE Centricity PACS RA1000 Workstation (for diagnostic image analysis); The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals. It is also intended for use as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print, and export images when connected with the Centricity PACS infrastructure.
CODE
Software versions PACS 1.0CSR5/2.0CSR1 and all
service packs
RECALLING FIRM/MANUFACTURER GE Healthcare
Integrated IT Solutions, Barrington, IL,
REASON There is a
potential patient safety issue involving the use of the F12 keyboard Approve
& Complete function. The issue may cause erroneous information to be added
to a report when trying to approve and complete it.
VOLUME OF
PRODUCT IN COMMERCE 138
units
DISTRIBUTION Nationwide and countries of Canada,
China, Denmark, Germany, Hong Kong, India, Ireland, Italy, Japan, Korea,
Lebanon, Mexico, Netherlands, Saudi Arabia, Spain, Taiwan and United
Kingdom
PRODUCT
a) Cook Urological Endoureterotomy Stent Set, 10Fr/13 cm, 6.0Fr/18
cm, sterile; Ref EUSCFS-060018-10-13; GPN REF G18116. Ureteral stent.
b) Cook Urological Endoureterotomy Stent Set, 10Fr/13 cm, 6.0Fr/20 cm, sterile; Ref EUSCFS-060020-10-13; GPN REF G18117. Ureteral stent.
c) Cook Urological Endoureterotomy Stent Set, 10Fr/13 cm, 6.0Fr/22 cm, sterile; Ref EUSCFS-060022-10-13; GPN REF G18118. Ureteral stent.
d) Cook Urological Endoureterotomy Stent Set, 10Fr/13 cm, 6.0Fr/24 cm, sterile; Ref EUSCFS-060024-10-13; GPN REF G18132. Ureteral stent. ;
e) Cook Urological Endoureterotomy Stent Set, 10Fr/13 cm, 6.0Fr/26 cm, sterile; Ref EUSCFS-060026-10-13; GPN REF G18119. Ureteral stent.
f) Cook Urological Endoureterotomy Stent Set, 10Fr/13 cm, 6.0Fr/28 cm, sterile; Ref EUSCFS-060028-10-13; GPN REF G18120. Ureteral stent.
g) Cook Urological Endoureterotomy Stent Set, 10Fr/13 cm, 6.0Fr/30 cm, sterile; Ref EUSCFS-060030-10-13; GPN REF G18121. Ureteral stent.
h) Cook Urological Endoureterotomy Stent Set, 14Fr/13 cm, 7.0Fr/18 cm, sterile; Ref EUSCFS-070018-14-13; GPN REF G18641. Ureteral stent.
i) Cook Urological Endoureterotomy Stent Set, 14Fr/13 cm, 7.0Fr/20 cm, sterile; Ref EUSCFS-070020-14-13; GPN REF G18123. Ureteral stent.
j) Cook Urological Endoureterotomy Stent Set, 14Fr/13 cm, 7.0Fr/22 cm,
sterile; Ref EUSCFS-070022-14-13; GPN REF G18133. Ureteral stent.
k) Cook Urological Endoureterotomy Stent Set, 14Fr/13
cm, 7.0Fr/24 cm, sterile; Ref EUSCFS-070024-14-13; GPN REF G18125. Ureteral
stent.
l) Cook Urological Endoureterotomy Stent Set, 14Fr/13 cm, 7.0Fr/26 cm, sterile; Ref EUSCFS-070026-14-13; GPN REF G18127. Ureteral stent.
m) Cook Urological Endoureterotomy Stent Set, 14Fr/13 cm, 7.0Fr/28 cm, sterile; Ref EUSCFS-070028-14-13; GPN REF G18134. Ureteral stent.
n) Cook Urological Endoureterotomy Stent Set, 14Fr/13 cm, 7.0Fr/30 cm, sterile; Ref EUSCFS-070030-14-13; GPN REF G18130. Ureteral stent.
o) Cook Urological Endoureterotomy Stent Set, 10Fr/13 cm, 6.0Fr/26 cm,
sterile; Ref EUSCFS-060026-10-13; GPN REF G18131. Ureteral stent
CODE
a) Lot number: UF1560100;
b) Lot numbers: UF1565087, UF1401313, and UF1418996;
c) Lot numbers: UF1422329, UF1418997, UF1577105, UF1588937, UF1569283, UF1548852, UF1526279, UF1395329, UF1406376, UF1410869, UF1413150 and UF1385573;
d) Lot numbers: U1528861, U1542982, U1402436, U1361233, U1551529, U1391927, U1364946, U1347423, U1328716, U1416939, U1356437, U1525569, U1391927, U1328716, U1377093 and U1532644;
e) Lot numbers: U1368875, U1532226, U1541632, U1402438, U1356439, U1380409, U1418931, U1538110, U1362570, U1408480, U1411865, U1426173, U1527133, U1357612, U1416045, U1421059, UF1383702, UF1391074, U1426173 and U1525571;
f) Lot numbers: U1358169, U1346509, U1423782, U1342391, U1530194, U1532646, U1354142, U1412596, U1535132 and U1401255;
g) Lot number: UF1424956;
h) Lot number: UF1578018;
i) Lot number: UF1418998;
j) Lot numbers: UF1591697, UF1537682,
UF1562307, UF1420730, UF1404622, UF1402891, UF1531085, UF1526725, UF1396538,
UF1406357, UF1533678, UF1542186, UF1570187, UF1575838, UF1586147, UF1581083,
UF1555888 and UF1389149;
k) Lot numbers: U1396476, U1352478, U1584786, U1342393, U1537634, U1530196, U1347425, U1398599, U1425757, U1391929, U1404166, U1360397, U1584786, U1412598, U1363042, U1391929, U1347425, UF1386555 and UF1387624;
l) Lot numbers: U1415656, U1392361, U1343299, U1416941, U1371126, U1363042, U1409768, U1405040, U1546157, U1414495, U1347427, U1551531, U1571113, U1533048, U1525573, U1535611 and U1409768S;
m) Lot numbers: UF1394511, UF1546568, UF1548011, UF1550302, UF1558381, UF1562308, UF1560070, UF1531086, UF1409854, UF1390546, UF1412673, UF1388726, UF1545756, UF1549848, UF1580601, UF1588492, UF1537146, UF1564616, UF1581084, UF1389605, UF1402894 and UF1404646;
n) Lot number: UF1406358;
o) Lot numbers: UF1537358 and UF1362608
RECALLING
FIRM/MANUFACTURER Vance Products, Inc., Spencer, IN, by letter dated
July 31, 2008.
REASON The bond may break where the two
materials meet, resulting in separation of the stent.
VOLUME OF
PRODUCT IN COMMERCE 1,739
sets
DISTRIBUTION Nationwide, Australia, Canada,
Colombia, Hong Kong, India, Ireland, Israel, Japan, Malaysia, Mexico, South
Korea, Taiwan, Venezuela and Viet
Nam
PRODUCT Integra Ruggles Surgical Instrument ACF Distraction Screws Sterile,
Rx only; Catalog number: R6357A (12 mm), Catalog number: R6397A (16 mm).
CODE
Lot number: 326081; and Lot number:
327081
RECALLING FIRM/MANUFACTURER
Recalling Firm:
Integra LifeSciences Corp., Plainsboro, NJ,
Manufacturer: Instrumed International, Inc., Schaumburg, IL.
REASON
A packaging
error had occurred where 16 mm screws were packaged and labeled as 12 mm screws
and vice versa.
VOLUME OF PRODUCT IN COMMERCE 351 -
12 mm
screws
DISTRIBUTION Nationwide
PRODUCT
a) CombiSet True Flow Series, Hemodialysis Blood Tubing with
Priming Set and Transducer Protectors, with Access Flow Reversing Connector
(Twister) Product Code: 03-2994-6 24 sets / case CombiSet Bloodlines Including
split-septum injection site, viral-retentive transducer protectors, &
priming set & Twister device.
b) CombiSet True Flow Series, Hemodialysis Blood Tubing with Priming Set and
Transducer Protectors, with Access Flow Reversing Connector (Twister) Product
Code: 03-2794-0 Bonded CombiSets - includes priming set w/ needleless access
port & Y-injection site bonded to arterial line, viral-retentive transducers
and Twister device,
CODE
a) Lot
number: 8LR902;
b) Lot numbers: 8ER273, 8HR038, 8JR021, 8KR013, 8LR013,
8ER283, 8HR066, 8JR075, 8KR032, 8LR020, 8HR114, 8JR136, 8KR060, 8LR048, 8HR125,
8JR142, 8KR113, 8LR074, 8HR189, 8JR167, 8KR131, 8LR123, 8HR207, 8JR193, 8KR159,
8LR209, 8HR225, 8JR209, 8KR170, 8LR909, 8HR241, 8JR222, 8KR200, 8HR266, 8JR266,
8KR210, 8HR285, 8KR228, 8KR250, and 8KR271
RECALLING
FIRM/MANUFACTURER Fresenius Medical Care North America, Waltham, MA
REASON Port may crack and
separate resulting in potential blood exposure/blood loss for the
patient.
VOLUME OF PRODUCT IN COMMERCE 351,800
units
DISTRIBUTION Nationwide and
Canada
PRODUCT
a) Zimmer NEXGEN Complete Knee Solution Cruciate Retaining (CR)
Provisional - Articular surface, Size Blue, 14 MM Height, nonsterile; REF
5971-50-14. Used during total knee arthroplasty surgery to determine the
required thickness of the tibial bearing surface to be implanted. Reusable, non-implantable.
b) Zimmer NEXGEN Complete Knee Solution Cruciate Retaining (CR) Provisional -
Articular surface, Size Blue, 12 MM Height, nonsterile; REF 5971-50-12. Used
during total knee arthroplasty surgery to determine the required thickness of
the tibial bearing surface to be implanted. Reusable, non-implantable.
CODE
a) Lot numbers: 60969810 and
60969811;
b) Lot numbers: 61007188, 60987590, 60969807, 60976981 and
60976982
RECALLING FIRM/MANUFACTURER Zimmer Inc., Warsaw,
IN,
REASON Some 12 mm thick
devices may have been etched and packaged as 14 mm thick, and vice
versa.
VOLUME OF PRODUCT IN COMMERCE 131
devices
DISTRIBUTION Nationwide, Canada, Denmark,
Finland, Germany, Poland, Spain, Sweden and United
Kingdom
PRODUCT
a) Expressaire, Disposable Tourniquet Cuff 34", 1 line (18),
sterile; Cuff type A70107010. Each cuff is marked ExpressAire by Sammons Preston
and A70107010. The tourniquet cuff is placed on a patient's thigh during an
operation procedure of the lower leg. The cuff is not operational until it is
attached to an air pump supply that is owned by the care giving facility
(hospital).
b) Expressaire, Disposable Tourniquet Cuff 34", 2 line (18), sterile; Cuff
type A70107011. Each cuff is marked ExpressAire by Sammons Preston and
A70107011. The tourniquet cuff is placed on a patient's thigh during an
operation procedure of the lower leg. The cuff is not operational until it is
attached to an air pump supply that is owned by the care giving facility
(hospital).
CODE
a) Lot numbers:
SO590, SO1205, and SO1300;
b) Lot numbers: SO1106, SO2202, SO2227, and
SO2232
RECALLING FIRM/MANUFACTURER Virtus, Inc.,
Batesville, IN,
REASON The inflation tube may
come loose at the connection to the cuff during use.
VOLUME OF
PRODUCT IN COMMERCE 5,958
units
DISTRIBUTION Nationwide
PRODUCT Somatom Definition AS, Model number: 8098027. Computed tomography
x-ray system.
CODE Serial numbers:
64064, 64065, 64067, 64068, 64071, 64074, 64077, 64078, 64081, 64085, 64087,
64091, 64097, 64103, 64104, 64105, 64111, 64112, 64113, 64114, 64115, 64117,
64118, 64122, 64123, 64138, 64140, 64142, 64145, 64154, 64164, 64166, 64168,
64169, 64170, 64171, 64172, 64177, 64179, 64180, and 64181
RECALLING
FIRM/MANUFACTURER
Recalling Firm: Siemens Medical Solutions USA,
Inc., Malvern, PA,
Manufacturer: Siemens
AG, Medical Solution, Forchheim, Germany.
REASON Possible improper installation of tilt transport
lock/tilt limiter.
VOLUME OF PRODUCT IN COMMERCE 42
units
DISTRIBUTION Nationwide
PRODUCT Obwegeser Ramus Clamp w/ wide shield, 20 cm, 8", Right.
ORAL-MAXILLO-CRANIO-FACIAL SURGERY, Original OBWEGESER Instruments. NOTE: None
distributed domestically. Cranio-maxillofacial manual instrumentation to be used
to aid in the treatment and care of patients, as determined appropriate, by a
trained medical professional. The Obwegeser Ramus Clamps are self-retaining
clamps, used on the ascending ramus of the mandible with an extension on one of
the blades (shields) to aid in retracting the soft tissue of the cheek.
CODE
Lot Number: 050801
RECALLING
FIRM/MANUFACTURER Biomet Microfixation, Inc., Jacksonville, FL,
REASON
The Biomet Microfixation
Obwegeser Ramus Clamp, Wide Right 01-7996 were incorrectly etched as the left
part number 01-7997.
VOLUME OF PRODUCT IN COMMERCE 4
units
DISTRIBUTION Ireland,
Japan
PRODUCT AMS InteXen LP Collagen Dermal Matrix, REF 72404066. Product Size:
8 x 12 cm. AMS InteXen LP is a sterile non-perforated processed porcine collagen
dermal matrix. Sterile EO. The AMS collagen dermal matric is intended for use in
the treatment of hernias where connective tissue has ruptured or for
implantation to reinforce soft tissues where weakness exists in the urological,
gynecological, or gastroenterological anatomy.
CODE Lot/serial numbers: 567398002, 567398003
RECALLING FIRM/MANUFACTURER American Medical Systems,
Inc., Minnetonka, MN
REASON The product size shown on the carton
labeling does not match the size on the product inside.
VOLUME OF
PRODUCT IN COMMERCE 5
units
DISTRIBUTION Nationwide
PRODUCT Bard 3D Max Mesh, Left large , Sterile Pre-formed Knitted
Polypropylene Product Code: 0151531. Bard 3D Max Mesh is an anatomically shaped
pre-formed polypropylene mesh indicated for use in the repair of hernias and
chest wall defects. It is designated in two orientations - "Right" and "Left"
for use to the right or left side of the human body. There is a blue
monofilament Medial marker "M" and a small arrow stitched into each orientation
to designate the right or left medial edge of the product.
CODE Lot number: 43JQD016
RECALLING
FIRM/MANUFACTURER
Recalling Firm: Davol, Inc., Sub. C. R. Bard,
Inc., Warwick, RI,
Manufacturer: Bard
Shannon Limited, Humacao, PR.
REASON Mislabeled; Product labeled
as left large, may contain a Right Large.
VOLUME OF PRODUCT IN
COMMERCE 232
units
DISTRIBUTION Nationwide
WEEK ENDING FEBRUARY 14
PRODUCT Medtronic O-Arm Intraoperative Imaging System, Catalog number:
Bl-700-0027, CODE Software versions
prior to version 3.0.2 S/N: 102-155
RECALLING
FIRM/MANUFACTURER
Recalling Firm: Medtronic Navigation, Inc.,
Louisville, CO,
Manufacturer: Medtronic
Navigation, Inc., Littleton, MA.
REASON Navigation accuracy problem: Accuracy
may fall outside of specification when used in combination with the Medtronic
Synergy Experience StealthStation System Spine Software with the O-Arm
Intraoperative Imaging System and tracker.
VOLUME OF PRODUCT IN
COMMERCE
54 units
DISTRIBUTION Nationwide,
Canada, Czech Republic, Denmark, Germany, Netherlands, South Korea, Poland,
South Africa, and
Singapore
PRODUCT Excelsior SL-10
Microcatheter, placed in sterile pouch. The
ExcelsiorTM SL-10 Microcatheter is intended to assist in the delivery of
diagnostic agents, such as contrast media, and therapeutic agents, such as
occlusion coils, into the peripheral, coronary, and neurovasculature; a) Model
Number M0031681890, Catalog Number 168189, b) Model Number, M0031681810, Catalog
Number 168181.
CODE
a) Domestic Lot/Batch Numbers: 9410580 11094006 9182538 9410580 9797825
11002715 11279121 9797825 11112088 9549061 9761925 9812892 11037582 11022392
11103887 11055248 9468202 11279121 9489075 9558487 11022392 11037582 9567011
9539686 11055248 11280366 9360513 11103722 11037582 11103887 9474771 9563837
9549061 9786599 8559155 9407572 9403111 9466592 9567133 9403111 9410580 9463008
9474934 9805069 9797825 9805069 11022392 11103722 11051431 11103889 11016347
11021069 8559155 9277244 9407574 9323623 9466592 9497150 9407574 9410580 9420194
9539686 9805069 9761923 9797825 11025202 11094006 11021070 9811899 11086559
11279696 11085511 11086559 9360513 11137647 11085284 9277244 9558487 11037582
9786599 9800091 11097465 9497151 9800091 11103886 11055248 9549849 9549062
9813014 9702841 9797824 9797726 11085282 11103886 11002715 11129185 11065054
11112088 9539686 9786599 11103887 11085282 11093861 9474771 9497151 9796878
11023381 11095396 11055248 9471043 11023381 9786599 11034065 11120294 9360513
9471043 9471044 9565700 9540909 9805069 9812499 11002716 9761925 11103722
11085282 11037582 9811899 11065054 11095396 11276374 11110298 11152045 11107242
9800091 9797825 11021072 9797825 11085284 11103886 11279696 11280366 11085511
9796878 9414130 9439020 9468202 9549062 11002715 11279696 9439020 9335697
8559155 9323623 9335694 9323623 9328893 9323623 9332195 9414130 9403111 9360513
9335694 9417932 9317444 9463008 9407572 9439020 9427379 9407574 9471043 9418154
9567011 9563838 9521760 11092289 11107242 9414130 9466590 9427378 9497151
11103722 11065056 11153871 9497151 11103889 8559155 9466592 9549060 11016347
8563856 9466592 9474771 9761925 9778433 9797824 11037582 11280366 11103889
9563837 9796878 11021069 11021072 11108176 11283875 11110298 11146441 9549062
9323623 9558487 9761925 11150467 11034065 9317444 9427379 9812497 9562124
9812497 11110298 9815767 8559155 9471044 9407574 11085511 9761923 11120293
9468202 11002715 11037582 9182538 9497151 9786599 11021070 11071036 11137647
11095396 11129185 11023381 11073526 11137647 11146238 11021069 9815767 9317444
9815768 11036567 8563856 9277244 9357310 9414130 9471044 9474934 9439931 9471044
9805069 9549061 9812497 9797824 11025202 11002430 11051431 11108176 9800091
11071037 9563838 11065054 9805069 9357310 9414130 9427379 9549060 9549061
11112088 9332195 9414130 9427379 9471043 9466591 9539686 9562126 9549849 9567133
9427379 9332195 11021072 11021069 11025202 11021069 11110298 11107242 9474934
11120294 9328893 9277244 9796878 9471043 9417932 9797825 9418154 9471047 9468202
9407574 9474771 9435037 9565700 9812497 9797824 9812891 9813014 11055248 9811899
11086780 11153871 9328893 9414130 11103722 9407572 9407574 9540909 9335697
9805069 11037582 11093861 11088219 11280366 9562124 11103889 11036567 11129185
11152045 11054197 11280366 9797824 11016347 11055248 9414130 9328893 9812361
9539334 9407572 9489075 9496078 9463008 9417932 9567133 9427379 9463008 9427379
9556944 9439020 9463008 9815767 9702841 9797824 11023381 9812891 9817640
11021072 11022392 11103886 11085284 11088219 11129185 11085511 11153871 11085284
9811899 11071037 11023381 11153871 11279121 11021072 9549849 9182538 9417932
9815768 9812891 9796878 9817640 11036567 9812499 11065054 11085511 11086780
9357310 9407574 9277244 9474935 9497150 9471044 11094006 11097465 11225675
8559155 9497151 11103722 9521763 9420194 9471043 9277244 9328893 9277244 9317444
9403111 9539686 11103889 9800091 11094006 11088219 9317444 9439020 9800091
11103886 11134246 11283875 9489075 9474935 9815768 9797824 9817640 11103887
11103722 11021070 11093860 9815767 9407574 11120293 9417931 9468202 9420194
9521763 9471044 9815767 9812891 11016346 11051431 9815768 9812361 9182538
9439020 9526083 9563837 11085282 9474771 9466591 9797824 9812892 9815768
11021069 11103887 9797825 11021069 11103722 9797824 11036567 11103886 11276374
11086559 11120293 9562126 9489075 9796877 9797824 11022392 11051431 11107242
9549062 9812497 11150468 11021072 11025202 11103722 11279121 9549062 11023381
11002430 9474771 9471043 9567133 9563837 9815767 9562124 9471043 11034065
11073526 9800091 11016346 9811899 9439020 9468202 9817640 11016346 11279121
11120293 9786599 11023381 11021072 11002430 9317444 9182538 9317444 9407572
9474771 9335697 9563837 9815767 11279121 11103887 9466592 9539334 11121598
9549062 9521763 9317444 9562124 9182538 9562126 9466592 11085284 9435037 9335697
8559155 9407574 9414130 9439020 9407572 9471043 9414130 9549060 9797825 11021069
11279121 11055248 11092289 11150468 9563837 9420194 9796878 11103887 9539686
9796878 9817388 9812891 11085284 11085282 11016347 11085511 9474934 9427379
11088219 9332195 9761923 9489075 9702841 9496078 11103722 11150468 9403111
9328893 9567011 9796878 9489075 11103886 9332195 9435037 9317444 9360513 9474771
9815768 8563856 9563838 9549062 11034065 11025202 11073526 9797824 11016347
11103887 11107242 9463008 9317444 9497151 9549061 9471044 9702841 11002715
11103886 11021069 11093861 11103886 8563856 9812497 9815767 11085282 11073526
9466590 9817640 11021069 9812361 11021070 8563856 9332195 9328893 9407574
9439020 9497150 9466591 9407574 9474771 9463008 9439929 9439020 9489075 9439929
9558487 9549061 9567133 9526083 9539688 9563837 9558486 9563837 9549060 9335697
9565700 11036567 9812891 9786599 9812891 11021069 11103889 11016346 11022392
9796878 11022392 11037582 11034065 11182789 11021069 11085282 11016346 11086559
11280366 11110298 11055248 11150468 11121598 11093860 11092289 11112088 11073526
11022392 9815768 11150467 9439020 9549061 9549062 9489075 9797824 11025202
11225675 9526083 9800091 9778171 11037582 9813014 11103722 9796878 11150468
11107242 9521760 9815767 11129185 11088219 9796878 9812497 9761925 11002430
8563856 9539334 9549062 9779107 11065056 11086559 9497150 9751645 9735635
9817640 11021070 11225676 9797825 11055248 9439020 9567133 9563837 9817640
11103887 11150468 11134246 9812361 11153871 11121598 9407572 9567133 9463008
9357310 9420194 9549062 9556943 9556944 9812891 11021069 11085511 11065054
11153871 11065054 11112088 9182538 9332195 9439020 9563837 9420194 9796878
9549060 9817640 9812891 11002430 9497152 11016346 9797824 11085511 11021069
11023381 8559155 9539686 9562126 9796877 9817640 11097465 11023381 9779107
11095396 11120293 11121598 11103887 9414130 11085284 11088219 9815767 9797726
11016346 8563856 9439020 9435037 9521763 9439020 9497150 9439020 9471043 9562126
9466590 9471044 9797824 9754894 9812892 9805069 11103889 9800091 11002716
11129938 9797825 9812497 11153871 11021072 11279696 9761923 11279121 9332195
9328893 9439932 9468202 9540909 9420194 9435037 9565700 9563837 9797726 11021069
11037582 11103722 11097465 9562126 9812892 11055248 11129185 11121598 9471043
9323623 9489075 9549062 9562126 9562124 11002716 11108176 11023381 11022392
8563856 8559155 9332195 9357310 9407574 9407572 9407574 9407572 9407574 9357310
9489075 9439020 9466592 9471044 9471043 9497150 9526083 9420194 9471044 9497151
9565700 9521760 9567133 9526083 9558487 9539686 9805069 9796878 9797824 9815768
9812891 9815767 9558487 9761925 9815767 11002715 9761925 9815768 9735635 9797824
11002430 9803722 11093861 9817388 11085282 11103722 11103889 9797824 11108176
11016347 11055248 11103722 9811899 11086780 11279121 11110298 11120293 11071036
11146238 11121598 11110298 11021072 9466590 11103887 8563856 9521763 9558487
11086780 9435037 9471044 9567133 9786599 11085282 9797824 11065054 8563856
9323623 9317444 9277244 9407574 9323623 9410580 9357310 9463008 9328893 9435037
9466590 9463008 9471047 9474771 9474934 9418154 9549061 9539686 9761923 9812497
9817640 9761925 9797825 9761925 9817640 11036567 9797726 9817640 11093861
11021069 9786599 11021070 11103887 11093861 11103886 9812497 11022392 11055248
11279121 9811899 11065056 11279121 11095396 11225675 11055248 11110298 11279121
11093860 11148046 11073526 11112088 9332195 9360513 9407572 9567011 11036567
11108176 11085282 11093861 11097465 11280366 9539686 11107242 11071037 9466591
9468202 9414130 9471044 11103887 9761925 11023381 9418154 11146441 11086559
8559155 9417932 9497151 9466592 9471043 9357310 9562124 9549061 9815768 11002716
9796878 11103887 11016347 11276374 11103886 11085284 11055248 11085282 9332195
9562126 9797824 11021072 9335697 9565700 9521760 11051431 11112088 11093861
9549061 9332195 9410580 11071037 11036567 9466592 8559155 9419872 9549062
8559155 9540908 9565700 9761925 9702841 9805069 11182789 9471047 9471044 9567011
11022392 11025202 11071036 11002430 11021069 9414130 9410580 9463008 9471043
9558487 9466592 9812891 9815767 11025202 9761923 9761925 9567133 9489075 9539686
11120294 9435037 9539686 11051431 9471047 9407574 9815768 8563856 9182538
9323623 9414130 9357310 9414130 9417932 9407574 9471043 9489075 9549062 9526083
9549062 9563837 9463008 9562124 9549849 9805069 11103887 11088219 9812361
11279696 11085511 11279121 11055248 11071037 11112088 11071037 9812892 8559155
9323623 9414130 9410580 9468202 9410580 9497151 9407574 9435037 11025202
11103889 11065054 11283714 9407574 9468202 9549060 9761923 9796878 11037582
11036567 11055248 11085511 11086559 11153871 9567133 9497151 9565700 9761923
9812891 9800091 9796878 11112088 9471043 11036567 9489075 11280366 9435037
9471043 9407574 9417932 11103886 9332195 9407574 9489075 9471043 9539686 9549060
9497150 9565700 9317444 9565700 9761923 9761925 11016346 9761925 11002715
11021072 11093860 11120294 11121598 9466591 9328893 9463008 9549061 9497150
9562126 9549061 9328893 9796878 11085282 11276374 9797824 11071036 9468202
9439020 9549062 11279696 11153871 11071037 11088219 11121598 9417932 11021070
9811899 9497151 9539686 9182538 9567011 11279121 9811899 8559155 9407572 9702841
11225675 9474771 9797825 9360513 9407574 11002430 9360513 9435037 9471044
9489075 11144685 11085284 9805069 9360513 9417931 11088219 9439931 9563838
9417932 9419872 9549062 11280366 11103884 9812891 11095396 11112088 9323623
11016347 11279121 11092289 9407572 9562126 11085282 11280366 11276374 9332195
9418154 9407572 9539334 9797824 11283875 9549060 11025202 9357310 9471044
9813014 9496078 11088219 9410580 9474934 9521763 9800091 9549849 9563838 9817640
11022392 11225675 11279121 11153871 9418154 9417932 11225675 11225676 9435037
9558486 9761925 9471043 9567133 9466591 9407572 9539686 9463001 11103886 9563837
9496078 9797825 11055248 9328893 9360513 9549060 9474934 9796878 9549060 9786599
9797825 11002430 11023381 11112088 11107242 9812497 9805069 9420194 9786599
11094006 11095396 9497151 11092289 11034065 11120293 9417932 9463001 11088219
9489075 9539686 9474771 9549060 9813014 9549060 11103889 11021072 11054197
9761925 9786599 11055248 11225676 11085511 11110298 11085284 11095396 9497150
9332195 11021070 9562124 9566693 11097465 11103886 11021069 9328893 9407572
9539334 11036567 11025202 11022392 11103886 11182789 11276374 11036567 11146441
8563856 9435037 9417932 9466590 9497151 9471043 9565700 9556944 9420194 11283714
11279696 11283714 11146441 11065054 11148046 9562124 9702841 11182789 11103886
9435037 9800091 11088219 9812497 9497150 11103889 11112088 9317444 9471043
9466590 9526083 9497151 11022392 9335697 9556943 9563838 9539334 9549061 9761925
11088219 11023381 11022392 11150468 9761925 11055248 11095396 11103884 11152045
11112088 11085284 9817640 9317444 9328893 9466590 8563856 8559155 9463008
9410580 9471044 9549849 9466591 9567011 9539334 9407572 9549849 9335697 9549849
9761923 9812497 9563838 11036567 11108176 11103889 11085282 11085284 11132217
11085282 11095396 11146441 11110298 11150467 11110298 11137647 11086559 11152045
9812891 9357310 9418153 9435037 9566519 9805069 11025202 11036567 11085284
8563856 9420194 9567133 9563837 9813014 9549849 9796878 11085282 11283714
9813014 9761925 8563856 9277244 9357310 9439020 9418154 9471044 9556944 9539334
9549060 9562124 9556944 9786599 11002430 9755342 9761924 9813014 11034065
9800091 11016346 11108176 9761923 9800091 11021069 11085282 11103887 9563837
11134246 11146441 11153871 11110298 11085511 11146441 11103884 11153871 11225675
11120293 11279121 11073526 11107242 9563837 11002430 9811899 9805069 9466590
11002715 9466590 9182538 9474771 9539334 9796878 9521764 11016346 11002715
11065056 11153871 11150467 11112088 9435037 9417932 9812891 9812497 11103722
11054197 11280366 11055248 11146238 9563837 11034065 9812361 11055248 9563837
9526083 11108176 11025202 11002715 11023381 11276374 11135840 11279121 9417932
11073526 9796878 9539688 11002715 9471043 9466590 9540909 9277244 9549061
9813014 11283875 9521760 9471047 9549062 9812892 9332195 9414130 9323623 9761923
9761925 9702841 9796877 11110298 11279696 11146238 11093860 9414130 9419872
9526083 9539334 9558486 9526083 9815768 9815767 9497151 9805069 9797825 11085284
11279696 11280366 11093861 11021072 11016346 11279696 9277244 11022392 9539334
9786599 9521760 11279696 9468202 9563837 11021069 11093860 9526083 9277244
11279121 8559155 11021069 11037582 9463008 9496078 9521760 9817388 9539334
9563837 9797825 9357310 9420194 8559155 9474771 9471043 9539334 11093861 9417932
9427379 9549061 9549060 9474771 9539688 9797825 11021072 11086559 11129185
9427379 9702841 11120294 9567133 9539688 9563837 9805069 11086559 11137647
9466590 9471043 11103889 9420194 9812892 11016346 11153871 11280366 9471043
9474934 9427379 9812497 9778171 11103889 11108176 11103889 11085282 11025202
11065054 11065056 9357310 9496078 9466591 9420194 9558487 9471043 11103722
11279121 11146238 11280366 11037582 11071036 11071037 9815768 11134246 9466591
11021072 11085284 9466592 9497150 8563856 9439929 9466592 9521763 9549849
9427379 9489075 9558487 9800091 11051431 9427379 11103886 11071037 11283875
11110298 11002430 9410580 9496078 9489075 9474771 9539334 9420194 9539334
9466590 9805069 9812499 11108176 11225675 11112088 9332195 9439020 9549060
9471044 11022392 11036567 11022392 11002715 9570584 11280366 9418154 11121598
9435037 11086559 11022392 11055248 11150467 11002430 11093861 9403111 9360513
9466592 9419872 9439929 9549060 9466590 9420194 9521760 9817640 11025202
11034065 11021069 11085282 11021069 11085282 11021069 11016346 11002715 11002716
11065054 11134246 11065056 11279696 11065056 11280366 11086559 11283875 11085511
9335694 9323623 9558486 9562126 9812892 11095396 11283875 11112088 11073526
9539686 9565700 9418154 9539334 9471044 9435037 11002430 9427377 11002715
9812497 9813014 9817640 11025202 9797824 11034065 11103889 11025202 11021070
11103889 11036567 11103722 11036567 11108176 11021070 11108176 11054197 11146441
11095396 11065056 11023380 11112088 11092289 9786599 11112088 9317444 11021072
11023381 8563856 9328893 9410580 9328893 9521763 9549060 9539334 9474935 9558486
9812497 9786599 11002430 11094006 9796878 11110298 9811899 9463001 11129185
11120294 11107242 11112088 9489075 11279696 11110298 11085511 11152045 8563856
9427379 9420194 9797726 11051431 11103889 9332195 9497151 9474771 9549060
9566693 11134246 9812497 9407572 9436977 9539334 9549062 9439931 11276374
11153871 9761923 9812497 11002430 8559155 9786599 9815768 9797824 9417932
9800091 11103886 9812892 9800091 9410580 9562126 9565700 11023381 11283714
11132217 11134246 9800091 9323623 8559155 9474934 9702841 9817640 11280366
9761923 9497151 9539334 9474935 9812497 9182538 9805069 9796878 9797825 11086780
11280366 11095396 11146238 11093860 11065054 9562124 9812891 11002430 11103886
9812497 11146238 9328893 11279121 9549849 11103889 11065054 11036567 9797824
11108176 9468202 9489075 9812891 9813014 11110298 9407572 9562124 9549060
11103722 11103889 11088219 9563837 9328893 9360513 9328893 9558487 11036567
9319889 9407574 11225675 11093860 9815767 11085282 9761924 9474771 9410580
9471044 9570584 9805069 11051431 11095396 9328893 11002430 9414130 9549062
9335697 9357310 11051431 9562124 9796878 9815768 11086559 8559155 9562124
9815768 9761923 11036567 11103722 11034065 11002716 9761925 11097465 9474771
9562124 9786599 9812892 11016346 11280366 9360513 9317444 9319889 9407572
9417932 9539334 9521760 9417932 9565700 9562126 9489075 9566693 9812891 9786599
9796878 9813014 11016346 9812891 9812892 9800091 11034065 11002430 11051431
11025202 11036567 11034065 9786599 11034065 11023381 11153871 11086780 11153871
11134247 11110298 11021070 9471043 9468202 9357310 9567133 9549060 9549062
9778171 11103886 11279121 11065054 11095396 8559155 8563856 9182538 9407572
9403111 9474771 9471044 9474771 9471044 9521760 9466590 9474934 9521763 9526083
9549062 9526083 9521760 9439020 9562126 9566693 9565700 9812892 9497152 9812497
9761925 9815767 9761923 9815767 9761925 9812499 9817388 11021070 11023381
9800091 11085284 11085282 11085284 11103889 11103886 11132217 11135840 11129185
9812361 11016347 11023381 11065056 11146441 11103722 11021072 11085284 9418154
9566519 11108176 9566519 11071037 11073526 11146238 11283875 9556943 9779107
9558487 11225676 9417932 8559155 9539688 9761923 9360513 11146238 11093860
9328893 11021072 9328893 9567011 9567133 9539686 9539688 9796878 11002715
11108176 11103887 11085282 9817640 11097465 11065056 11283714 11085511 11036567
9360513 9471043 9539686 9761923 9815767 9797726 11150468 9812361 8563856 9182538
8563856 9812497 11021072 8563856 9489075 9471044 9489075 9566519 9556944 9558487
9779107 9800091 11065054 9815767 9403111 11085511 9410580 9407572 11065056
9549849 9796877 11002715 9328893 9403111 9407574 11134246 11051431 11002430
11023381 9549062 11021072 11225676 9562126 9357310 9403111 8563856 9418154
9463008 9357310 9417931 9427379 9521763 9471043 9335697 9521760 9549062 9570584
9567011 9539686 9567011 9796878 9800091 9817640 11225676 11002716 11021070
9526083 11025202 11022392 11071037 11093860 9563838 8563856 9539686 9407572
9410580 9796878 9468202 9497150 9812891 11022392 9813014 11129185 11120294
11225675 9567133 9335694 11021070 11071036 9466590 9735635 9800091 11021072
11021070 11021069 11037582 11279121 9277244 9360513 9357310 9417932 9497151
9418154 9410580 9323623 9466590 9474935 9539334 9549062 9565700 9332195 9403111
9558487 9565700 9562124 9797726 9761923 11021069 11025202 11103722 11103887
11103889 11022392 11103886 11103887 11086559 11283714 11283875 9797825 11086780
11095396 11092289 11016347 11037582 8559155 9466590 9539334 9549849 9563837
9563838 9418154 9562126 9563837 9761925 11021069 9778433 9812361 11071036
11016347 9562124 11182789 11146441 9549062 11016346 11225675 11134246 9786599
9549062 11085282 9410580 9466590 11095396 11182789 11103886 9417932 9466592
11025202 8559155 9357310 9497150 11025202 9786599 11065054 9435037 9562126
8559155 9800091 9761925 11097465 9332195 9549849 9549060 9539334 9471044 9549060
9474771 9466590 9420194 9549062 9566693 9539334 9521760 9558487 9815768 9817640
9796878 11002430 11025202 11103887 11085284 11103886 11037582 9797825 11134246
9812891 9800091 11065056 11283714 11055248 11085511 11071036 9540909 9800091
11025202 9420194 9521760 9567133 9778171 11016346 9558486 9562126 9812497
11073526 11021069 9489075 9526083 9549060 11016346 9328893 9521760 11137647
11016346 9410580 11002715 9815768 11002430 9407574 9563838 9317444 9474771
9786599 11021072 9332195 9468202 9463008 9761923 11103889 11022392 11095396
11225675 9496077 11023380 9796877 9277244 9319889 9562124 9797726 11021070
11065056 11034065 9466590 9526083 9558487 9812497 9812891 9562126 11036567
11108176 11150468 11280366 9407572 9403111 9407572 9563837 9526083 9702841
9786599 9702841 9812892 9812497 9797825 9817640 9812891 11055248 11095396
11085511 11095396 11153871 11283875 9797824 8563856 9474771 9435037 11065054
9328893 9410580 9805069 9357310 11097465 9817388 11022392 11134246 9563838
11036567 11137647 9563837 11051431 11103722 11051431 11065054 11103722 11283875
11120294 9466590 9466591 9778171 11023381 11153871 9565700 11108176 11150468
9812361 11016347 11120294 9463008 9410580 9414130 9735635 9332195 9414130
9549060 9414130 9497150 9549060 9521760 9815768 11108176 9778433 11103887
9786599 11146238 9328893 11023381 9539334 9474771 9566693 11021072 9812891
11103722 9539686 11276374 9812361 11153871 11097465 9786599 11103884 11110298
9562124 11283714 9761923 11071036 11034065 9439020 11055248 11034065 11103886
11086559 11094006 11085282 11025202 11088219 11225675 9317444 9797825 11002430
11002715 11153871 11023381 11134246 9497151 9811899 11034065 11088219 11051431
9549060 9332195 11065056 11279696 11150468 11120293 11110298 11279696 11085284
9781935 11146238 11093860 9539334 11153871 11103889 11103887 9468202 11279696
11148046 9817640 9468202 9497151 11146238 9565700 11073526 11129185 9410580
9497151 9466592 9466590 11152045 11121598 11021072 11103722 11085284 11002715
11085282 9463008 11095396 11085282 11279696 9468202 11150468 9328893 9521760
11103722 9797726 0007987352 0007987916 0007991670 0007991671 0008351757
0008359860 0008410983 0008410984 0008410985 0008414795 0008414796 0008423679
0008431471 0008431472 0008436777 0008436953 0008449703 0008449704 0008514417
0008514418 0008514987 0008514988 0008517577 0008517578 0008517775 0008517776
0008521898 0008521899 0008539759 0008547983 0008547984 0008548255 0008548256
0008555957 0008555958 0008556864 0008556865 0008559151 0008559152 0008559155
0008563856 0008568095 0008568096 0008568097 0008571447 0008571448 0008572916
0008572917 0008575546 0008576798 0008577472 0008577473 0008579039 0008579040
0008583924 0008585988 0008586369 0008590090 0008590091 0008592018 0008646347
0008646348 0008648851 0008648852 0008650135 0008650137 0008650138 0008652528
0008652529 0008652530 0008654201 0008654202 0008656704 0008656705 0008656706
0008657634 0008657635 0008666368 0008668334 0008668335 0008671327 0008671328
0008671329 0008671330 0008675642 0008675644 0008685131 0008685133 0008685142
0008688905 0008688906 0008688907 0008693602 0008696260 0008696268 0008697171
0008721135 0008721136 0008721137 0008727528 0008735644 0008736896 0008737561
0008737562 0008739499 0008739747 0008739749 0008740563 0008740996 0008744006
0008744007 0008744702 0008744704 0008771765 0008775227 0008775229 0008775399
0008775400 0008779375 0008779381 0008793105 0008793107 0008793634 0008793635
0008797093 0008805781 0008805782 0008810294 0008810295 0008814696 0008820279
0008820280 0008822815 0008844202 0008844422 0008844424 0008847910 0008848697
0008848699 0008849981 0008849984 0008852365 0008853186 0008853734 0008853735
0008885119 0008885120 0008901591 0008905213 0008906045 0008907532 0008907533
0008943816 0008943817 0008952093 0008952094 0008956682 0008956683 0009008410
0009016029 0009019389 0009027788 0009027789 0009031429 0009032269 0009032270
0009036286 0009038891 0009038992 0009038993 0009040224 0009040305 0009040306
0009040797 0009042595 0009042597 0009043272 0009043273 0009049033 0009052986
0009052987 0009052989 0009052990 0009078216 0009079855 0009081985 0009179388
0009179492 0009179940 0009179941 0009182538 0009182539 0009182545 0009185028
0009186996 0009190032 0009213135 0009213136 0009213138 0009215767 0009216185
0009216186 0009222993 0009222994 0009222995 0009268462 0009268463 0009268465
0009268469 0009271988 0009271989 0009272469 0009272470 0009275288 0009275289
0009275290 0009275291 0009275653 0009275654 0009277242 0009277243 0009277244
0009283044 0009287357 0009287358 0009287359 0009291362 0009308800 0009308801
0009308802 0009309007 0009309008 0009309009 0009317441 0009317442 0009317443
0009317444 0009319888 0009319889 0009319890 0009319891 0009323622 0009323623
0009323624 0009325403 0009325404 0009328672 0009328893 0009328894 0009328895
0009328896 0009330959 0009332193 0009332194 0009332195 0009332196 0009335694
0009335695 0009335696 0009335697 0009340011 0009340232 0009340233 0009340234
0009357310 0009357416 0009358449 0009360513 0009360514 0009360515 0009360516
0009403061 0009403109 0009403110 0009403111 0009403192 0009403193 0009403194
0009403195 0009406697 0009407572 0009407574 0009407575 0009407576 0009410580
0009410581 0009410582 0009410583 0009414130 0009414131 0009417931 0009417932
0009418153 0009418154 0009419871 0009419872 0009420193 0009420194 0009427376
0009427377 0009427378 0009427379 0009432062 0009432063 0009432064 0009435034
0009435035 0009435036 0009435037 0009436976 0009436977 0009439020 0009439929
0009439930 0009439931 0009439932 0009462997 0009463001 0009463008 0009466589
0009466590 0009466591 0009466592 0009468202 0009468203 0009471043 0009471044
0009471047 0009471275 0009474771 0009474933 0009474934 0009474935 0009482771
0009489075 0009489076 0009496076 0009496077 0009496078 0009496079 0009497149
0009497150 0009497151 0009497152 0009509403 0009509406 0009509407 0009509408
0009511142 0009511581 0009512748 0009517073 0009517075 0009517160 0009517161
0009521760 0009521762 0009521763 0009521764 0009526081 0009526082 0009526083
0009526084 0009532847 0009532848 0009532849 0009532850 0009535472 0009535534
0009535536 0009535537 0009539334 0009539686 0009539688 0009539689 0009540908
0009540909 0009549060 0009549061 0009549062 0009549849 0009556943 0009556944
0009558486 0009558487 0009560542 0009562124 0009562126 0009563836 0009563837
0009563838 0009565700 0009566518 0009566519 0009566693 0009567011 0009567133
0009570217 0009570584 0009574151 0009574152 0009574428 0009574713 0009574715
0009574716 0009574718 0009575864 0009575956 0009575958 0009578709 0009578710
0009579246 0009582304 0009582305 0009582306 0009582724 0009588125 0009589654
0009589655 0009590098 0009590099 0009592091 0009592563 0009592564 0009592842
0009594087 0009594088 0009594089 0009597489 0009597490 0009597491 0009597492
0009600389 0009600783 0009602726 0009602727 0009602803 0009602804 0009602805
0009603577 0009603578 0009603580 0009603582 0009613100 0009613102 0009614225
0009614226 0009630783 0009630784 0009631316 0009631317 0009632762 0009632764
0009634171 0009634253 0009637291 0009637493 0009641670 0009641810 0009641811
0009650580 0009650582 0009652573 0009653193 0009660925 0009663735 0009663736
0009663894 0009664488 0009664489 0009667021 0009667022 0009667881 0009677236
0009678148 0009678149 0009678878 0009678879 0009681028 0009681029 0009681030
0009681736 0009682622 0009682623 0009682624 0009688495 0009688496 0009688968
0009688969 0009688970 0009702841 0009702842 0009703870 0009705192 0009705313
0009705673 0009706354 0009706355 0009706356 0009706357 0009731020 0009731021
0009731022 0009733186 0009733192 0009735635 0009735636 0009735637 0009736802
0009739199 0009748080 0009748081 0009748083 0009750526 0009751645 0009751646
0009751649 0009751960 0009751961 0009754893 0009754894 0009755340 0009755341
0009755342 0009761923 0009761924 0009761925 0009762076 0009776913 0009776914
0009778171 0009778172 0009778433 0009779106 0009779107 0009781917 0009781934
0009781935 0009783017 0009786599 0009786601 0009786602 0009789334 0009796876
0009796877 0009796878 0009797040 0009797047 0009797726 0009797727 0009797824
0009797825 0009800091 0009801243 0009801818 0009801819 0009801947 0009801948
0009802314 0009803721 0009803722 0009803723 0009805069 0009811899 0009811900
0009812357 0009812358 0009812361 0009812497 0009812499 0009812891 0009812892
0009813013 0009813014 0009815767 0009815768 0009816961 0009817387 0009817388
0009817640 0009817641 0009831888 0009832576 0009832577 0009832578 0011002430
0011002715 0011002716 0011003706 0011016241 0011016346 0011016347 0011016348
0011016349 0011016350 0011019058 0011019482 0011021069 0011021070 0011021072
0011022392 0011022393 0011022884 0011022885 0011022886 0011023380 0011023381
0011025202 0011025203 0011025981 0011026366 0011033353 0011033354 0011033356
0011033496 0011033605 0011034065 0011035798 0011035799 0011035980 0011036565
0011036566 0011036567 0011037581 0011037582 0011037584 0011037843 0011037844
0011040092 0011040093 0011040094 0011040876 0011040877 0011040878 0011044159
0011048148 0011048149 0011048150 0011050778 0011050779 0011051040 0011051425
0011051428 0011051430 0011051431 0011051432 0011051433 0011054195 0011054196
0011054197 0011055247 0011055248 0011055249 0011056350 0011056351 0011056353
0011063224 0011063339 0011063360 0011065054 0011065055 0011065056 0011065438
0011066818 0011069902 0011070049 0011070703 0011070704 0011071036 0011071037
0011073525 0011073526 0011073527 0011085282 0011085283 0011085284 0011085511
0011085512 0011086559 0011086780 0011086781 0011088219 0011088300 0011088301
0011092288 0011092289 0011092290 0011093860 0011093861 0011094006 0011094007
0011094008 0011095396 0011095397 0011097463 0011097464 0011097465 0011102222
0011103465 0011103466 0011103467 0011103721 0011103722 0011103884 0011103885
0011103886 0011103887 0011103888 0011103889 0011107241 0011107242 0011108175
0011108176 0011110298 0011110299 0011110381 0011112088 0011120293 0011120294
0011121597 0011121598 0011121599 0011122286 0011124802 0011124803 0011124804
0011129184 0011129185 0011129186 0011129936 0011129937 0011129938 0011130709
0011130710 0011132215 0011132216 0011132217 0011134246 0011134247 0011134253
0011134255 0011135840 0011135841 0011135843 0011136694 0011136695 0011137647
0011144684 0011144685 0011146238 0011146239 0011146441 0011148045 0011148046
0011148047 0011150467 0011150468 0011152044 0011152045 0011152046 0011153870
0011153871 0011153872 0011155179 0011155417 0011155418 0011157288 0011157293
0011174734 0011174882 0011176930 0011176931 0011176932 0011178957 0011178958
0011178959 0011181607 0011182784 0011182789 0011182955 0011182956 0011182957
0011186666 0011186667 0011188240 0011188241 0011188703 0011189571 0011189572
0011189574 0011191803 0011191804 0011192228 0011192229 0011193285 0011193286
0011193287 0011193288 0011194029 0011194030 0011194032 0011194034 0011194035
0011207811 0011207813 0011207814 0011208035 0011208036 0011208037 0011215043
0011215046 0011215047 0011217049 0011217050 0011217365 0011218781 0011218783
0011218784 0011219157 0011219462 0011220086 0011220386 0011220389 0011220992
0011222189 0011222190 0011222191 0011223588 0011223589 0011224118 0011224346
0011224347 0011224348 0011225675 0011225676 0011232315 0011232316 0011232317
0011233199 0011233360 0011234142 0011235389 0011235390 0011236219 0011251410
0011253891 0011254351 0011254972 0011255340 0011257073 0011272806 0011273163
0011273601 0011273602 0011273616 0011275535 0011275977 0011276374 0011277316
0011277317 0011279121 0011279693 0011279696 0011280366 0011280585 0011280586
0011281146 0011282377 0011282548 0011282994 0011283713 0011283714 and
0011283875;
b) Lot/Batch Numbers: 9703869 9582725 11032901 9563839 9703869
11017848 11018893 9582725 11017847 11018893 11067023 11032901 9800280 9750583
9446914 9637671 11257964 9703869 11017847 9499078 9750583 11067023 9750583
9827896 9499078 11018895 11017848 9499078 11019057 9827896 9582725 11007160
9446914 9703869 9750582 9582725 11018893 9827896 9446914 11019057 9499078
11017847 9499078 9800280 9446914 9499078 9582725 9446914 9582725 9827896
11017848 9827896 11017847 9637671 11018895 9827896 11017847 9563839 11007160
9800280 9563839 11018893 11067023 11032901 11017847 11019057 11032901 9703869
9499078 9637671 9582725 9750583 9582725 9637671 9446914 11019057 9827896 9563839
11018893 11032901 9750582 9827896 9499078 11018893 9563839 11019057 11067023
9750583 9703869 11032902 9582725 9827896 11019057 11018895 9582725 9563839
9750583 9800280 11018893 11067023 11017848 11018893 9446914 9563839 9582725
11032902 11007160 11257964 11067023 11018893 9499078 9800280 9750582 9800280
9750582 11018895 11019057 9446914 9750582 9582725 11017847 9703869 9637671
11017848 11018893 11032901 9800280 9499078 9637671 11018895 11067023 11018893
11032902 11067023 9499078 11017848 11017847 11067023 11019057 9750582 9637671
9827896 11017847 9827896 9750582 11032902 11018895 11257964 9563839 11017847
11067023 9750582 11067023 9563839 11017848 11018893 9499078 11018894 9563839
9499078 9750583 9827896 11067023 9827896 9637671 9703869 9827896 9703869
11017848 9446914 11019057 9499078 9446914 9582725 11017848 11032902 9827896
9800280 11019057 9800280 9499078 9750582 9446914 11017847 11018893 11067023
9750582 9499078 11017847 9800280 11018893 9750582 9446914 11018895 9750582
11032901 11017847 9827896 9563839 11018893 9750582 11007160 9750583 9582725
9563839 9827896 11018893 11067023 9827896 11018895 11032901 11032902 11019057
9499078 9446914 9750583 9750582 9637671 9446914 9800280 9446914 11017848
11257964 9800280 9637671 9499078 9750582 9499078 9563839 11019057 9750582
9827896 9800280 9563839 11017848 11017847 9637671 9499078 9582725 9637671
9750583 9637671 9800280 9750582 9446914 9499078 9582725 11032901 9800280
11067023 11018893 11017847 11019057 9563839 11018894 9750582 9750583 9563839
9446914 9800280 11017848 9827896 11018893 9750582 9800280 9827896 11257964
9637671 9827896 11257964 11018895 11017847 11018893 9703869 11017848 11017847
11067023 9827896 11018895 9637671 9582725 0009213455 0009224253 0009224267
0009368091 0009368638 0009368640 0009446914 0009499078 0009563839 0009582725
0009613990 0009637671 0009645145 0009668007 0009678226 0009678228 0009703869
0009750582 0009750583 0009800280 0009827896 0011007160 0011007161 0011017847
0011017848 0011018893 0011018894 0011018895 0011019057 0011032901 0011032902
0011067023 0011257964 and 0011261956.
For International Lot Numbers, contact
your Boston Scientific Corporation representative.
RECALLING
FIRM/MANUFACTURER Boston Scientific Corporation, Fremont, CA,
REASON
Packaging issue may result in failure to
maintain sterile barrier.
VOLUME OF PRODUCT IN
COMMERCE
21,423 units
DISTRIBUTION Nationwide
and
Internationally
PRODUCT
a) Stryker Medical Stair-PRO Stair-Chair with wheels, patient
transport device, Model 6251.
b) Stryker Medical Stair-PRO Stair-Chair with wheels and a track system, patient transport device, Model 6252.
c) Stryker Medical Stair-PRO Evacuation Chair civilian use stair chair,
patient transport device, Model 6253.
CODE
a) Serial numbers: 080940844, 080940845,
080940858, 080940819, 080940820 and 080940640;
b) Serial numbers: 080940553, 080940768, 080940600, 080940601, 080940831, 080940592, 080940899, 080940821, 080940822, 080940766, 080940767, 080940597, 080940642, 080940643, 080940832, 080940833, 080940834, 080940835, 080940836, 080940580, 080940581, 080940825, 080940639, 080940595, 080940596, 080940589, 080940765, 080940838, 080940839, 080940840, 080940607, 080940608, 080940829, 080940594, 080940582, 080940583, 080940584, 080940611, 080940612, 080940613, 080940830, 080940763, 080940826, 080940827, 080940871, 080940644, 080940859, 080940860, 080940861, 080940862, 080940863, 080940864, 080940865, 080940866, 080940867, 080940868, 080940869, 080940828, 080940593, 080940760, 080940761, 080940898, 080940837, 080940551, 080940552, 080940752, 080940769, 080940873, 080940874, 080940875, 080940876, 080940877, 080940641, 080940770, 080940843, 080940247, 080940248, 080940762, 080940842, 080940753, 080940754, 080940755, 080940756, 080940757, 080940758, 080940759, 080940841, 080940587, 080940588, 080940585, 080940586, 080940771, 080940550, 080940645, 080940646, 080940647, 080940648, 080940649, 080940750, 080940751, 080940823, 080940824, 080940870, 080940764 and 080940872;
c) Serial numbers: 080940895, 080940896, 080940578, 080940614, 080940615,
080940616, 080940618, 080940619, 080940620, 080940621, 080940623, 080940624,
080940625, 080940626, 080940628, 080940880, 080940571, 080940572, 080940573,
080940574, 080940575, 080940576, 080940629, 080940630, 080940599, 080940554,
080940555, 080940887, 080940888, 080940889, 080940890, 080940891, 080940892,
080940893, 080940894, 800940881, 080940772, 080940773, 080940774, 080940564,
080940565, 080940566, 080940850, 080940856, 080940556, 800940557, 080940558,
080940559, 080940560, 080940561, 800940562, 080940563, 080940633, 080940634,
080940635, 080940636, 080940245, 080940246, 080940851, 080940852, 080940853,
080940854, 080940846, 080940847, 080940848, 080940849, 080940810, 080940811,
080940812, 080940813, 080940814, 080940815, 080940577, 080940631, 080940632,
080940617, 080940622, 080940627, 080940818, 080940857, 080940882, 080940897,
080940816, 080940817, 080940567, 080940568, 080940569, 080940570, 080940579,
080940800, 080940801, 080940802, 080940878, 080940879, 080940855, 080940883,
080940884, 080940885, 080940886, 080940637, 080940638 and
080940803
RECALLING FIRM/MANUFACTURER Stryker Medical
Division of Stryker Corp., Portage, MI
REASON Handle could come
off. The bolts may back out over time, resulting in a risk of dropping the
patient and/or injury to transport personnel, a delay in medical treatment, or a
delay in evacuation.
VOLUME OF PRODUCT IN COMMERCE
213
chairs
DISTRIBUTION Nationwide, India, Netherlands and
United
Kingdom
PRODUCT
a) Terumo Advanced Perfusion System 1 Air Bubble Detection Module;
Catalog number 802110. For use for up to 6 hours in the extracorporeal
circulation of blood for arterial perfusion, regional perfusion and
cardiopulmonary bypass procedures.
b) Terumo Advanced Perfusion System 1 Pressure Monitoring Unit; Catalog number 802112. For use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion and cardiopulmonary bypass procedures.
c) Terumo Advanced Perfusion System 1 Temperature Monitoring Systems; Catalog number 802114. For use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion and cardiopulmonary bypass procedures.
d) Terumo Advanced Perfusion System 1 Pressure Venous Occluder; Catalog number 803480. For use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion and cardiopulmonary bypass procedures.
e) Terumo Advanced Perfusion System 1, 6 inch diameter Roller Pump; Model
801041. For use for up to 6 hours in the extracorporeal circulation of blood for
arterial perfusion, regional perfusion and cardiopulmonary bypass procedures.
CODE
a) Serial number: 0727;
b) Serial numbers: 00001 through 01246, 01257 through 01277, 01279 through 01286, 01288 through 01295. 01298 through 01308, 01310 through 01315 and 01319 through 01320;
c) Serial number: 0637;
d) Serial number: 0220;
e) Serial number: 0188
RECALLING
FIRM/MANUFACTURER Terumo Cardiovascular Systems Corp., Ann Arbor,
MI,
REASON The system may
malfunction due to t-filter corrosion on an internal circuit
board.
VOLUME OF PRODUCT IN COMMERCE
1,219
units
DISTRIBUTION Nationwide and countries of Australia,
Belgium, Canada, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India,
Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Russia,
Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey,
United Arab Emirates and Viet
Nam
PRODUCT Atlantis PV Peripheral Imaging Catheter, model number H749364560,
catalog number 36456. Medical device, intended for ultrasound examination of
peripheral pathology. Intravascular ultrasound imaging is indicated in patients
who are candidates for transluminal interventional procedures.
CODE
Batch/Lot numbers: 11420600, 11420601,
11420602, 11611665, 11716746, 11720663, 11721240, 11722752, 11727107, 11727873,
11727874, 11728458, 11731354, 11732665, 11735762, 11736559, 11739563, 11742590,
11754455, 11758051, 11764281, 11777994, 11788045, 11790848, 11820112, 11897802,
11897904, and 11904295
RECALLING
FIRM/MANUFACTURER Boston Scientific Corp., Fremont, CA,
REASON
Product is mislabeled as having a shelf
life of 3 years, where the actual shelf life is 1 year.
VOLUME OF
PRODUCT IN COMMERCE
1,033
units
DISTRIBUTION Nationwide and
Panama
PRODUCT
a) REF: 915-1006 Lactosorb System, Lacto Endobrow Lift Complete,
10X13CM. (3.75X5.25") (WXL), Polypropylene, Radel. Non Sterile Product. Aid in
the alignment and stabilization of bone in the oral cranio-maxillofacial
skeletal system.
b) REF: 915-1000 Lactosorb, Office Fixation Kit, 1 Screwdriver, 1 Tap Handle, 1 Drill Bit, 1 Screw Extractor, 1 Self Drilling Tap, S/S. Aluminum , Teflon, Ra del. Non Sterile Product. Aid in the alignment and stabilization of bone in the oral cranio-maxillofacial skeletal system.
c) REF: 01-6871 IMF Set 254 X 127 X34MM (10"X1"X1.2") Polypropylene, Radel, AL, SS, T.C. Non Sterile Product. Aid in the alignment and stabilization of bone in the oral cranio-maxillofacial skeletal system.
d) REF 01-6851 Office System 1.5/IMF Small Office Set, 156MMX125MM,
(6.4"X4.9") Polypropylene, Ultem, Radel, AL, SS. Non Sterile Product. Aid
in the alignment and stabilization of bone in the oral cranio-maxillofacial
skeletal system.
CODE
a) Lot
numbers: 026660, 044820, 044830, 044900, 069630, 119940, 175710, 207320, 286830,
323270, 373220, 375190, 393190, 396390, 464030, 620270, and 718350;
b) Lot numbers: 056960, 060930, 071130, 085650, 155540, 181390, 2247100, 248520, 368120, 584110, 716090, 716100, 716110, 717510, 782200, 851620, 990550, and 999999;
c) Lot numbers: 907650, 252341, and 300870;
d) Lot numbers: 108450, and 103010
RECALLING
FIRM/MANUFACTURER Biomet Microfixation, Inc.,
Jacksonville, FL,
REASON The finished trays were
packaged and shipped to facilities without the instructions for use and without
the single use symbol on the packaged tray label.
VOLUME OF PRODUCT
IN COMMERCE
420 units
DISTRIBUTION Nationwide
and countries of Argentina, Australia, Canada, Columbia, Hong Kong, Japan,
Mexico, New Zealand, Puerto Rico, Taiwan, Thailand, Denmark, France,
Netherlands, Greece, Italy, Norway, Finland, Spain, Czech Republic, and
Venezuela
PRODUCT Restoration Modular Conical Distal Stem; Sterile, Titanium Alloy.
One unit per package. This system is intended to be used for primary or revision
total hip arthroplasty, as well as in the presence of severe proximal bone loss.
Catalog No. 6276-7-017 (17mm), Catalog No. 6276-7-014, (14mm).
CODE
Lot Code CAXHA68D; Lot Code CAXHB18A
RECALLING FIRM/MANUFACTURER
Recalling Firm: Stryker
Howmedica Osteonics Corp., Mahwah, NJ,
Manufacturer: Stryker Ireland, Ltd., Orthopedics, Carrigtohill, County Cork,
Ireland.
REASON Stryker became aware of a lot for lot mix
up.
VOLUME OF PRODUCT IN COMMERCE
4
units
DISTRIBUTION Nationwide
PRODUCT Regenesis Biomedical PROVANT Wound Therapy System, Model 4201 - The
Provant Wound Therapy System is a solid-state, fixed-power output radio
frequency generator and transmitter designed to operate at the FCC authorized
medical device frequency of 27.12 MHz. Provant transmits a fixed dose of
non-ionizing, non-thermal RF energy via an applicator pad that is placed
adjacent to the patient's dressed wound. The Provant Wound Therapy System is
indicated for adjunctive use in the palliative treatment of postoperative pain
and edema in superficial soft tissue.
CODE Serial Numbers: 704, 920, 460, 709, 711,
714, and 727
RECALLING FIRM/MANUFACTURER
Regenesis
Biomedical Inc., Scottsdale, AZ,
REASON Under certain
unusual clinical circumstance, the affected devices may be unable to generate a
therapeutic dose of radio frequency energy. As such, for these few devices, the
effectiveness of Provant therapy may be impaired.
VOLUME OF PRODUCT
IN COMMERCE
7 units
DISTRIBUTION WV, VA and
NC
PRODUCT
a) Solar Shoulder Humeral Stem. These devices are modular
components of a total shoulder system. These humeral stems are intended for use
with humeral heads and glenoid components in primary or revision total shoulder
arthroplasty. Catalog Numbers, Description 5351-4103 SHOULDER - HUMERAL STEM
7mm, 5351-4105 SHOULDER - HUMERAL STEM 9mm, 5351-4105 SHOULDER - HUMERAL STEM
9mm, 5351-4107 SHOULDER - HUMERAL STEM 11mm, 5351-4108 SHOULDER - HUMERAL STEM
12mm, 5351-4108 SHOULDER - HUMERAL STEM 12mm, 5351-4109 SHOULDER - HUMERAL STEM
13mm VWHMKE, 5351-4110 SHOULDER - HUMERAL STEM 14mm HTXMKE, 5351-4110 SHOULDER -
HUMERAL STEM 14mm, 5351-4113 SHOULDER - HUMERAL STEM 17mm, and 5351-4205
SHOULDER - HUMERAL 200mmX9mm.
b) Triathlon Cemented Stem - Knee These devices are modular components of a total knee system. These modular stems are intended for use with femoral and tibial components in primary or revision total knee arthroplasty. Item Number, Product Name: 5560-S-209 TRI CEMENTED STEM 9MMX100MM and 5560-S-209 TRI CEMENTED STEM 9MMX100MM.
c) Triathlon-Press-Fit Stem These devices are modular components of a total knee system. These modular stems are intended for use with femoral and tibial components in primary or revision total knee arthroplasty. Catalog numbers, Description: 5565-S-015 TRI PRESS-FIT STEM 15MM X 100MM, 5565-S-016 TRI PRESS-FIT STEM 16MM X 100MM, 5565-S-016 TRI PRESS-FIT STEM 16MM X 100MM, 5565-S-017 TRI PRESS-FIT STEM 17MM X 100MM, 5566-S-012 TRI PRESS-FIT STEM 12X150MM, 5566-S-013 TRI PRESS-FIT STEM 13X150MM, and 5566-S-018 TRI PRESS-FIT STEM 18X150MM.
d) Omnifit HA Hip Stem These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty. Catalog number, Description: 6017-0525A OMNIFIT HA HIP STEM #5 and 6017-1035A OMNIFIT HA HIP STEM #10.
e) Omnifit Normalized Hip Stem These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty. Catalog number, Description: 6034-0730 OMNIFIT NORMALIZED HIP STEM.
f) Secur-Fit Max Hip Stem. These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty. Catalog numbers, Description: 6051-0625S SECUR-FIT MAX 132 HIP STEM #6, 6051-0730S SECUR-FIT MAX 132 HIP STEM #7, 6051-0730S SECUR-FIT MAX 132 HIP STEM #7, 6051-0830S SECUR-FIT MAX 132 HIP STEM #8, 6051-0830S SECUR-FIT MAX 132 HIP STEM #8, 6051-0935S SECUR-FIT MAX 132 HIP STEM #9, 6051-1035S SECUR-FIT MAX 132 HIP STEM #10, 6051-1140S SECUR-FIT MAX 132 HIP STEM #11, 6051-1140S SECUR-FIT MAX 132 HIP STEM #11, 6052-0830S SECUR-FIT MAX 127 HIP STEM #8, 6052-0935S SECUR-FIT MAX 127 HIP STEM #9, 6052-1035S SECUR-FIT MAX 127 HIP STEM #10, 6052-1035S SECUR-FIT MAX 127 HIP STEM #10 and 6052-1240S SECUR-FIT MAX 127 HIP STEM #12.
g) Primary Secur-fit Plus Max Hip Stem. These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty. Catalog numbers, Description: 6054-0812S PRIMARY SECUR-FIT PLUS MAX #8/12, 6054-0812S PRIMARY SECUR-FIT PLUS MAX #8/12, 6054-0915S PRIMARY SECUR-FIT PLUS MAX #9/15, 6054-1014S PRIMARY SECUR-FIT PLUS MAX #10/14, 6054-1014S PRIMARY SECUR-FIT PLUS MAX #10/14, 6054-1014S PRIMARY SECUR-FIT PLUS MAX #10/14 and 6054-1218S PRIMARY SECUR-FIT PLUS MAX #12/18.
h) Accolade C CS. These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty. Catalog number, Description: 6057-0335D ACCOLADE C CS 127 NK, 6057-0537D ACCOLADE C CS 127 NK, 6057-0740D ACCOLADE C CS 127 NK and 6058-0335D ACCOLADE C CS 132 NK.
i) Omnifit HFX Hip Stem. These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty. Catalog numbers, Description: 6070-0730A OMNIFIT HFX HIP STEM SIZE #07 132, 6070-0730A OMNIFIT HFX HIP STEM SIZE #07 132, 6070-0730A OMNIFIT HFX HIP STEM SIZE #07 132, 6070-0830A OMNIFIT HFX HIP STEM SIZE #08 132, 6070-0830A OMNIFIT HFX HIP STEM SIZE #08 132, 6070-0830A OMNIFIT HFX HIP STEM SIZE #08 132, 6070-0830A OMNIFIT HFX HIP STEM SIZE #08 132, 6070-0935A OMNIFIT HFX HIP STEM SIZE #09 132, 6070-0935A OMNIFIT HFX HIP STEM SIZE #09 132, 6070-0935A OMNIFIT HFX HIP STEM SIZE #09 132, 6070-0935A OMNIFIT HFX HIP STEM SIZE #09 132, 6070-0935A OMNIFIT HFX HIP STEM SIZE #09 132, 6070-0935A OMNIFIT HFX HIP STEM SIZE #09 132, 6070-0935A OMNIFIT HFX HIP STEM SIZE #09 132 and 6070-1140A OMNIFIT HFX HIP STEM SIZE #11 132.
j) Accolade HFx. These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty. Catalog # Description: 6077-0230 Accolade HFx Size #2, 6077-0230 Accolade HFx Size #2, 6077-0335 Accolade HFx Size #3, 6077-0435 Accolade HFx Size #4, 6077-0435 Accolade HFx Size #4 and 6077-0435 Accolade HFx Size #4.
k) Omnifit FC Cemented Stem. These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty. Catalog number, Description: 6093-1035 OMNIFIT-FC CEMENTED STEM and 6093-1035 ONMIFIT FC CEMENTED STEM.
l) Omnifit Eon. These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty. Catalog number, Description: 6098-0630 OMNIFIT EON 132, 6098-0630 OMNIFIT EON 132, 6098-0735 OMNIFIT EON 132, 6098-0940 OMNIFIT EON 132 and 6098-1140 OMNIFIT EON 132.
m) Japanese Omnifit Eon 127 NK. These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty. Item number: J6097-0425, Item number J6097-05390.
n) Reunion HA Fracture Stem. These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty. Catalog numbers, Description: R5351-4507 REUNION HA FRACTURE STEM 7MM, R5351-4508 ReUnion HA Fracture Stem 8MM, R5351-4508 ReUnion HA Fracture Stem 8MM and R5351-4509 REUNION HA FRACTURE STEM 9MM.
o) Secur-Fit ARC/HA Collar Stem. These devices are modular components of a
total hip system. These femoral stems are intended for use with femoral heads
and acetabular components in primary or revision total hip arthroplasty. Catalog
number, Description: S-2337-HF09 SECUR-FIT ARC/HA COLLAR STEM #9, S-2337-HF09
SECUR-FIT ARC/HA COLLAR STEM#9 and S-2337-HF10 SECURFIT ARC/HA COLLAR
STEM#10.
CODE
a) Lot Numbers: XKRMHE, 8ETMJE, 8EYMJE, V45MKE, VV3MKE, VV4MKE,
VWHMKE, HTKMKE, T7TMKE, KKHMKE
and MT9MKE;
b) Lot numbers: N3N31K, N3N31W;
c) Lot numbers: M3M07B, M3M01B, M3M01E, M3M03A, M3L11F, M3L12B, M3L17D;
d) Lot numbers: 3PMMNE, NKAMME;
e) Lot number: PJ9MLE;
f) Lot numbers: HKNMME, K03MME, K08MME, E56MLE, J41MLE, EN1MME, 70PMME, 4PMMME, 901MME, V3PMLE, 5ERMME, 69WMME, AL8MME, EP4MME;
g) Lot numbers: 72MMME, WA5MLE, 11AMME, AKKMME, AKXMME, DHKMME, EP2MME;
h) Lot numbers: JYLMME, JYNMME, E5EMME, JYVMME;
i) Lot numbers: P7MMME, RHWMME, VJMMME, 0J0MNE, D9VMME, VK0MME, T9RMME, 0HDMNE, 0HEMNE, P7XMME, RE8MME, REDMME, VK2MME, VK3MME, 0HKMNE;
j) Lot numbers: 8XAMLE, 8XEMLE, 8PAMME, 8PEMME, PJLMLE, R27MLE;
k) Lot numbers: 28RMLE, 3DHME;
l) Lot numbers: 48HMME, TAMMME, 0W9MNE, TN3MLE, 2HNMME;
m) Lot number: TN7MLE and lot number: 41WMME;
n) Lot numbers: T2T72836-J, T2N71522-I, T2T72422-C, T2N71523-G;
o) Lot numbers: 8T3MME, 99EMME, 2H0MME
RECALLING
FIRM/MANUFACTURER
Stryker Howmedica Osteonics Corp., Mahwah, NJ,
REASON Stryker Orthopedics became aware of a visual
defect in certain blister packages of certain devices.
VOLUME OF
PRODUCT IN COMMERCE
322
units
DISTRIBUTION Nationwide and countries of Australia,
Canada, Japan, Argentina, Brazil, Germany, Spain, Columbia, Guatemala and
Sweden
WEEK ENDING FEBRUARY 21
PRODUCT DeVilbiss Model 7310PR-D VacuAide
Compact Suction Unit. A portable, AC-powered or compressed air-powered suction
pump. The device may include a microbial filter.
CODE
Serial numbers: PRD000006, PRD000012, PRD000013,
PRD000014, PRD000046, PRD000082, PRD000083, PRD000084, PRD000088, PRD000090,
PRD000091, PRD000099, PRD000100, PRD000101, PRD000102, PRD000152, PRD000174,
PRD000175, PRD000177, PRD000184, PRD000186, PRD000189, PRD000191, PRD000193,
PRD000201, PRD000202, PRD000203, PRD000206, PRD000207, PRD000209, PRD000225,
PRD000226, PRD000229, PRD000230, PRD000262, PRD000269, PRD000323, PRD000341,
PRD000355, PRD000363, PRD000364, PRD000368, PRD000370, PRD000371, PRD000373,
PRD000374, PRD000375, PRD000383, PRD000385, PRD000386, PRD000388, PRD000389,
PRD000390, PRD000392, PRD000393, PRD000399, PRD000400, PRD000401, PRD000403,
PRD000405, PRD000406, PRD000407, PRD000409, PRD000411, PRD000415, PRD000417,
PRD000419, PRD000420, PRD000422, PRD000425, PRD000426, PRD000428, PRD000432,
PRD000443, PRD000447, PRD000448, PRD000449, PRD000450, 0425198002000, and
04275195001000
RECALLING FIRM/MANUFACTURER Sunrise
Medical, Somerset , PA
REASON Improper suction: Regulator knob does not turn properly which can
affect the suction level.
VOLUME OF PRODUCT IN COMMERCE 78 units
DISTRIBUTION
Nationwide and
countries of Mexico , Guatemala , Columbia , and Costa Rica
PRODUCT SERVO Ventilator 300A (SV300A). The device is intended for general
and critical ventilatory care of adult, pediatric, infant and neonatal patients.
The unit is designed to be used at the bedside and for intra-hospital transport.
It is not intended for transport use in ambulances or in helicopters in the US
market.
CODE All units of SV300A
RECALLING FIRM/MANUFACTURER
Recalling Firm:
Maquet,
Inc., Bridgewater , NJ ,
Manufacturer:
Maquet Critical Care AB , Solna , Sweden .
REASON SERVO Ventilator 300A could cause
hyperventilation if the use of the PC or PRVC mode with the Automode function
set to "On" in Neonatal and Pediatric patient ranges with small tidal volumes
and short set inspiratory rise time.
VOLUME OF PRODUCT IN COMMERCE
3,688 units
DISTRIBUTION Nationwide
PRODUCT
a) Zimmer TIV Self Tap Screw, 40 mm. Item number: 48231904000.
NOTE: This product is not sold in the United States . The device is used to
attach healing plates to long bones for stabilization of fractures during the
normal healing process. Devices are individually packaged and sold sterile.
b) Zimmer TIV Self Tap Screw, 42 mm. Item
number: 48231904200. NOTE: This product is not sold in the United States . The
device is used to attach healing plates to long bones for stabilization of
fractures during the normal healing process. Devices are individually packaged
and sold sterile. CODE
a) Lot
number: 60557215;
b) Lot numbers: 60548987, 60548986, 60548999, 60557220 and
60557219
RECALLING FIRM/MANUFACTURER
Recalling Firm:
Zimmer, Inc., Warsaw , IN ,
Manufacturer:
Zimmer Manufacturing B.V., Mercedita, PR.
REASON The package may contain a 40 mm long screw
instead of the labeled 42 mm long screw, or vice versa.
VOLUME OF
PRODUCT IN COMMERCE 508 units
DISTRIBUTION Japan
PRODUCT
a) Boston Scientific EndoVive 20F (6.7
mm) Safety PEG Kit, Push Method for use with sterile Xylocaine ampule, latex
free, sterile; Cat. no./REF 6647, a) inner pouch UPN M00566470 and b) outer
carton UPN M00566471. The devices are indicated for providing nutrition directly
into the stomach in adult and pediatric patients who are unable to consume
nutrition by conventional means.
b) Boston
Scientific EndoVive 24F (8 mm) Safety PEG Kit, Push Method for use with sterile
Xylocaine ampule, latex free, sterile; Cat. no./REF 6649, a) inner pouch UPN
M00566490 and b) outer carton UPN M00566491. The devices are indicated for
providing nutrition directly into the stomach in adult and pediatric patients
who are unable to consume nutrition by conventional means.
c) Boston Scientific EndoVive 20Fr (6.7 mm) Safety PEG Kit, Push
Method, latex free, sterile; Cat. no./REF 6653, a) inner pouch UPN M00566530 and
b) outer carton UPN M00566531. The devices are indicated for providing nutrition
directly into the stomach in adult and pediatric patients who are unable to
consume nutrition by conventional means.
d)
Boston Scientific EndoVive 20F (6.7 mm) Safety PEG Kit, Push Method for use with
pre-filled lidocaine syringe, latex free, sterile; Cat. no./REF 6673, a) inner
pouch UPN M00566730 and b) outer carton UPN M00566731. The devices are indicated
for providing nutrition directly into the stomach in adult and pediatric
patients who are unable to consume nutrition by conventional means.
CODE
a) a) Lot numbers: 12004380, 12017735,
12039127, 12068485, 12074073, 1209443, 12115001, 12127100 and 12159831; b) lot
numbers of 12069493, 12069494, 12112331, 12123598 and 12145424;
b) a) Lot numbers: 12007899 and b) lot numbers: 12065904 and 12148450;
c) Lot number: 12080466 on both pouches;
d) a) Lot
numbers: 12015222, 12015223, 12017140, 12035930, 12037032, 12038621, 12056814,
12074992, 12096466, 12110667, 12111382, 12111566, 12112012, 12113348, 12129137,
12133028, 12133803 and 12133804; b) lot numbers: 12065719, 12065900, 12065903,
12088673, 12105251, 12115743, 12119946, 12145423 and 12148447
RECALLING FIRM/MANUFACTURER Boston Scientific, Corp.,
Spencer , IN
REASON The user may be unable to pass the
device over the guidewire for placement in the patient, resulting in an
inability to use the device.
VOLUME OF PRODUCT IN COMMERCE 2,460 units
DISTRIBUTION
Nationwide
PRODUCT
a) Stryker
PainPump, Two Site Infusion
Kit with ExFen 2.5", 120 mL Infusion Kit, sterile, 5 per box. Product number:
0500-166-025, REF 500-166-25. A disposable syringe infuser which uses sustained
vacuum pressure to deliver a continuous infusion of medications or fluids at a
controlled rate. It delivers controlled amounts of medication directly to the
intraoperative site for pain management and/or antibiotic administration. It is
also intended for controlled delivery of local anesthetics in close proximity to
nerves for postoperative regional anesthesia and pain management. Routes of
administration may be intraoperative, perineural, or percutaneous.
b) Stryker PainPump Two Site Infusion Kit with ExFen
5.0", 120 mL Infusion Kit, sterile, 5 per box. Product number: 0500-166-050, REF
500-166-50. A disposable syringe infuser which uses sustained vacuum pressure to
deliver a continuous infusion of medications or fluids at a controlled rate. It
delivers controlled amounts of medication directly to the intraoperative site
for pain management and/or antibiotic administration. It is also intended for
controlled delivery of local anesthetics in close proximity to nerves for
postoperative regional anesthesia and pain management. Routes of administration
may be intraoperative, perineural, or percutaneous. ;
c) Stryker PainPump Two Site Infusion Set, sterile, 10 per box.
Product number: 0500-200-000, REF 500-200. A disposable syringe infuser which
uses sustained vacuum pressure to deliver a continuous infusion of medications
or fluids at a controlled rate. It delivers controlled amounts of medication
directly to the intraoperative site for pain management and/or antibiotic
administration. It is also intended for controlled delivery of local anesthetics
in close proximity to nerves for postoperative regional anesthesia and pain
management. Routes of administration may be intraoperative, perineural, or
percutaneous.
CODE
a) Lot number:
2008011403, 2008020502, 2008061701 and 2008072102;
b) Lot number: 2008011404
and 2008050201;
c) Lot number: 2008011601, 2008040204 and 2008061801;
RECALLING FIRM/MANUFACTURER
Recalling Firm: Stryker
Instruments Div. of Stryker Corp., Portage , MI ,
Manufacturer: Medisize Donegal Healthcare, Donegal , Ireland .
REASON
There is a
potential for the pump to overinfuse.
VOLUME OF PRODUCT IN COMMERCE
118 boxes
DISTRIBUTION Nationwide
PRODUCT Stryker PainPump 2
Blockaid, 400 mL PainPump for Continuous Nerve
Block, sterile, 5 per box. REF 575-100, Product Number 0575-100-000. The device
is used to deliver controlled amounts of medication and narcotics directly to
the intraoperative site for pain management; the pump infuses the medication at
an hourly flow rate and provides the option for patient controlled bolus doses.
Medications and narcotics are infused through intramuscular, subcutaneous and
epidural routs. It is also intended for controlled delivery of local anesthetics
and narcotics in close proximity to nerves for postoperative regional anesthesia
and pain management, in which case routes of administration may be
intraoperative, perineural or percutaneous.
CODE
Lot numbers: 08219012, 08220012, 08221012,
08222012, 08225012, 08226012, 08229012, 08231022, 08233012, 08234012, 08235012,
08238022, 08239012, 08241012, 08242012, 08243012, 08246022, 08247012, 08249032,
08250012, 08252022, 08252032, 08253012, 08254012, 08256032, 08256042, 08257012
and 08259022
RECALLING FIRM/MANUFACTURER
Recalling
Firm: Stryker Instruments, Div. of Stryker Corp., Portage , MI ,
Manufacturer: Stryker Puerto Rico, Inc., Arroyo,
PRng.
REASON The pump may
display an E-4 error message, which causes the pump to stop dispensing the pain
medication.
VOLUME OF PRODUCT IN COMMERCE 2,009 boxes
DISTRIBUTION Nationwide and Canada
PRODUCT Artiste MV SA Linear Accelerator with syngo RT Therapist Version
4.1 build 67 and build 122. Part Number 8139789.
CODE
Serial numbers: 5190, 5350, 5355, 5364, 5367,
5374, 5377, 5398, 5405, 5410, 5419, 5424, 5165, 5166, 5202, 5238, 5298, 5306,
5342, 5352, 5365, 5366, 5373, 5378, 5379, 5380, 5397, 5399, 5400, 5401, 5402,
5409, 5414, 5416, 5423 and 5425
RECALLING FIRM/MANUFACTURER
Siemens Medical Solutions USA, Inc., Concord , CA ,
REASON
If a plan contains a sequence of beams, including at least 1 ARC
treatment beam and if during the setup stage of a beam following an ARC beam in
the delivery sequence a termination occurs, and the user selects the immediate
resumption feature, syngo RT Therapist Express systems will display and record a
second delivery of a prior beam and not the delivery of the current beam in
certain circumstances.
VOLUME OF PRODUCT IN COMMERCE 36
units
DISTRIBUTION Nationwide and countries of
Netherlands , Germany , Poland , Norway , Australia , Hungary , Belgium , New
Zealand , France , Malaysia , Spain and Canada
PRODUCT GE Centricity PACS RA1000 Workstation (for diagnostic image
analysis). The Centricity PACS Workstation is intended for use as a primary
diagnostic and analysis tool for diagnostic images by trained healthcare
professionals, including radiologists, physicians, technologists, clinicians and
nurses. It is also intended for use as a clinical review workstation throughout
the healthcare facility. The workstation interface provides the user with a
means to display, manipulate, archive, print, and export images when connected
with the Centricity PACS infrastructure.
CODE Software version 2.1.5.5 and 3.0.4 and higher
RECALLING
FIRM/MANUFACTURER GE Healthcare Integrated IT Solutions, Barrington
, IL ,
REASON There is a potential patient safety issue
involving patient jacket content, which may result in unintended historical
image(s) being displayed.
VOLUME OF PRODUCT IN COMMERCE 476 units
DISTRIBUTION
Nationwide and
Canada
PRODUCT Low Profile U-Joint Driver for the
SynFix Mini-Open System, Catalog number 03.802.231. The Synthes SynFix LR is
indicated for use in patients with degenerative disc disease (DDD) at one or two
contiguous levels from L2 to S1. These DDD patients may also have up to Grade I
spondylolisthesis at the involved level(s). The interior of the spacer component
of the SynFix LR can be packed with autograft.
CODE Lot numbers: 2317363, 2331806, 2331810, 2387975,
2387976, 2401982, 5637286, 5644605, 5660377, 5664629, 5826243, and 5826244
RECALLING FIRM/MANUFACTURER
Recalling Firm: Synthes
Spine, West Chester , PA ,
Manufacturer: Synthes Gmbh, Oberdorf
, Switzerland .
REASON Screwdriver tip may break off and lodge in the screw head.
VOLUME OF PRODUCT IN COMMERCE 381 units
DISTRIBUTION Nationwide and countries of Canada and
Switzerland
PRODUCT Ergolift 400 Lbs. Floor Lift,
model/part numbers ERGOLIFT and ERGOLIFT-2. The equipment is intended to be used
in hospitals, nursing homes, or other health care facilities by trained
caregivers for the transfer of patients.
CODE Serial numbers: ERLI-0001 through ERLI-1717, manufactured between
1995 and 1999
RECALLING FIRM/MANUFACTURER
Recalling
Firm: Arjo, Inc., Roselle , IL , .
Manufacturer: B.H.M. Medical, Inc.,
Magog , Canada .
REASON Report of structural failure of the mast during use.
VOLUME OF PRODUCT IN COMMERCE 220 lifts
DISTRIBUTION Nationwide
PRODUCT Posey Connecting Straps and Belts/Locking Cuffs. The device is used
to limit the wearer's movement by securing cuffs to a connecting strap or belt.
Catalog numbers: 2320, 2330, 2346, 2350, 2351, 2355, 2360, 2208, 2209, 2210,
2211, 2212 and 2213.
CODE Unknown
RECALLING FIRM/MANUFACTURER J. T. Posey Company,
Arcadia , CA
REASON A patient may be able to catch the belt
on a stationary object and use their body force to tear or break the strap,
creating freedom of arm movement.
VOLUME OF PRODUCT IN COMMERCE
51,200 devices (approximately)
DISTRIBUTION Nationwide and countries of Australia , Belgium , Canada , France ,
Germany , Ireland , Italy , Korea , Lithuania and Netherlands
PRODUCT Stryker AVS AL Spacer Trial Inserter; Non Sterile. Catalog number:
48329800.
CODE Lot numbers: 056398 and 056399
RECALLING
FIRM/MANUFACTURER
Recalling Firm: Stryker Spine, Allendale ,
NJ ,
Manufacturer: Stryker Spine, Cestas,
Aquitane , France .
REASON DESIGN FLAW: Issues were reported concerning fracture of handle,
difficulty in removing inserter and inserter difficult to lock.
VOLUME OF PRODUCT IN COMMERCE 37 units
DISTRIBUTION Nationwide
PRODUCT
a) VOYAGER NC Coronary Dilatation Catheter, 4.5 X 8.0 mm. The
device is used for in the following: (1) Balloon dilatation of the stenotic
portion of a coronary artery or bypass graft stenosis for the purpose of
improving myocardial perfusion, (2) Balloon dilatation of a coronary artery
occlusion for the purpose of restoring coronary flow in patients with ST-segment
elevation myocardial infarction, and (3) Balloon dilatation of a stent after
implantation. Part Number: 1011759-08. ;
b)
VOYAGER NC Coronary Dilatation Catheter, 4.5 X 12.0 mm. The device is used for
in the following: (1) Balloon dilatation of the stenotic portion of a coronary
artery or bypass graft stenosis for the purpose of improving myocardial
perfusion, (2) Balloon dilatation of a coronary artery occlusion for the purpose
of restoring coronary flow in patients with ST-segment elevation myocardial
infarction, and (3) Balloon dilatation of a stent after implantation. Part
Number: 1011759-12, 8061961 and Part Number: 1011750-12.
CODE
a) Lot numbers: 8050961 and 8062361;
b) Lot
Numbers: 8041051, 8050561, Lot Number 8100761
RECALLING
FIRM/MANUFACTURER Abbott Vascular-Cardiac Therapies dba Guidant
Corp., Temecula , CA ,
REASON The recall was initiated after
Abbott Vascular received reports the protective, yellow balloon sheath of the
4.5mm diameter VOYAGER NC Coronary Dilatation Catheter has been tight. If
unnoticed by the user, there is a potential to cause damage to the balloon
dilatation catheter during removal of the sheath. The balloon sheath covers the
balloon, maintaining the fold configuration and protects the balloon during
shipment. It is disposed of during device preparation.
VOLUME OF
PRODUCT IN COMMERCE 224 devices
DISTRIBUTION Nationwide and Internationally
PRODUCT IMPAX Cardiovascular Polling Procedure version 1.00.09, used in
IMPAX 6.2.1 and 6.3.1.; Catalog numbers: E18R4000, L5NQJ000, L57OP000, LBCDB000,
and E2ZCK000. IMPAX Cardiovascular is a picture archiving and communications
system intended to be used as a networked cardiovascular information management
system. The IMPAX Centric Polling Procedure provides a mechanism to ensure that
patient demographic and merge/split operations performed on the IMPAX database
utilizing the IMPAX Service Tool are captured and applied to the IMPAX CV
database, in order to keep patient updates from the IMPAX database synchronized
with the IMPAX CV database. Equipment number: 15013696; Equipment number:
10281800; Equipment number: 10262174; Equipment number: 15003839; and Equipment
number: 15009724.
CODE Serial
number: KQRCPF9; Serial number: BQWBZ81; Serial number: FB5C561; Serial number:
USE708NBP3 and Serial number: GN6Z1E1
RECALLING FIRM/MANUFACTURER
AGFA Corp., Greenville , SC ,
REASON Polling
procedure may fail when multiple merge/split is carried out.
VOLUME
OF PRODUCT IN COMMERCE 5 units
DISTRIBUTION Nationwide
PRODUCT Epicor LP Connecting Cable,
CC-6-LP-CE, Catalog Number 12263, an accessory to the Epicor Cardiac Ablation
Control System, Disposable connecting cable, accessory to the Epicor Cardiac
Ablation Control System. The cable is used to connect the Epicor disposable
tissue device (UltraCinch or UltraWand) to the Epicor Cardiac Ablation control
System.
CODE
All serial numbers
affected, lot numbers 08051201, 08051504, 08052202, 08062605, 08070702,
08070704, 08071604, 08071801, 08072505, 08072506, 08080119, 08080801, 08082702,
08091112, 08092210, 08092608, 08100202, 08100703, 08101005, 08102006, 08102325,
08102805, 08110530, and 08111412. ***UPDATE 01/09/2009*** Serial numbers
NS-10156, NS-10158, NS-10159, NS-10160 and NS-10161 were originally not included
in the original submission. The lot number assigned to prior SN assignment adds
08052901, 08081801, 08091801, 08040701, 08071806, 08101326, and 08103001
RECALLING FIRM/MANUFACTURER St Jude Medical, Inc.,
Sunnyvale , CA ,
REASON Firm has received multiple reports of
difficulties connecting the product and/or the presence of error code messages
occurring with the use of the connecting cable and the Ablation Control System
unit.
VOLUME OF PRODUCT IN COMMERCE 152 units
DISTRIBUTION Nationwide and country of Belgium
PRODUCT iSite PACS (Picture Archiving and Communications System) Versions
4.1.x up to and including 4.1.51.3. Device is an image management system
intended to be used by trained professionals, and is a software package used
with general purpose computing hardware to acquire, store, distribute, process
and display images and associated data throughout a clinical environment. The
software performs digital image processing, measurement, communication and
storage.
CODE Versions 4.1.x up to
and including 4.1.51.3
RECALLING FIRM/MANUFACTURER Philips Healthcare Informatics, Inc., Foster City , CA ,
REASON
Product may display clinical reports with the study date and time
in GMT format, which may not be obvious to the end user, who may expect to see
the local time in the report.
VOLUME OF PRODUCT IN COMMERCE
40 units
DISTRIBUTION Italy, Denmark,
Germany, Netherlands, Sweden, Spain, Portugal, Switzerland, Spain, Turkey,
France, Saudi Arabia, Australia, Thailand, New Zealand, Japan, Brazil
PRODUCT Artiste MV Digital Linear Accelerator System equipped with 160 MCL
Accessory Holder. Medical charged-particle radiation therapy system.
CODE
All units of Part number: 8139789
RECALLING FIRM/MANUFACTURER Siemens Medical Solutions
USA, Inc., Concord , CA
REASON The range of tolerance is too high
for the accessory holder; therefore, Wedges and other accessories do not latch
to the locking mechanism correctly. This issue represents a potential serious
injury to a patient or operator.
VOLUME OF PRODUCT IN COMMERCE
43 units
DISTRIBUTION Nationwide and
Internationally
PRODUCT
a) HR90K Platinum Series Headpiece
Model Number AB-7300-001. The HiRes90K headpiece is used in conjunction with
speech processors. The Harmony HiResolution Bionic Ear System and Clarion
Multi-Strategy Cochlear Implant are cochlear implant systems designed to provide
useful hearing to individuals with severe-to-profound sensorineural hearing
loss.
b) Platinum Series Headpiece Model
Number AB-7300. The headpiece is used in conjunction with speech processors. The
Harmony HiResolution Bionic Ear System and Clarion Multi-Strategy Cochlear
Implant are cochlear implant systems designed to provide useful hearing to
individuals with severe-to-profound sensorineural hearing loss.
c) S-Series Long Range Headpiece a) Model Number
AB-5301-00 (beige) and b) AB-5301-10 (brown). The headpiece is used in
conjunction with speech processors manufactured by Advanced Bionics.
d) S-Series Short Range Headpiece,
a) Model Number MMT-5300-00 (beige) and b) MMT-5300-10 (brown). The headpiece is
used in conjunction with speech processors.
CODE
a) Serial Numbers: 167419, 167420, 169309, 169310,
169311, 170983, 170984, 170985, 172709, 172710, 174272, 174274, 174275, 174298,
175139, 175988, 175989, 175990, 175991, 177627, 177628, 177629, 177630, 177631,
177671, 177672, 177673, 178494, 179252, 181303, 181304, 183489, 184405 and
184694;
b) Serial Numbers: 36241, 36242, 36243, 36244, 36245,
36246, 36247, 36248, 36249, 36250, 36251, 36252, 36253, 36254, 36255, 36256,
36257, 36258, 36259, 36260, 36261, 36262, 36263, 36264, 36265, 36266, 36267,
36268, 36269, 36270, 36271, 36272, 36273, 36274, 36275, 36276, 36291, 36292,
36293, 36295, 36297, 36298, 36299, 36300, 36324, 36325, 36326, 36327, 36328,
36329, 36330, 36331, 36332, 36333, 36334, 36335, 36336, 36338, 36339, 36340,
36341, 36342, 36343, 36344, 36345, 36346, 36347, 36348, 36349, 36350, 36351,
36352, 36353, 36354, 36355, 36356, 36357, 36358, 36359, 36360, 36481, 36482,
36483, 36484 36485, 36486, 36487, 36488, 36489, 36490, 36491, 36493, 36494,
36495, 36496, 36497, 36498, 36499, 36500, 36501, 36502, 36503, 36504, 36506,
36507, 36508, 36509, 36510, 36511, 36512, 36513, 36514, 36515 36516, 36517,
36518, 36519, 36520, 36521, 36522, 36523, 36524, 36525, 36526, 36527, 36528,
36529 36530, 36531, 36532, 36533, 36534, 36535, 36536, 36537, 36538, 36539,
36540, 36568, 36569, 36570, 36571, 36572, 36573, 36574, 36575, 36576, 36577,
36578, 36579, 36580, 36581, 36582, 36583 36584, 36585, 36586, 36587, 36588,
36589, 36590, 36591, 36592, 36593, 36594, 36595, 36596, 36597, 36598, 36599,
36600, 36601, 36602, 36603, 36604, 36605, 36606, 36607, 36608, 36609, 36610,
36611, 36612, 36613, 36614, 36615, 36616, 36617, 36618 36620, 36621, 36622,
36623, 36624, 36625, 36626, 36627, 36628, 36629, 36630, 36631, 36632, 36633,
36634, 36635, 36636, 36637, 36638, 36639, 36640, 36641, 36642, 36643, 36644,
36645, 36646, 36647, 36648, 36649, 36650, 36651, 36652, 36653, 36654, 36655,
36656, 36657, 36658, 36659, 36660, 36661, 36662, 36663, 36664, 36665, 36666,
36667, 36668, 36669, 36670, 36671, 36672, 36673, 36674, 36675, 36676, 36677,
36678, 36679, 36680, 36718, 36719, 36720, 36721, 36722, 36723, 36724, 36725,
36726, 36727, 36728, 36729, 36731, 36732, 36733, 36735, 36736, 36737, 36738,
36739, 36740, 36742, 36743, 36744, 36745, 36746, 36751, 36752, 36755, 36756,
36757, 36758, 36759, 36760, 36761, 36762, 36763, 36764, 36765, 36766, 36767,
36769, 36770, 36771, 36772, 36773, 36774, 36775, 36776, 36777, 36778, 36779,
36780, 36781, 36782, 36784, 36785, 36786, 36787, 36789, 36790, 36791, 36792,
36793, 36794, 36795, 36796, 36797, 36798, 36799, 36800, 36801, 36802, 36803,
36804, 36805, 36806, 36807, 36808, 36809, 36810, 36811, 36812, 36813, 36814,
36815, 36816, 36817, 36818, 36819, 36820, 36821, 36822, 36823, 36824, 36825,
36827, 36828, 36829, 36830, 36831, 36832, 36833, 36834, 36835, 36836, 36837,
36838, 36839, 36840, 36841, 36842, 36843, 36844, 36845, 36846, 36847, 36848,
36849, 36850, 36851, 36852, 36853, 36854, 36855, 36856, 36857, 36858, 36859,
36860, 36868, 36892, 36893, 36895, 36898, 36900, 36901, 36902, 36904, 36905,
36906, 36907, 36908, 36909, 36910, 36921, 36922, 36923, 36924, 36925, 36926,
36927, 36928, 36929, 36931, 36932, 36933, 36934, 36935, 36936, 36937, 36938,
36939, 36940, 36941, 36942, 36943, 36944, 36945, 36946, 36947, 36948, 36949,
36950, 36952, 36965, 36966, 36968, 36971, 36972, 36973, 36975, 36976, 36977,
36979, 36981, 36982, 36983, 36985, 36986, 36987, 36989, 36990, 36991, 36992,
36993, 36994, 36995, 36996, 36997, 36999, 37000, 37001, 37002, 37003, 37004,
37005, 37006, 37007, 37008, 37009, 37010, 37041, 37042, 37043, 37044, 37045,
37046, 37047, 37048, 37049, 37050, 37051, 37052, 37053, 37054, 37055, 37056,
37057, 37058, 37059, 37060, 37061, 37062, 37063, 37064, 37065, 37066, 37068,
37069, 37101, 37102, 37103, 37104, 37105, 37106, 37107, 37108, 37109, 37110,
37111, 37112, 37113, 37114, 37131, 37132, 37133, 37134, 37135, 37136, 37138,
37139, 37140, 37141, 37142, 37143, 37144, 37145, 37146, 37147, 37148, 37149,
37150, 37151, 37152, 37153, 37154, 37155, 37156, 37158, 37159, 37160, 37177,
37178, 37180, 37181, 37182, 37184, 37185, 37190, 37191, 37192, 37193, 37194,
37195, 37196, 37197, 37198, 37199, 37200, 37206, 37207, 37208, 37209, 37210,
37211, 37212, 37213, 37214, 37215, 37216, 37217, 37218, 37219, 37228, 37236,
37242, 37245, 37247, 37251, 37252, 37253, 37254, 37255, 37257, 37258, 37259,
37260, 37261, 37262, 37263, 37264, 37265, 37266, 37268, 37269, 37270, 37271,
37273, 37274, 37277, 37279, 37284, 37288, 37290, 37293, 37295, 37296, 37297,
37298, 37299, 37300, 37302, 37304, 37308, 37310, 37323, 37324, 37325, 37326,
37327, 37328, 37329, 37330, 37331, 37332, 37333, 37334, 37335, 37336, 37337,
37338, 37339, 37340, 37341, 37342, 37343, 37344, 37345, 37346, 37347, 37348,
37349, 37350, 37351, 37353, 37354, 37356, 37357, 37358, 37359, 37360, 37361,
37362, 37363, 37364, 37365, 37366, 37367, 37368, 37369, 37373, 37378, 37380,
37381, 37382, 37383, 37384, 37385, 37386, 37387, 37388, 37389, 37390, 37391,
37392, 37393, 37394, 37395, 37396, 37397, 37398, 37401, 37404, 37406, 37413,
37430, 37431, 37432, 37434, 37437, 37438, 37439, 37440, 37441, 37442, 37443,
37444, 37445, 37446, 37447, 37448, 37449, 37450, 37451, 37452, 37453, 37454,
37455, 37456, 37457, 37458, 37459, 37460, 37461, 37462, 37463, 37464, 37465,
37466, 37467, 37468, 37469, 37470, 37471, 37472, 37473, 37474, 37475, 37479,
37480, 37481, 37482, 37483, 37484, 37488, 37490, 37493, 37496, 37497, 37498,
37501, 37502, 37503, 37504, 37505, 37506, 37507, 37508 37509, 37510, 37511,
37512, 37513, 37514, 37515, 37516, 37517, 37518, 37519, 37520, 37522, 37523,
37524, 37525, 37526, 37527, 37528, 37532, 37534, 37535, 37536, 37537, 37540,
37542, 37543, 37544, 37546, 37547, 37549, 37551, 37552, 37553, 37554, 37555,
37556, 37557, 37558, 37559, 37572, 37573, 37574, 37575, 37576, 37577, 37578,
37579, 37580, 37581, 37582, 37583, 37584, 37585, 37586, 37587, 37588, 37589,
37592 37600, 37615, 37616, 37620, 37621, 37622, 37626, 37630, 37632, 37633,
37634, 37636, 37638, 37639, 37640, 37641, 37642, 37643, 37644, 37645, 37646,
37647, 37648, 37649, 37650, 37651, 37652, 37655, 37658, 37659, 37663, 37664,
37665, 37667, 37670, 37671, 37672, 37673, 37674, 37676, 37677, 37678, 37679,
37681, 37682, 37683, 37684, 37685, 37686, 37687, 37688, 37689, 37690, 37691,
37692, 37693, 37694, 37695, 37696, 37697, 37698, 37699, 37700, 37701, 37702,
37703, 37704, 37705, 37706, 37707, 37708, 37709, 37710, 37711, 37712, 37715,
37716, 37717, 37718, 37719, 37720, 37721, 37723, 37724, 37726, 37727, 37728,
37729, 37730, 37731, 37732, 37733, 37734, 37735, 37736, 37737, 37738, 37739,
37740, 37741, 37742, 37743, 37744, 37745, 37746, 37747, 37748, 37749, 37750,
37751, 37752, 37753, 37754, 37755, 37756, 37757, 37758, 37760, 37761, 37763,
37764, 37765 37766, 37767, 37768, 37770, 37771, 37772, 37774, 37775, 37776,
37777, 37778, 37779, 37780, 37781, 37782, 37783, 37784, 37785, 37786, 37787,
37788, 37789, 37791, 37792, 37793, 37795, 37796, 37797, 37798, 37800, 37801,
37802, 37803, 37804, 37805, 37806, 37807, 37808, 37809, 37810, 37811, 37812,
37813, 37814, 37815, 37816, 37817, 37818, 37819, 37820, 37821, 37822, 37823,
37824, 37825, 37826, 37827, 37828, 37829, 37830, 37832, 37833, 37834, 37835,
37836, 37837, 37838, 37839, 37840, 37841, 37842, 37843, 37844, 37845, 37846,
37847, 37848, 37849, 37850, 37851, 37852, 37853, 37854, 37855, 37856, 37857,
37858, 37859, 37860, 37861, 37862, 37863, 37864, 37865, 37866, 37867, 37868,
37869, 37870, 37871, 37872, 37873, 37874, 37875, 37876, 37877, 37878, 37879,
37880, 37881, 37882, 37883, 37884, 37885, 37886, 37887, 37888, 37889, 37890,
37891, 37892, 37893, 37894, 37895, 37896, 37897, 37898, 37899, 37900, 37901,
37902, 37903, 37904, 37905, 37906, 37907, 37908, 37909, 37910, 37912, 37913,
37914, 37915, 37916, 37917, 37918, 37919, 37920, 37921, 37922, 37923, 37924,
37925, 37926, 37929, 37930, 37931, 37932, 37933, 37934, 37935, 37936, 37937,
37938, 37939, 37940, 37941, 37942, 37943, 37944, 37945, 37946, 37947, 37948,
37949, 37950, 37951, 37952, 37953, 37954, 37955, 37956, 37957, 37958, 37959,
37960, 37961, 37962, 37963, 37964, 37965, 37966, 37967, 37968, 37969, 37970,
37971, 37972, 37973, 37974, 37975, 37976, 37977, 37978, 37979, 37980, 37981,
37982, 37983, 37984, 37985, 37986, 37987, 37988, 37989, 37990, 37991, 37993,
37994, 37995, 37996, 37997, 37998, 37999, 38000, 38001, 38002, 38003, 38004,
38005, 38006, 38007, 38010, 38011, 38014, 38015, 38016, 38017, 38018, 38019,
38020, 38021, 38022, 38023, 38024, 38025, 38026, 38027, 38028, 38029, 38030,
38031, 38032, 38033, 38034, 38035, 38036, 38037, 38038, 38039, 38040, 38041,
38042, 38043, 38044, 38045, 38046, 38047, 38048, 38049, 38050, 38051, 38052,
38053, 38054, 38055, 38056, 38057, 38058, 38059, 38060, 38061, 38062, 38063,
38064, 38065, 38066, 38067, 38068, 38069, 38070, 38071, 38072, 38073, 38074,
38075, 38076, 38077, 38078, 38079, 38080, 38081, 38082, 38083, 38084, 38085,
38086, 38087, 38088, 38089, 38090, 38091, 38092, 38093, 38094, 38095, 38096,
38097, 38098, 38099, 38100, 38101, 38102, 38103, 38105, 38106, 38109, 38110,
38111, 38112, 38113, 38114, 38115, 38116, 38117, 38119, 38120, 38121, 38122,
38123, 38124, 38125, 38126, 38127, 38128, 38129, 38132, 38133, 38134, 38135,
38137, 38138, 38139, 38140, 38142, 38143, 38144, 38146, 38147, 38148, 38149,
38150, 38151, 38153, 38154, 38155, 38156, 38157, 38159, 38160, 38186, 38187,
38188, 38189, 38190, 38191, 38192, 38194, 38195, 38196, 38197, 38198, 38199,
38200, 38201, 38202, 38203, 38204, 38205, 38206, 38207, 38208, 38209, 38210,
38211, 38212, 38213, 38214, 38215, 38216, 38217, 38218, 38220, 38221, 38222,
38223, 38225, 38226, 38227, 38228, 38229, 38230, 38231, 38232, 38233, 38234,
38235, 38236, 38237, 38238, 38239, 38240, 38242, 38243, 38244, 38245, 38246,
38247, 38248, 38249, 38250, 38251, 38252, 38253, 38254, 38255, 38256, 38257,
38258, 38259, 38260, 38261, 38262, 38263, 38264, 38265, 38266, 38267, 38268,
38269, 38270, 38271, 38272, 38273, 38274, 38275, 38276, 38277, 38278, 38279,
38280, 38281, 38283, 38284, 38285, 38286, 38287, 38288, 38289, 38290, 38292,
38293, 38294, 38295, 38296, 38297, 38299, 38300, 38301, 38302, 38303, 38304,
38305, 38308, 38309, 38310, 38311, 38312, 38313, 38314, 38315, 38316, 38317,
38318, 38319, 38320, 38321, 38322, 38323, 38324, 38325, 38326, 38327, 38328,
38329, 38330, 38331, 38332, 38333, 38334, 38335, 38336, 38337, 38338, 38339,
38340, 38341, 38342, 38343, 38344, 38345, 38346, 38347, 38348, 38349, 38350,
38351, 38352, 38353, 38354, 38355, 38356,38357, 38358, 38359 38360, 38361,
38362, 38363, 38364, 38365, 38366, 38367, 38368, 38369, 38370, 38371, 38372,
38373, 38374, 38375, 38376, 38377, 38378, 38379, 38380, 38381, 38382, 38383,
38384, 38385, 38386, 38387, 38388, 38390, 38391, 38392, 38393, 38394, 38395,
38396, 38397, 38398, 38399, 38400, 38401, 38402, 38403, 38404, 38405, 38406,
38407, 38408, 38409, 38410, 38411, 38412, 38413, 38414, 38415, 38416, 38417,
38418, 38419, 38420, 38421, 38422, 38423, 38424, 38425, 38426, 38427, 38428,
38429, 38430, 38456, 38457, 38458, 38459, 38460, 38461, 38462, 38463, 38464,
38465, 38466, 38467, 38469, 38470, 38471, 38472, 38473, 38474, 38475, 38476,
38477, 38478, 38479, 38480, 38481, 38482, 38484, 38485, 38486, 38487, 38488,
38489, 38490, 38491, 38492, 38493, 38494, 38495, 38496, 38497, 38498, 38499,
38500, 38501, 38502, 38503, 38504, 38505, 38506, 38507, 38508, 38509, 38510,
38511, 38512, 38513, 38514, 38515, 38516, 38517, 38518, 38519, 38520, 38521,
38522, 38523, 38524, 38525, 38526, 38527, 38528, 38529, 38530, 38531, 38532,
38533, 38534, 38535, 38536, 38537, 38538, 38539, 38540, 38541, 38542, 38543,
38544, 38545, 38547, 38548, 38549, 38550, 38551, 38552, 38553, 38554, 38555,
38556, 38557, 38558, 38559, 38560, 38561, 38562, 38563, 38564, 38565, 38566,
38567, 38568, 38569, 38570, 38571, 38572 38573, 38574, 38575, 38576, 38577,
38578, 38579, 38580, 38581, 38582, 38583, 38584, 38585, 38586 38587, 38589,
38590, 38591, 38592, 38593, 38594, 38595, 38596, 38597, 38598, 38599, 38601,
38602, 38603, 38604, 38605, 38606, 38607, 38608, 38609, 38610, 38611, 38612,
38613, 38614, 38615, 38616, 38617, 38618, 38619, 38620, 38621, 38622, 38623,
38625, 38626, 38627, 38628, 38629, 38630, 38631, 38632, 38633, 38634, 38635,
38636, 38637, 38638, 38639, 38640, 38641, 38642, 38643, 38644, 38645, 38646,
38647, 38648, 38649, 38650, 38651, 38652, 38653, 38654, 38655, 38656, 38657,
38658, 38659, 38660, 38661, 38662, 38663, 38664, 38665, 38666, 38667, 38668,
38669, 38671, 38672, 38673, 38674, 38675, 38676, 38678, 38679, 38680, 38681,
38682, 38683, 38684, 38685, 38686, 38687, 38688, 38691, 38692, 38693, 38695,
38696, 38697, 38698, 38699, 38700, 38701, 38702, 38703, 38704, 38705, 38706,
38707, 38708, 38709, 38710, 38711, 38712, 38713, 38714, 38715, 38716, 38717,
38718, 38719, 38720, 38721, 38722, 38723, 38724, 38725, 38726, 38727, 38728,
38729, 38730, 38732, 38733, 38734, 38735, 38736, 38737, 38738, 38739, 38740,
38741, 38743, 38744, 38745, 38746, 38747, 38748, 38749, 38752, 38753, 38754,
38756, 38759, 38761, 38762, 38763, 38764, 38765, 38766, 38767, 38768, 38769,
38770, 38771, 38772, 38773, 38774, 38775, 38777, 38778, 38779, 38780, 38781,
38782, 38783, 38784, 38785, 38786, 38787, 38788, 38789, 38790, 38791, 38792,
38793, 38794, 38795, 38796, 38797, 38798, 38799, 38800, 38801, 38802, 38803,
38804, 38805, 38806, 38807, 38808, 38809, 38810, 38811, 38812, 38813, 38814,
38815, 38816, 38817, 38818, 38819, 38820, 38821, 38822, 38823, 38824, 38825,
38826, 38827, 38828, 38829, 38830, 38831, 38832, 38833, 38834, 38835, 38836,
38837, 38838, 38840, 38841, 38842 38843, 38844, 38845, 38846, 38847, 38848,
38849, 38850, 38851, 38852, 38853, 38854, 38855, 38856, 38857, 38858, 38859,
38860, 38861, 38862, 38863, 38864, 38865, 38866, 38867, 38868, 38869, 38870,
38871, 38872, 38873, 38874, 38875, 38876, 38877, 38878, 38879, 38880, 38881,
38882, 38883, 38884, 38885, 38886, 38887, 38888, 38889, 38890, 38891, 38892,
38893, 38894, 38960, 38961, 38962, 38963, 38965, 38966, 38968, 38969, 38970,
38971, 38972, 38973, 38974, 38976, 38977, 38978, 38979, 38980, 38981, 38982,
38983, 38984, 38985, 38986, 38987, 38988, 38989, 39128, 39138, 39171, 39172,
39173, 39174, 39175, 39176, 39177, 39181, 39182, 39184, 39185, 39191, 39194,
39196, 39198, 39199, 39200, 39201, 39202, 39215, 39216, 39217, 39219, 39223,
39224, 39348, 39349, 39350, 39351, 39352, 39353, 39360, 39375, 39378, 39379,
39380, 39381, 39382, 39385, 39388, 39389, 39390, 39391, 39392, 39393, 39394,
39395, 39405, 39406, 39407, 39468, 39469, 39470, 39471, 39472, 39473, 39475,
39476, 39477, 39478, 39479, 39480, 39482, 39484, 39485, 39486, 39487, 39489,
39490, 39493, 39494, 39495, 39496, 39497, 39498, 39499, 39500, 39504, 39512,
39513, 39515, 39518, 39519, 39520, 39521, 39525, 39527, 39560, 39561, 39562,
39563, 39564, 39565, 39566, 39567, 39569, 39571, 39572, 39573, 39574, 39577,
39578, 39579, 39581, 39582, 39583, 39584, 39586, 39587, 39588, 39590, 39591,
39593, 39594, 39597, 39598, 39599, 39600, 39601, 39606, 39607, 39609, 39610,
39612, 39613, 39614, 39615, 39616, 39617, 39632, 39654, 39656, 39657, 39658,
39659, 39661, 39662, 39663, 39664, 39666, 39668, 39670, 39671, 39672, 39674,
39675, 39677, 39678, 39679, 39681 and 39682;
c) a) Serial
Numbers: 14532, 14544, 14546, 14590, 14591, 14592, 14593, 14594, 14595, 14596,
14597, 14598, 14599, 14600, 14601, 14602, 14603, 14604, 14605, 14606, 14607,
14608, 14609, 14610, 14611, 14612, 14613, 14614, 14615, 14616, 14617, 14618,
14619, 14650, 14654, 14655, 14657, 14663, 14666, 14669, 14670, 14671, 14678,
14679, 14680, 14681, 14682, 14683, 14684, 14685, 14687, 14688, 14690, 14691,
14694, 14695, 14696, 14697, 14698, 14699, 14700, 14702, 14703, 14704, 14705,
14706, 14707, 14708, 14709, 15269, 15272, 15303, 15304, 15305, 15306, 15307,
15308, 15309, 15310, 15311, 15312, 15313, 15314, 15315, 15316, 15317, 15318,
15319, 15320, 15321, 15322, 15323, 15324, 15325, 15326, 15327, 15328, 15329,
15330, 15331, 15332, 15483, 15484, 15485, 15486, 15487, 15488, 15489, 15490,
15491, 15492, 15493, 15496, 15497, 15499, 15500, 15501, 15502 and 15504. b)
Serial Numbers: 11112, 11113, 11114, 11115, 11121, 11123, 11124, 11125, 11126,
11129, 11160, 11161, 11163, 11164, 11165, 11166, 11167, 11169, 11170, 11171,
11172, 11173, 11174, 11175, 11176, 11177, 11178, 11179, 11180, 11181, 11182,
11183, 11184, 11185, 11186, 11188, 11189, 11190, 11191, 11196, 11201, 11208,
11215, 11216, 11217, 11218, 11250, 11251, 11254, 11255, 11256, 11257, 11259,
11260, 11261, 11262, 11263, 11265, 11267, 11269, 11270, 11271, 11272, 11273,
11274, 11275, 11277, 11278, 11280, 11281, 11282, 11283, 11284, 11285, 11286,
11287, 11288, 11289, 11290, 11291, 11292, 11293, 11294, 11295, 11296, 11297,
11298, 11299, 11300, 11301, 11302, 11303, 11304, 11305, 11306, 11307, 11308,
11309, 11324, 11325, 11326, 11327, 11328, 11329, 11330, 11331, 11332, 11333,
11334, 11406, 11413, 11414, 11416, 11418, 11430, 11432, 11435, 11437, 11438,
11439, 11443, 11445, 11446, 11449, 11452, 11453, 11454, 11499, 11501, 11508,
11510 and 11517;
d) a) Serial
Numbers:14122, 14124, 14125, 14126, 14127, 14128, 14129, 14130, 14132, 14133,
14134, 14135, 14137, 14138, 14139, 14141, 14142, 14143, 14144, 14145, 14147,
14148, 14149, 14150, 14151, 14152, 14153, 14155, 14156, 14157, 14158, 14159,
14160, 14161, 14162, 14163, 14164, 14166, 14168, 14169, 14170, 14171, 14715,
14716, 14717, 14718, 14719, 14720, 14721, 14722, 14723, 14724, 14725, 14726,
14727, 14730, 14732, 14735, 14736, 14741, 14742, 14743, 14800, 14955, 14956,
14964, 14971, 15105, 15106, 15107, 15110, 15111, 15112, 15117, 15118 and 15120.
b) Serial Numbers: 10250, 10251, 10252, 10254, 10255, 10256, 10289, 10290,
10291, 10292, 10293, 10294, 10295, 10296, 10297, 10298, 10299, 10300, 10301,
10302, 10303, 10304, 10305, 10306, 10307, 10308, 10309, 10310, 10311, 10312,
10313, 10314, 10315, 10316, 10317, 10318, 10319, 10320, 10321, 10322, 10323,
10324, 10325, 10326, 10327, 10328, 10329, 10330, 10331, 10332, 10333, 10334,
10335, 10336, 10337, 10338, 10339, 10340, 10341, 10342, 10344, 10345, 10346,
10348, 10352, 10354, 10362, 10367, 10374, 10375, 10378, 10559, 10561, 10566,
10572, 10573, 10577, 10578, 10582, 10583, 10586, 10587, 10588, 10649, 10650,
10651, 10652, 10653, 10654, 10655, 10656, 10657, 10658, 10659, 10660, 10661,
10663, 10664, 10665, 10666, 10668, 10669, 10670, 10673, 10674, 10675, 10676,
10677, 10678, 10682, 10684, 10702, 10705, 10769, 10770, 10771, 10773 and 10774
RECALLING FIRM/MANUFACTURER Advanced Bionics LLC,
Sylmar , CA ,
REASON Mistuned headpiece - Advanced Bionics recently
became aware of a tuning issue with some Platinum Series and S-Series headpieces
shipped from May 2007 through April 2008. Tuning refers to the process through
which the headpiece's communication frequency is adjusted to match that of the
internal device. Potential clinical symptoms include: -No lock -Intermittent
lock -Implant locks, but cannot measure impedances -Reduction in battery life,
primarily for C1 patients with thick flaps.
VOLUME OF PRODUCT IN
COMMERCE Approximately 4,617 units
DISTRIBUTION Nationwide
PRODUCT
a) PROMOS Releasing Clamp (3 parts),
REF 420 189, S&N 75006600, Rx only, QTY (1), non-sterile Instrument for
prosthesis, shoulder, semi-constrained, metal/polymer cementer.
b) PROMOS Reverse, Glenosphere Loosening Adapter,
REF 24 000 047, S&N 75005271, Rx only, QTY (1), non-sterile Instrument for
prosthesis, shoulder, semi-constrained, metal/polymer cementer.
CODE
a) Lot numbers: A1, A1B1107, A2,
A3B1107, A3, A3B1107, A4, A4/I, A51767, A51767B706, B57815, B58460, B60400, and
B62141;
b) Lot numbers: A55827, A59222, A62005, A62127, A62349,
A63526, and A63779
RECALLING FIRM/MANUFACTURER
Recalling Firm: Smith & Nephew Inc., Memphis , TN ,
Manufacturer: Smith & Nephew Orthopaedics
Ag, Rotkreuz , Switzerland .
REASON The product was breaking during use.
VOLUME OF PRODUCT IN COMMERCE 159 Clamps and 51
Adapters
DISTRIBUTION International distribution only.
Germany , France , Italy Austria , Switzerland , Belgium , United Kingdom ,
Greece , Australia , South Africa , Spain , and Cypress
PRODUCT Biomet Oxford Unicompartmental Knee, Phase 3, Right Medical Tibial
Tray, Standard Size B, CoCr, sterile. Unicompartmental knee joint replacement
system. Designed to allow for motion between the plastic tibial bearing and
metal tibial baseplate components of a unicompartmental knee prosthesis.
CODE
Lot number: 1563796
RECALLING FIRM/MANUFACTURER
Recalling Firm:
Biomet,
Inc., Warsaw , IN ,
Manufacturer: Biomet U.K. , Ltd, Bridgend, South Wales, UK . ng.
REASON
Packages labeled as “right”
actually contain a “left” component.
VOLUME OF PRODUCT IN COMMERCE
2 units
DISTRIBUTION MA and WV
PRODUCT
a) NFix II System 150 mm, Catalog number NX60150-1 spinal fusion
surgery.
b) NFix II System 200 mm, Catalog
number NX60200-1. Spinal fusion surgery.
c) N-Hance System 150 mm, Catalog number NH60150-1. Spinal fusion surgery,
d) TI
N-Hance Rod 150 mm Sterile, Catalog number
04.600.715.02S Spinal fusion surgery.
e)
NFlex Stabilization System 200 mm, Catalog number NF60020-1. Spinal fusion
surgery.
f) NFlex Stabilization System 150 mm, Catalog number NF60150-1.
Spinal fusion surgery.
CODE
a) Lot
numbers: P06J01 exp 9/30/07; P06J02 exp 9/30/07; P06M01 exp 12/31/07; P06M02 exp
12/31/07; P07D02 exp 4/30/08; P07H01 exp 8/31/08; and P07J01 exp 9/30/08;
b) Lot number: P07G01 exp 7/31/08;
c) Lot number:
P07M01 exp 12/31/08;
d) Lot number: 5764185 exp 5/20/2013;
e) Lot number: NF60003 exp 3/31/07;
f) Lot numbers: P07M01
exp 12/31/08; P07M03 exp 12/31/08; P06H01 exp 8/31/07; P06M01 exp 12/31/07; and
P06J01 exp 9/30/07
RECALLING FIRM/MANUFACTURER
Recalling Firm: Synthes Spine, West Chester , PA ,
Manufacturer:
Synthes ( USA ), Monument, CO. .
REASON The dimensional specifications are incorrect.
VOLUME OF PRODUCT IN COMMERCE 1,508 units
DISTRIBUTION Nationwide and countries of Germany ,
Denmark , Switzerland and Korea
PRODUCT GE Centricity PACS RA1000 Workstation
(for diagnostic image analysis). The Centricity PACS (Picture Archiving and
Communication System) Workstation is intended for use as a primary diagnostic
and analysis tool for diagnostic images by trained healthcare professionals,
including radiologists, physicians, technologists, clinicians and nurses. It is
also intended for use as a clinical review workstation throughout the healthcare
facility. The workstation interface provides the user with a means to display,
manipulate, archive, print, and export images when connected with the Centricity
PACS infrastructure.
CODE Software
versions 2.1.X and 3.0.X and higher with Talk Extend
RECALLING
FIRM/MANUFACTURER GE Healthcare Integrated IT Solutions, Barrington
, IL ,
REASON Unread exams may inadvertently get marked as
dictated in PACS workstation when the user signs off the report in Talkstation
and opens the next exam.
VOLUME OF PRODUCT IN COMMERCE 22 units
DISTRIBUTION
CA, FL, IL, IA, LA,
OR, PA, TX, UT, VA, and WI
WEEK ENDING FEBRUARY 28
PRODUCT
a) MicroScan Synergies Plus Dried Overnight Gram Negative MIC/Combo Panels, 20
panels per box. Medical device designed for use in determining quantitative and/or
qualitative antimicrobial agent susceptibility and/or identification to the species
level of colonies, grown on solid media, of rapidly growing aerobic and facultative
anaerobic gram-negative bacteria. Catalog numbers B1016-100, B1016-101, B1016-102,
B1016-105, B1016-107, B1016-108, B1016-109, B1016-119, B1016-120, B1016-121,
B1016-122, B1016-126, B1016-127, B1016-128, B1016-129, B1016-131, B1016-132,
B1016-61a, B1016-62a, B1016-64a, B1016-65a, B1016-80, B1016-82, B1017-300, B1017-301,
B1017-302, B1017-303, B1017-304, B1017-305, B1017-306, B1017-307, B1017-308,
B1017-309, B1017-312, B1017-313, B1017-315, B1017-400, B1017-401, B1017-402,
B1017-403, B1017-404, B1017-405, B1017-406, B1017-407, J1016-311c, J1016-312c,
J1016-611c, J1016-611j, J1016-612b, J1016-612J, J1016-632J, J1016-623m, J1016-631J,
J1016-632J, J1016-65c, J1016-66J, J1025-311, J1025-312, J1025-333, and J1025-611.
b) MicroScan Synergies Plus Negative Panels, 20 panels per box Medical device
designed for use in determining quantitative and/or qualitative antimicrobial
agent susceptibility and/or identification to the species level of colonies,
grown on solid media, of rapidly growing aerobic and facultative anaerobic
gram-negative bacteria. Catalog numbers: B1016-200, B1016-201, B1016-202, B1016-203,
B1025-103, B1025-104, B1025-105, B1025-106, B1025-107, B1025-108, B1025-109,
B1025-112, and B1027-101.
CODE
Unknown
RECALLING FIRM/MANUFACTURER
Siemens Microscan, West Sacramento, CA,
REASON
Falsely lowered imipenem MICs may result, leading to major errors in results.
VOLUME OF PRODUCT IN COMMERCE
569,375 panels
DISTRIBUTION
Nationwide and Internationally
PRODUCT
a) Ossur 3'" Titanium Skull Pin, 1 pack, Model Number: 516V2S.
b) Ossur 3" Ceramic Tip Pin, 4 pack, Model Number: 517DS.
c) Ossur 2 1/4" Ceramic Tip Pin, 4 pack, Model Number: 522CS.
CODE
a) Lot Number: 08060410;
b) Lot Number: 08060408;
c) Lot Number: 08060305
RECALLING FIRM/MANUFACTURER
Recalling Firm: Ossur, Reykjavik, Ireland,
Manufacturer: Ossur Engineering, Inc., Albion, MI.
REASON
The firm is unable to guarantee the sterility of the affected products.
VOLUME OF PRODUCT IN COMMERCE
43 pins
DISTRIBUTION
Nationwide
PRODUCT
a) ReSolve Hydrophilic Coated Locking Biliary Catheter, REF RBC-8-038, 8.5
F, Sterile EO. Used for drainage of bile within the biliary system.
b) ReSolve Hydrophilic Coated Locking Biliary Catheter, REF RBC-10-038, 10F, Sterile EO. Used for drainage of bile within the biliary system.
c) ReSolve Hydrophilic Coated Locking Biliary Catheter, 12F, REF RBC-12-038, Sterile EO. Used for drainage of bile within the biliary system.
d) ReSolve Hydrophilic Coated Locking Biliary Catheter, REF RBC-14-038,
14F, Sterile EO. Used for drainage of bile within the biliary system
CODE
a) Lot numbers: G543418, MR604367, and MR607897;
b) Lot numbers: G543416, G625227, MR604361, MR604362, MR604363, MR604364, and MR607895;
c) Lot numbers: G543417, MR604365, and MR607896;
d) Lot numbers: G543419, G625228, and MR604366
RECALLING FIRM/MANUFACTURERN
Merit Medical Systems, Inc., South Jordan, UT,
REASON
Biliary drainage catheter tips may separate from the catheter shaft after placement.
VOLUME OF PRODUCT IN COMMERCE
1,164 units
DISTRIBUTION
Nationwide and countries of Canada, Colombia, France, Germany, Mexico, Ireland,
Netherlands, South Korea, and UK
PRODUCT
a) Terumo TenderFlow Pediatric Arterial Cannulae, 2.0 mm (6Fr) OD, with 1/4
inch vented connector, 9 inch (22.9 cm) long; Catalog no. 813567. The TenderFlow
(TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending
aorta during cardiopulmonary bypass surgery. These devices are indicated for
up to 6 hours of use.
b) Terumo TenderFlow Pediatric Arterial Cannulae, 2.0mm (6Fr) OD, with 1/4 inch non-vented connector, 9 inch (22.9 cm) long; Catalog no. 813568. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
c) Terumo TenderFlow Pediatric Arterial Cannulae, 2.7mm (8 Fr) OD, with 1/4 inch vented connector, 9 inch (22.9 cm) long; Catalog no. 813569. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
d) Terumo TenderFlow Pediatric Arterial Cannulae, 2.7mm (8Fr) OD, with 1/4 inch non-vented connector, 9 inch (22.9 cm) long; Catalog no. 813570. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
e) Terumo TenderFlow Pediatric Arterial Cannulae, 3.3 mm (10Fr) OD, with 1/4 inch vented connector, 9 inch (22.9 cm) long; Catalog no. 813571. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
f) Terumo TenderFlow Pediatric Arterial Cannulae, 3.3mm (10 Fr) OD, with 1/4 inch non-vented connector, 9 inch (22.9 cm) long; Catalog no. 813572. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
g) Terumo TenderFlow Pediatric Arterial Cannulae, 4.0 mm (12 FR) OD, with 1/4 inch vented connector, 9 inch (22.9 cm) long; Catalog no. 813573. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
h) Terumo TenderFlow Pediatric Arterial Cannulae, 4.0mm (12 Fr) OD, with 1/4 inch non-vented connector, 9 inch (22.9 cm) long; Catalog no. 813574. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
i) Terumo TenderFlow Pediatric Arterial Cannulae, 4.7 mm (14Fr) OD, with 1/4 inch vented connector, 9 inch (22.9 cm) long; Catalog no. 813575. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
j) Terumo TenderFlow Pediatric Arterial Cannulae, 4.7mm (14 Fr) OD, with 1/4 inch non-vented connector, 9 inch (22.9 cm) long; Catalog no. 813576. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
k) Terumo TenderFlow Pediatric Arterial Cannulae, 5.3 mm (16 Fr) OD, with 1/4 inch vented connector, 9 inch (22.9 cm) long; Catalog no. 813577. The TenderFlow (TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
l) Terumo TenderFlow Pediatric Arterial
Cannulae, 5.3mm (16 Fr) OD, with 1/4
inch non-vented connector, 9 inch (22.9 cm) long; Catalog no. 813578. The TenderFlow
(TM) Pediatric Arterial Cannula is indicated for use in perfusion of the ascending
aorta during cardiopulmonary bypass surgery. These devices are indicated for
up to 6 hours of use.
CODE
a) Lot numbers: 0501557 and 0512008;
b) Lot numbers: 0501639 and 0512002;
c) Lot numbers: 0502199, 0513068, and 0517852;
d) Lot numbers: 0502200 and 0513069;
e) Lot numbers: 0507120, 0512919, and 0518680;
f) Lot numbers: 0507121 and 0512918;
g) Lot numbers: 0509238 and 0517874;
h) Lot number: 0509240;
i) Lot numbers: 0500759 and 0513331;
j) Lot number: 0509241;
k) Lot number: 0502878;
l) Lot number: 0500760
RECALLING FIRM/MANUFACTURER
Terumo Cardiovascular Systems Corp., Ann Arbor, MI,
REASON
The introducer or stylet is difficult to remove from the pediatric arterial
cannula body, which may result in the need to remove the cannula and replace
it.
VOLUME OF PRODUCT IN COMMERCE
2,892 units
DISTRIBUTION
Nationwide and countries of Australia, Belgium and Canada
PRODUCT
Drive Belt for Radiation Therapy Simulation systems SIMVIEW 3000 and SIMVIEW
NT with CT option or IDI. Radiation therapy simulation system. NT with model
numbers 1957153, 1957161, 5496992, 5497016, 5496984.
CODE
All drive belts in SIMVIEW 3000 and NT Models.
RECALLING FIRM/MANUFACTURER
Siemens Medical Solutions USA, Inc., Concord, CA,
REASON
Excessive wear on drive belts used for Image Intensifier and/or X-ray head
drive mechanisms may result in uncontrolled movement, which may cause injury.
VOLUME OF PRODUCT IN COMMERCE
223 units
DISTRIBUTION
Nationwide and Internationally
PRODUCT
Philips M8000-64100 Table Mount Shipped standard with all IntelliVue Patient
Monitors MP40-70, Gas Modules Gl and G5 and certain options of the IntelliVue
Patient Monitors MP5 MPST and MP20/30. Product shipped between 6/03/08 and
9/12/08 as follows: M1013A IntelliVue G1 Gas Module M1019A IntelliVue GS Gas
Module M8001A option E22 Quick release mount for IntelliVue MP20 M8002A option
E22 Quick release mount for IntelliVue MP30 M8003A IntelliVue MP40 M8004A IntelliVue
MPSO M8005A IntelliVue MP60 M8007A IntelliVue MP70 M8105A option E22 IntelliVue
MP5 Quick Release Mount M8105AT option E22 IntelliVue MPST Quick Release Mount
M8040A option U02 U02 Table top mount kit M8040A option E22 Quick release kit
for MP20MP30 The Philips IntelliVue Patient Monitors are intended for monitoring
and recording of and to generate alarms for, multiple physiological parameters
of adults, pediatrics and neonates in hospital environments. The MP2, X2, MP5,
MP20, MP30, MP40, and MP50 are additionally intended for use in transport situations
within hospital environments.
CODE
Product shipped between 6/03/08 and 9/12/08. A list of S/N's can be obtained
from the recalling firm.
RECALLING FIRM/MANUFACTURER
Recalling Firm: Philips Medical Systems, Andover, MA,
Manufacturer: Philips Medizin Systeme Boblingen Gmbh, Boblingen, Germany.
REASON
Table Mounts assembled with incorrect length hardware may fall.
VOLUME OF PRODUCT IN COMMERCE
14,292 table mounts
DISTRIBUTION
Nationwide, Canada, Argentina, Aruba, Australia, Austria, Bahain, Bangladesh,
Belgium, Bosnia, Brazil, Chile, China, Colombia, Czech Republic, Denmark, Estonia,
Egypt, Finland, France, Gabon, Germany, Great Britain, Greece, Hong Kong, Iceland,
India, Iraq, Ireland , Italy, Japan, Korea, Libyan Arab Jamahiriya, Lithuania,
Luxembourg, Malaysia, Namibia, Netherlands, Norway, Peru, Philippines, Poland,
Portugal, Russia, Saudi Arabia, Singapore, Slovenia South Africa, Spain,
Srilinka,
Sweden, Switzerland, Thailand, Tunisia, Turkey, Uganda, UAE, Uzbekistan, Venezuela,
Vietnam, and Zambia
PRODUCT
Atlantis PV Peripheral Imaging Catheter, model number H749364560, catalog number
36456, manufactured by Boston Scientific Corporation, Fremont, CA. Medical
device, intended for ultrasound examination of peripheral pathology. Intravascular
ultrasound imaging is indicated in patients who are candidates for transluminal
interventional procedures.
CODE
Batch/lot numbers: 11420600, 11420601, 11420602, 11611665, 11716746, 11720663,
11721240, 11722752, 11727107, 11727873, 11727874, 11728458, 11731354, 11732665,
11735762, 11736559, 11739563, 11742590, 11754455, 11758051, 11764281, 11777994,
11788045, 11790848, 11820112, 11897802, 11897904, and 11904295
RECALLING FIRM/MANUFACTURER
Boston Scientific Corp., Fremont, CA,
REASON
Product is mislabeled as having a shelf life of 3 years, where the actual shelf
life is 1 year.
VOLUME OF PRODUCT IN COMMERCE
1,033 units
DISTRIBUTION
Nationwide, PR and Panama
PRODUCT
Varian Medical Systems Eclipse treatment planning system, Model Number H48,
versions 6.5/7.x and Inspiration 8.0.
CODE
Serial numbers: H480002 H481912 H483369 H484526 H485316 H485593 H485859 H486079
H486328 H486602 H480005 H482181 H483429 H484532 H485334 H485594 H485862 H486086
H486330 H486606 H480016 H482186 H483471 H484587 H485338 H485603 H485867 H486089
H486331 H486613 H480092 H482192 H483485 H484593 H485340 H485615 H485871 H486092
H486332 H486617 H480112 H482194 H483488 H484635 H485342 H485629 H485873 H486094
H486336 H486618 H480114 H482250 H483560 H484661 H485345 H485632 H485874 H486097
H486338 H486629 H480132 H482270 H483566 H484663 H485347 H485633 H485875 H486108
H486342 H486633 H480169 H482294 H483570 H484692 H485352 H485642 H485883 H486120
H486346 H486639 H480293 H482324 H483586 H484709 H485359 H485646 H485901 H486121
H486348 H486641 H480294 H482427 H483715 H484763 H485362 H485648 H485912 H486123
H486349 H486642 H480300 H482462 H483724 H484767 H485363 H485656 H485913 H486127
H486356 H486653 H480427 H482523 H483731 H484791 H485370 H485668 H485918 H486128
H486369 H486665 H480433 H482528 H483756 H484813 H485374 H485669 H485926 H486130
H486385 H486667 H480446 H482541 H483764 H484838 H485375 H485670 H485933 H486132
H486411 H486680 H480467 H482557 H483769 H484841 H485380 H485671 H485934 H486134
H486412 H486685 H480481 H482579 H483797 H484847 H485382 H485672 H485936 H486136
H486440 H486686 H480514 H482677 H483799 H484857 H485388 H485676 H485941 H486138
H486444 H486689 H480570 H482678 H483800 H484866 H485391 H485681 H485946 H486140
H486445 H486694 H480583 H482699 H483808 H484875 H485393 H485694 H485947 H486143
H486451 H486700 H480670 H482706 H483827 H484876 H485414 H485699 H485948 H486188
H486452 H486701 H480723 H482710 H483879 H484950 H485421 H485701 H485949 H486198
H486457 H486703 H480759 H482722 H483921 H484954 H485433 H485703 H485953 H486202
H486473 H486707 H480780 H482725 H483937 H484970 H485437 H485709 H485963 H486205
H486474 H486708 H480789 H482727 H483964 H484975 H485441 H485715 H485964 H486206
H486481 H486715 H480809 H482761 H483980 H484977 H485443 H485717 H485966 H486207
H486488 H486717 H480822 H482769 H484013 H484979 H485447 H485739 H485967 H486223
H486490 H486721 H480889 H482821 H484103 H484988 H485449 H485749 H485971 H486224
H486498 H486724 H480890 H482848 H484186 H485014 H485454 H485751 H485972 H486228
H486500 H486756 H480996 H482892 H484200 H485024 H485463 H485753 H485981 H486230
H486501 H486758 H481003 H482920 H484235 H485026 H485468 H485771 H485983 H486233
H486503 H486759 H481154 H482934 H484250 H485035 H485472 H485776 H485987 H486236
H486509 H486766 H481165 H482970 H484256 H485039 H485475 H485778 H485988 H486247
H486520 H486793 H481193 H482997 H484263 H485043 H485477 H485789 H486003 H486257
H486522 H486803 H481203 H483060 H484283 H485081 H485485 H485794 H486008 H486260
H486525 H486804 H481215 H483070 H484294 H485098 H485488 H485795 H486010 H486271
H486531 H486805 H481220 H483086 H484308 H485129 H485489 H485798 H486016 H486273
H486541 H486806 H481296 H483109 H484311 H485136 H485491 H485802 H486023 H486276
H486543 H486809 H481308 H483146 H484317 H485143 H485517 H485804 H486024 H486278
H486549 H486820 H481309 H483176 H484323 H485148 H485519 H485805 H486025 H486288
H486550 H486826 H481445 H483178 H484332 H485165 H485533 H485806 H486029 H486289
H486555 H486831 H481497 H483182 H484336 H485193 H485537 H485812 H486032 H486292
H486556 H486840 H481551 H483200 H484346 H485236 H485538 H485815 H486034 H486294
H486558 H486848 H481562 H483203 H484377 H485240 H485541 H485816 H486038 H486295
H486561 H486856 H481667 H483210 H484419 H485243 H485543 H485820 H486040 H486298
H486566 H486866 H481685 H483243 H484430 H485291 H485545 H485823 H486046 H486299
H486567 H486871 H481737 H483248 H484439 H485292 H485546 H485827 H486058 H486300
H486570 H486872 H481744 H483273 H484450 H485297 H485550 H485828 H486062 H486301
H486579 H486875 H481787 H483279 H484472 H485301 H485551 H485831 H486066 H486313
H486589 H486893 H481835 H483321 H484475 H485308 H485556 H485834 H486067 H486316
H486592 H486899 H481870 H483327 H484506 H485309 H485564 H485844 H486072 H486317
H486597 H486902 H481906 H483329 H484510 H485310 H485568 H485850 H486075 H486327
H486600 H486907 H486908 H486914 H486924 H486949 H486973 H486975
RECALLING FIRM/MANUFACTURER
Varian Medical Systems oncology Systems, Palo Alto, CA,
REASON
Possible incorrect wedge orientation in treatment: Software labeling was issued
without warning message relating to preservation of user selected values for
Siemens wedge orientation, wedge limits and SAD values for Elekta machines,
and collimator operating limits for Saturne Machines when the software version
is upgraded.
VOLUME OF PRODUCT IN COMMERCE
516 units
DISTRIBUTION
Nationwide, Austria, Belgium, Canada, Czech Republic, Finland, France, Germany,
Greece, Hungary, Italy, Ireland, Lithuania, Norway, Portugal, Romania, Slovakia,
Spain, Sweden, Switzerland, Netherlands, Turkey and the United Kingdom.
PRODUCT
Stainless Steel Surgical Blades Part number: 4-312B. The product is supplied
sterile, individually packaged in boxes containing 100 blades.
CODE
Lot number: S07C06
RECALLING FIRM/MANUFACTURER
Recalling Firm: Miltex, Inc., York, PA,
Manufacturer: Kai Surgical Instrument Co., Ltd., Seki City, Japan.
REASON
Blade may break; Over milling of the blade resulted in tip more narrow than
specified.
VOLUME OF PRODUCT IN COMMERCE
200 boxes
DISTRIBUTION
CA, FL, GA, IN, NE, NJ, NY, PA, SC, TN, TX and Canada
PRODUCT
Monitor portion of the LifeVest device – specifically the Defibrillator
printed circuit assembly (PCA) within the monitor. The LifeVest is indicated
for adult patients who are at risk for sudden cardiac arrest and are not candidates
for or refuse an implantable defibrillator. Model numbers: WCD3000 and WCD3100.
CODE
Unknown
RECALLING FIRM/MANUFACTURER
Zoll Lifecor Corp., Pittsburgh, PA,
REASON
The charging circuit will not shut down properly.
VOLUME OF PRODUCT IN COMMERCE
1,685 units
DISTRIBUTION
Nationwide, Germany, United Kingdom, Italy and France
PRODUCT
Regenesis Biomedical PROVANT Wound Therapy System, Model 4201 - The Provant
Wound Therapy System is a solid-state, fixed-power output radio frequency generator
and transmitter designed to operate at the FCC authorized medical device frequency
of 27.12 MHz. Provant transmits a fixed dose of non-ionizing, non-thermal RF
energy via an applicator pad that is placed adjacent to the patient's dressed
wound. The Provant Wound Therapy System is indicated for adjunctive use in
the palliative treatment of postoperative pain and edema in superficial soft
tissue.
CODE
Serial numbers: 704, 920, 460, 709, 711, 714, and 727
RECALLING FIRM/MANUFACTURER
Regenesis Biomedical, Inc., Scottsdale, AZ, .
REASON
In the course of device testing, the firm determined that the affected devices
may, under certain unusual clinical circumstance, be unable to generate a therapeutic
dose of radio frequency energy. As such, for these few devices, the effectiveness
of Provant therapy may be impaired.
VOLUME OF PRODUCT IN COMMERCE
7 units
DISTRIBUTION
WY, VA and NC
PRODUCT
a) Femoral Array, Model number: 0150ARR00001 and b) Tibial Array, Model number:
0150ARR00002. The devices are accessories used with MAKO's Tactile Guidance
System (TGS). The Tactile Guidance System is intended to assist the surgeon
in providing software defined spatial boundaries for orientation and reference
information to anatomical structures during orthopedic procedures.
CODE
a) Lot Number/Quantity: 0609038 (6), 06051107 (4), 01010806 (4), 09010108 (2),
06040707-01 (1), 01010107 (3), 01010607 (1), 01010106 (1) and 06040707 (2)
and
b) Lot Number/Quantity: 06060308 (6), 06061107 (7), 01020866 (2), 01010507
(1), 01020806 (1), 01020107 (3), 09081107 (2) and 01020607 (2)
RECALLING FIRM/MANUFACTURER
Mako Surgical Corp., Davie, FL,
REASON
The product in its current state may fail during use, which would cause the
instrument to be in-accurate.
VOLUME OF PRODUCT IN COMMERCE
44 units
DISTRIBUTION
Nationwide