AUGUST 2006

WEEK ENDING AUGUST 5

PRODUCT Welch Allyn PIC 50 Portable Intensive Care System Multi-Parameter Monitor/Defibrillator; a portable, 12V internal battery powered defibrillator, Part numbers 971081, 971082, 971083 and 971084, 
CODE
Serial numbers 80104, 80119, 80120, 80204, 80211, 80212, 80429, 80432, 80569, 80627, 80661, 80685, 80880, 80964, 81056, 81134, 81263, 81293/81295, 81303, 81445/ 81448, 81733, 81738, 81739, 81744/81752, 81757, 81761/81781, 81802/81807, 81824, 81825, 81835, 81845, 81846, 81852, 81864, 81865, 81868, 81869, 81884/81886, 81892/81897, 81918/81955, 81957, 81983/81990, 81996/82025, 82027/82033, 82035/82039, 82043, 82044, 82061/82080, 82084, 82097/82111, 82122/82128, 82133, 82136/82143, 82150/82169, 82172/82204, 82245, 82252/82273, 82283/82286, 82307/82330, 82340/82349, 82405/82426, 82430, 82460/82490, 82498/82553, 82555/82565, 82568/82570, 82612/82631, 82654/82673, 82689, 82690, 82773/82813, 82827/82830, 82852, 82856/82859, 82906, 82936, 82937, 82958/82977, 82983/83013, 83060/83071, 83083/83090, 83093, 83142/83147, 83153/83172, 83198/83239, 83255, 83368/83370, 83400/83407, 83414/83418, 83420/83429, 83440/83443, 83499/83517, 83578/83580, 83633, 83644, 83645, 83654, 85001/85005, 85009/85013, 85026/85029, 85032/85052, 85073/85079, 85093, 85094, 85110, 85111, 85143/85157, 85187/85201, 85223, 85224, 85226/85238, 85244/85249, 85329, 85332/85334, 85337/85350, 85364, 85365, 85368/85370, 85381, 85388/85398, 85402/85404, 85410/85414, 85420/85433, 85441, 85505/85515, 85517/85520, 85522/85530, 85534, 85552, 85554/85557, 85587/85589, 85602/85607, 85619/85623, 85630, 85645/85655, 85658/85667, 85686, 85728/85734, 85755/85757, 85761/85766, 85771/85785, 85787/85793, 85810, 85812/85823, 85825/85838, 85856/85872, 85877/85884, 85888/85902, 85931/85933, 85945, 85955, 85964, 85967/85970, 86017, 86018, 86024/86029, 86055/86057, 86103, 86104, 86118, 86141/86147, 86149, 86150, 86222/86228, 86230, 86245, 86256/86258, 86261/86268, 86281/86283, 86289/86293, 86295, 86296, 86299, 86305, 86306, 86314, 86315, 86320/86322, 86324/86328, 86334, 86335, 86358, 86359, 86385, 86387, 86389/86399, 86400, 86411, 86412, 86419, 86420, 86439, 86440
RECALLING FIRM/MANUFACTURER
Recalling Firm: Welch Allyn Inc., Skaneateles Falls, NY, 
Manufacturer: MRL, Inc., A Welch Allyn Company, Buffalo, IL. 
REASON The Welch Allyn PIC 50 Defibrillators may display a "Defib Comm" or "Pace Comm" error message on the device display during use which may result in a terminal failure of the device to analyze the patient’s ECG and deliver the appropriate therapy.
VOLUME OF PRODUCT IN COMMERCE 1184 units
DISTRIBUTION Nationwide and Internationally

PRODUCT
a) MD3 X-ray system, 
b) MD4 X-ray system, 
CODE
a) Site numbers 59515, 38929, 76329, 83654, 82673, 35471, 59524,
86146, 38922, 101151, 38714, 38716, 44865, 6847, 103855, 26071,
103856, 40941, 83047, 62502, 38266, 38422, X1198, 50317, 38289,
17669, 6931, X0917, 26019, 26020, 13661, 38418, 59571, 41462,
82721, 101782, 103209;
b) Site numbers 102550 (20279385), 103023, 38783, 102562 (20279430),
83957, 101155, 82628, 38442, 83038, 13721, 85034, 102583, 80306,
41156, 38993, 84930, X1321, 41032, X1077, 100014, 50328, 17656,
83367, 38586, 100954, 76834, 87176, 41426, 104137, 82800, 103651,
10388, 41111, 17622, 13918, 103377, 101038, X0934, 102034, 52471,
76804, 50251, 44898, 13258, 86545, 101863, 38587, 103043 (20283876),
82688, 38635, 47160, 14099, 45042, 38917, 101156 (20278521), 86328,
38501, 13879, 41439, 41069, 76975, X1277, 100637, X1078, 35484,
101759, 84463, 105022, X1751, 76877, 13172, 38944, 41143, 85694,
87140, 59546, 13704, 10458, 45039, 76869, 52455, 84892, 41271,
87302, 6842, 41031, 41000, X0662, 87030, 76808, 6838, 62938, 86218,
85491, 50340, 82687, 38912, 101153, 103210
RECALLING FIRM/MANUFACTURER
Recalling Firm: Philips Medical Systems North America Co. Phillips, Bothell, WA
Manufacturer: Philips Medical Systems, Best, Netherlands. 
REASON Potential for unexpected table movement.
VOLUME OF PRODUCT IN COMMERCE 137 units
DISTRIBUTION Nationwide

PRODUCT
a) Kimberly-Clark/Ballard medical R2 triangle shaped non-radiolucent
multifunction electrode products for pediatrics. Catalog number 30-595,
Catalog number 3100-1700, Catalog number 3111-1720,
Catalog number 3112-1730, Catalog number 3114-1740,
Catalog number 3200-1711, 
b) Kimberly-Clark/Ballard medical R2 triangle shaped non-radiolucent
multifunction electrode products for adults, Catalog number 30-585,
Catalog number 30-590, Catalog number 30-591, Catalog number 30-593,
Catalog number 3100-1702, Catalog number 3100-703,
Catalog number 3111-1721, Catalog number 3112-1731,
Catalog number 3114-1741, Catalog number 3200-1715,
Catalog number 3250-1780, 
CODE
a) Lot number 288003; Lot numbers 284026 and 299215;
Lot numbers 288963, 288969, 290317, 291534, 293216, 294205,
295161, 296269, 298031, 300851, 301438, 303002, 304752, 305508;
Lot number 282182, 286334, 286341, 290319, 290906, 291536, 293219,
294865, 295602, 296272, 298034, 299217, 301440, 302660, 303005, 305511;
Lot numbers 290321, 297265 and 302661; Lot numbers 282187, 282941,
288973, 290910, 291540, 292666, 293223, 294209, 294869, 295606,
297268, 298037, 299990, 300856, 301443, 302664, 303007, 305514;
b) Lot number 300847; Lot numbers 28802, 297261, 304750;
Lot number 299214; Lot number 288962, 293213, 294860, 302657;
Lot number 297735; Lot number 282936, 293214, 300849, 303000;
Lot numbers 282181, 282938, 286332, 286339, 288964, 288970,
290318, 290905, 291535, 292660, 293217, 294206, 294862, 295600,
296270, 297263, 298032, 299985, 302659, 303003, 304091, 304753,
305509; Lot numbers 282183, 282939, 285955, 286335, 286342, 290320,
290907, 291537, 292662, 293220, 294208, 294866, 295603, 296273,
299218, 299987, 300853, 304093, 305512; Lot numbers 282185, 284030,
290322, 291538, 292664, 295604, 297266, 300854, 304095;
Lot numbers 282188, 282942, 284031, 288004, 288005, 288966,
288967, 288974, 288975, 290911, 291541, 292667, 293224,
294210 294870, 295607, 296276, 297269, 298038, 299221,
299991, 301444, 302665, 303008, 304097, 304757, 305515;
Lot numbers 292086, 297270, 301445, 302666
RECALLING FIRM/MANUFACTURER Kimberly-Clark Corporation, Roswell, GA
REASON The tinfoil in the triangle electrodes has the potential to develop cracks, which may hinder electrical flow and product performance during defibrillation.
VOLUME OF PRODUCT IN COMMERCE 95,639 electrode sets
DISTRIBUTION Nationwide and Internationally

PRODUCT SoftPath ASXII Software Releases 2.3.0 and 4.3.7. The software is used in pathology labs for administrative and clinical documentation and data processing, 
CODE Software Releases 2.3.0 and 4.3.7
RECALLING FIRM/MANUFACTURER SCC Soft Computer, Palm Harbor, FL
REASON Text from one case is overwriting another. This could cause an incorrect diagnostic report to be sent to a clinician/surgeon.
VOLUME OF PRODUCT IN COMMERCE 4 units
DISTRIBUTION Nationwide and Canada

PRODUCT
a) Hamilton brand RAPHAEL Ventilator (Software Version 2.2x),
b) Hamilton brand RAPHAEL Silver Ventilator (Software Version 2.2xS),
c) Hamilton brand RAPHAEL Color Ventilator (Software Version 2.2xC, 2.2xCU),
CODE a), b) and c) Serial Numbers: 2975 to 5360
RECALLING FIRM/MANUFACTURER
Recalling Firm: Hamilton Medical, Inc., Reno, Nevada, 
Manufacturer: Hamilton Medical AG, Bonaduz, Switzerland. 
REASON Alarm Failure -- Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.
VOLUME OF PRODUCT IN COMMERCE 2,385 units
DISTRIBUTION Nationwide and Internationally

PRODUCT SoftPath ASCII Software, used in pathology labs to track and report specimens and diagnostic results. Releases 1.2, 2.1, and 2.3, 
CODE Software Releases 1.2, 2.1, and 2.3
RECALLING FIRM/MANUFACTURER SCC Soft Computer, Palm Harbor, FL
REASON The client had made a decision to use the SoftPath Module in a nonstandard manner. User specific circumstances, incorrect text may appear on a patient report.
VOLUME OF PRODUCT IN COMMERCE 171
DISTRIBUTION Nationwide and Canada

PRODUCT ExacTrac Robotics, sold alone as Robotic Tilt Module catalog 49720 and in a bundle as catalog 49700, for installation on the Varian Exact Couch, a component of the Novalis Shaped Beam Surgery System; a table tilt device - powered radiation therapy patient support assembly, 
CODE Bundle #06-69448-49700, tilt module #3520-05-011-49720; bundle #06-69445-49700, tilt module #3520-05-006-49720; bundle #06-78573-49700, tilt module #3520-05-017-49720; bundle #06-69573-49700, tilt module #3520-05-002-49720; bundle #06-78149-49700, tilt module #3520-05-015-49720; bundle #06-66427-49700, tilt module #3520-05-013-49720; bundle #06-62416-49700, tilt module #3520-05-008-49720; tilt module #3520-05-018-49720; bundle #06-77290-49700, tilt module #3520-05-019-49720; bundle #06-77669-49700, tilt module #3520-05-020-49720; bundle #06-78440-49700, tilt module #3520-05-014-49720; bundle #06-74601-49700, tilt module #3520-05-010-49720; bundle #06-64377-49700, tilt module #7520-05-005-49720
RECALLING FIRM/MANUFACTURER Brainlab AG, Kirchheim B. Muenchen, Germany, 
REASON During ExacTrac Robotics installation on the Varian Exact Couch, the couch height position indication is re-calibrated to read out the correct positions. This re-calibration could cause the eventual failure of the vertical lift mechanism of the couch, which might result in patient injury or death.
VOLUME OF PRODUCT IN COMMERCE 19 devices
DISTRIBUTION Nationwide

PRODUCT Soft Cosmetic Contact Lenses sold under the brand names: Circle Lens, Europa, Pacifica, Gothika and Water Color, 
CODE All lots
RECALLING FIRM/MANUFACTURER
Recalling Firm: Hebron International, Inc., Atlanta, GA, 
Manufacturer: Mi Gwang Contact Lens Co., Ltd., Kyungsan, Korea. 
REASON Contact lenses were distributed and sold to inappropriate retail establishments without professional eye care involvement.
VOLUME OF PRODUCT IN COMMERCE 334,187 lenses
DISTRIBUTION Nationwide

PRODUCT St. Francis Medical Technologies brand X STOP® Interspinous Process Decompression (IPD) System, 14 mm X STOP implants in peel pouch, Catalog #1-2210; The X STOP consists of a main body, a spacer, and a tissue expander that is secured using a universal wing assembly consisting of a wing and a setscrew, 
CODE Lot numbers 051122, 060126, 060213, 060306, 060322, 060405, 060413, 060420, 060504, 060509
RECALLING FIRM/MANUFACTURER St. Francis Medical Technologies, Inc., Alameda, CA,
REASON Some units may be mislabeled as 10 mm X STOP implant that actually contain 14 mm X STOP implant.
VOLUME OF PRODUCT IN COMMERCE 440 units
DISTRIBUTION Nationwide and Internationally

PRODUCT Beckman Coulter Cytomics FC 500 Flow Cytometry System, Part Numbers 626553 (TN; FC500 with UPS), 6605627 (TN; 5 CLR, FC500 (100V), 6605628 (TN; 5CLR, FC500 (120V), 6605629 (TN; 5 CLR, FC500 (220V), 6605630 (TN; 5 CLR, FC500 (240V) with CXP Software Versions 2.0 and 2.1,
CODE Software Versions 2.0 and 2.1
RECALLING FIRM/MANUFACTURER
Recalling Firm: Beckman Coulter, Inc., Brea, CA, 
Manufacturer: Applied Cytometry, Sheffield, United Kingdom.
REASON Beckman Coulter has determined that under certain conditions a system error generated during a Pause and rotate routine may result in a sample misidentification with a risk of reporting erroneous results.
VOLUME OF PRODUCT IN COMMERCE 424 devices
DISTRIBUTION Nationwide and Canada

PRODUCT Waste tubing, Model # 842-326 (pump tube for waste) and Model # 842-327 (pump tube for solution), Waste Tubing used on ABL 7xx and ABL8xx component of the ABL700/800 Blood Gas Analyzers, also included in service kits part no. 905-671, 
CODE No lot numbers supplied.
RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America, Inc., Westlake, OH, 
Manufacturer: Radiometer A/S, Copenhagen, Denmark. 
REASON An increased failure rate was observed with the yellow rubber tubing. The tubing was found to be leaking blood and solutions after distribution.
VOLUME OF PRODUCT IN COMMERCE 2,576 units
DISTRIBUTION Nationwide and Canada

PRODUCT Dri-Tex BG-OX Cartridge, Model # D7120, Catalog # 946-003. Cartridges used in the NPT7 Analyzer, an in vitro diagnostic device, 
CODE Lot # R0299, Expiration Date: 02/2008
RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America, Inc., Westlake, OH,
Manufacturer: Radiometer A/S, Copenhagen, Denmark. 
REASON Air bubbles may be trapped on or close to the O2 sensor in the pO2 chamber of the cuvettes in cartridges used for blood gas analysis on NPT7 analyzer. This produces too high pO2 readings without a question mark.
VOLUME OF PRODUCT IN COMMERCE 185 units
DISTRIBUTION Nationwide

PRODUCT Radiance Data Management System Software Version 2.42 (Stand-alone software package) for the Radiance STAT Analyzer Management System, in vitro diagnostic, Model 914-317 Radiance Basic Kit, 
CODE Version 2.42 operating software
RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America, Inc., Westlake, OH, 
Manufacturer: Radiometer A/S, Copenhagen, Denmark. 
REASON Software anomaly. Data generated by blood gas equipment did not correlate with data shown in the Radiance Data Management System.
VOLUME OF PRODUCT IN COMMERCE 306 units
DISTRIBUTION Nationwide

PRODUCT
a) Microcuff Endotracheal Tube; Pediatric Oral/Nasal Magill: ID Size: 3,0mm,
Microcuff Product Reference: I-MPEDC-30; ID Size: 3,5mm, Product
Reference: I-MPEDC-35; ID Size: 4,0mm, Product Reference: I-MPEDC-40;
ID Size: 4.5mm, Product Reference: I-MPEDC-45; ID Size: 5,0mm, Product
Reference: I-MPEDC-50; ID Size: 5,5mm, Product Reference: I-MPEDC-55;
ID Size: 6,0mm, Product Reference: I-MPEDC-60; ID Size: 6,5mm, Product
Reference: I-MPEDC-65 and ID Size: 7,0mm, Product Reference: I-MPEDC-70
* Sterile * Single use only * , 
b) Microcuff Pediatric Endotracheal Tubes - Oral Pre-Curved: ID Size: 3,0mm,
Microcuff Product Reference: I-MPEDC-30; ID Size: 3,5mm, Product Reference:
I-MPEOC-35; ID Size: 4,0mm, Product Reference: I-MPEOC-40; ID Size: 4.5mm,
Product Reference: I-MPEOC-45; ID Size: 5,0mm, Product Reference: I-MPEOC-50;
ID Size: 5,5mm, Product Reference: I-MPEOC-55; ID Size: 6,0mm, Product
Reference: I-MPEOC-60; ID Size: 6,5mm, Product Reference: I-MPEOC-65 and ID
Size: 7,0mm, Product Reference: I-MPEOC-70 * Sterile * Single use only *,
c) Microcuff Adult Endotracheal Tubes - Oral/Nasal Magill with Murphy Eye *
ID Size: 5,0mm, Microcuff Product Reference: I-HMICU-50; ID Size: 5.5mm,
Product Reference: I-HMICU- 55; ID Size: 6.0mm, Product Reference:
I-HMICU- 60; ID Size: 6.5mm, Product Reference: I-HMICU- 65; ID
Size: 7.0mm, Product Reference: I-HMICU- 70; ID Size: 7.5mm, Product
Reference: I-HMICU-75; ID Size: 8.0mm, Product Reference: I-HMICU-80;
ID Size: 8.5mm, Product Reference: I-HMICU-85; ID Size: 9.0mm, Product
Reference: I-HMICU-90; and ID Size: 10.0mm Product Reference: I-HMICU-10 *
Sterile * Single use only *,
CODE All lots
RECALLING FIRM/MANUFACTURER
Recalling Firm: Microcuff Gmbh, Weinheim, Germany,
Manufacturer: Unomedical Industries, Kedah, Malaysia. 
REASON Inadequate seals and/or holes in the pouch may compromise the sterility of the product.
VOLUME OF PRODUCT IN COMMERCE 1,730 units
DISTRIBUTION Nationwide

PRODUCT
a) Terumo Advanced Perfusion System 4 inch diameter Roller Pump;
Model 801040, 
b) Terumo Advanced Perfusion System 1 Large Roller Pump; Model 801041,
CODE
a) Serial numbers: 0031 through 1188;
b) Serial numbers: 0033 through 1339
RECALLING FIRM/MANUFACTURER Terumo Cardiovascular Systems Corp., Ann Arbor, MI,
REASON The pump may stop during the autodose delivery without completion of the dose delivery to the patient.
VOLUME OF PRODUCT IN COMMERCE 1,898 devices
DISTRIBUTION Nationwide and Internationally

PRODUCT
a) Gamma Win software, part number GM11019110, versions up to and
including 1.62, for the GammaMedplus radionuclide applicator system,
for radiation therapy, 
b) Gamma Win software, part number GM11019110, versions up to and
including 1.62, for the GammaMedplus 3/24 radionuclide applicator system,
for radiation therapy, 
c) Gamma Win software, part number GM11019110, versions up to and
including 1.62, for the Mammo Source radionuclide applicator system,
for radiation therapy, 
CODE
All lots are recalled. Serial numbers are described: 

a) H64Y001 H64A032 H64A052 H64B074 H64C096 H64D120 H64Y002 H64A033 H64A053 H64B075 H64C097 H64D121 H64Y003 H64A034 H64A054 H64B076
H64C098 H64D122 H64Y006 H64A035 H64A055 H64B077 H64C099 H64D123 H64Y007 H64A036 H64A056 H64B078 H64C100 H64D124 H64Y009 H64A037 H64A057 H64B079 H64C101 H64D127 H64Y012 H64A038 H64A058 H64B080 H64C102 H64D128 H64Y013 H64A039 H64A059 H64B081 H64C103 H64D129 H64Z015 H64A040 H64A060 H64B082 H64C104 H64D130 H64Z018 H64A041 H64A061 H64B083 H64C106 H64D131 H64Z019 H64A042 H64A062 H64B084 H64C107 H64D132 H64Z021 H64A043 H64A063 H64B085 H64C108 H64D133 H64Z022 H64A044 H64A064 H64B087 H64C109 H64D134 H64Z024 H64A045 H64A066 H64B088 H64C110 H64D135 H64Z025 H64A046 H64A067 H64B089 H64C112 H64D136 H64Z026 H64A047 H64A068 H64B090 H64C113 H64E137 H64Z027 H64A048 H64B070 H64B091 H64D115 H64E138 H64Z028 H64A049 H64B071 H64C093 H64D116 H64E140 H64Z029 H64A050 H64B072 H64C094 H64D117 H64E141 H64Z030 H64A051 H64B073 H64C095 H64D118 H64E142 H64E143 H64E166 H64F191 H64F211 H640234 H640254 H64E147 H64E167 H64F192 H64F212 H640235 H640255 H64E148 H64E168 H64F193 H64F213 

H640236 H640256 H64E149 H64E169 H64F194 H64F216 H640237 H640257 H64E150 H64E170 H64F195 H640217 H640238 H640258 H64E151 H64E171 H64F196 H640218 H640239 H640259 H64E152 H64E172 H64F197 H640219 H640240 H640260 H64E153 H64E173 H64F198 H640220 H640241 H640261 H64E154 H64E175 H64F199 H640221 H640242 H640262 H64E155 H64E176 H64F200 H640222 H640243 H640264 H64E156 H64E177 H64F201 H640224 H640244 H640265 H64E157 H64E178 H64F202 H640225 H640245 H640266  H64E158 H64E179 H64F203 H640226 H640246 H640267 H64E159 H64E180 H64F204 H640227 H640247 H640268 H64E160 H64F181 H64F205 H640228 

H640248 H640269 H64E161 H64F182 H64F206 H640229 H640249 H640270 H64E162 H64F183 H64F207 H640230 H640250 H640271 H64E163 H64F184 H64F208 H640231 H640251 H640273 H64E164 H64F185 H64F209 H640232 H640252 H640275 H64E165 H64F189 H64F210 H640233 H640253;
b) All lots are recalled. Serial numbers are described: H64Y001 H64A032 H64A052 H64B074 H64C096 H64D120 H64Y002 H64A033 H64A053 H64B075 H64C097 H64D121 H64Y003 H64A034 H64A054 H64B076 H64C098 H64D122 H64Y006 H64A035 H64A055 H64B077 H64C099 H64D123 H64Y007 H64A036 H64A056 H64B078 H64C100 H64D124 H64Y009 H64A037 H64A057 H64B079 H64C101 H64D127 H64Y012 H64A038 H64A058 H64B080 H64C102 H64D128 H64Y013 H64A039 H64A059 H64B081 H64C103 H64D129 H64Z015 H64A040 H64A060 H64B082 H64C104 H64D130 H64Z018 H64A041 H64A061 H64B083 H64C106 H64D131 H64Z019 H64A042 H64A062 H64B084 H64C107 H64D132 H64Z021 H64A043 H64A063 H64B085 H64C108 H64D133 H64Z022 H64A044 H64A064 H64B087 H64C109 H64D134 H64Z024 H64A045 H64A066 H64B088 H64C110 H64D135 H64Z025 H64A046 H64A067 H64B089 H64C112 H64D136 H64Z026 H64A047 H64A068 H64B090 H64C113 H64E137 H64Z027 H64A048 H64B070 H64B091 H64D115 H64E138 H64Z028 H64A049 H64B071 H64C093 H64D116 H64E140 H64Z029 H64A050 H64B072 H64C094 H64D117 H64E141
H64Z030 H64A051 H64B073 H64C095 H64D118 H64E142 H64E143 H64E166 H64F191 H64F211 H640234 H640254 H64E147 H64E167 H64F192 H64F212 H640235 H640255 H64E148 H64E168 H64F193 H64F213 H640236 H640256 H64E149 H64E169 H64F194 H64F216 H640237 H640257 H64E150 H64E170 H64F195 H640217 H640238
H640258 H64E151 H64E171 H64F196 H640218 H640239 H640259 H64E152 H64E172 H64F197 H640219 H640240 H640260 H64E153 H64E173 H64F198 H640220 H640241 H640261 H64E154 H64E175 H64F199 H640221 H640242 H640262 H64E155 H64E176 H64F200 H640222 H640243 H640264 H64E156 H64E177 H64F201 H640224 H640244 H640265 H64E157 H64E178 H64F202 H640225 H640245 H640266 H64E158 H64E179 H64F203 H640226 H640246 H640267 H64E159 H64E180 H64F204 H640227 H640247 H640268 H64E160 H64F181 H64F205 H640228 H640248 H640269 H64E161 H64F182 H64F206 H640229 H640249 H640270 H64E162 H64F183 H64F207 H640230 H640250 H640271 H64E163 H64F184 H64F208 H640231 H640251 H640273 H64E164 H64F185 H64F209 H640232 H640252 H640275 H64E165 H64F189 H64F210 H640233 H640253;
c) All lots are recalled. Serial numbers are described: H64Y001 H64A032 H64A052 H64B074 H64C096 H64D120 H64Y002 H64A033 H64A053 H64B075 H64C097 H64D121 H64Y003 H64A034 H64A054 H64B076 H64C098 H64D122 H64Y006 H64A035 H64A055 H64B077 H64C099 H64D123 H64Y007 H64A036 H64A056 H64B078 H64C100
H64D124 H64Y009 H64A037 H64A057 H64B079 H64C101 H64D127 H64Y012 H64A038 H64A058 H64B080 H64C102 H64D128 H64Y013 H64A039 H64A059 H64B081 H64C103 H64D129 H64Z015 H64A040 H64A060 H64B082 H64C104 H64D130 H64Z018 H64A041 H64A061 H64B083 H64C106 H64D131 H64Z019 H64A042 H64A062 H64B084 H64C107 H64D132 H64Z021 H64A043 H64A063 H64B085 H64C108 H64D133 H64Z022 H64A044 H64A064 H64B087 H64C109 H64D134 H64Z024 H64A045 H64A066 H64B088 H64C110 H64D135 H64Z025 H64A046 H64A067 H64B089 H64C112 H64D136 H64Z026 H64A047 H64A068 H64B090 H64C113 H64E137 H64Z027 H64A048 H64B070 H64B091 H64D115 H64E138 H64Z028 H64A049 H64B071 H64C093 H64D116 H64E140 H64Z029 H64A050 H64B072 H64C094 H64D117 H64E141 H64Z030 H64A051 H64B073 H64C095 H64D118 H64E142 H64E143 H64E166 H64F191 H64F211 H640234 H640254 H64E147 H64E167 H64F192 H64F212 H640235 H640255 H64E148 H64E168 H64F193 H64F213 H640236 H640256 H64E149 H64E169 H64F194 H64F216 H640237 H640257 H64E150 H64E170 H64F195 H640217 H640238H640258 H64E151 H64E171 H64F196 H640218 H640239 H640259 H64E152 H64E172 H64F197 H640219 H640240 H640260 H64E153 H64E173 H64F198 H640220 H640241 H640261 H64E154 H64E175 H64F199 H640221 H640242 H640262 H64E155 H64E176 H64F200 H640222 H640243 H640264 H64E156 H64E177 H64F201 H640224 H640244 H640265 H64E157 H64E178 H64F202 H640225 H640245 H640266 H64E158 H64E179 H64F203 H640226 H640246 H640267 H64E159 H64E180 H64F204 H640227 H640247 H640268
H64E160 H64F181 H64F205 H640228 H640248 H640269 H64E161 H64F182 H64F206 H640229 H640249 H640270 H64E162 H64F183 H64F207 H640230 H640250 H640271 H64E163 H64F184 H64F208 H640231 H640251 H640273 H64E164 H64F185 H64F209 H640232 H640252 H640275 H64E165 H64F189 H64F210 H640233 H640253
RECALLING FIRM/MANUFACTURER
Recalling Firm: Varian Medical Systems, Charlottesville, VA, 
Manufacturer: Varian Medical Systems, Haan, Germany. 
REASON Medical device software for brachytherapy may cause erroneous data to be recorded and affect patient radiation treatments. Erroneous data may be listed on the treatment history report, and the default step size may lead to a misadministration if treatment data is entered manually.
VOLUME OF PRODUCT IN COMMERCE 254 units
DISTRIBUTION Nationwide and Internationally

PRODUCT
a) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 8 mm diameter, 130 mm length non-sterile, catalog no. 4309-08-13 (00-4309-008-13), 
b) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 9 mm diameter, 130 mm length non-sterile, catalog no. 4309-09-13 (00-4309-009-13), 
c) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 10 mm diameter, 130 mm length non-sterile, catalog no. 4309-10-13 (00-4309-010-13), 
d) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 11 mm diameter, 130 mm length non-sterile, catalog no. 4309-11-13 (00-4309-011-13), 
e) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 12 mm diameter, 130 mm length non-sterile, catalog no. 4309-12-13 (00-4309-012-13), 
f) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 13 mm diameter, 130 mm length non-sterile, catalog no. 4309-13-13 (00-4309-013-13), 
g) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 14 mm diameter, 130 mm length non-sterile, catalog no. 4309-14-13 (00-4309-014-13), 
h) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 15 mm diameter, 130 mm length non-sterile, catalog no. 4309-15-13 (00-4309-015-13),
i) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 16 mm diameter, 130 mm length non-sterile, catalog no. 4309-16-13 (00-4309-016-13), 
j) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 17 mm diameter, 130 mm length non-sterile, catalog no. 4309-17-13 (00-4309-017-13), 
k) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 18 mm diameter, 130 mm length non-sterile, catalog no. 4309-18-13 (00-4309-018-13), 
l) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 8 mm diameter, 170 mm length non-sterile, catalog no. 4309-08-17 (00-4309-008-17), 
m) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 10 mm diameter, 170 mm length non-sterile, catalog no. 4309-10-17 (00-4309-010-17), 
n) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 12 mm diameter, 170 mm length non-sterile, catalog no. 4309-12-17 (00-4309-012-17), 
o) Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 14 mm diameter, 170 mm length non-sterile, catalog no. 4309-14-17 (00-4309-014-17), 
CODE All lots
RECALLING FIRM/MANUFACTURER Zimmer Inc., Warsaw, IN
REASON During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
VOLUME OF PRODUCT IN COMMERCE 324 units
DISTRIBUTION Nationwide and Internationally

PRODUCT
a) Guidant CONTAK RENEWAL 3 (models H170, H175) CONTAK RENEWAL 3 HE (models H177, H179) Cardiac Resynchronization Therapy Defibrillator (CRT-D). Cardiac resynchronization therapy defibrillators (CRT-D) provide ventricular tachyarrhythmia and cardiac resynchronization therapies. Ventricular tachyarrhythmia therapy is for the treatment of ventricular tachycardia (VT) and ventricular fibrillation (VF), rhythms that are associated with sudden cardiac death (SCD). Cardiac resynchronization therapy is for the treatment
of heart failure (HF) and uses biventricular electrical stimulation to synchronize ventricular contractions. Cardioversion/defibrillation therapies include a range of low-and high-energy shocks using either a biphasic or monophasic waveform, 
b) These models are not available in the US. Guidant CONTAK RENEWAL 4 (models H190, H195). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Cardiac resynchronization therapy defibrillators (CRT-D) provide ventricular tachyarrhythmia and cardiac resynchronization therapies. Ventricular tachyarrhythmia therapy is for the treatment of ventricular tachycardia (VT) and ventricular fibrillation (VF), rhythms that are associated with sudden cardiac death (SCD). Cardiac resynchronization therapy is for the treatment of heart failure (HF) and uses biventricular electrical stimulation to synchronize ventricular contractions. Cardioversion/defibrillation therapies include a range of low-and high-energy shocks using either a biphasic or monophasic waveform, 
c) CONTACT RENEWAL 4 AVT HE model numbers M177. Cardiac Resynchronization Therapy Defibrillator. CONTAK RENEWAK AVT models provide both atrial and ventricular tachyarrhythmia and cardiac resynchronization therapies, 
d) VITALITY AVT, model A155. Implantable Cardioverter, Defibrillator, Atrial and Ventricular Therapies, Model A155, Recall # Z-1291-06;
e) Guidant VITALITY 2 (models T165, T175), VITALITY DS (model T125), VITALITY EL (model T135), Implantable Cardioverter Defibrillators (ICD), are designed to detect and terminate ventricular tachycardia (VT) and ventricular fibrillation (VF) and provide bradycardia therapy (atrial and ventricular pacing). Therapies include both low- and high-energy shocks
using either a biphasic or monophasic waveform. Vitality 2 devices also offer a wide variety of antitachycardia pacing schemes to terminate slower, more stable ventricular tachyarrhythmias. Bradycardia pacing, including adaptive-rate features, is available to detect and treat bradyarrhythmias and to support the cardiac rhythm after defibrillation therapy. Devices denoted with DR offer dual-chamber bradycardia features (atrial and/or ventricular pacing and sensing), and the devices denoted with VR offer single-chamber bradycardia features (ventricular pacing and sensing),
CODE
a) Model H170, Serial numbers: 509659, 509677, 509678, 509679, 509685, 509694, 509695, 509696, 509762, 509884, 509974, 510012 

model H175, serial numbers: 512665, 512666, 512667, 512668, 512708, 512709, 512710, 512711, 512713, 512719, 512721, 512722, 512723, 512759, 512791, 512793, 513062, 513064, 513073, 513085, 513086, 513227
model H177, serial numbers: 506951, 506953, 506954, 506955, 506956, 506957, 506959, 506960, 506967, 506969, 506970, 506973, 506974, 507013, 507014, 507472 

model H179, serial numbers: 508312, 508316, 508322, 508325, 508326, 508336, 508342, 508349, 508404;
b) Model H190, Serial numbers: 301196, 301198, 301204, 301206
Model H195, Serial numbers: 306327, 306332, 306333, 306334, 306337, 306339, 306348, 306349, 306351, 306367, 306368, 306378, 306380, 306447;
c) Serial numbers: 100160, 100161, 100162;
d) Serial numbers: 113567, 113568, 113569, 113570, 113572, 113574, 113585, 113586, 113587, 113591, 113612, 113613, 113616, 113617, 113618, 113619, 113620, 113621, 113623, 113624, 113626, 113627, 113628, 113629, 113631, 113633, 113635, 113637, 113639, 113640, 113641, 113654, 113659, 113660, 113672, 113674, 113675, 113676,
113678, 113679, 113680, 113681, 113682, 113684, 113686, 113687, 113690, 113693, 113694, 113695, 113696, 113778, 113779, 113781, 113783, 113784, 113789, 113791, 113792, 113814, 113908, 113910, 113912, 113913, 113914, 114083;
e) Model T165, Serial numbers: 106641, 106643, 106644, 106645, 106646, 106647, 106648, 106649, 106651, 106652, 106676, 106677, 106678, 106679, 106680, 106681, 106683, 106685, 106686, 106687, 106691, 106692, 106693, 106694, 106696, 106697, 106700, 106701, 106702, 106778, 106779, 106780, 106781, 106782, 106788, 106789, 106790, 106792, 106793, 106794, 106795, 106796, 106797, 106798, 106799, 106803, 106804, 106805, 106806, 106807, 106808, 106810, 106813, 106815, 106816, 106817, 106818, 106819, 106820, 106821, 106822, 106823, 106824, 106825, 106826, 106827, 106828, 106829, 106830, 106831, 106832, 106833, 106834, 106835, 106836, 106837, 106838, 106839, 106840, 106841, 106842, 106843, 106844, 106845, 106846, 106847, 106849, 106850, 106851, 106852, 106853, 106854, 106855, 106856, 106857, 106858, 106860, 106861, 106862, 106863, 106864, 106865, 106867, 106868, 106869, 106870, 106871, 106872, 106873, 106874, 106875, 106876, 106877, 106878, 106879, 106880, 106881, 106882, 106883, 106884, 106885, 106886, 106888, 106889, 106890, 106891, 106892, 106894, 106895, 106896, 106898, 106899, 106900, 106901, 106903, 106905, 106906, 106907, 106908, 106909, 106910, 106911, 106914, 106915, 106916, 106917, 106919, 106920, 106922, 106923, 106925, 106926, 106929, 106932, 106935, 106936, 106937, 106939, 106940, 106941, 106942, 106943, 106945, 106946, 106947, 106948, 106949, 106952, 106953, 106954, 106955, 106956, 106957, 106959, 106960, 106962, 106963, 106964, 106965, 106968, 106969, 106970, 106971, 106972, 106973, 106974, 106975, 106976, 106977, 106978, 106979, 106980, 106981, 106982, 106983, 106984, 106985, 106987, 106990, 106991, 106992, 106993, 106996, 106997, 106998, 106999, 107000, 107001, 107002, 107003, 107004, 107007, 107008, 107009, 107010, 107011, 107012, 107013, 107015, 107017, 107018, 107019, 107020, 107021, 107033, 107034, 107035, 107036, 107037, 107039, 107043, 107044, 107045, 107046, 107047, 107057, 107058, 107059, 107060, 107061, 107062, 107064, 107065, 107066, 107067, 107068, 107069, 107070, 107072, 107075, 107077, 107078, 107079, 107082, 107087, 107091, 107092, 107099, 107101, 107102, 107103, 107106, 107110, 107116, 107123, 107124, 107141, 107208, 107210, 107254, 107263, 107264, 107265, 107266, 107267, 107298, 107299, 107300, 107301, 107303, 107304, 107306, 107307, 107308, 107315, 107316, 107317, 107318, 107320, 107321, 107322, 107389, 107718, 108310, 108376, 115934
Model T175, serial numbers: 105118, 105119, 105120, 105122, 105126, 105130, 105131, 105140, 105141, 105156, 105157, 105158, 105161, 105162, 105165, 105166, 105167, 105169 105170, 105172, 105173, 105174, 105176, 105177, 105178, 105179, 105180, 105181, 105182, 105183, 105184, 105186, 105187, 105189, 105191, 105192, 105194,
105196, 105197, 105200, 105207, 105208, 105209, 105210, 105211, 105212, 105213, 105217, 105304, 105305, 105306, 105307, 105308, 105364, 105365, 105366, 105367, 105368, 105369, 105370, 105371, 105372, 105373, 105374, 105375, 105376, 105377, 105378, 105379, 105380, 105381, 105382, 105383, 105384, 105385, 105386, 105387,
105388, 105389, 105390, 105391, 105393, 105395, 105397, 105398, 105399, 105400, 105401, 105402, 105403, 105404, 105405, 105406, 105407, 105409, 105410, 105411, 105412, 105413, 105414, 105415, 105416, 105417, 105418, 105419, 105420, 105421, 105422, 105423, 105424, 105425, 105426, 105427, 105428, 105429, 105431, 105432,
105433, 105434, 105435, 105436, 105437, 105438, 105439, 105440, 105442, 105443, 105444, 105445, 105446, 105447, 105448, 105449, 105450, 105451, 105452, 105453, 105454, 105455, 105456, 105458, 105459, 105460, 105461, 105462, 105463, 105464, 105465, 105466, 105467, 105468, 105469, 105470, 105471, 105472, 105473, 105474,
105476, 105477, 105478, 105479, 105480, 105481, 105482, 105483, 105484, 105485, 105487, 105488, 105573, 105575, 105576, 105577, 105578, 105579, 105580, 105581, 105584, 105586, 105587, 105588, 105589, 105590, 105591, 105592, 105593, 105594, 105595, 105596, 105598, 105599, 105600, 105602, 105603, 105604, 105605, 105606,
105607, 105611, 105613, 105624, 105625, 105627, 105628, 105629, 105630, 105632, 105633, 105634, 105635, 105636, 105638, 105639, 105640, 105642, 105643, 105644, 105645, 105646, 105647, 105648, 105649, 105650, 105651, 105652, 105654, 105655, 105656, 105657, 105658, 105659, 105660, 105661, 105662, 105663, 105664, 105665,
105666, 105667, 105668, 105669, 105670, 105671, 105672, 105673, 105674, 105675, 105676, 105677, 105678, 105679, 105680, 105681, 105682, 105683, 105687, 105688, 105689, 105690, 105691, 105694, 105705, 105706, 105707, 105712, 105713, 105714, 105715, 105716, 105717, 105718, 105720, 105721, 105722, 105723, 105724, 105725,
105726, 105727, 105728, 105739, 105740, 105741, 105743, 105744, 105745, 105746, 105747, 105748, 105749, 105750, 105752, 105753, 105754, 105755, 105756, 105757, 105758, 105759, 105760, 105762, 105763, 105764, 105765, 105766, 105767, 105768, 105785, 105786, 105790, 105792, 105796, 105820, 105857, 105881, 105884, 105885,
105895, 105898, 105914, 105915, 105917, 105919, 105923, 105924, 105925, 105926, 105927, 105928, 105929, 105930, 105931, 105932 105933, 105934, 105935, 105936, 105937, 105940, 105944, 105945, 105946, 105947, 105948, 105949, 105950, 105951, 105952, 105953, 106074, 106155, 106156, 106157, 106171, 106172, 106173, 106174,
106177, 106178, 106186, 106187, 106189, 106191, 106199, 106200, 106201, 106212, 106213, 106235, 106282, 106361, 106503, 107047, 107301, 107593, 108285
Model T125, Serial numbers: 113211, 113221, 113222, 113223, 113224, 113241, 113243, 113244, 113272, 113273, 113274, 113275, 113276, 113277, 113278, 113279, 113282, 113283, 113284, 113285, 113286, 113287, 113288, 113291, 113292, 113294, 113295, 113296, 113297, 113298, 113299, 113300, 113301, 113317, 113318, 113319, 113320,
113322, 113323, 113324, 113325, 113326, 113327, 113328, 113329, 113330, 113331, 113333, 113334, 113335, 113336, 113337, 113339, 113340, 113341, 113342, 113343, 113344, 113345, 113346, 113347, 113348, 113349, 113350, 113351, 113352, 113353, 113355, 113356, 113359, 113360, 113362, 113363, 113364, 113365, 113366, 113367,
113368, 113369, 113370, 113372, 113373, 113374, 113375, 113378, 113379, 113380, 113381, 113383, 113384, 113385, 113386, 113387, 113388, 113389, 113390, 113391, 113393, 113394, 113396, 113397, 113398, 113399, 113400, 113403, 113408, 113414, 113417, 113420, 113422, 113425, 113427, 113428, 113429, 113430, 113431, 113433,
113434, 113435, 113487, 113492, 113495, 113496, 113500, 113557, 113643, 113847, 113850, 114102
Model T135, Serial numbers: 941570, 941572, 941573, 941574, 941578, 941585, 941589, 941590, 941591, 941592, 941593, 941595, 941597, 941601, 941602, 941603, 941604, 941605, 941606, 941607, 941608, 941609, 941610, 941613, 941614, 941615, 941616, 941618, 941619, 941620, 941621, 941622, 941623, 941624, 941625, 941626, 941627,
941628, 941629, 941630, 941632, 941633, 941634, 941652, 941653, 941654, 941680, 941681, 941682, 941683, 941688, 941689, 941690, 941691 941692, 941695, 941696, 941697, 941698, 941699, 941700, 941701, 941702, 941703, 941704, 941709, 941710, 941712, 941715, 941716, 941717, 941718, 941719, 941721, 941722, 941723, 941724,
941725, 941727, 941728, 941732, 941734, 941735, 941736, 941737, 941738, 941739, 941740, 941741, 941742, 941743, 941745, 941747, 941748, 941749, 941750, 941752, 941753, 941754, 941756, 941757, 941759, 941762, 941767, 941775, 941776, 941777, 941778, 941781, 941784, 941786, 941787, 941788, 941789, 941800, 941802, 941803,
941804, 941805, 941807, 941810, 941812, 941813, 941814, 941819, 941829, 941830, 941831, 941845, 941846, 941929, 941930, 941933, 941940, 941948, 941953, 941956, 941962, 941970, 941975, 941978, 941979, 942017
RECALLING FIRM/MANUFACTURER
Recalling Firm: Guidant Corporation, Saint Paul, MN, 
Manufacturer: Guidant-Ireland, Clomel, Ireland. 
REASON Guidant had identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.
VOLUME OF PRODUCT IN COMMERCE 996 devices
DISTRIBUTION Nationwide and Internationally

PRODUCT Boston Scientific Urovac Bladder Evacuator, sterile, single use only, REF/Catalog no. 730-125 and sold individually as UPN M0067301250 and in packaged of ten as UPN M0067301251,
CODE All lots expiring on or before June 12, 2010
RECALLING FIRM/MANUFACTURER Boston Scientific Corp., Spencer, IN
REASON Sterility is compromised because the product is labeled as sterile, but the packages may have holes in them.
VOLUME OF PRODUCT IN COMMERCE 446,097 units
DISTRIBUTION Nationwide and Internationally

PRODUCT Ultra-Filtration (UF) Removal Regulators(Rgltr Assy UF REM Molded Spare); a spare part used with Baxter’s System 1000, Tina, Arena, Altratouch and Aurora Hemodialysis Instruments; item numbers 6001276024 and 6001276028, 
CODE Lot numbers 01095126 and 01092479; Lot number 01095127
RECALLING FIRM/MANUFACTURER
Recalling Firm: Baxter Healthcare Renal Div., Mc Gaw Park, IL,
Manufacturer: Baxter Healthcare Corporation, Largo, FL. 
REASON The UF Removal Regulators were assembled incorrectly. The mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy.
VOLUME OF PRODUCT IN COMMERCE 569 units
DISTRIBUTION Nationwide and Internationally

PRODUCT VIASYS Healthcare Disposable Bar Electrodes. Re-order No. 019-435600. Single Use only, 
CODE Lot 604664
RECALLING FIRM/MANUFACTURER
Recalling Firm: Nicolet Biomedical Div. of Viasys Healthcare, Madison, WI,
Manufacturer: Tyco Healthcare Uni-Patch, Wabasha, MN. 
REASON The Bar Electrodes do not meet performance specification for electrode impedance. A manufacturing defect was discovered in disposable bar electrodes, part number 019-435600, lot 604664, that has the potential to reduce the amplitude of a motor or sensory nerve action potential. These possible lower amplitudes will be measured only if the bar electrode is used as a bar.
VOLUME OF PRODUCT IN COMMERCE 53 boxes, 10 electrodes per pouch, 10 pouches per box
DISTRIBUTION Nationwide and Internationally

PRODUCT IMPAX 4.5 Systems, used in the acceptance, transfer, display, storage and digital processing of medical images, 
CODE All Impax 4.5 sites running Solaris 9 (Sun OS 5.9) that do not have Sun Patch ID 118305-08 installed.
RECALLING FIRM/MANUFACTURER AGFA Corp., Greenville, SC, 
REASON Possible corrupted image appearing after System Start.
VOLUME OF PRODUCT IN COMMERCE 2 units
DISTRIBUTION NY

PRODUCT
a) Terumo Advanced Perfusion System 1; 100V -- 120V, 15 A (circuit breaker), 50/60 Hz (20A power source required); Models 801763 and 801763 (Japan unit).
Units using Power Manager software versions 1.10, Recall # Z-1286-06;
b) Terumo Advanced Perfusion System 1; 220V -- 240V, 7 A (circuit breaker),50/60 Hz (10A power source required); Model 801764. Not distributed within
the United States. Units using Power Manager software versions 1.10,
CODE
a) Serial numbers 11 through 339;
b) Serial numbers 6 through 156
RECALLING FIRM/MANUFACTURER Terumo Cardiovascular System Corp., Ann Arbor, MI, 
REASON While on battery power, the system may flash a low battery warning even though the battery is adequately charged.
VOLUME OF PRODUCT IN COMMERCE 442 units
DISTRIBUTION Nationwide and Internationally

WEEK ENDING AUGUST 12

PRODUCT TriActiv ProGuard Embolic Protection System, Part number 61000-01. The product is shipped as a procedure kit. It is used with a standard 6F guide catheter using the standard femoral approach, 
CODE Lot number 48275, exp. 1/07
RECALLING FIRM/MANUFACTURER Kensey Nash Corp., Exton, PA, 
REASON Alarm activation-A priming issue involving the flow control unit due to a software problem has caused false positive Extraction Line Block (ELB) alarming. Use of this product may result in a health hazard resulting from early termination of the flush and extraction procedure after stenting.
VOLUME OF PRODUCT IN COMMERCE 69 kits
DISTRIBUTION Germany and Italy

PRODUCT Beckman Coulter Access Immunoassay system CEA, in vitro diagnostic, Part No. 33200,
CODE All CODEs
RECALLING FIRM/MANUFACTURER
Recalling Firm: Beckman Coulter, Inc., Brea, CA, 
Manufacturer: Beckman Coulter, Inc., Chaska, MN.
REASON Firm has confirmed that the Access CEA Assay may generate decreased values with certain frozen platelet specimens.
VOLUME OF PRODUCT IN COMMERCE 775 units
DISTRIBUTION Nationwide and Canada

PRODUCT Medplus ER Drug Screen Test, Model # 5800KAB, packaged as 25 individually pouched devices in a labeled kit box,
CODE
a) Lot number: 85699 and
b) Lot number: 85671
RECALLING FIRM/MANUFACTURER Applied Biotech, Inc., San Diego, CA,
REASON The firm has discovered that a number of distributed devices of the Medplus ER Drug Screen Test may be incorrectly assembled which may give incorrect results. The product is an in vitro test for rapid detection of various drugs such as Amphetamines, Barbiturates, Phencyclidine, Cocaine, Methamphetamine, Tricyclic Antidepressants, Morphine, Tetrahydrocannabinol, & Benzodiazepines in human urine.
VOLUME OF PRODUCT IN COMMERCE a) 71 kits (1775 devices) & b) 44 kits (1100 devices)
DISTRIBUTION Nationwide

PRODUCT D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer, 
CODE Lot number: R0053
RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America, Inc., Westlake, OH, 
Manufacturer: Radiometer Medical, Bronshoj, Denmark. 
REASON The CI membrane units generate to low values for status and sensitivity during calibration.
VOLUME OF PRODUCT IN COMMERCE 85/4 units per box
DISTRIBUTION Nationwide

PRODUCT Package Insert which accompanies the Hematocrit and Metabolite QUALICHECK LEVEL 2 Quality Control of Hematocrits and Metabolites, used with the ABL 77/70/555 Blood Gas & Electrolyte Analyzers. The QC reagents are packaged in a box of 30 glass vials of (1.5 ml) per box. The package insert is packed one per box. Model #.S7180, Part Number: 944040, Recall # Z-1316-06 Package Insert which accompanies the Hematocrit and Metabolite QUALICHECK LEVEL 2 Quality Control of Hematocrits and Metabolites, used with the ABL 77/70/555 Blood Gas & Electrolyte Analyzers. The QC reagents are packaged in a box of 30 glass vials of (1.5 ml) per box. The package insert is packed one per box. Model #.S7180, Part Number: 944040, 
CODE Lot # 15
RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America, Inc., Westlake, OH, 
Manufacturer: Radiometer Medical, Bronshoj, Denmark. 
REASON The package insert, which accompanied the QC reagents used with ABL blood gas analyzers, referenced incorrect control ranges for Hct (hematocrit) testing. The range stated on the package insert was too low to generate accurate Hct test results (The control ranges were acceptable for the glucose and lactate testing).
VOLUME OF PRODUCT IN COMMERCE 55 boxes
DISTRIBUTION Nationwide, London, and Canada

WEEK ENDING AUGUST 20

PRODUCT Fugo Blade Incising Tip Assembly. Incising tips to be used with the Fugo Blade M1000 Anterior Capsulotomy Unit. For Ophthalmic Use Only. The product is shipped 12 units per carton
CODE Lot numbers 01013786, 01017917, 01017918, 01017684 and 01017442
RECALLING FIRM/MANUFACTURER Medisurg Research & Management Corp, Norristown PA,
REASON No documentation to support shelf life/sterility
VOLUME OF PRODUCT IN COMMERCE 155 bags
DISTRIBUTION Nationwide and Canada

PRODUCT
a) Abbott brand FreeStyle Blood Glucose Meter, 
b) Abbott brand FreeStyle Flash Blood Glucose Meter, 
CODE All codes
RECALLING FIRM/MANUFACTURER
Recalling Firm: Abbott Diabetes Care, Inc. AlamedaCA, 

Manufacturer: Flextronics International, Shenzhen, China, 
REASON The products may encounter display problem, 'Er 4' message, during prolonged use when the low battery symbol is displayed. The situation can render the meter either inoperable or operable with invalid user configuration data including selectable unit of measure, and strip calibration code.
VOLUME OF PRODUCT IN COMMERCE Approx. 2,365,000 meters
DISTRIBUTION Nationwide and Internationally

PRODUCT TEKLENS II (TEKIA label) and Polylens A60 (Polytech label) Model 614, intraocular lens (IOL) :
1a) Products labeled with TEKIA as manufacturer: IOLs with power of 17.0 D;
1b) IOLs with power of 22.5 D;
2a) Products labeled with Polytech as manufacturer: IOLs with power of 17.0 D:
2b)IOLs with power of 22.5 D,
CODE
1a) 616170E031002180207, 616170E031002190207, 616170E031002200207, 616170E031002210207;
1b) 616225E031001460207, 616225E031001470207, 616225E031001480207, 616225E031001490207, 616225E031001500207, 616225E031001510207; 616225E031001520207, 616225E031001530207.
2a) 613170E031000640207, 613170E031000650207, 613170E031000660207, 613170E031000670207, 613170E031000680207, 613170E031000690207, 613170E031000700207, 613170E031000710207,613170E031000720207, 613170E031000730207, 613170E031000740207, 613170E031000750207, 613170E031000760207, 613170E031000770207, 613170E031000780207, 613170E031000790207, 613170E031000800207, 613170E031000810207, 613170E031000820207, 613170E031000830207, 613170E031000840207, 613170E031000851207, 613170E031000860207, 613170E031000870207, 613170E031000880207, 613170E031000890207;
2b)613225E031000990207, 613225E031001000207, 613225E031001010207, 613225E031001020207, 613225E031001030207, 613225E031001040207, 613225E031001050207, 613225E031001060207, 613225E031001070207, 613225E031001080207, 613225E031001090207, 613225E031001100207, 613225E031001110207, 613225E031001120207, 613225E031001130207, 613225E031001140207, 613225E031001150207, 613225E031001160207, 613225E031001170207, 613225E031001180207, 613225E031001190207, 613225E031001200207
RECALLING FIRM/MANUFACTURER Tekia, Inc., Irvine, CA
REASON Intraocular lenses (IOLs) with power of 17.0 D and 22.5 D are mislabeled with each other's labels.
VOLUME OF PRODUCT IN COMMERCE 60 IOLs
DISTRIBUTION Germany

PRODUCT OPUS SpeedStitch Suture Cartridge with Magnumwire, a component of the OPUS MiniPlus Implant Set, for use in placement of sutures through soft tissue in endoscopic and other limited access procedures., Catalog number OM-6006. This catalog number is referred to as the OPUS MiniPlus Implant Set, 
CODE Lot number: 110399
RECALLING FIRM/MANUFACTURER
Recalling Firm: Arthrocare, Corp., Sunnyvale, CA, 
Manufacturer: Arthrocare, Corp., Aurora de Heredia, Costa Rica. 
REASON The product may lose sterility due to omission of the final pouch seal.
VOLUME OF PRODUCT IN COMMERCE 170 units
DISTRIBUTION Nationwide

PRODUCT
a) Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzer, ABL-700 series, 
b) Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzer, ABL-800 series,
CODE
a) All analyzers manufactured from 1/1999-12/2005. The devices contain serial numbers 902-441RxxxNxxx. The Rxxx number denotes manufacturing run, and the Nxxx number denotes the individual instrument within the run;
b) All analyzers manufactured from 1/2005-10/31/2005. The devices contain serial numbers 902-441RxxxNxxx. The Rxxx number denotes manufacturing run, and the Nxxx number denotes the individual instrument within the run.
RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America Inc, WestlakeOH, 
Manufacturer: Radiometer Medical ApS, Bronshoj, Denmark. 
REASON Software defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase.
VOLUME OF PRODUCT IN COMMERCE
Series 700 -- 1,635 Analyzers;
Series 800 -- 108 Analyzers
DISTRIBUTION Nationwide

PRODUCT ABL 5 Capillary Adaptor (a connection that fits the capillary tube on to the aspirator on the ABL 5 & BPH 5 Analyzers), in vitro diagnostic accessory, 
CODE Part #906-018
RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America Inc, WestlakeOH, 
Manufacturer: Radiometer Medical ApS, Bronshoj, Denmark. 
REASON The dimensions and shape of the analyzer adaptor do not fit the analyzer tubes or probe, causing air to be aspirated with the blood sample and causing a deviation in the p02 sample results.
VOLUME OF PRODUCT IN COMMERCE 155 bags
DISTRIBUTION Nationwide and Canada

PRODUCT
a) Capillary (end) caps, accessory for use with blood gas analyzers included in
Radiometer's CLINITUBES kits. The end caps are packaged in bags.
These bags are placed in the capillary tube kit which also contains: capillary tubes,
end caps, mixing wires, and a magnet. Each kit may contain 1 or 5 bags depending
on the kit model. Each bag contains: 110 Green caps per bag and 160 Red caps per bag,
b) Capillary (end) caps, accessory for use with blood gas analyzers included in
Radiometer's CLINITUBES kits. The end caps are packaged in bags.
These bags are placed in the capillary tube kit which also contains: capillary tubes,
end caps, mixing wires, and a magnet. Each kit may contain 1 or 5 bags depending on
the kit model. Each bag contains: 110, 
CODE
a) Model # 904-439;
b) Model # 904-453
RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America Inc, WestlakeOH, 
Manufacturer: Radiometer Medical ApS, Bronshoj, Denmark. 
REASON Capillary end caps, an accessory for ABL blood gas analyzers, leak blood from capillary tubes during transport.
VOLUME OF PRODUCT IN COMMERCE 386 kits
DISTRIBUTION Nationwide, Canada and Istanbul

PRODUCT
a) Pointe Scientific Liquid Glucose HEX (R1) Reagent Set, for the quantitative
determination of glucose in serum, 5 x 600 ml R1 glucose HEX, for use with Hitachi
analyzer; in vitro diagnostic, Catalog number HG420-R1,
b) Pointe Scientific Liquid Glucose HEX (R2) Reagent Set, for the quantitative
determination of glucose in serum, 5 x 600 ml R2 glucose HEX, for use with Hitachi
analyzer; in vitro diagnostic, Catalog number HG420-R2,
c) Pointe Scientific Liquid Glucose (HEX) Reagent Set, for the quantitative determination
of glucose in serum, 10 x 5 ml R1 and 5 x 20 ml R2, or 9 x 67 ml R1and 9 x 17 ml R2
glucose reagent, for use with Hitachi analyzer; in vitro diagnostic, Catalog numbers
HG720-600 and HG920-756, 
d) Pointe Scientific Liquid Glucose (HEXO) Reagent Set, for the quantitative
determination of glucose in serum, 10 x 100 ml R1, 10 x 20 ml R2 glucose reagent,
or 1 x 2 L R1 x 500 ml R2, for use with Hitachi analyzer; in vitro diagnostic, Catalog
numbers HG920-1200 and HG920-2500, 
e) Glucose Hex reagent sold by Pointe Scientific; in vitro diagnostic, Catalog no.
3-HG920-L, 
f) Glucose HEX reagent sold by Pointe Scientific; in vitro diagnostic, Catalog no.
3-HG920-L, 
g) Glucose HEX R1 reagent sold by Pointe Scientific; in vitro diagnostic, Catalog no’s
3-HG920-R1, 7-HG920-R1-67, 7-HG920-R1-600 and 7-HG920-R1-800,
h) Glucose HEX R2 reagent sold by Pointe Scientific; in vitro diagnostic,
Catalog no’s 3-HG920-R2, 7-HG920-R2-17, 7-HG920-R2-200 and 7-HG920-R2-300,
CODE
a) and b) Lots 511901, exp. 4/07 and 525101, exp. 9/07;
c), d), e) and f) Lots 431704, exp. 11/06 and 511901, exp. 4/07;
g) and h) Lots 431704, exp. 11/06, 511901, exp. 4/07 and 525101, exp. 9/07
RECALLING FIRM/MANUFACTURER Pointe Scientific, Inc., Lincoln Park, MI
REASON Product does not meet performance specification through its labeled expiration period.
VOLUME OF PRODUCT IN COMMERCE 13 boxes
DISTRIBUTION Nationwide and Canada

PRODUCT
a) Zimmer M/DN Intramedullary Fixation 4.5 dia. cortical screw, 3.5 mm hex head,
52.5 mm length, 22-13-5 stainless steel, sterile bone screw;
Cat. no. 2253-52-45 (00225305545), 
b) Zimmer M/DN Intramedullary Fixation 4.5 dia. cortical screw, 3.5 mm
hex head, 55 mm length, 22-13-5 stainless steel, sterile bone screw;
Cat. no. 2253-55-45 (00225305545), 
c) Zimmer M/DN Intramedullary Fixation 4.5 dia. cortical screw, 3.5 mm hex head,
65 mm length, 22-13-5 stainless steel, sterile bone screw; Cat. no. 2253-65-45
(00225306545), 
d) Zimmer M/DN Intramedullary Fixation 4.2 dia. cortical screw, 3.5 mm hex head,
70 mm length, 22-13-5 stainless steel, sterile bone screw; Cat. no. 2253-70-42
(00225307042), 
e) Zimmer M/DN Intramedullary Fixation 5.5 dia. cortical screw, 3.5 mm hex head,
70 mm length, 22-13-5 stainless steel, sterile bone screw; Cat. no. 2253-70-55
(00225307055), 
f) Zimmer M/DN Intramedullary Fixation 4.2 dia. cortical screw, 3.5 mm hex head,
75 mm length, 22-13-5 stainless steel, sterile bone screw; Cat. no. 2253-75-42
(00225307542), 
g) Zimmer M/DN Intramedullary Fixation 4.2 dia. cortical screw, 3.5 mm hex head,
85 mm length, 22-13-5 stainless steel, sterile bone screw; Cat. no. 2253-85-42
(00225308542). 
h) Zimmer M/DN Intramedullary Fixation 5.5 dia. cortical screw, 3.5 mm hex head,
95 mm length, 22-13-5 stainless steel, sterile bone screw; Cat. no. 2253-95-55
(00225309555), 
i) Zimmer M/DN Intramedullary Fixation 5.5 dia. cortical screw, 3.5 mm hex head,
100 mm length, 22-13-5 stainless steel, sterile bone screw; Cat. no. 2253-100-55
(00225310055), 
j) Zimmer Bone Screw, self tapping, 6.5 mm dia., 20 mm length, tivanium TI-6AL-4V
alloy, sterile, Catalog no. 6250-65-20 (00625006520), 
k) Zimmer HGP II Acetabular Components Bone Screw, self-tapping, 6.5 mm dia.,
30 mm length, tivanium TI-6AL-4V alloy, sterile, Catalog no. 6624-65-30
(00662406530),
l) Zimmer HGP II Acetabular Components Bone Screw, self-tapping, 6.5 mm dia.,
35 mm length, tivanium TI-6AL-4V alloy, sterile, Catalog no. 6624-65-35
(00662406535); 
m) Zimmer HGP II Acetabular Components Bone Screw, self-tapping, 6.5 mm dia.,
40 mm length, tivanium TI-6AL-4V alloy, sterile, Catalog no. 6624-65-40
(00662406540); 
n) Zimmer Herbert Cannulated Bone Screw System Bone Screw, 4.5 mm dia.,
25 mm length, tivanium TI-6AL-4V alloy, sterile, Catalog no. 47-1155-25-05
(47115502505), 
o) Zimmer Herbert Cannulated Bone Screw System Bone Screw, 6.5 mm dia.,
35 mm length, tivanium TI-6AL-4V alloy, sterile, Catalog no. 47-1155-35-07
(47115503507), 
p) Zimmer Herbert Cannulated Bone Screw System Bone Screw, 4.5 mm dia.,
45 mm length, tivanium TI-6AL-4V alloy, sterile, Catalog no. 47-1155-45-05
(47115504505); 
q) Zimmer Herbert Cannulated Bone Screw System Bone Screw, 4.5 mm dia.,
50 mm length, tivanium TI-6AL-4V alloy, sterile, Catalog no. 47-1155-50-05
(47115505005), 
CODE
a) Lots 24122200, 24730600, 25342400, 25765300, 60025157 and 60034732;
b) Lots 24122600, 24122700, 24123000, 24667300, 24730700, 24730800,
25769900, 26003200, 26113700, 26435200, 60005229, 60019022,
60034734, 60034735, 60043306, 60043307 and 60043308;
c) Lots 24667700, 24732400, 24732500, 24732600, 24732700, 25770200,
26003400, 26435500, 60005230, 60019029, 60030638, 60034743,
60034748, 60043067 and 60043311;
d) Lots 17035000, 17308200, 17600700, 17733300, 18277400, 19001000,
19148900 and 19149000;
e) Lot 60310724;
f) Lots 14865700, 17308800, 17733600, 17733700, 19001400, 24543400
and 25105000;
g) Lots 15887800, 16568000, 17310000, 17310100, 17657500, 17734300
and 23083300;
h) Lots 17036400, 17658000, 17734800, 17887700, 18279700, 21846200,
22983000 and 23083900;
i) Lots 16568200, 16706800, 16706900, 17036500, 17658100, 17887900,
18129600, 18448200, 19501900, 19682100, 19805800, 22708900
and 23843500;
j) Lot 60330168;
k) Lots 12921700, 14414400, 14915600, 15878600, 70234200;
l) Lots 12662700, 12718800, 12718900, 12757600, 12808600, 13331900,
13776600, 13776700, 13935600, 14196100, 14238900, 15590000, 15590100,
15610700, 16419200, 16579200, 16718400, 16718500, 16718600, 43764000,
48387800, 49078400, 49539800, 51295600, 52044900, 57690600, 57690800,
58905500, 59114400, 59476000, 66527800, 66527900, 66816800, 68890400,
68890500, 68890600 and 69606300;
m) Lots 42330300, 56131900, 71517900, 73057500 and 77838200;
n) Lots 15550200 and 19495400;
o) Lot 15751900:
p) Lot 14878500;
q) Lot 19908900
RECALLING FIRM/MANUFACTURER
Recalling Firm: Zimmer Caribe, Inc., Warsaw, IN,
REASON Lack of assurance of sterility, as the sterile barrier packaging may have been compromised.
VOLUME OF PRODUCT IN COMMERCE 43
DISTRIBUTION Nationwide and Internationally

PRODUCT D711, Reference Membranes, Part #942-058, used with the ABL835 Blood Gas Analyzers,
CODE Lot Code: R0308, Expiration Date: 4/07; Lot Code: R0309, Expiration Date: 5/07; and Lot Code: R0310, Expiration Date: 5/07.
RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America Inc, WestlakeOH, 
Manufacturer: Radiometer Medical ApS, Bronshoj, Denmark. 
REASON The recalled membranes cause the status value of the pH and electrolyte electrodes to be offset causing readings outside of the acceptable range.
VOLUME OF PRODUCT IN COMMERCE 13 boxes
DISTRIBUTION Nationwide and Canada

PRODUCT ABL800 Series Blood Gas Analyzer equipped with Software Version 5.21, 
CODE Software Versions 5.21 i.e. ABL8xx 754R00xx N0xx
RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America, Inc., Westlake, OH,
Manufacturer: Radiometer Medical Aps, Bronshoj, Denmark. 
REASON ABL800 Series Blood Gas Analyzer became inoperable when the restore default setup button was touched.
VOLUME OF PRODUCT IN COMMERCE 36 analyzers
DISTRIBUTION Nationwide and Internationally