OCTOBER 2005

WEEK ENDING OCTOBER 8

PRODUCT Hemashield Vantage Vascular Graft. Vascular graft prosthesis.
CODE Catalog Number, Product Description, Lot Number: M002023750060, HVVG 6mm x 100cm, 5649003; M002023750060, HVVG 6mm x 100cm, 5948031; M002023750060, HVVG 6mm x 100cm, 6639821; M002023750060, HVVG 6mm x 100cm, 6647296; M002023750060, HVVG 6mm x 100cm, 7030066; M002023750070, HVVG 7mm x 100cm, 5669601; M002023750070, HVVG 7mm x 100cm, 5728283; M002023750070, HVVG 7mm x 100cm, 5821740; M002023750070, HVVG 7mm x 100cm, 6032950; M002023750070, HVVG 7mm x 100cm, 6962284; M002023750080, HVVG 8mm x 100cm, 6007978; M002023750080, HVVG 8mm x 100cm, 6279546; M002023750080, HVVG 8mm x 100cm,6722180; M002023750080, HVVG 8mm x 100cm, 6919131; M002023750080, HVVG 8mm x 100cm, 7161034; M002023750080, HVVG 8mm x 100cm, 7249926; M002023750080, HVVG 8mm x 100cm, 7402224; M002023750100, HVVG 10mm x 100cm, 6876321; M002023750100, HVVG 10mm x 100cm, 6962286; M002023754060, HVVG 6mm x 40cm, 5721446; M002023754060, HVVG 6mm x 40cm, 5728284; M002023754060, HVVG 6mm x 40cm, 5780318; M002023754060, HVVG 6mm x 40cm, 5860237; M002023754060, HVVG 6mm x 40cm, 5899944; M002023754060, HVVG 6mm x 40cm, 6011811; M002023754060, HVVG 6mm x 40cm, 6239445; M002023754060, HVVG 6mm x 40cm, 6268215; M002023754060, HVVG 6mm x 40cm, 6304504; M002023754060, HVVG 6mm x 40cm, 6320785; M002023754060, HVVG 6mm x 40cm, 6364795; M002023754060, HVVG 6mm x 40cm, 6440114; M002023754060, HVVG 6mm x 40cm, 6615660; M002023754060, HVVG 6mm x 40cm, 6628696; M002023754060, HVVG 6mm x 40cm, 6647104; M002023754060, HVVG 6mm x 40cm, 6696492; M002023754060, HVVG 6mm x 40cm, 6807737; M002023754060, HVVG 6mm x 40cm, 6917823; M002023754060, HVVG 6mm x 40cm, 6979854; M002023754060, HVVG 6mm x 40cm, 7000352; M002023754060, HVVG 6mm x 40cm, 7071546; M002023754060, HVVG 6mm x 40cm, 7200982; M002023754070, HVVG 7mm x 40cm, 5691792; M002023754070, HVVG 7mm x 40cm, 5726049; M002023754070, HVVG 7mm x 40cm, 5779847; M002023754070, HVVG 7mm x 40cm, 5797094; M002023754070, HVVG 7mm x 40cm, 5955887; M002023754070, HVVG 7mm x 40cm, 6011812; M002023754070, HVVG 7mm x 40cm, 6037086; M002023754070, HVVG 7mm x 40cm, 6136721; M002023754070, HVVG 7mm x 40cm, 6611338; M002023754070, HVVG 7mm x 40cm, 6657664; M002023754070, HVVG 7mm x 40cm, 6691603; M002023754070, HVVG 7mm x 40cm, 6851214; M002023754070, HVVG 7mm x 40cm, 7161578; M002023754070, HVVG 7mm x 40cm, 7175592; M002023754070, HVVG 7mm x 40cm, 7249015; M002023754070, HVVG 7mm x 40cm, 7402368; M002023754070, HVVG 7mm x 40cm, 7403197; M002023754070, HVVG 7mm x 40cm, 7535410; M002023754070, HVVG 7mm x 40cm, 7565108; M002023754070, HVVG 7mm x 40cm, 7631134; M002023754080, HVVG 8mm x 40cm, 5669605; M002023754080, HVVG 8mm x 40cm, 5728288; M002023754080, HVVG 8mm x 40cm, 5783271; M002023754080, HVVG 8mm x 40cm, 5797165; M002023754080, HVVG 8mm x 40cm, 5841181; M002023754080, HVVG 8mm x 40cm, 5912738; M002023754080, HVVG 8mm x 40cm, 5976346; M002023754080, HVVG 8mm x 40cm, 6008050; M002023754080, HVVG 8mm x 40cm, 6107641; M002023754080, HVVG 8mm x 40cm, 6138907; M002023754080, HVVG 8mm x 40cm, 6238549; M002023754080, HVVG 8mm x 40cm, 6320790; M002023754080, HVVG 8mm x 40cm, 6361655; M002023754080, HVVG 8mm x 40cm, 6427209; M002023754080, HVVG 8mm x 40cm, 6598660; M002023754080, HVVG 8mm x 40cm, 6617325; M002023754080, HVVG 8mm x 40cm, 6646329; M002023754080, HVVG 8mm x 40cm, 6647152; M002023754080, HVVG 8mm x 40cm, 6691608; M002023754080, HVVG 8mm x 40cm, 6876382; M002023754080, HVVG 8mm x 40cm, 7114368; M002023754080, HVVG 8mm x 40cm, 7138064; M002023754080, HVVG 8mm x 40cm, 7161580; M002023754080, HVVG 8mm x 40cm, 7175587; M002023754080, HVVG 8mm x 40cm, 7276583; M002023754080, HVVG 8mm x 40cm, 7402360; M002023754080, HVVG 8mm x 40cm, 7535409; M002023754100, HVVG 10mm x 40cm, 5691791; M002023754100, HVVG 10mm x 40cm, 5731519; M002023754100, HVVG 10mm x 40cm, 5891897; M002023754100, HVVG 10mm x 40cm, 6052007; M002023754100, HVVG 10mm x 40cm, 6418381; M002023754100, HVVG 10mm x 40cm, 6926326; M002023754100, HVVG 10mm x 40cm, 7000766; M002023754100, HVVG 10mm x 40cm, 7504489; M002023756060, HVVG 6mm x 60cm, 5707200; M002023756060, HVVG 6mm x 60cm, 5724793; M002023756060, HVVG 6mm x 60cm, 5745982; M002023756060, HVVG 6mm x 60cm, 5767491; M002023756060, HVVG 6mm x 60cm, 5767495; M002023756060, HVVG 6mm x 60cm, 5852837; M002023756060, HVVG 6mm x 60cm, 5853224; M002023756060, HVVG 6mm x 60cm, 5891798; M002023756060, HVVG 6mm x 60cm, 5947960; M002023756060, HVVG 6mm x 60cm, 5968420; M002023756060, HVVG 6mm x 60cm, 5988577; M002023756060, HVVG 6mm x 60cm, 6050593; M002023756060, HVVG 6mm x 60cm, 6096033; M002023756060, HVVG 6mm x 60cm, 6268220; M002023756060, HVVG 6mm x 60cm, 6347541; M002023756060, HVVG 6mm x 60cm, 6388858; M002023756060, HVVG 6mm x 60cm, 6418565; M002023756060, HVVG 6mm x 60cm, 6611350; M002023756060, HVVG 6mm x 60cm, 6628703; M002023756060, HVVG 6mm x 60cm, 6654862; M002023756060, HVVG 6mm x 60cm, 6655453; M002023756060, HVVG 6mm x 60cm, 6770384; M002023756060, HVVG 6mm x 60cm, 6807126; M002023756060, HVVG 6mm x 60cm, 6850094; M002023756060, HVVG 6mm x 60cm, 6919450; M002023756060, HVVG 6mm x 60cm, 7010071; M002023756060, HVVG 6mm x 60cm, 7057154; M002023756060, HVVG 6mm x 60cm, 7075488; M002023756060, HVVG 6mm x 60cm, 7161035; M002023756060, HVVG 6mm x 60cm, 7173034; M002023756060, HVVG 6mm x 60cm, 7230491; M002023756060, HVVG 6mm x 60cm, 7259726; M002023756060, HVVG 6mm x 60cm, 7528090; M002023756070, HVVG 7mm x 60cm, 5633991; M002023756070, HVVG 7mm x 60cm, 5691790; M002023756070, HVVG 7mm x 60cm, 5745981; M002023756070, HVVG 7mm x 60cm, 5773786; M002023756070, HVVG 7mm x 60cm, 5793283; M002023756070, HVVG 7mm x 60cm, 5853511; M002023756070, HVVG 7mm x 60cm, 5903684; M002023756070, HVVG 7mm x 60cm, 5988575; M002023756070, HVVG 7mm x 60cm, 6011813; M002023756070, HVVG 7mm x 60cm, 6037090; M002023756070, HVVG 7mm x 60cm, 6109301; M002023756070, HVVG 7mm x 60cm, 6282450; M002023756070, HVVG 7mm x 60cm, 6383841; M002023756070, HVVG 7mm x 60cm, 6440361; M002023756070, HVVG 7mm x 60cm, 6611337; M002023756070, HVVG 7mm x 60cm, 6667242; M002023756070, HVVG 7mm x 60cm, 6754019; M002023756070, HVVG 7mm x 60cm, 6930145; M002023756070, HVVG 7mm x 60cm, 7137399; M002023756070, HVVG 7mm x 60cm, 7276579; M002023756070, HVVG 7mm x 60cm, 7403200; M002023756070, HVVG 7mm x 60cm, 7545048; M002023756070, HVVG 7mm x 60cm, 7545052; M002023756070, HVVG 7mm x 60cm, 7579389; M002023756080, HVVG 8mm x 60cm, 5633990; M002023756080, HVVG 8mm x 60cm, 5726141; M002023756080, HVVG 8mm x 60cm, 5856535; M002023756080, HVVG 8mm x 60cm, 5891889; M002023756080, HVVG 8mm x 60cm, 5948034; M002023756080, HVVG 8mm x 60cm, 6073565; M002023756080, HVVG 8mm x 60cm, 6185146; M002023756080, HVVG 8mm x 60cm, 6339117; M002023756080, HVVG 8mm x 60cm, 6388288; M002023756080, HVVG 8mm x 60cm, 6415673; M002023756080, HVVG 8mm x 60cm, 6598671; M002023756080, HVVG 8mm x 60cm, 6610550; M002023756080, HVVG 8mm x 60cm, 6628701; M002023756080, HVVG 8mm x 60cm, 6722144; M002023756080, HVVG 8mm x 60cm, 6754031; M002023756080, HVVG 8mm x 60cm, 6820115; M002023756080, HVVG 8mm x 60cm, 6861408; M002023756080, HVVG 8mm x 60cm, 6926327; M002023756080, HVVG 8mm x 60cm, 6948142; M002023756080, HVVG 8mm x 60cm, 7008949; M002023756080, HVVG 8mm x 60cm, 7071547; M002023756080, HVVG 8mm x 60cm, 7115438; M002023756080, HVVG 8mm x 60cm, 7528226; M002023756100, HVVG 10mm x 60cm, 5856641; M002023756100, HVVG 10mm x 60cm, 6051871; M002023756100, HVVG 10mm x 60cm, 6193600; M002023756100, HVVG 10mm x 60cm, 6524295; M002023756100, HVVG 10mm x 60cm, 6620113; M002023756100, HVVG 10mm x 60cm, 6691596; M002023756100, HVVG 10mm x 60cm, 6960306; M002023756100, HVVG 10mm x 60cm, 6961911; M002023758060, HVVG 6mm x 80cm, 5669380; M002023758060, HVVG 6mm x 80cm, 5779844; M002023758060, HVVG 6mm x 80cm, 5922480; M002023758060, HVVG 6mm x 80cm, 5964495; M002023758060, HVVG 6mm x 80cm, 6032948; M002023758060, HVVG 6mm x 80cm, 6216685; M002023758060, HVVG 6mm x 80cm, 6304505; M002023758060, HVVG 6mm x 80cm, 6347707; M002023758060, HVVG 6mm x 80cm, 6364796; M002023758060, HVVG 6mm x 80cm, 6388362; M002023758060, HVVG 6mm x 80cm, 6416301; M002023758060, HVVG 6mm x 80cm, 6615663; M002023758060, HVVG 6mm x 80cm, 6628554; M002023758060, HVVG 6mm x 80cm, 6646521; M002023758060, HVVG 6mm x 80cm, 6654376; M002023758060, HVVG 6mm x 80cm, 6696487; M002023758060, HVVG 6mm x 80cm, 6722189; M002023758060, HVVG 6mm x 80cm, 6774183; M002023758060, HVVG 6mm x 80cm, 6816154; M002023758060, HVVG 6mm x 80cm, 6930142; M002023758060, HVVG 6mm x 80cm, 7000351; M002023758060, HVVG 6mm x 80cm, 7010061; M002023758060, HVVG 6mm x 80cm, 7161036; M002023758060, HVVG 6mm x 80cm, 7230490; M002023758060, HVVG 6mm x 80cm, 7630848; M002023758070, HVVG 7mm x 80cm, 5724827; M002023758070, HVVG 7mm x 80cm, 5728287; M002023758070, HVVG 7mm x 80cm, 5955888; M002023758070, HVVG 7mm x 80cm, 6050597; M002023758070, HVVG 7mm x 80cm, 6230257; M002023758070, HVVG 7mm x 80cm, 6288273; M002023758070, HVVG 7mm x 80cm, 6440359; M002023758070, HVVG 7mm x 80cm, 6691602; M002023758070, HVVG 7mm x 80cm, 6770192; M002023758070, HVVG 7mm x 80cm, 6851207; M002023758070, HVVG 7mm x 80cm, 7161577; M002023758070, HVVG 7mm x 80cm, 7175590; M002023758070, HVVG 7mm x 80cm, 7248691; M002023758070, HVVG 7mm x 80cm, 7402229; M002023758070, HVVG 7mm x 80cm, 7402371; M002023758070, HVVG 7mm x 80cm, 7631133; M002023758080, HVVG 8mm x 80cm, 5700634; M002023758080, HVVG 8mm x 80cm, 5765239; M002023758080, HVVG 8mm x 80cm, 5815016; M002023758080, HVVG 8mm x 80cm, 5911567; M002023758080, HVVG 8mm x 80cm, 5972377; M002023758080, HVVG 8mm x 80cm, 6068815; M002023758080, HVVG 8mm x 80cm, 6238548; M002023758080, HVVG 8mm x 80cm, 6256385; M002023758080, HVVG 8mm x 80cm, 6320652; M002023758080, HVVG 8mm x 80cm, 6347706; M002023758080, HVVG 8mm x 80cm, 6416627; M002023758080, HVVG 8mm x 80cm, 6598409; M002023758080, HVVG 8mm x 80cm, 6617271; M002023758080, HVVG 8mm x 80cm, 6647222; M002023758080, HVVG 8mm x 80cm, 6691605; M002023758080, HVVG 8mm x 80cm, 6876377; M002023758080, HVVG 8mm x 80cm, 6962282; M002023758080, HVVG 8mm x 80cm, 7114366; M002023758080, HVVG 8mm x 80cm, 7138066; M002023758080, HVVG 8mm x 80cm, 7161579; M002023758080, HVVG 8mm x 80cm, 7220186; M002023758080, HVVG 8mm x 80cm, 7276584; M002023758080, HVVG 8mm x 80cm, 7317245; M002023758080, HVVG 8mm x 80cm, 7402219; M002023758100, HVVG 10mm x 80cm, 5783268; M002023758100, HVVG 10mm x 80cm, 5891786; M002023758100, HVVG 10mm x 80cm, 6165924; M002023758100, HVVG 10mm x 80cm, 6524301; M002023758100, HVVG 10mm x 80cm, 6598574; M002023758100, HVVG 10mm x 80cm, 7031876; M002023758100, HVVG 10mm x 80cm, 7317854; M002023758100, HVVG 10mm x 80cm, 7504480.
RECALLING FIRM/MANUFACTURER
Recalling Firm: Boston Scientific Corporation, Natick MA
Manufacturer: Boston Scientific Corporation, Wayne, NJ. 
REASON During surgery, several reports of the grafts fraying and tearing upon suturing by the surgeon.
VOLUME OF PRODUCT IN COMMERCE 3,487 devices
DISTRIBUTION Nationwide and Internationally

PRODUCT LifeScan brand OneTouch Ultra Meter – Owner's Booklet
CODE All System Kits shipped during June 2003 until December 2003
RECALLING FIRM/MANUFACTURER Lifescan Inc., Milpitas, CA
REASON LifeScan brand One Touch Ultra Meter – Owner's Booklets are missing instructions on how to set the Units of Measurement (UOM)—mg/dL versus mmol/L. This error has the postential to lead to acute hyperglycemic complications if the user fails to realize that 18mg/dL of glucose is equal to 1 (one) mmol/L.
VOLUME OF PRODUCT IN COMMERCE 572,000 units
DISTRIBUTION Nationwide

PRODUCT
a) Baxter Lock Box for 100 mL or 250 mL Bag Set, for use with 6060 Pump Family; Baxter product code 2L9354 and previously marketed by Sabratek under product code 606194
b) Baxter Lock Box for 250 mL Container, for use with 6060 Pump Family; Baxter product code 2L9363 and previously marketed by Sabratek under product code 606160; Product code 2L9363 is sold separately and
as a component of the 6060 Pain kit, product code 2M9832P
CODE
a) All lot numbers/serial numbers of product codes 2L9354 and 606194;
b) All lot numbers/serial numbers of product codes 2L9363, 606160 and 2M9832P
RECALLING FIRM/MANUFACTURER
Recalling Firm: Baxter Healthcare Corp., Round Lake, IL
Manufacturer: Baxter Healthcare Corp., Andover, MA. 
REASON The pump may free-flow if the door to the 6060 pump is not closed prior to loading the pump in to the lock box. The pump will alarm “Door Open", but if the pump is turned off to silence the alarm while remaining in the lock box, the fluid will free-flow.
VOLUME OF PRODUCT IN COMMERCE 1,486 units
DISTRIBUTION Nationwide and Internationally

PRODUCT
a) Bard® Luminexx® Endoscopic Biliary Stent and Delivery System, 7.5 FR, 8 mm x 60 mm Catalog Number: LXB08060
b) Bard® Luminexx® Endoscopic Biliary Stent and Delivery System, 7.5 FR, 8 mm x 80 mm Catalog Number: LXB08080,
c) Bard® Luminexx® Endoscopic Biliary Stent and Delivery System, 7.5 FR, 8 mm x 100 mm Catalog Number: LXB080100, 
d) Bard® Luminexx® Endoscopic Biliary Stent and Delivery System, 7.5 FR, 10 mm x 40 mm Catalog Number: LXB10040,
e) Bard® Luminexx® Endoscopic Biliary Stent and Delivery System, 7.5 FR, 10 mm x 60 mm Catalog Number: LXB10060, 
f) Bard® Luminexx® Endoscopic Biliary Stent and Delivery System, 7.5 FR, 10 mm x 80 mm Catalog Number: LXB10080,
g) Bard® Luminexx® Endoscopic Biliary Stent and Delivery System, 7.5 FR, 10 mm x 100 mm Catalog Number: LXB10100, 
CODE
a) Lot Number: 96CP1992;
b) Lot Number: 96CP1385;
c) Lot Number: 96CP2406;
d) Lot Numbers: 96CP1388, 96CP1389, 96CP2244;
e) Lot Numbers: 96CP1390, 96CP1391, 96CP1392, 96CP1393, 96CP1988, 96CP2497, 96CP2498, 96CP1996;
f) Lot Numbers: 96CP1394, 96CP1395, 96CP1396, 96CP1397, 96CP1995;
g) Lot Number: 96CP2508
RECALLING FIRM/MANUFACTURER Angiomed Gmbh, Karlsruhe, Germany, Federal Republic  
REASON Biliary Stents may malfunction and cause failure to expand, fracture, incomplete deployment, inaccurate positioning, migration, or lumen perforation.
VOLUME OF PRODUCT IN COMMERCE 204 units
DISTRIBUTION Nationwide

PRODUCT Cook brand Fascial Dilator; 10FR/37cm, sterile, Rx only, Ref G14191; Catalog No. 0073710, CODE Lots 15-U1320351 and 15-U1320349
RECALLING FIRM/MANUFACTURER Cook Urological, Inc., Spencer, IN
REASON Mislabeled as to size.
VOLUME OF PRODUCT IN COMMERCE 3 units
DISTRIBUTION CA and FL

PRODUCT Bard Luminexx 3 Biliary Stent and Delivery System
CODE Catalog # Lot # LUB06040 96DP1000; LUB06060 96CP1044 & 96DP2404; LUB14040 96CP2110; LUB14060 96CP0771, 96CP2307, & 96CP2771; LUB14080 96DP2900; LUC05060 96DP2937; LUC06060 96CP2111 & 96DP2939; LUC08040 96CP2310 & 96DP2321
RECALLING FIRM/MANUFACTURER
Recalling Firm: Brad Peripheral Vascular Inc., Tempe, AZ
Manufacturer: Angiomed Gmbh, Karlsruhe, Germany, Federal Republic. 
REASON Broken stent struts of nitinol stents were noted in manufacturing.
VOLUME OF PRODUCT IN COMMERCE 102 units
DISTRIBUTION Nationwide

PRODUCT Guidant OTW OMNILINK.018 & .035 Biliary Stent Systems
CODE Model number of the device and manufacturing lots involved: OTW OmniLink 18 1007942-18 OTW OMNI 18 5.0MMX18MMX135CM 1007944-18 OTW OMNI 18 6.0MMX18MMX135CM 1007946-18 OTW OMNI 18 7.0MMX18MMX135CM 1007948-18 OTW OMNI 18 8.0MMX18MMX135CM 1007949-18 OTW OMNI 18 9.0MMX18MMX135CM 1007950-18 OTW OMNI 18 10.0MMX18MMX135CM 1007980-18 OTW OMNI 18 5.0MMX18MMX80CM 1007982-18 OTW OMNI 18 6.0MMX18MMX80CM 1007984-18 OTW OMNI 18 7.0MMX18MMX80CM 1007986-18 OTW OMNI 18 8.0MMX18MMX80CM 1007987-18 OTW OMNI 18 9.0MMX18MMX80CM 1007988-18 OTW OMNI 18 10.0MMX18MMX80CM 1007942-28 OTW OMNI 18 5.0MMX28MMX135CM 1007944-28 OTW OMNI 18 6.0MMX28MMX135CM 1007946-28 OTW OMNI 18 7.0MMX28MMX135CM 1007948-28 OTW OMNI 18 8.0MMX28MMX135CM 1007949-28 OTW OMNI 18 9.0MMX28MMX135CM 1007950-28 OTW OMNI 18 10.0MMX28MMX135CM 1007980-28 OTW OMNI 18 5.0MMX28MMX80CM 1007982-28 OTW OMNI 18 6.0MMX28MMX80CM 1007984-28 OTW OMNI 18 7.0MMX28MMX80CM 1007986-28 OTW OMNI 18 8.0MMX28MMX80CM 1007987-28 OTW OMNI 18 9.0MMX28MMX80CM 1007988-28 OTW OMNI 18 10.0MMX28MMX80CM 1007942-38 OTW OMNI 18 5.0MMX38MMX135CM 1007944-38 OTW OMNI 18 6.0MMX38MMX135CM 1007946-38 OTW OMNI 18 7.0MMX38MMX135CM 1007948-38 OTW OMNI 18 8.0MMX38MMX135CM 1007949-38 OTW OMNI 18 9.0MMX38MMX135CM 1007950-38 OTW OMNI 18 10.0MMX38MMX135CM 1007980-38 OTW OMNI 18 5.0MMX38MMX80CM 1007982-38 OTW OMNI 18 6.0MMX38MMX80CM 1007984-38 OTW OMNI 18 7.0MMX38MMX80CM 1007986-38 OTW OMNI 18 8.0MMX38MMX80CM 1007987-38 OTW OMNI 18 9.0MMX38MMX80CM 1007988-38 OTW OMNI 18 10.0MMX38MMX80CM 1007942-58 OTW OMNI 18 5.0MMX58MMX135CM 1007944-58 OTW OMNI 18 6.0MMX58MMX135CM 1007946-58 OTW OMNI 18 7.0MMX58MMX135CM 1007948-58 OTW OMNI 18 8.0MMX58MMX135CM 1007949-58 OTW OMNI 18 9.0MMX58MMX135CM 1007950-58 OTW OMNI 18 10.0MMX58MMX135CM 1007980-58 OTW OMNI 18 5.0MMX58MMX80CM 1007982-58 OTW OMNI 18 6.0MMX58MMX80CM 1007984-58 OTW OMNI 18 7.0MMX58MMX80CM 1007986-58 OTW OMNI 18 8.0MMX58MMX80CM 1007987-58 OTW OMNI 18 9.0MMX58MMX80CM 1007988-58 OTW OMNI 18 10.0MMX58MMX80CM 1007940-12 OTW OMNI 18 4.0MMX12MMX135CM 1007942-12 OTW OMNI 18 5.0MMX12MMX135CM 1007944-12 OTW OMNI 18 6.0MMX12MMX135CM 1007946-12 OTW OMNI 18 7.0MMX12MMX135CM 1007978-12 OTW OMNI 18 4.0MMX12MMX80CM 1007980-12 OTW OMNI 18 5.0MMX12MMX80CM 1007982-12 OTW OMNI 18 6.0MMX12MMX80CM 1007984-12 OTW OMNI 18 7.0MMX12MMX80CM 1007940-16 OTW OMNI 18 4.0MMX16MMX135CM 1007942-16 OTW OMNI 18 5.0MMX16MMX135CM 1007944-16 OTW OMNI 18 6.0MMX16MMX135CM 1007946-16 OTW OMNI 18 7.0MMX16MMX135CM 1007978-16 OTW OMNI 18 4.0MMX16MMX80CM 1007980-16 OTW OMNI 18 5.0MMX16MMX80CM 1007982-16 OTW OMNI 18 6.0MMX16MMX80CM 1007984-16 OTW OMNI 18 7.0MMX16MMX80CM 1007978-18 OTW OMNI 18 4.0MMX18MMX80CM 1007940-18 OTW OMNI 18 4.0MMX18MMX135CM US Part Numbers Omnilink .035 1008159-18 OTW OMNI 35 4.0MMX18MMX135CM 1008161-18 OTW OMNI 35 5.0MMX18MMX135CM 1008163-18 OTW OMNI 35 6.0MMX18MMX135CM 1008165-18 OTW OMNI 35 7.0MMX18MMX135CM 1008167-18 OTW OMNI 35 8.0MMX18MMX135CM 1008168-18 OTW OMNI 35 9.0MMX18MMX135CM 1008169-18 OTW OMNI 35 10.0MMX18MMX135CM 1008174-18 OTW OMNI 35 4.0MMX18MMX80CM 1008176-18 OTW OMNI 35 5.0MMX18MMX80CM 1008178-18 OTW OMNI 35 6.0MMX18MMX80CM 1008180-18 OTW OMNI 35 7.0MMX18MMX80CM 1008182-18 OTW OMNI 35 8.0MMX18MMX80CM 1008183-18 OTW OMNI 35 9.0MMX18MMX80CM 1008184-18 OTW OMNI 35 10.0MMX18MMX80CM 1008161-28 OTW OMNI 35 5.0MMX28MMX135CM 1008163-28 OTW OMNI 35 6.0MMX28MMX135CM 1008165-28 OTW OMNI 35 7.0MMX28MMX135CM 1008167-28 OTW OMNI 35 8.0MMX28MMX135CM 1008168-28 OTW OMNI 35 9.0MMX28MMX135CM 1008169-28 OTW OMNI 35 10.0MMX28MMX135CM 1008176-28 OTW OMNI 35 5.0MMX28MMX80CM 1008178-28 OTW OMNI 35 6.0MMX28MMX80CM 1008180-28 OTW OMNI 35 7.0MMX28MMX80CM 1008182-28 OTW OMNI 35 8.0MMX28MMX80CM 1008183-28 OTW OMNI 35 9.0MMX28MMX80CM 1008184-28 OTW OMNI 35 10.0MMX28MMX80CM 1008161-38 OTW OMNI 35 5.0MMX38MMX135CM 1008163-38 OTW OMNI 35 6.0MMX38MMX135CM 1008165-38 OTW OMNI 35 7.0MMX38MMX135CM 1008167-38 OTW OMNI 35 8.0MMX38MMX135CM 1008168-38 OTW OMNI 35 9.0MMX38MMX135CM 1008169-38 OTW OMNI 35 10.0MMX38MMX135CM 1008176-38 OTW OMNI 35 5.0MMX38MMX80CM 1008178-38 OTW OMNI 35 6.0MMX38MMX80CM 1008180-38 OTW OMNI 35 7.0MMX38MMX80CM 1008182-38 OTW OMNI 35 8.0MMX38MMX80CM 1008183-38 OTW OMNI 35 9.0MMX38MMX80CM 1008184-38 OTW OMNI 35 10.0MMX38MMX80CM 1008161-58 OTW OMNI 35 5.0MMX58MMX135CM 1008163-58 OTW OMNI 35 6.0MMX58MMX135CM 1008165-58 OTW OMNI 35 7.0MMX58MMX135CM 1008167-58 OTW OMNI 35 8.0MMX58MMX135CM 1008168-58 OTW OMNI 35 9.0MMX58MMX135CM 1008169-58 OTW OMNI 35 10.0MMX58MMX135CM 1008176-58 OTW OMNI 35 5.0MMX58MMX80CM 1008178-58 OTW OMNI 35 6.0MMX58MMX80CM 1008180-58 OTW OMNI 35 7.0MMX58MMX80CM 1008182-58 OTW OMNI 35 8.0MMX58MMX80CM 1008183-58 OTW OMNI 35 9.0MMX58MMX80CM 1008184-58 OTW OMNI 35 10.0MMX58MMX80CM 1008159-12 OTW OMNI 35 4.0MMX12MMX135CM 1008161-12 OTW OMNI 35 5.0MMX12MMX135CM 1008163-12 OTW OMNI 35 6.0MMX12MMX135CM 1008165-12 OTW OMNI 35 7.0MMX12MMX135CM 1008174-12 OTW OMNI 35 4.0MMX12MMX80CM 1008176-12 OTW OMNI 35 5.0MMX12MMX80CM 1008178-12 OTW OMNI 35 6.0MMX12MMX80CM 1008180-12 OTW OMNI 35 7.0MMX12MMX80CM 1008159-16 OTW OMNI 35 4.0MMX16MMX135CM 1008161-16 OTW OMNI 35 5.0MMX16MMX135CM 1008163-16 OTW OMNI 35 6.0MMX16MMX135CM 1008165-16 OTW OMNI 35 7.0MMX16MMX135CM 1008174-16 OTW OMNI 35 4.0MMX16MMX80CM 1008176-16 OTW OMNI 35 5.0MMX16MMX80CM 1008178-16 OTW OMNI 35 6.0MMX16MMX80CM 1008180-16 OTW OMNI 35 7.0MMX16MMX80CM
RECALLING FIRM/MANUFACTURER Guidant Endovascular Solutions, Inc., Temecula, CA
REASON Stents may become loose or dislodged from the delivery catheter.
VOLUME OF PRODUCT IN COMMERCE 69,700 units
DISTRIBUTION Nationwide and Internationally

PRODUCT
a) CryoValve, Pulmonary Valve & Conduit, Model # PV00
b) CryoValve, Aortic Valve & Conduit SG, Model # SGAV00
c) CryoValve, Pulmonary Valve & Conduit SG, Model # SGPPV00
CODE
a) Donor #58842, Serial #7426089;
b) Donor #65389, Serial #7943760;
c) Donor #62963, Serial #7819366
RECALLING FIRM/MANUFACTURER Cryolife, Inc., Kennesaw, GA
REASON Tissues associated with complaints of alleged recipient infection.
VOLUME OF PRODUCT IN COMMERCE 1 tissue
DISTRIBUTION MO, TX and Australia

PRODUCT CryoValve, Pulmonary Valve & Conduit, Donor #68538, Model # PV00
CODE Serial #8045214
RECALLING FIRM/MANUFACTURER Cryolife, Inc., Kennesaw, GA
REASON Mislabeling; An allograft labeled as an aortic valve & conduit, actually contained a pulmonary valve and conduit.
VOLUME OF PRODUCT IN COMMERCE 1 tissue
DISTRIBUTION CA

PRODUCT Elx800 microtiter plate reader software program written for use with ORTHO HCV Version 3.0 ELISA Test System. 
CODE Version 12 and older software
RECALLING FIRM/MANUFACTURER
Recalling Firm: Diasorin Inc., Stillwater, MN
Manufacturer: Ortho Clinical Diagnostics, Inc., Raritan, NJ. 
REASON An error in the acceptance specification for the substrate blank optical density was discovered in the Bio-Tek Elx800 programming specifications for the ORTHO HCV Version 3.0 ELISA Test System Assay.
VOLUME OF PRODUCT IN COMMERCE 128
DISTRIBUTION Nationwide

WEEK ENDING OCTOBER 15

PRODUCT Prisma continuous renal replacement system. Model/Catalog Numbers: 018080100, 018080101, 018080001D
CODE All serial numbers
RECALLING FIRM/MANUFACTURER
Recalling Firm: Gambro Renal Products, Lakewood, CO
Manufacturer: Gambro Dasco S.p.A. Monitor Division, Medolla, Italy. 
REASON Critically ill patients undergoing continuous renal replacement therapy may suffer excessive fluid loss. Failure of the operator to adequately address the cause for an ‘Incorrect Weight Change Detected” alarm and choose to continue the procedure could resulting excessive ultrafiltration.
VOLUME OF PRODUCT IN COMMERCE 1,964 units
DISTRIBUTION Nationwide

PRODUCT FreeStyle CoPilot Web Based Data Management System, and FreeStyle CoPilot Health Management System, Versions 1 and 2. Web Based Data Management System, no packaging.  
CODE Versions 1 and 2. Built 1041 and 1048. Product is a downloadable software package
RECALLING FIRM/MANUFACTURER Abbott Laboratories Medisense Products, Alameda, CA
REASON Downloadable software obtained from the firm’s website displays user data (including but not limited to insulin levels and pump statistics data) in a matter that may be misinterpreted and result in mistreatment of therapy.
VOLUME OF PRODUCT IN COMMERCE N/A. Product is a downloadable software package.
DISTRIBUTION Nationwide

PRODUCT OASIS Medical Vidaurri LASIK Flap Irrigator, Product Specification 4036V. The Vidaurri LASIK Flap Irrigator is a 16g x 1/2" stainless steel cannula with two 25g x 9/16" tip extensions. There are eight .25mm ports (four per tip).
CODE A. Product Specification 4036V Lot D1201AD Lot D0702A Y Lot D0802J Lot D09028 Lot D0103V Lot D0303S Lot D1203AH Lot D0104AC Lot D0304AQ Lot D0504AG Lot D0105AV Lot D0205J Lot D0305AH Lot D03058D Lot D0605AE Lot D0705Y Lot D0805A8 B. Product Specification 8065403604 Lot D0304S Lot D0405A Y Lot D05058H C. Product Specification MK-036V Lot D0305Y D. Product Specification 8-4036V-NS Lot D060488 Lot D0804T Lot D0904A Lot D1004AV Lot D12048X
RECALLING FIRM/MANUFACTURER Oasis Medical, Inc., Glendora, CA
REASON Foreign deposit present on the surface of the Vidaurri Lasik Flap Irrigators.
VOLUME OF PRODUCT IN COMMERCE 9,859 cannulas
DISTRIBUTION Nationwide and Internationally

WEEK ENDING OCTOBER 22

PRODUCT
a) Stackhouse FreedomAire Lens/Hood; a sterile disposable protective head garment for use with the FreedomAire Surgical Helmet System, providing a clear plastic bubble lens and protective head hood to connect to the surgical helmet; individually packaged hoods, 12 per case. Part #10450, Catalog # SA-500/US,
b) Stackhouse Lens/Hood, Mark III Universal; a sterile disposable protective head garment for use with the Steakhouse Mark III Helmet, providing a clear plastic bubble lens and protective head hood to connect to the
surgical helmet; individually packaged hoods, 12 per case. Part #10567, Catalog # SA-300/US,
CODE
a) Lot number: 16963;
b) Lot number: 19253
RECALLING FIRM/MANUFACTURER
Recalling Firm: VIASYS Med Systems, Wheeling, IL
Manufacturer: Servicios De Ensamble Internationales S.A. de C.V., Juarez, Mexico. 
REASON The packaging may be compromised such that sterility of the hoods cannot be guaranteed.
VOLUME OF PRODUCT IN COMMERCE 606 cases
DISTRIBUTION Nationwide, Japan, Chile and Switzerland

PRODUCT D-Start Dry Hemostatic Bandage. The product is an adhesive bandage. 
CODE Lot numbers: 301323A and 301393
RECALLING FIRM/MANUFACTURER
Recalling Firm: B. Braun Medical, Inc, Allentown, PA
Manufacturer: Vascular Solutions, Minneapolis, MN. 
REASON Package sealing defect -- product may not be sterile.
VOLUME OF PRODUCT IN COMMERCE 60 units
DISTRIBUTION Nationwide

PRODUCT
a) Hard shell carry case for the Lifepak 500 AED. Catalog number 3005384-000 through 002,
b) Hard shell carry case for the Lifepak CR Plus. Catalog number 3201582-000 through --002,
CODE Cases do not contain serial numbers or lot numbers.
RECALLING FIRM/MANUFACTURER Medtronic Emergency Response Systems, Inc., Redmond,WA
REASON Some hard shell carry case for Lifepak 500 and Lifepak CR Plus automated external defibrillators may have blocked pressure vent which may make it difficult to open the case after a sudden decrease in altitude (such as in aircraft landing or mountain descent).
VOLUME OF PRODUCT IN COMMERCE
a) 4,608 cases;
b) 133 cases
DISTRIBUTION Nationwide and Internationally

PRODUCT
a) Nichols Advantage Invasive Trophoblast Antigen (ITA) Assay Cartridges. Catalog #62-7052
b) Hyperglycosylated Human Chorionic Gonadotropin (H-hCG) Assay Cartridges. Catalog # 62-7039,
CODE
a) Lot # 62-500261;
b) Lot # 62-500262
RECALLING FIRM/MANUFACTURER Nichols Institute Diagnostics, San Clemente, CA
REASON Cartridge does not maintain its stability over the 12 month expiration dating of the product, and that stability is assured only through 6 months.
VOLUME OF PRODUCT IN COMMERCE 196 units
DISTRIBUTION Nationwide, China, Germany and Australia

PRODUCT Architect c800 for Conventional Units and SI Units, Version 5.00,
CODE
List number 08G98-07 for Conventional Units;
List number 04J62-07 for SI Units
RECALLING FIRM/MANUFACTURER Abbott laboratories, Inc., Irving, TX,
REASON Device has incorrect notation associated with therapeutic drug monitoring assays. These assays have an "IUO" notation after the results on the printed reports indicating that the results are for "Investigational Use Only". For both Conventional Units and SI Units, the ARCHITECT c8000 Assay Disk, Version 5.00 has the reagent name for Phenobarbital identified incorrectly as PHEN0 with a zero (()) instead of with the letter "O". This causes an error code to be generated during installation.
VOLUME OF PRODUCT IN COMMERCE 310 units
DISTRIBUTION Nationwide and Internationally

PRODUCT Synchron LX Systems Version 4.5 Operating Software
CODE Synchron LX 20 and PRO Version 4.5 Operating Software Part Number A21127 (Upgrade Kit) Software Versions V4.5N4.6 CDB (Chemistry Database). Synchron LXi Version 4.5 Operating Software A21128 (Upgrade Kit) Software Versions V4.5N4.6 CDB(Chemistry Database)
RECALLING FIRM/MANUFACTURER Beckman Coulter, Inc., Brea, CA
REASON Synchron LX Systems Operating Software Version 4.5/v4.6 CBD (Chemistry Database) introduced an adjustment to the urine Uric Acid database calculation scheme which for some patients, depending on the volume of 24 hour urine sample, results may be reduced from above the upper limit of the laboratory's reference interval to below the limit.
VOLUME OF PRODUCT IN COMMERCE 868 units
DISTRIBUTION Nationwide and Canada

PRODUCT Architect c8000 Analyzer System Processing Module, a fully automated, random access, clinical chemistry analyzer which utilizes spectrophotometry (monochromatic and bichromatic modes of measurement) for photometric based determinations. List number 1G06-01
CODE All serial numbers
RECALLING FIRM/MANUFACTURER
Recalling Firm: Abbott Laboratories, Inc., Irving, TX
Manufacturer: Toshiba Medical Systems Company, Tochigi-Ken, Japan. 
REASON Software deficiency causing potential shift in absorbance readings for test results.
VOLUME OF PRODUCT IN COMMERCE 1,107 units
DISTRIBUTION Nationwide and Internationally

PRODUCT Allocraft TLIF Bone Inserter. Catalog No. 483600100
CODE Lot Code: 04C570
RECALLING FIRM/MANUFACTURER
Recalling Firm: Stryker Howmedica Osteonics Corp., Mahwah, NJ
Manufacturer: Stryke, Cestas, France. 
REASON The short arm of the bone inserter may fracture during use and potentially result in an adverse health consequence.
VOLUME OF PRODUCT IN COMMERCE 9 units
DISTRIBUTION Nationwide

PRODUCT EBI Trochanteric 11mm Long Nail
CODE Catalog No./Lot Nos: 28226/164522; 28228/164523; 28230/175576; 28234/175583; 28238/175589; 28248/164527; 28324/164529; 28326/164531; 28328/164532; 28330/175614; 28332/170182; 28334/175621; 28338/175630; 28342/166415; 28344/164540; 28348/164545
RECALLING FIRM/MANUFACTURER EBI, L.P., Parsippany, NJ
REASON Upon evaluation, the firm found that the fatigue strength of the nails was below the required value as specified in the design specification.
VOLUME OF PRODUCT IN COMMERCE 102 nails
DISTRIBUTION Nationwide

PRODUCT
a) TC-PLUS Tibial Insert Ultra-Congruent Size 10; 9 mm, Model 25456
b) TC-PLUS Tibial Insert 2; 15 mm, Model 25309,
c) TC-PLUS Tibial Insert UC 4; 11 mm, Model 25427,
d) TC-PLUS Tibial Insert UC 4; 13 mm, Model 25428,
e) TC-PLUS Tibial Insert UC 6; 9 mm, Model 25436,
f) TC-PLUS Tibial Insert UC 6; 13 mm, Model 25438,
g) TC-PLUS Tibial Insert UC 10; 11 mm, Model 25457,
CODE
a) Lot code: 0204.13.2355;
b) Lot code: 0303.13.1623;
c) Lot code: 0501.13.0198;
d) Lot code: 0501.13.0263;
e) Lot code: 0501.13.0264;
f) Lot code: 0501.13.0199, 0303.13.1616;
g) Lot code: 0501.13.0200
RECALLING FIRM/MANUFACTURER
Recalling Firm: Plus Orthopedics USA, San Diego, CA
Manufacturer: Plus Endoprothetik AG, Rotkreuz, Switzerland. 
REASON Products did not meet specifications. The insert was produced with a smaller length dimension, resulting in micromotion experienced in preoperative assembly of a tibial insert and tibial baseplate.
VOLUME OF PRODUCT IN COMMERCE 108 units
DISTRIBUTION AZ, CA, ID, NY, TX, and VA

PRODUCT
a) Pioneer Surgical brand Quantum Spinal Rod System Polyaxial Screw, length 45 mm; Catalog number
10-575-PA-45. 
b) Pioneer Surgical brand Quantum Spinal Rod System Polyaxial Screw, length 40 mm; Catalog number
10-575-PA-40
CODE
a) Lot code: 306507;
b) Lot code: 308623
RECALLING FIRM/MANUFACTURER Pioneer Surgical Technology, Marquette, MI
REASON Screw may be incorrectly etched as to length. Screw that is etched "45mm" may be 40mm rather than 45mm. Manufacturer is recalling both lots because of a "potential of lot contamination".
VOLUME OF PRODUCT IN COMMERCE 234 screws
DISTRIBUTION Nationwide

PRODUCT Interis Allura BiPlane X-Ray System
CODE The firm utilizes site numbers: 520582, 531926, 535761, 535775, 520949, 530660, 530654, 522136, 521325, 532241, 531939, 520721, 104949, 530396, 532152, 530653, 532383, 531649, 534756, 533290, 521525, 533397, 534786, 520688, 522456, 520964, 532218, 521149, 520275, 532240, 531758, 520666, 519196, 531799, 519981, 520756, 76992, 534131
RECALLING FIRM/MANUFACTURER
Recalling Firm: Philips Medical Systems North America Co. Phillips, Bothell, WA
Manufacturer: Philips Medical Systems, Netherlands. 
REASON
1) During positioning of the Lateral Stand the rotation and/or skew movement may travel unexpectedly in opposite direction; 

2) movement may be faster than normal speed; 3) the stand may not stop immediately upon release of the joystick.
VOLUME OF PRODUCT IN COMMERCE 38 units
DISTRIBUTION Nationwide

PRODUCT
a) Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 40 mm dia. articular surface, small, sterile, for cemented use only in the USA, Cat. no. 00430604000.
b) Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 46 mm dia. articular surface, small, sterile, for cemented use only in the USA, Cat. no. 00430604046
c) Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 40 mm dia. articular surface,med, sterile, for cemented use only in the USA, Cat. no.00430604640
d) Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 46 mm dia. articular surface, med, sterile, for cemented use only in the USA, Cat. no. 00430604600
e) Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 52 mm dia. articular surface, med, sterile, for cemented use only in the USA, Cat. no. 00430604652,
f) Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 46 mm dia. articular surface, large, sterile, for cemented use only in the USA, Cat. no. 00430605246
g) Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 52 mm dia. articular surface, large, sterile, for cemented use only in the USA, Cat. no. 00430605200
h) Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 56 mm dia. articular surface, small, sterile, for cemented use only in the USA, Cat. no. 00430605256
CODE All lots
RECALLING FIRM/MANUFACTURER
Recalling Firm: Zimmer Inc., Warsaw, IN
Manufacturer: Zimmer Trabecular, Allendale, NJ. 
REASON The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement.
VOLUME OF PRODUCT IN COMMERCE 723 units
DISTRIBUTION Nationwide and Internationally

PRODUCT Rita Brand Habib 4X Electrosurgical Coagulator, Catalog Number: 4401 US
CODE Lot number: 0011
RECALLING FIRM/MANUFACTURER
Recalling Firm: Rita Medical Systems, Fremont, CA
Manufacturer: Rita Medical Systems, Manchester, GA. 
REASON Devices' sterility may be compromised as evidenced by a loss of package integrity.
VOLUME OF PRODUCT IN COMMERCE 262 units
DISTRIBUTION Nationwide

PRODUCT Baxter's Patient Care System (BPCS) software, composed of Pharmacy Management - Inpatient - product code 2M5013, Point of Care - product code 2M5014 and Management Reports - product code 2M5015, 
CODE Software version 4.1-77, 4.0-66 and 3.4-18.24
RECALLING FIRM/MANUFACTURER
Recalling Firm: Baxter Healthcare, Corp., Round Lake, IL5.
Manufacturer: Baxter Healthcare Corp./Patient Care System, Toronto, Canada.
REASON If a pharmacy end user alters the Administration Unit of Measure (AUOM) of an ordered drug, the AUOM will display incorrectly on the Medication Administration Record for the patient, possibly leading to medication errors.
VOLUME OF PRODUCT IN COMMERCE 5 units
DISTRIBUTION MO, MA, VT, CA and Canada

PRODUCT Stryker brand Intra-Compartmental (STIC) Pressure Monitor, Model 295-001-000
CODE Serial numbers: 0428702183 through 0518008193
RECALLING FIRM/MANUFACTURER Stryker Instruments, Division of Stryker Corporation, Kalamazoo, MI
REASON The pressure indicated may be inaccurate, and this inaccuracy may be not detectable by the user.
VOLUME OF PRODUCT IN COMMERCE 473 units
DISTRIBUTION Nationwide and Internationally

PRODUCT
a) Fetalgard Lite Fetal Monitors, Model APIS (Single Fetus Monitoring)
b) Fetalgard Lite Fetal Monitors, Model HCS (Home Care Single Fetus Monitoring)
CODE
a) System Serial Numbers: FA002021 FA002022 FA002025 FA002027 FA002033 FA002020 FA002029 FA002005 FA002041 FA002044 FA002046 FA002112 FA002097 FA002117 FA002028 FA002047 FA002048 FA002118 FA002034 FA002093 FA002095 FA002116 FA002008 FA002023 FA002030 FA002031 FA002009 FA002098 FA002040 FA002092 FA002024 FA002114 FA002113 FA002115 FA002026 FA002032 FA001328;
b) Serial Numbers: FA002010 FA002051 FA002054 FA002059 FA002061 FA002062 FA002064 FA002011 FA002013 FA002014 FA002056 FA002065 FA002058
RECALLING FIRM/MANUFACTURER Analogic Corporation, Peabody, MA
REASON Unit may not power up due to faulty circuit board.
VOLUME OF PRODUCT IN COMMERCE 50 units
DISTRIBUTION CT, PR, and Internationally

PRODUCT
a) Datex-Ohmeda S/5 Aespire Anesthesia System 7100,
b) Datex-Ohmeda S/5 Aespire Anesthesia System 7900,
c) Datex-Ohmeda Aespire Carestation. This device is not marketed in the US
CODE
AMXJ00203, AMXJ00204, AMXJ00248 thru AMXJ00257, AMXJ00259, AMXJ00260, AMXJ00262 thru AMXJ00305, AMXJ00308 thru AMXJ00312, AMXJ00317, AMXJ00318, AMXJ00333 thru AMXJ00335, AMXJ00338 thru AMXJ00342, AMXJ00350 thru AMXJ00354, AMXJ00380 thru AMXJ00411, AMXJ00416, AMXJ00419 thru AMXJ00425, AMXJ00429 thru AMXJ00442, AMXJ00451 thru AMXJ00473, AMXJ00485 thru AMXJ00576, ZMXJ00581 thru AMXJ00629, AMXJ00631 thru AMXJ00639, AMXJ00641 thru AMXJ00699;
b) ANCJ00100 thru ANCJ00113, ANCJ00116, ANCJ00118;
c) AMZJ00111 thru AMZJ00146
RECALLING FIRM/MANUFACTURER Datex-Ohmeda, Inc., Madison, WI
REASON Two conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired supplemental oxygen which could lead to serious injury if undetected. Also the O2 switch may leak or fail.
VOLUME OF PRODUCT IN COMMERCE 419 units
DISTRIBUTION Nationwide and Internationally

WEEK ENDING OCTOBER 29

PRODUCT VeriCal Calibrator Set,
CODE Lot numbers: 161813, 161907, 161908, and 161909
RECALLING FIRM/MANUFACTURER Biomerieux, Inc., Durham, NC,
REASON The product is being recalled due to mis-assignment of ISI values associated with VeriCal use. This could lead to inaccurate patient INR results.
VOLUME OF PRODUCT IN COMMERCE 2,334 kits
DISTRIBUTION Nationwide

PRODUCT
a) FreeStyle FLASH Blood Glucose Monitoring System,
b) Tracker Blood Glucose Monitoring System,
c) Xceed Blood Glucose Monitoring System,
d) ReliOn Ultima Glucose Monitoring System,
e) Kroger Glucose Monitoring System
f) RiteAid Glucose Monitoring System
g) Liberty Glucose Monitoring System
h) Boots Glucose Monitoring System
i) Optium Glucose Monitoring System
j) Xtra Classic Glucose Monitoring System,
k) Easy Glucose Monitoring System
l) SofTrac Glucose Monitoring System
m) FreeStyle Glucose Monitoring System
CODE
a) All lot numbers beginning with R and all lot numbers T-G158-62981 and below are included.
b) All lot numbers beginning with E are included in the recall;
c) All lot numbers beginning with XC0999 and below are included in this recall;
d), e), f), g), h), i), j), k), and l) All lot numbers are included in this recall;
m) For lot numbers beginning with A, B, C, D, G, H, K, and M, all lot numbers are included. All lot numbers
from V-G157-31584 and below are included, and also all lot numbers Y-G018-32983 and below are included in the recall.
RECALLING FIRM/MANUFACTURER Abbott Diabetes Care, Inc., Alameda, CA
REASON User may inadvertently change the Units of Measure from mg/dL to mmol/L and the blood glucose results could be misinterpreted. This may lead to under treatment and potential for hyperglycemia.
VOLUME OF PRODUCT IN COMMERCE Approximately 10,005,206 meters
DISTRIBUTION Nationwide and Internationally

PRODUCT Nichols Advantage QuiCk-Intraoperative Bio-Intact PTH Assay, Catalog No. 62-7045
CODE Lot 62-402651
RECALLING FIRM/MANUFACTURER Nichols Institute Diagnostics, San Clemente, CA,
REASON Controls were out of specification range.
VOLUME OF PRODUCT IN COMMERCE 323 units
DISTRIBUTION Nationwide, Japan, Spain, Germany, and Sweden

PRODUCT Thermocool Irrigated Catheters Celsius LABEL US Part Number# EU CATALOG # L-1189-01-S D17TCBLRT 35Q-13R PMA; L-1189-02-S D17TCDLRT 35Q-33R PMA; L-1189-03-S D17TCFLRT 35Q-53R PMA; L-1189-04-S D17TCJLRT 35Q-73R PMA; L-1189-15-S D71TCB5L252RT 35Q-17Q EXPORT; L-1189-16-S D71TCD5L252RT 35Q-37Q EXPORT; L-1189-17-S D71TCF5L252FT 35Q-57Q EXPORT; L-1189-03-S D71TFL252RT 35R-53R EXPORT; L-1192-03-S D71TFL252RT 35R-53R EXPORT; L-1192-15-S D71TB5L252RT 35R-17R EXPORT; L-1192-16-S D7ITD5L252RT 35R-37R EXPORT; L-1189-02-SI D17TCDLRTI N/A IDE; L-1189-15-SI D71TCB5L252RTI N/A IDE; L-1189-16-SI D71TCD5L252RTI N/A IDE; L-1189-17-SI D71CF5L252RTI N/A IDE Navistar LABEL US Part Number# EU CATALOG # L-1192-17-S D71TF5L252RT 35R-57R EXPORT; L-1197-14-S NI75TCBH 34H-17M PMA; L-1197-15-S NI75TCCH 34H-27M PMA; L-1197-16-S NI75TCDH 34H-37M PMA; L-1197-17-S NI75TCFH 34H-57M PMA; L-1197-18-S NI75TCJH 34H-J7M PMA; L-1208-05-S NS75TBCT252HS 34G-17M PMA; L-1208-06-S NS75TCCT252HS 34G-27M PMA; L-1208-07-S NS75TDCT252HS 34G-37M PMA; L-1208-08-S NS75TFCT252HS 34G-57M PMA; L-1197-14-SI NI75TCBHI N/A IDE; L-1197-15-SI NI75TCCHI N/A IDE; L-1197-16-SI NI75TCDHI N/A IDE; L-1197-17-SI NI75TCFHI N/A IDE; L-1197-18-SI NI75TCJHI N/A IDE; L-1208-05-SI NS75TBCT252HSI N/A IDE; L-1208-06-SI NS75TCCT252HSI N/A IDE; L-1208-07-SI NS75TDCT252HSI N/A IDE; L-1208-08-SI NS75TFCT252HSI N/A IDE
CODE Lot numbers: 13027030 to 13044734
RECALLING FIRM/MANUFACTURER Biosense Webster, Inc., Irwindale, CA
REASON Increased complaints regarding catheters with leaks and flow obstruction.
VOLUME OF PRODUCT IN COMMERCE 9,679 units
DISTRIBUTION FL, IL, MA, MI, OH, OK and Internationally

PRODUCT
a) Scorpio Total Knee Cruciate Retaining Femoral Component, #7, left, Catalog No. 70-4107L. 1 each to a package
b) Scorpio Total Knee Cruciate Retaining Femoral Component, #7, right, Catalog No. 70-4107R. 1 each to a package,
CODE
a) Lot Code: K04W537;
b) Lot code: K04W896
RECALLING FIRM/MANUFACTURER Stryker Howmedica Osteonics, Corp., Mahwah, NJ
REASON The Left Scorpio Total Knee Cruciate Retaining Waffle Femoral Component #7 packaged and labeled as left may contain a Right and vice versa.
VOLUME OF PRODUCT IN COMMERCE 15 units
DISTRIBUTION Nationwide and Internationally

PRODUCT T2 Recon Nail System
CODE Catalog No./Description/Lot Code: 18976065S-- T2 Recon Lag Screw 6.5x65MM--K454322; T2 Recon Lag Screw 6.5x65MM--K518614; T2 Recon Lag Screw 6.5x65MM--K518615; T2 Recon Lag Screw 6.5x65MM--K518616; T2 Recon Lag Screw 6.5x65MM--K518617; T2 Recon Lag Screw 6.5x65MM--K678251; T2 Recon Lag Screw 6.5x65MM--K678252. 18976070S-- T2 Recon Lag Screw 6.5x70MM--K412886; T2 Recon Lag Screw 6.5x70MM--K412887; T2 Recon Lag Screw 6.5x70MM--K454324; T2 Recon Lag Screw 6.5x70MM--K497821; T2 Recon Lag Screw 6.5x70MM--K518728; T2 Recon Lag Screw 6.5x70MM--K678253; T2 Recon Lag Screw 6.5x70MM--K678254. 18976075S-- T2 Recon Lag Screw 6.5x75MM--K400117; T2 Recon Lag Screw 6.5x75MM--K454326; T2 Recon Lag Screw 6.5x75MM--K518854; T2 Recon Lag Screw 6.5x75MM--K518855; T2 Recon Lag Screw 6.5x75MM--K518857; T2 Recon Lag Screw 6.5x75MM--K518858; T2 Recon Lag Screw 6.5x75MM--K678255; T2 Recon Lag Screw 6.5x75MM--K678256. 18976080S-- T2 Recon Lag Screw 6.5x80MM--K400118; T2 Recon Lag Screw 6.5x80MM--K454327; T2 Recon Lag Screw 6.5x80MM--K519005; T2 Recon Lag Screw 6.5x80MM--K519008; T2 Recon Lag Screw 6.5x80MM--K519009; T2 Recon Lag Screw 6.5x80MM--K678257; T2 Recon Lag Screw 6.5x80MM--K678258. 18976085S-- T2 Recon Lag Screw 6.5x85MM--K400119; T2 Recon Lag Screw 6.5x85MM--K412895; T2 Recon Lag Screw 6.5x85MM--K454328; T2 Recon Lag Screw 6.5x85MM--K488715; T2 Recon Lag Screw 6.5x85MM--K519003; T2 Recon Lag Screw 6.5x85MM--K519176; T2 Recon Lag Screw 6.5x85MM--K519177; T2 Recon Lag Screw 6.5x85MM--K519178; T2 Recon Lag Screw 6.5x85MM--K519182; T2 Recon Lag Screw 6.5x85MM--K519183; T2 Recon Lag Screw 6.5x85MM--K519184; T2 Recon Lag Screw 6.5x85MM--K519185; T2 Recon Lag Screw 6.5x85MM--K519186; T2 Recon Lag Screw 6.5x85MM--K678259; T2 Recon Lag Screw 6.5x85MM--K678260; T2 Recon Lag Screw 6.5x85MM--K757952. 18976090S-- T2 Recon Lag Screw 6.5x90MM--K400120; T2 Recon Lag Screw 6.5x90MM--K454329; T2 Recon Lag Screw 6.5x90MM--K488717; T2 Recon Lag Screw 6.5x90MM--K519286; T2 Recon Lag Screw 6.5x90MM--K519288; T2 Recon Lag Screw 6.5x90MM--K519289; T2 Recon Lag Screw 6.5x90MM--K519290; T2 Recon Lag Screw 6.5x90MM--K519291; T2 Recon Lag Screw 6.5x90MM--K519292; T2 Recon Lag Screw 6.5x90MM--K519293; T2 Recon Lag Screw 6.5x90MM--K678262; T2 Recon Lag Screw 6.5x90MM--K757953. 18976095S-- T2 Recon Lag Screw 6.5x95MM--K400121; T2 Recon Lag Screw 6.5x95MM--K412911; T2 Recon Lag Screw 6.5x95MM--K412914; T2 Recon Lag Screw 6.5x95MM--K412920; T2 Recon Lag Screw 6.5x95MM--K454330; T2 Recon Lag Screw 6.5x95MM--K488720; T2 Recon Lag Screw 6.5x95MM--K519401; T2 Recon Lag Screw 6.5x95MM--K519402; T2 Recon Lag Screw 6.5x95MM--K519405; T2 Recon Lag Screw 6.5x95MM--K519406; T2 Recon Lag Screw 6.5x95MM--K519407; T2 Recon Lag Screw 6.5x95MM--K519408; T2 Recon Lag Screw 6.5x95MM--K669388; T2 Recon Lag Screw 6.5x95MM--K678264; T2 Recon Lag Screw 6.5x95MM--K757954. 18976100S-- T2 Recon Lag Screw 6.5x100MM--K400122; T2 Recon Lag Screw 6.5x100MM--K454331; T2 Recon Lag Screw 6.5x100MM--K488721; T2 Recon Lag Screw 6.5x100MM--K519501; T2 Recon Lag Screw 6.5x100MM--K519502; T2 Recon Lag Screw 6.5x100MM--K519503; T2 Recon Lag Screw 6.5x100MM--K519504; T2 Recon Lag Screw 6.5x100MM--K519505; T2 Recon Lag Screw 6.5x100MM--K519508; T2 Recon Lag Screw 6.5x100MM--K519509; T2 Recon Lag Screw 6.5x100MM--K678265; T2 Recon Lag Screw 6.5x100MM--K678266; T2 Recon Lag Screw 6.5x100MM--K757955. 18976105S-- T2 Recon Lag Screw 6.5x105MM--K400123; T2 Recon Lag Screw 6.5x105MM--K454332; T2 Recon Lag Screw 6.5x105MM--K488751; T2 Recon Lag Screw 6.5x105MM--K519604; T2 Recon Lag Screw 6.5x105MM--K519605; T2 Recon Lag Screw 6.5x105MM--K519606; T2 Recon Lag Screw 6.5x105MM--K519607; T2 Recon Lag Screw 6.5x105MM--K519608; T2 Recon Lag Screw 6.5x105MM--K519609; T2 Recon Lag Screw 6.5x105MM--K519611; T2 Recon Lag Screw 6.5x105MM--K678267; T2 Recon Lag Screw 6.5x105MM--K678268; T2 Recon Lag Screw 6.5x105MM--K757974. 18976110S-- T2 Recon Lag Screw 6.5x110MM--K400124; T2 Recon Lag Screw 6.5x110MM--K454333; T2 Recon Lag Screw 6.5x110MM--K488758; T2 Recon Lag Screw 6.5x110MM--K519730; T2 Recon Lag Screw 6.5x110MM--K519731; T2 Recon Lag Screw 6.5x110MM--K519736; T2 Recon Lag Screw 6.5x110MM--K678269; T2 Recon Lag Screw 6.5x110MM--K757979. 18976115S-- T2 Recon Lag Screw 6.5x115MM--K400125; T2 Recon Lag Screw 6.5x115MM--K454334; T2 Recon Lag Screw 6.5x115MM--K488759; T2 Recon Lag Screw 6.5x115MM--K497822; T2 Recon Lag Screw 6.5x115MM--K519899; T2 Recon Lag Screw 6.5x115MM--K519900; T2 Recon Lag Screw 6.5x115MM--K519901; T2 Recon Lag Screw 6.5x115MM--K678271; T2 Recon Lag Screw 6.5x115MM--K678293. 18976120S-- T2 Recon Lag Screw 6.5x120MM--K400126; T2 Recon Lag Screw 6.5x120MM--K454335; T2 Recon Lag Screw 6.5x120MM--K520037; T2 Recon Lag Screw 6.5x120MM--K520038; T2 Recon Lag Screw 6.5x120MM--K520039; T2 Recon Lag Screw 6.5x120MM--K678294; T2 Recon Lag Screw 6.5x120MM--K678295; T2 Recon Lag Screw 6.5x120MM--K757980. 18976125S-- T2 Recon Lag Screw 6.5x125MM--K454336; T2 Recon Lag Screw 6.5x125MM--K488786; T2 Recon Lag Screw 6.5x125MM--K520139; T2 Recon Lag Screw 6.5x125MM--K520140; T2 Recon Lag Screw 6.5x125MM--K678296; T2 Recon Lag Screw 6.5x125MM--K678297; T2 Recon Lag Screw 6.5x125MM--K808190. 18976130S-- T2 Recon Lag Screw 6.5x130MM--K454337; T2 Recon Lag Screw 6.5x130MM--K520156; T2 Recon Lag Screw 6.5x130MM--K520157; T2 Recon Lag Screw 6.5x130MM--K520158; T2 Recon Lag Screw 6.5x130MM--K520159; T2 Recon Lag Screw 6.5x130MM--K520160; T2 Recon Lag Screw 6.5x130MM--K678298; T2 Recon Lag Screw 6.5x130MM--K678299; T2 Recon Lag Screw 6.5x130MM--K757982
RECALLING FIRM/MANUFACTURER
Recalling Firm: Stryker Howmedica Osteonics, Corp., Mahwah, NJ
Manufacturer: Stryker Trauma, Kiel, Germany. 
REASON The blister pack containing the Recon 6.5 Lag Screws may become damage during shipment resulting in loss of sterility.
VOLUME OF PRODUCT IN COMMERCE 4,698 units
DISTRIBUTION Nationwide

PRODUCT Baxter's Patient Care System (BPCS) software, Pharmacy Management - Inpatient - product code 2M5013 and Point of Care - product code 2M5014; made in Canada
CODE All units of software versions 4.0 (66) and 4.1 (77)
RECALLING FIRM/MANUFACTURER
Recalling Firm: Baxter Healthcare Corp., Round Lake, IL
Manufacturer: Baxter Healthcare Corp/Patient Care System, Toronto, Canada. 
REASON If an administrative user modifies any of the seven pre-defined Monitoring Parameters when they are attached to a medication in the drug file, a Point of Care end user will be unable to enter monitoring parameter results on the handheld.
VOLUME OF PRODUCT IN COMMERCE 5 units
DISTRIBUTION MO, MA, VT, CA and Canada

PRODUCT Boston Scientific brand Position Acquisition Module (PAM), a component of Realtime Position Management (RPM) System, Catalog/Model Numbers: 8200010, 8200TWA0
CODE All lots
RECALLING FIRM/MANUFACTURER Boston Scientific, San Jose, CA
REASON The device has the potential for overheating its transformer and has the potential for fire hazard.
VOLUME OF PRODUCT IN COMMERCE 94 units
DISTRIBUTION Nationwide, EU and Canada

PRODUCT
a) Remington Medical Ten Disposable Tear-Away Introducer Sets Size 7 French, Sterile: Single Use/Disposable Radiopaque Contents Ethylene Oxide Sterilized,
b) Remington Medical, Ten Disposable Tear-Away Introducer Sets, Size 8 French, Sterile: Single Use/Disposable Radiopague Contents Ethylene Oxide Sterilized,
c) Remington Medical, 8 French-Sheath Only, Size 8 French, Single Use. Ethylene Oxide sterilized;
CODE
a) Lot No. 040508, 041111, 042116, 042367, 043146, 050077, 041464;
b) Lot No. 043003, 042674, 042313;
c) Lot No. 041603, 041731, 050201, 050274
RECALLING FIRM/MANUFACTURER
Recalling Firm: Remington Medical Inc., Alpharetta, GA
Manufacturer: B. Braum Medical, Inc., Allentown, PA. 
REASON The Tear-Away introducer sheath in the Disposable Tear-Away Vascular Introducer kits may exhibit cracked hubs/handles and or improper peeling of the sheath during use.
VOLUME OF PRODUCT IN COMMERCE 4,800 units
DISTRIBUTION Nationwide

PRODUCT Gemini GXL PET/CT Systems, 6, 10, 16 slice configurations. (Product Numbers/manufacturer numbers: 16 slice: 4535 679 71891, 10 slice: 4535 679 75691, 6 slice: 4535 679 75691. Catalog numbers: 16 slice: 882410, 10 slice: 882400, 6 slice: 882390.)
CODE Serial numbers: 4001-4019
RECALLING FIRM/MANUFACTURER Philips Medical Systems (Cleveland), Inc., Cleveland, OH
REASON The image balance center point correction algorithm used in CT image reconstruction may cause a disappearance or alteration of small circular anatomical structures that are both near the image center and perpendicular to the image plane. This artifact may lead to the misinterpretation of anatomy at the center of the image.
VOLUME OF PRODUCT IN COMMERCE 19 devices
DISTRIBUTION Nationwide and Internationally

PRODUCT
a) DTX(tm) Plus DT-DL60 Continuous flush catheter, Part Number 682028;
b) CRITIFLOW(tm) TA4004 continuous flush catheter, Part Number 682060
c) TAKHPQ60R continuous flow catheter, Part Number 682117.
d) CRITIFLO(tm) TAK0460 continuous flow catheter, Part number 682118.
e) DT-DL60-1 W/Bifurcated, Part Number 682147,
f) CRITIFLO(tm) flush kit. Part Number 687198,
g) CRITIFLO(tm) Flush Kit. Part Number 687566,
CODE
a) Lot numbers: 102034, 103270, 106049, 107034, 108073, 109040, 107034, 110023, 110037, 111023, 203021, 205038, 206055, 207079, 207525, 208110, 208476, 209101, 210095, 211082, 212097, 301139, 301666, 303123, 303616, 305027, 306242, 308059, 309106, 310143, 311160, 312121, 401036, 402022, 403063, 404043, 405079;
b) Lot numbers: 012485, 101226, 102036, 103011, 103370, 106015, 107041, 108033, 109060, 110004, 111037, 201056, 204002, 205041, 206026, 207062, 208073, 211376, 212407, 301104, 302256, 303085, 303246, 304562, 306211, 306443, 308071, 308169, 308518, 309086, 310124, 311143, 312418, 403315, 404050, 405097, 407055, 407468, 408018;
c) Lot numbers 107612, 109012, 110019, 112035, 210060, 212386, 301150;
d) Lot Numbers 402035, 409152;
e) Lot numbers 010019, 102055, 105054, 103029, 106054, 107040, 108079, 109045, 110042, 201072, 203025, 207529, 310143, 310147, 311166, 312127, 401042, 401589, 403069;
f) Lot Numbers 301850, 302267, 303012, 305055;
g) Lot Number 302520
RECALLING FIRM/MANUFACTURER
Recalling Firm: Becton Dickinson Medical Systems, Sandy, UT,
Manufacturer: Bd Critical Care Systems P. L., Singapore. 
REASON Device could permit an over-infusion of IV solutions during pressure monitoring procedures.
VOLUME OF PRODUCT IN COMMERCE 27,085 units
DISTRIBUTION Nationwide

PRODUCT J-VAC Reservoir is a component of the J-VAC Closed Wound Drainage System. This system is a sterile, portable system used for closed wound drainage. This system is composed of two parts: the J-VAC Reservoir and Suction Drains
CODE Product code 2179 Lot 520198, exp June 2010; Lot 520200, exp June 2010
RECALLING FIRM/MANUFACTURER
Recalling Firm: Ethicon, Inc., Somerville, NJ
Manufacturer: Avil Medical Products, Inc., San Diego, CA. 
REASON Two lots of J-VAC Drain Reservors were shipped to Japan prior to sterilization and were released.
VOLUME OF PRODUCT IN COMMERCE 4,630 units
DISTRIBUTION Japan

PRODUCT
a) Merit Custom Procedure Kit, Part number: K08-01894, 
b) Merit Custom Procedure Kit, Part number: K09-08347,
c) Merit Custom Procedure Kit, Part number: K09-05248P, 
d) Merit Custom Procedure Kit, Part number: K09-05248AP,
e) Merit Custom Procedure Kit, Part number: K09-06702,
CODE
a) Lot number: A306856;
b) Lot numbers: A304591, A322679;
c) Lot numbers: A304826, A314447, A316544, A322290, A325742, A336903, A345355;
d) Lot numbers: A348563, A360799, A365773, A370274, A372525, A377402, A382733, A388638,
F392744;
e) Lot number: C229306
RECALLING FIRM/MANUFACTURER Merit Medical Systems, Inc., South Jordan, UT
REASON Component of procedure kits may allow higher than desired amount of IV solution to be infused in to patient during pressure monitoring procedure.
VOLUME OF PRODUCT IN COMMERCE 917 kits
DISTRIBUTION DC, MD, VA, the Netherlands and Switzerland

PRODUCT
a) Medrad Avanta Single Patient Disposable Sets. Catalog number AVA 500 SPAT,
b) Medrad Avanta Multi-patient Disposable Sets. Catalog number AVA 500 MPAT, 
CODE
a) Lot number 50577, 50578, 51219, 51220, 51221, and 52208;
b) Lot numbers 50627, 50628, 51069, 51070, 51071, and 51747
RECALLING FIRM/MANUFACTURER Medrad, Inc., Indianola, PA
REASON Stress crack on the stopcock.
VOLUME OF PRODUCT IN COMMERCE 2,425 sets
DISTRIBUTION Nationwide, Netherlands, Denmark, France, and Italy

PRODUCT Serum Blood Collection Tubes
CODE Item #455071, Lot #B040513 and Item #455092, Lot #B040503
RECALLING FIRM/MANUFACTURER Greiner VACUETTE North America, Monroe, NC
REASON Serum Blood Collection tubes found with excess of Blood Clot Accelerator (BCA) solution.
VOLUME OF PRODUCT IN COMMERCE 1,623 cases (1,947,600 pieces)
DISTRIBUTION Nationwide and Internationally

PRODUCT Y-valve, catalog number 7427-01 and T-valve, catalog number 7990-01, PSD/4 with Valve and Syringe Drive Module parts, catalog number 51863-01-IL
CODE For 7427-01 (Lot number, date of manufacture): 277702, 4/29/2003; 282550A, 8/11/2003; 282550, 8/13/2003; 286630, 10/1/2003; 287449, 10/20/2003; 287137, 10/20/2003; 287949, 12/18/2003; 301809, 6/30/2004; 305599, 8/19/2005; 309328, 11/5/2004; 313276, 12/17/2004; 314662, 3/7/2005; 317940, 4/6/2005; 319506, 4/11/2005; 321121, 5/6/2005; 324204, 6/23/2005; 325483, 7/20/2005. For 7990-01: 277703, 4/30/2003; 283119, 9/30/2003 291279, 1/7/2004; 303942, 7/19/2004; 303687, 7/19/2004; 318833, 3/24/2005; 319684, 5/5/2005
RECALLING FIRM/MANUFACTURER Hamilton, CO. Reno, NV
REASON The product valve has a high probability of deforming and may result in leakage during use.
VOLUME OF PRODUCT IN COMMERCE 171 Y valves and 30 Y valves were distributed.
DISTRIBUTION CA. MA. MI, WI, Japan and Switzerland

PRODUCT Misys Laboratory version 6.1 - Microbiology Module,
CODE All serial numbers for these models
RECALLING FIRM/MANUFACTURER Misys Healthcare Systems, Tucson, AZ
REASON Software problem when the user has indicated to enter a coded comment instead of using the defined keyboard, then enters 100,000 the system will file the number 100 instead of number 100,000.
VOLUME OF PRODUCT IN COMMERCE 67 units
DISTRIBUTION Nationwide and Internationally

PRODUCT Dual Lumen Cannula, 12Fr., sterile, disposable, Model V V 12F,
CODE Lot number: 36BC20121
RECALLING FIRM/MANUFACTURER Origen Biomedical, Inc., Austin, TX
REASON Leak between two lumens causes reduced performance.
VOLUME OF PRODUCT IN COMMERCE 170 catheters
DISTRIBUTION Nationwide, United Kingdom, Canada, and Germany

PRODUCT HoverMatt Patient Transfer Solution/Air Transfer Mattress, Model numbers: HM28DC, HM34HS, HM34DC, HM39HS, HM39DC
CODE ???
RECALLING FIRM/MANUFACTURER D. T. Davis Enterprises, Ltd., Bethlehem, PA
REASON Instructions for use are not adhering to the product.
VOLUME OF PRODUCT IN COMMERCE 866 units
DISTRIBUTION Nationwide, United Kingdom, Canada, and Australia

PRODUCT Look Brand Black Monofilament Nylon Suture. 2xS Taper Straight 61 mm 2/0 Black Monofilament Nylon 30”. Model number 933B. The product is shipped in boxes containing 12 units.
CODE Lot number: M732780
RECALLING FIRM/MANUFACTURER Surgical Specialties Corp, Reading, PA
REASON Mislabeled as double armed straight needle.
VOLUME OF PRODUCT IN COMMERCE 27 boxes
DISTRIBUTION Nationwide