FEBRUARY 2012
WEEK ENDING FEBRUARY 25
PRODUCT GE Healthcare
Signa, Discovery and
Optima. Signa HDe 1.5T; Signa Infinity TwinSpeed MR System, 1.5T Infinity
TwinSpeed, Signa Lx; Signa Infinity MR System with EXCITE Technology, Excite
II, Signa Excite 1.5T Expert; Signat 3T Infinity with Excite; Signa 3.0T with
EXCITE; Signa Excite 1.5T and Signa Excite 3T, Signa Excite HD, Signa HD 1.5T
and Signa HD 3.0T; Signa HDx; Signa HDxt; Signa Vibrant; Signa HDi; Signa
Horizon Cx aka MR/I; Discovery MR450; Discovery MR750; Optima MR450w; Signa
Cvi, and Signa Nvi. The Optima MR45Ow is a whole body magnetic resonance
scanner designed to support high resolution and high signal-to-noise ratio
images in short exam times. it is indicated for use as a diagnostic imaging
device to produce axial, sagittal, coronal, and oblique anatomical images,
spectroscopic data, parametric maps, or dynamic images of the structures or
functions of the entire body. The indication for use includes, but is not
limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints,
prostate, blood vessels, and musculoskeletal regions of the body. Depending on
the region of interest being imaged, contrast agents may be used. The images
produced by the Optima MR450w reflect the spatial distribution or molecular
environment of nuclei exhibiting magnetic resonance. These images and spectra,
when interpreted by a trained physician yield information that may assist in
diagnosis.
CODE Lot #’s: 5147788-5, 5139394, and
5139394-2; 2199859 and 2202900-6; 5148810 and 5373011; 5148810-2 and
5373011-2; 5148810-3, 5148810, and 5373011
RECALLING FIRM/MANUFACTURER
GE Healthcare, LLC, Waukesha, WI,
REASON GE Healthcare has become aware of
missing labels on the alignment lasers.
VOLUME OF PRODUCT IN COMMERCE
4614 units
PRODUCT The EnCor biopsy probe, designed for use
with EnCor and EnCor Enspire Vacuum Assisted Biopsy (VAB) driver systems. The
EnCor Biopsy Device is indicated for use in acquiring tissue for diagnosing
breast abnormalities and is supplied sterile and intended for single use.
CODE Units: ECP0110G, ECP017G, ECP0112G,
ECPMR0110G, ECP017GV, ECPMR0110GBT, ECP0112GV, ECPMR017G
RECALLING FIRM/MANUFACTURER
Bard Peripheral Vascular Inc., Tempe,
AZ,
REASON Product’s damaged packaging trays.
VOLUME OF PRODUCT IN COMMERCE
64,570 units
DISTRIBUTION Nationwide, Canada, Europe,
LAPAC, and
Australia
PRODUCT
1) Brilliance 64 Computed Tomography
System, Model #728231, 510(k) #K 033326, Serial Numbers: 95504, 91003, 300030,
95130, and 95157. The Brilliance 64 is a Computed Tomography X-Ray System
intended to produce cross-sectional images of the body by computer
reconstruction of x-ray transmission data taken at difference angles and
planes. This device may include signal analysis and display equipment,
patient, and equipment supports, components and accessories.
2) Ingenuity CT computed Tomography
System, Model #728326, 510(k) #KK 033326, Serial Numbers: 300010 and 300018.
The Ingenuity CT is a Computed Tomography X-Ray System intended to produce
cross-sectional images of the body by computer reconstruction of x-ray
transmission data taken at difference angles and planes. This device may
include signal analysis and display equipment, patient, and equipment
supports, components and accessories.
CODE
1) Serial Numbers: 95504, 91003, 300030,
95130, and 95157;
2) Serial Numbers: 300010 and 30
RECALLING FIRM/MANUFACTURER
Philips Medical Systems (Cleveland)
Inc., Cleveland, OH,
REASON Software errors. A number of
non-conformances were logged during in-house testing and external validation
testing of Brilliance 64 and Ingenuity CT software version 4.0 Beta 2. The
non-conformances are due to software errors identified in these device
systems.
VOLUME OF PRODUCT IN COMMERCE
7 units
DISTRIBUTION Nationwide, of Australia, China,
Belgium, Israel, and Sweden
PRODUCT Abbott m2000sp Instrument; an automated
system for performing sample preparation for nucleic acid testing; list 09K14.
CODE All serial numbers
RECALLING FIRM/MANUFACTURER
Abbott Molecular, Des Plaines, IL,
REASON Abbott Molecular has received reports in
which the liquid waste sensor and components in the m2000sp lower instrument
cabinet overheated and/or caught fire.
VOLUME OF PRODUCT IN COMMERCE
432 units
DISTRIBUTION Nationwide and Internationally
PRODUCT Vitrea (available in various
configuration as Vitrea Enterprise Suite). Vitrea Enterprise Suite 1.2, Vitrea
Enterprise Suite 1.3, Vitrea enterprise Suite 6.0 and Vitrea Enterprise Suite
6.1. A medical diagnostic system that allows the processing, review, analysis,
communication and media interchange of multi-dimensional digital images
acquired from variety of imaging devices.
CODE
Vital ID 10013 10079 10082 10092 10099
10104 10112 10117 10126 10127 10141 10149 10150 10158 10161 10174 10186 10199
10200 10210 10224 10225 10226 10227 10236 10250 10276 10300 10305 10306 10327
10351 10355 10359 10364 10380 10394 10395 10396 10400 10409 10412 10413 10415
10416 10418 10432 10437 10478 10487 10491 10499 10533 10539 10574 10575 10592
10607 10610 10613 10633 10635 10637 10656 10664 10665 10671 10685 10691 10697
10698 10709 10729 10732 10737 10739 10741 10748 10753 10767 10796 10798 10800
10802 10806 10827 10834 10859 10870 10878 10879 10880 10898 10921 10968 10991
11033 11088 11093 11119 11129 11136 11150 11165 11168 11178 11185 11233 11235
11252 11260 11262 11263 11269 11271 11280 11287 11289 11310 11316 11333 11342
11359 11366 11401 11404 11412 11444 11483 11513 11544 11555 11597 11613 11623
11684 11696 11776 11827 11888 11922 11941 11991 12043 12118 12122 12226 27622
27658 27674 27694 27729 27766 27774 27833 27849 27879 27889 27893 27908 27933
27958 27992 27994 28041 28076 28179 28194 28228 28236 28243 28263 28356 28405
28455 28470 28510 28516 28547 28552 28559 28567 28582 28603 28667 28668 28716
28724 28746 28808 28882 28914 28915 28917 28920 28921 28939 28956 28958 29007
29009 29010 29017 29024 29025 29026 29087 29088 29097 29154 29155 29156 29157
29158 29167 29229 29271 29281 29290 29340 29341 29351 29372 29373 29375 29376
29507 29546 29551 29557 29569 29570 29573 29574 29583 29586 29601 29603 29606
29607 29609 29610 29623 40108 40117 40131 40163 40176 40192 40284 40295 40325
40513 40565 40614 40617 40647 40706 40714 40722 40739 40771 40926 41109 41145
41194 41230 41256 41439 41461 41507 41539 41596 41674 41981 42093 42153 42521
42873 42934 43369 43503 43536 43557 43635 43669 43720 45759 45960 46274 46672
47454 47920 48344 52042 52976 53018 53025 53033 53055 53117 53175 53194 53196
53197 53443 53535 53537 53544 53672 53673 53744 53745 53748 53897 53898 54012
54022 54102 54318 54821 54822 54827 54830 54863 54945 55005 55008 55063 55302
55327 55375 55901 56015 56185 57343 28591 40391
RECALLING FIRM/MANUFACTURER
Vital Images, Inc., Plymouth, MN,
REASON Vital Images has found two issues that
may display incorrect Standard Update Values (SUV) when viewing certain PET/CT
images. This issue affects the VitreaCore PET/CT feature for Vitrea Enterprise
Suite 1.2, 1.3, 6.0, 6.1, including all updates to those versions.
VOLUME OF PRODUCT IN COMMERCE
338 (233 US)
DISTRIBUTION Nationwide and Internationally
PRODUCT Kotex Natural Balance Security tampons,
regular absorbency, unscented. packaged in 18 count or 36 count retail
containers. To absorb menstrual or other vaginal discharge.
CODE Lot numbers (found in the red box
located on bottom of package): AC129321X, AC129421X, AC129521X, AC129621X,
AC129721X, AC129821X, AC129921X 36 count: AC129325X, AC129425X, AC129525X,
AC129625X, AC129725X, AC129825X, AC129925X
RECALLING FIRM/MANUFACTURER
Recalling Firm: Kimberly-Clark Corp.,
Roswell, GA,
REASON A limited number of Kotex Natural
Balance Security Unscented Tampons, regular absorbency, are being recalled
after a raw material contaminated with a bacterium, Enterobacter sakazakii,
was found when testing the plunger portion of the applicator.
VOLUME OF PRODUCT IN COMMERCE
1,400 cases (216 tampons per case)
PRODUCT Box labeled in part, "REF:
72-1000***IQ SYSTEM IQ INTELLIGENT DRIVER 16.51CM (6.5")(L) RABALON,
STAINLESS STEEL, GOLD PLATED***NON-STERILE PRODUCT ***" The Biomet
Microfixation IQ Intelligent Driver is a handheld cordless device designed for
three functions: drill pilot holes, high-speed insertion of 1.5x4mm titanium
screws, and high-speed insertion of 1.5x4mm and 1.5x5mm Lactosorb fixation
screws provided by Biomet Microfixation. The IQ Intelligent Driver is
indicated to aid in the implantation of screws (Drilling holes and inserting
screws). Model #- 72-1000.
CODE
Serial #'s: 000007, 000009, 000010,
000011, 000012, 000013, 000017, 0G0014, 0G0018, 0G0019, 0K0001, 0K0003,
0K0006, 0K0007, 0K0008, 0K0009, 0K0011, 0K0012, 0K0013, 0K0014, 0K0015,
0K0016, 0K0017, 0K0019, 0K0020, 0K0021, 0K0022, 0L0001, 0L0002, 0L0003,
0L0004, 0L0005, 0L0006, 0L0007, 0L0008, 0L0009, 0L0010, 0L0011, 0L0014,
0L0015, 0L0016, 0L0017, 0L0018, 0L0019, 0L0020, 0L0022, 0L0023, 0L0024,
0L0026, 0L0028, 0L0029, 0L0030, 0L0031, 0L0032, 0L0033, 0L0034, 0L0035,
0L0037, 0L0038, 0L0039, 0L0041, 0L0042, 0L0044, 0L0045, 0L0046, 1A0004,
1A0005, 1A0012, 1A0014, 1A0016, 1A0019, 1A0020, 1A0021, 1A0022, 1A0024,
1A0026, 1A0028, 1A0031, 1A0033, 1A0034, 1A0036, 1A0037, 1A0041, 1A0042,
1A0043, 1A0044, 1A0046, 1A0047, 1A0048, 1A0051, 1A0052, 1A0053, 1A0055,
1A0056, 1A0057, 1A0059,1A0062, 1A0064, 1A0067, 1A0069, 1A0070, 1A0072, 1A0073,
1A0079, 1A0081, 1A0084, 1A0085, 1A0086, 1A0088, 1A0089, 1A0090, 1A0091,
1A0093, 1A0095, 1A0096, 1A0097, 1A0098, 1A0099, 1A0100, 1A0101, 1A0102,
1A0103, 1A0105, 1A0106, 1A0107, 1A0108, 1A0113, 1A0117, 1A0118, 1A0119,
1A0120, 1A0123, 1A0125, 1A0126, 1A0127, 1A0128, 1A0129, 1A0130, 1A0131,
1A0133, 1A0134, 1A0135, 1A0136, 1A0137, 1A0138, 1A0142, 1A0144, 1A0145,
1A0148, 1A0149, 1A0152, 1A0153, 1A0156, 1A0158, 1A0171, 1A0174, 1A0176,
1A0198, 1B0010, 1C0004, 1C0014, 1C0023, 1C0025, CK 0030, CK0011, CK0024,
CK0025, CK0026, CK0027, CK0028, CK0029, CK0032, CK0033
RECALLING FIRM/MANUFACTURER
Biomet Microfixation, Inc.,
Jacksonville, FL,
REASON Recall due to failure of the front bit
sensor to detect the correct bit that is installed. When a screw is inserted
with the device in drill mode, it is possible for the screw or bone to strip.
VOLUME OF PRODUCT IN COMMERCE
161 units
DISTRIBUTION Nationwide and Internationally
PRODUCT ThermoSuit. Temperature reduction in
patients where clinically indicated; Monitoring of patient temperature.
CODE Units: 510 K K061023 N0110PD N0108NI
N0068JH N0109PD N0103NE N0106NI N0065JH N0104NI N0105NI
RECALLING FIRM/MANUFACTURER
Recalling Firm: Life Recovery Systems
HD, LLC, Waldwick, NJ,
Manufacturer: Nexcore Technology, Inc.,
Waldwick, NJ.
REASON ThermoSuit System Model TSP-50
Pump/Controller may have corrosion on the temperature circuit board which may
result in incorrect patient temperature reading.
VOLUME OF PRODUCT IN COMMERCE
28 units
PRODUCT
1) Varian Unique Single Energy Linear
Accelerator, C-Series Clinac or Trilogy, versions 7.x and 8.x, Reference/FSCA
Identifier: CP-0661; Model Numbers: H14, H18, H27, H29; Varian Unique Single
Energy Linear Accelerator: The UNIQUE is a Single Energy Linear Accelerator
intended to be used for conventional radiotherapy and includes modifications
to the previously cleared Varian Trilogy. The UNIQUE provides additional
features, safety improvements, and usability improvements. The UNIQUE is
intended to provide stereotactic radiosurgery and precision radiotherapy for
lesions, tumors and conditions anywhere in the body when radiation treatment
is indicated.
2) Varian High Energy Linear
Accelerator, C-Series Clinac, Reference/FSCA Identifier: CP-06611; Models
Numbers: 600C, 600CD, 6EX, DBX, 2100C, 2100CD, 2300CD, 21EX, 23EX, DMX, DHX
versions 2.x through 6.x. The Varian High Energy Linear Accelerator is
intended to provide stereotactic radiosurgery and precision radiotherapy for
lesions, tumors and conditions anywhere in the body when radiation treatment
is indicated.
CODE
1) H140001 ,H140145 ,H140711 ,H140772
,H140832 ,H140893 ,H140958 ,H271265 ,H271742 ,H272014 ,H272235 ,H272416
,H272858 ,H140011 ,H140146 ,H140712 ,H140773 ,H140833 ,H140895 ,H140959
,H271277 ,H271755 ,H272016 ,H272236 ,H272418 ,H272859 ,H140018 ,H140147
,H140713 ,H140774 ,H140834 ,H140896 ,H140962 ,H271293 ,H271760 ,H272023
,H272237 ,H272421 ,H272860 ,H140025 ,H140148 ,H140714 ,H140775 ,H140835
,H140897 ,H140963 ,H271308 ,H271762 ,H272029 ,H272239 ,H272422 ,H272862
,H140026 ,H140150 ,H140715 ,H140776 ,H140836 ,H140898 ,H180455 ,H271334
,H271763 ,H272031 ,H272240 ,H272423 ,H272864 ,H140027 ,H140151 ,H140716
,H140777 ,H140837 ,H140899 ,H182000 ,H271362 ,H271773 ,H272033 ,H272243
,H272425 ,H272866 ,H140029 ,H140152 ,H140717 ,H140778 ,H140838 ,H140900
,H182001 ,H271370 ,H271780 ,H272035 ,H272250 ,H272426 ,H272868 ,H140031
,H140153 ,H140718 ,H140779 ,H140839 ,H140901 ,H182002 ,H271378 ,H271783
,H272043 ,H272259 ,H272427 ,H272872 ,H140038 ,H140154 ,H140719 ,H140780
,H140840 ,H140902 ,H182003 ,H271385 ,H271785 ,H272045 ,H272260 ,H272428
,H272873 ,H140039 ,H140201 ,H140720 ,H140781 ,H140841 ,H140903 ,H182005
,H271393 ,H271797 ,H272049 ,H272271 ,H272432 ,H272876 ,H140043 ,H140202
,H140721 ,H140782 ,H140842 ,H140904 ,H182006 ,H271407 ,H271801 ,H272050
,H272277 ,H272434 ,H272885 ,H140045 ,H140203 ,H140722 ,H140783 ,H140843
,H140905 ,H182007 ,H271413 ,H271802 ,H272052 ,H272283 ,H272438 ,H272889
,H140046 ,H140204 ,H140723 ,H140784 ,H140844 ,H140906 ,H182008 ,H271414
,H271811 ,H272062 ,H272290 ,H272440 ,H272890 ,H140047 ,H140205 ,H140724
,H140785 ,H140845 ,H140907 ,H182009 ,H271420 ,H271825 ,H272064 ,H272291
,H272444 ,H272893 ,H140048 ,H140206 ,H140725 ,H140786 ,H140846 ,H140908
,H182010 ,H271425 ,H271831 ,H272069 ,H272293 ,H272445 ,H272899 ,H140058
,H140207 ,H140726 ,H140787 ,H140847 ,H140909 ,H182011 ,H271429 ,H271833
,H272078 ,H272299 ,H272446 ,H272905 ,H140062 ,H140208 ,H140727 ,H140788
,H140848 ,H140910 ,H182012 ,H271438 ,H271835 ,H272081 ,H272301 ,H272454
,H272908 ,H140067 ,H140209 ,H140728 ,H140789 ,H140849 ,H140911 ,H182013
,H271439 ,H271838 ,H272085 ,H272302 ,H272456 ,H272909 ,H140070 ,H140210
,H140729 ,H140790 ,H140850 ,H140912 ,H182014 ,H271440 ,H271845 ,H272086
,H272306 ,H272458 ,H272910 ,H140071 ,H140211 ,H140730 ,H140791 ,H140851
,H140913 ,H182016 ,H271443 ,H271853 ,H272089 ,H272307 ,H272460 ,H272911
,H140073 ,H140212 ,H140731 ,H140792 ,H140852 ,H140914 ,H182017 ,H271452
,H271856 ,H272111 ,H272309 ,H272462 ,H272912 ,H140076 ,H140213 ,H140732
,H140793 ,H140853 ,H140915 ,H182018 ,H271490 ,H271858 ,H272115 ,H272310
,H272464 ,H272914 ,H140077 ,H140214 ,H140733 ,H140794 ,H140854 ,H140916
,H182019 ,H271497 ,H271860 ,H272118 ,H272311 ,H272466 ,H272922 ,H140078
,H140215 ,H140734 ,H140795 ,H140855 ,H140917 ,H182020 ,H271502 ,H271862
,H272129 ,H272313 ,H272469 ,H272926 ,H140079 ,H140216 ,H140735 ,H140796
,H140856 ,H140918 ,H182021 ,H271507 ,H271866 ,H272134 ,H272318 ,H272470
,H272932 ,H140088 ,H140217 ,H140736 ,H140797 ,H140857 ,H140919 ,H182022
,H271515 ,H271870 ,H272136 ,H272321 ,H272473 ,H272933 ,H140090 ,H140218
,H140737 ,H140798 ,H140858 ,H140920 ,H182023 ,H271519 ,H271872 ,H272137
,H272324 ,H272476 ,H272936 ,H140092 ,H140219 ,H140738 ,H140799 ,H140859
,H140921 ,H182030 ,H271523 ,H271875 ,H272138 ,H272325 ,H272479 ,H272938
,H140096 ,H140220 ,H140739 ,H140800 ,H140860 ,H140922 ,H182032 ,H271526
,H271877 ,H272143 ,H272326 ,H272480 ,H272941 ,H140097 ,H140221 ,H140740
,H140801 ,H140861 ,H140923 ,H260905 ,H271531 ,H271879 ,H272153 ,H272327
,H272488 ,H272946 ,H140098 ,H140222 ,H140742 ,H140802 ,H140862 ,H140924
,H260909 ,H271534 ,H271881 ,H272155 ,H272328 ,H272491 ,H272951 ,H140101
,H140223 ,H140743 ,H140803 ,H140863 ,H140925 ,H260922 ,H271559 ,H271890
,H272156 ,H272329 ,H272492 ,H272953 ,H140107 ,H140224 ,H140744 ,H140804
,H140864 ,H140926 ,H260924 ,H271582 ,H271893 ,H272157 ,H272333 ,H272494
,H272956 ,H140109 ,H140225 ,H140745 ,H140805 ,H140865 ,H140927 ,H260925
,H271587 ,H271895 ,H272158 ,H272336 ,H272497 ,H272960 ,H140111 ,H140226
,H140746 ,H140806 ,H140866 ,H140929 ,H260928 ,H271588 ,H271896 ,H272159
,H272339 ,H272500 ,H272961 ,H140113 ,H140227 ,H140747 ,H140807 ,H140867
,H140930 ,H260931 ,H271593 ,H271903 ,H272162 ,H272341 ,H272801 ,H272963
,H140114 ,H140228 ,H140748 ,H140808 ,H140868 ,H140931 ,H260936 ,H271594
,H271905 ,H272163 ,H272343 ,H272803 ,H272964 ,H140115 ,H140229 ,H140749
,H140809 ,H140869 ,H140932 ,H260940 ,H271602 ,H271906 ,H272168 ,H272345
,H272804 ,H272968 ,H140116 ,H140230 ,H140750 ,H140810 ,H140870 ,H140933
,H260942 ,H271607 ,H271918 ,H272169 ,H272346 ,H272805 ,H272971 ,H140117
,H140231 ,H140751 ,H140811 ,H140871 ,H140934 ,H260946 ,H271618 ,H271921
,H272172 ,H272349 ,H272806 ,H272973 ,H140118 ,H140232 ,H140752 ,H140812
,H140872 ,H140935 ,H260949 ,H271624 ,H271926 ,H272173 ,H272354 ,H272811
,H272978 ,H140119 ,H140233 ,H140753 ,H140813 ,H140873 ,H140936 ,H260950
,H271626 ,H271931 ,H272174 ,H272355 ,H272816 ,H272981 ,H140121 ,H140234
,H140754 ,H140814 ,H140874 ,H140937 ,H260951 ,H271628 ,H271935 ,H272178
,H272359 ,H272820 ,H272982 ,H140122 ,H140235 ,H140755 ,H140815 ,H140875
,H140938 ,H260952 ,H271635 ,H271941 ,H272179 ,H272360 ,H272822 ,H272985
,H140124 ,H140236 ,H140756 ,H140816 ,H140876 ,H140939 ,H260953 ,H271640
,H271942 ,H272180 ,H272362 ,H272824 ,H272986 ,H140125 ,H140237 ,H140757
,H140817 ,H140877 ,H140940 ,H270767 ,H271662 ,H271951 ,H272184 ,H272365
,H272825 ,H272987 ,H140129 ,H140238 ,H140758 ,H140818 ,H140878 ,H140941
,H270788 ,H271663 ,H271954 ,H272187 ,H272370 ,H272826 ,H272990 ,H140132
,H140239 ,H140759 ,H140819 ,H140879 ,H140942 ,H270794 ,H271664 ,H271961
,H272188 ,H272372 ,H272827 ,H272993 ,H140133 ,H140240 ,H140760 ,H140820
,H140881 ,H140943 ,H271041 ,H271669 ,H271964 ,H272195 ,H272377 ,H272828
,H272996 ,H140134 ,H140700 ,H140761 ,H140821 ,H140882 ,H140944 ,H271066
,H271678 ,H271972 ,H272197 ,H272380 ,H272834 ,H272999 ,H140135 ,H140701
,H140762 ,H140822 ,H140883 ,H140945 ,H271105 ,H271681 ,H271973 ,H272202
,H272381 ,H272837 ,H273003 ,H140136 ,H140702 ,H140763 ,H140823 ,H140884
,H140947 ,H271157 ,H271691 ,H271974 ,H272203 ,H272388 ,H272841 ,H273004
,H140137 ,H140703 ,H140764 ,H140824 ,H140885 ,H140948 ,H271158 ,H271693
,H271975 ,H272204 ,H272394 ,H272842 ,H273006 ,H140138 ,H140704 ,H140765
,H140825 ,H140886 ,H140949 ,H271159 ,H271711 ,H271976 ,H272209 ,H272397
,H272846 ,H273011 ,H140139 ,H140705 ,H140766 ,H140826 ,H140887 ,H140950
,H271167 ,H271712 ,H271993 ,H272216 ,H272401 ,H272847 ,H273016 ,H140140
,H140706 ,H140767 ,H140827 ,H140888 ,H140951 ,H271194 ,H271713 ,H271994
,H272219 ,H272405 ,H272848 ,H273026 ,H140141 ,H140707 ,H140768 ,H140828
,H140889 ,H140952 ,H271214 ,H271714 ,H272000 ,H272220 ,H272407 ,H272849
,H273027 ,H140142 ,H140708 ,H140769 ,H140829 ,H140890 ,H140953 ,H271227
,H271720 ,H272001 ,H272223 ,H272410 ,H272854 ,H273029 ,H140143 ,H140709
,H140770 ,H140830 ,H140891 ,H140956 ,H271242 ,H271724 ,H272003 ,H272226
,H272412 ,H272856 ,H273030 ,H140144 ,H140710 ,H140771 ,H140831 ,H140892
,H140957 ,H271258 ,H271729 ,H272009 ,H272231 ,H272413 ,H272857 ,H273031
,H273032 ,H273133 ,H273197 ,H290501 ,H291030 ,H291094 ,H291158 ,H291220
,H293306 ,H293369 ,H293432 ,H293494 ,H293558 ,H273035 ,H273134 ,H273198
,H290503 ,H291031 ,H291095 ,H291159 ,H291221 ,H293307 ,H293370 ,H293433
,H293495 ,H293559 ,H273036 ,H273135 ,H273199 ,H290504 ,H291032 ,H291096
,H291160 ,H291222 ,H293308 ,H293371 ,H293434 ,H293496 ,H293560 ,H273039
,H273136 ,H273200 ,H290505 ,H291034 ,H291097 ,H291161 ,H291223 ,H293309
,H293372 ,H293435 ,H293497 ,H293561 ,H273040 ,H273137 ,H273201 ,H290508
,H291035 ,H291098 ,H291162 ,H291224 ,H293310 ,H293373 ,H293436 ,H293498
,H293562 ,H273041 ,H273138 ,H273202 ,H290509 ,H291036 ,H291099 ,H291163
,H291225 ,H293311 ,H293374 ,H293437 ,H293499 ,H293563 ,H273042 ,H273139
,H273203 ,H290510 ,H291037 ,H291101 ,H291164 ,H291226 ,H293312 ,H293375
,H293438 ,H293500 ,H293564 ,H273048 ,H273140 ,H273204 ,H290511 ,H291038
,H291102 ,H291165 ,H291227 ,H293313 ,H293376 ,H293439 ,H293501 ,H293565
,H273049 ,H273141 ,H273205 ,H290512 ,H291039 ,H291103 ,H291166 ,H291228
,H293314 ,H293377 ,H293440 ,H293502 ,H293566 ,H273052 ,H273142 ,H273206
,H290517 ,H291040 ,H291104 ,H291167 ,H291229 ,H293315 ,H293378 ,H293441
,H293503 ,H293567 ,H273053 ,H273143 ,H273208 ,H290519 ,H291041 ,H291105
,H291168 ,H291230 ,H293316 ,H293379 ,H293442 ,H293504 ,H293568 ,H273055
,H273144 ,H273209 ,H290520 ,H291042 ,H291106 ,H291169 ,H291231 ,H293317
,H293380 ,H293443 ,H293505 ,H293569 ,H273056 ,H273145 ,H273210 ,H290530
,H291043 ,H291107 ,H291170 ,H291232 ,H293318 ,H293381 ,H293444 ,H293506
,H293570 ,H273057 ,H273146 ,H273547 ,H290533 ,H291044 ,H291108 ,H291171
,H291233 ,H293319 ,H293382 ,H293445 ,H293507 ,H293571 ,H273063 ,H273147
,H273711 ,H290538 ,H291045 ,H291109 ,H291172 ,H291234 ,H293320 ,H293383
,H293446 ,H293508 ,H293572 ,H273064 ,H273148 ,H273929 ,H290540 ,H291046
,H291111 ,H291173 ,H291235 ,H293321 ,H293384 ,H293447 ,H293509 ,H293573
,H273065 ,H273149 ,H290186 ,H290541 ,H291047 ,H291112 ,H291174 ,H291236
,H293323 ,H293385 ,H293448 ,H293510 ,H293574 ,H273066 ,H273150 ,H290192
,H290547 ,H291048 ,H291113 ,H291175 ,H291237 ,H293324 ,H293386 ,H293449
,H293511 ,H293575 ,H273072 ,H273151 ,H290201 ,H290548 ,H291049 ,H291114
,H291176 ,H291238 ,H293325 ,H293387 ,H293450 ,H293512 ,H293576 ,H273075
,H273152 ,H290208 ,H290549 ,H291050 ,H291115 ,H291177 ,H291239 ,H293326
,H293388 ,H293451 ,H293513 ,H293577 ,H273077 ,H273153 ,H290222 ,H290550
,H291051 ,H291116 ,H291178 ,H291240 ,H293327 ,H293389 ,H293452 ,H293514
,H293578 ,H273080 ,H273155 ,H290224 ,H290551 ,H291052 ,H291117 ,H291179
,H291241 ,H293328 ,H293390 ,H293453 ,H293515 ,H293579 ,H273082 ,H273156
,H290233 ,H290552 ,H291053 ,H291118 ,H291180 ,H291242 ,H293329 ,H293391
,H293454 ,H293516 ,H293581 ,H273083 ,H273157 ,H290241 ,H290553 ,H291054
,H291119 ,H291181 ,H291243 ,H293330 ,H293392 ,H293455 ,H293517 ,H293583
,H273084 ,H273158 ,H290243 ,H290554 ,H291055 ,H291120 ,H291182 ,H291244
,H293331 ,H293393 ,H293456 ,H293518 ,H293584 ,H273086 ,H273159 ,H290248
,H290555 ,H291056 ,H291121 ,H291183 ,H291245 ,H293332 ,H293394 ,H293457
,H293519 ,H293585 ,H273088 ,H273160 ,H290249 ,H290556 ,H291057 ,H291122
,H291184 ,H291246 ,H293333 ,H293395 ,H293458 ,H293520 ,H293586 ,H273089
,H273162 ,H290256 ,H290557 ,H291059 ,H291123 ,H291185 ,H291247 ,H293334
,H293396 ,H293459 ,H293521 ,H293587 ,H273092 ,H273163 ,H290262 ,H290558
,H291060 ,H291124 ,H291186 ,H291248 ,H293335 ,H293397 ,H293460 ,H293522
,H293588 ,H273093 ,H273164 ,H290272 ,H290559 ,H291061 ,H291125 ,H291187
,H291249 ,H293336 ,H293398 ,H293461 ,H293523 ,H293589 ,H273100 ,H273165
,H290274 ,H290560 ,H291062 ,H291126 ,H291188 ,H291250 ,H293337 ,H293399
,H293462 ,H293524 ,H293590 ,H273101 ,H273166 ,H290278 ,H290561 ,H291063
,H291127 ,H291189 ,H291251 ,H293338 ,H293401 ,H293463 ,H293525 ,H293591
,H273102 ,H273167 ,H290282 ,H290600 ,H291064 ,H291128 ,H291190 ,H291252
,H293339 ,H293402 ,H293464 ,H293527 ,H293592 ,H273103 ,H273168 ,H290287
,H291001 ,H291065 ,H291129 ,H291191 ,H291253 ,H293340 ,H293403 ,H293465
,H293528 ,H293593 ,H273104 ,H273169 ,H290295 ,H291002 ,H291066 ,H291130
,H291192 ,H291254 ,H293341 ,H293404 ,H293466 ,H293529 ,H293594 ,H273105
,H273170 ,H290304 ,H291003 ,H291067 ,H291131 ,H291193 ,H291255 ,H293342
,H293405 ,H293467 ,H293530 ,H293595 ,H273106 ,H273171 ,H290308 ,H291004
,H291068 ,H291132 ,H291194 ,H291256 ,H293343 ,H293406 ,H293468 ,H293531
,H293596 ,H273107 ,H273172 ,H290313 ,H291005 ,H291069 ,H291133 ,H291195
,H291257 ,H293344 ,H293407 ,H293469 ,H293532 ,H293597 ,H273108 ,H273173
,H290317 ,H291006 ,H291070 ,H291134 ,H291196 ,H291258 ,H293345 ,H293408
,H293470 ,H293533 ,H293598 ,H273109 ,H273174 ,H290321 ,H291007 ,H291071
,H291135 ,H291197 ,H291259 ,H293346 ,H293409 ,H293471 ,H293534 ,H293599
,H273110 ,H273175 ,H290326 ,H291008 ,H291072 ,H291136 ,H291198 ,H291260
,H293347 ,H293410 ,H293472 ,H293535 ,H293600 ,H273111 ,H273176 ,H290341
,H291009 ,H291073 ,H291137 ,H291199 ,H291261 ,H293348 ,H293411 ,H293473
,H293536 ,H293601 ,H273112 ,H273177 ,H290352 ,H291010 ,H291074 ,H291138
,H291200 ,H291262 ,H293349 ,H293412 ,H293474 ,H293537 ,H293602 ,H273113
,H273178 ,H290358 ,H291011 ,H291075 ,H291139 ,H291201 ,H291263 ,H293350
,H293413 ,H293475 ,H293538 ,H293603 ,H273114 ,H273179 ,H290362 ,H291012
,H291076 ,H291140 ,H291202 ,H291264 ,H293351 ,H293414 ,H293476 ,H293539
,H293604 ,H273115 ,H273180 ,H290371 ,H291013 ,H291077 ,H291141 ,H291203
,H291265 ,H293352 ,H293415 ,H293477 ,H293540 ,H293605 ,H273116 ,H273181
,H290372 ,H291014 ,H291078 ,H291142 ,H291204 ,H291266 ,H293353 ,H293416
,H293478 ,H293541 ,H293606 ,H273117 ,H273182 ,H290376 ,H291015 ,H291079
,H291143 ,H291205 ,H291267 ,H293354 ,H293417 ,H293479 ,H293542 ,H293607
,H273118 ,H273183 ,H290379 ,H291016 ,H291080 ,H291144 ,H291206 ,H291268
,H293355 ,H293418 ,H293480 ,H293543 ,H293608 ,H273119 ,H273184 ,H290381
,H291017 ,H291081 ,H291145 ,H291207 ,H291269 ,H293356 ,H293419 ,H293481
,H293544 ,H293609 ,H273120 ,H273185 ,H290386 ,H291018 ,H291082 ,H291146
,H291208 ,H291270 ,H293357 ,H293420 ,H293482 ,H293545 ,H293610 ,H273121
,H273186 ,H290392 ,H291019 ,H291083 ,H291147 ,H291209 ,H291271 ,H293358
,H293421 ,H293483 ,H293546 ,H293611 ,H273122 ,H273187 ,H290397 ,H291020
,H291084 ,H291148 ,H291210 ,H291272 ,H293359 ,H293422 ,H293484 ,H293548
,H293612 ,H273123 ,H273188 ,H290399 ,H291021 ,H291085 ,H291149 ,H291211
,H291273 ,H293360 ,H293423 ,H293485 ,H293549 ,H293613 ,H273124 ,H273189
,H290400 ,H291022 ,H291086 ,H291150 ,H291212 ,H293009 ,H293361 ,H293424
,H293486 ,H293550 ,H293614 ,H273125 ,H273190 ,H290401 ,H291023 ,H291087
,H291151 ,H291213 ,H293207 ,H293362 ,H293425 ,H293487 ,H293551 ,H293615
,H273126 ,H273191 ,H290402 ,H291024 ,H291088 ,H291152 ,H291214 ,H293300
,H293363 ,H293426 ,H293488 ,H293552 ,H293616 ,H273127 ,H273192 ,H290403
,H291025 ,H291089 ,H291153 ,H291215 ,H293301 ,H293364 ,H293427 ,H293489
,H293553 ,H293617 ,H273128 ,H273193 ,H290405 ,H291026 ,H291090 ,H291154
,H291216 ,H293302 ,H293365 ,H293428 ,H293490 ,H293554 ,H293618 ,H273130
,H273194 ,H290414 ,H291027 ,H291091 ,H291155 ,H291217 ,H293303 ,H293366
,H293429 ,H293491 ,H293555 ,H293619 ,H273131 ,H273195 ,H290422 ,H291028
,H291092 ,H291156 ,H291218 ,H293304 ,H293367 ,H293430 ,H293492 ,H293556
,H293620 ,H273132 ,H273196 ,H290423 ,H291029 ,H291093 ,H291157 ,H291219
,H293305 ,H293368 ,H293431 ,H293493 ,H293557 ,H293621 ,H293622 ,H293686
,H293749 ,H293812 ,H293875 ,H293938 ,H294002 ,H294065 ,H294129 ,H294192
,H294254 ,H294317 ,H294381 ,H293623 ,H293687 ,H293750 ,H293813 ,H293876
,H293940 ,H294003 ,H294066 ,H294130 ,H294193 ,H294255 ,H294318 ,H294382
,H293624 ,H293688 ,H293751 ,H293814 ,H293877 ,H293941 ,H294004 ,H294067
,H294131 ,H294194 ,H294256 ,H294319 ,H294383 ,H293625 ,H293689 ,H293752
,H293815 ,H293878 ,H293942 ,H294005 ,H294068 ,H294132 ,H294195 ,H294257
,H294320 ,H294384 ,H293626 ,H293690 ,H293753 ,H293816 ,H293879 ,H293943
,H294006 ,H294069 ,H294133 ,H294196 ,H294258 ,H294321 ,H294385 ,H293627
,H293691 ,H293754 ,H293817 ,H293880 ,H293944 ,H294007 ,H294070 ,H294134
,H294197 ,H294259 ,H294322 ,H294386 ,H293628 ,H293692 ,H293755 ,H293818
,H293881 ,H293945 ,H294008 ,H294071 ,H294135 ,H294198 ,H294260 ,H294323
,H294387 ,H293629 ,H293693 ,H293756 ,H293819 ,H293882 ,H293946 ,H294009
,H294072 ,H294136 ,H294199 ,H294261 ,H294324 ,H294388 ,H293630 ,H293694
,H293758 ,H293820 ,H293883 ,H293947 ,H294010 ,H294073 ,H294137 ,H294200
,H294262 ,H294325 ,H294389 ,H293631 ,H293695 ,H293759 ,H293821 ,H293884
,H293948 ,H294011 ,H294074 ,H294139 ,H294201 ,H294263 ,H294326 ,H294390
,H293632 ,H293696 ,H293760 ,H293822 ,H293885 ,H293949 ,H294012 ,H294075
,H294140 ,H294202 ,H294264 ,H294328 ,H294391 ,H293633 ,H293697 ,H293761
,H293823 ,H293886 ,H293950 ,H294013 ,H294076 ,H294141 ,H294203 ,H294265
,H294329 ,H294392 ,H293634 ,H293698 ,H293762 ,H293824 ,H293887 ,H293951
,H294014 ,H294077 ,H294142 ,H294204 ,H294266 ,H294330 ,H294393 ,H293635
,H293699 ,H293763 ,H293825 ,H293888 ,H293952 ,H294015 ,H294078 ,H294143
,H294205 ,H294267 ,H294331 ,H294394 ,H293636 ,H293700 ,H293764 ,H293826
,H293889 ,H293953 ,H294016 ,H294079 ,H294144 ,H294206 ,H294268 ,H294332
,H294395 ,H293637 ,H293701 ,H293765 ,H293827 ,H293890 ,H293954 ,H294017
,H294080 ,H294145 ,H294207 ,H294269 ,H294333 ,H294396 ,H293638 ,H293702
,H293766 ,H293828 ,H293891 ,H293955 ,H294018 ,H294081 ,H294146 ,H294208
,H294270 ,H294334 ,H294397 ,H293639 ,H293703 ,H293767 ,H293829 ,H293892
,H293956 ,H294020 ,H294083 ,H294147 ,H294209 ,H294271 ,H294335 ,H294398
,H293640 ,H293704 ,H293768 ,H293830 ,H293893 ,H293957 ,H294021 ,H294084
,H294148 ,H294210 ,H294272 ,H294336 ,H294399 ,H293641 ,H293705 ,H293769
,H293831 ,H293894 ,H293958 ,H294022 ,H294085 ,H294149 ,H294211 ,H294273
,H294337 ,H294400 ,H293642 ,H293706 ,H293770 ,H293832 ,H293895 ,H293959
,H294023 ,H294086 ,H294150 ,H294212 ,H294274 ,H294338 ,H294401 ,H293643
,H293707 ,H293771 ,H293833 ,H293896 ,H293960 ,H294024 ,H294087 ,H294151
,H294213 ,H294275 ,H294339 ,H294402 ,H293644 ,H293708 ,H293772 ,H293834
,H293897 ,H293961 ,H294025 ,H294088 ,H294152 ,H294214 ,H294276 ,H294340
,H294403 ,H293645 ,H293709 ,H293773 ,H293835 ,H293898 ,H293962 ,H294026
,H294089 ,H294153 ,H294215 ,H294277 ,H294341 ,H294404 ,H293646 ,H293710
,H293774 ,H293836 ,H293899 ,H293963 ,H294027 ,H294090 ,H294154 ,H294216
,H294278 ,H294342 ,H294405 ,H293648 ,H293712 ,H293775 ,H293837 ,H293900
,H293964 ,H294028 ,H294091 ,H294155 ,H294217 ,H294279 ,H294343 ,H294406
,H293649 ,H293713 ,H293776 ,H293838 ,H293901 ,H293965 ,H294029 ,H294092
,H294156 ,H294218 ,H294280 ,H294345 ,H294407 ,H293650 ,H293714 ,H293777
,H293839 ,H293902 ,H293966 ,H294030 ,H294093 ,H294157 ,H294219 ,H294281
,H294346 ,H294408 ,H293651 ,H293715 ,H293778 ,H293840 ,H293903 ,H293967
,H294031 ,H294094 ,H294158 ,H294220 ,H294283 ,H294347 ,H294409 ,H293652
,H293716 ,H293779 ,H293841 ,H293904 ,H293968 ,H294032 ,H294095 ,H294159
,H294221 ,H294284 ,H294348 ,H294410 ,H293653 ,H293717 ,H293780 ,H293842
,H293905 ,H293969 ,H294033 ,H294097 ,H294160 ,H294222 ,H294285 ,H294349
,H294411 ,H293654 ,H293718 ,H293781 ,H293843 ,H293906 ,H293970 ,H294034
,H294098 ,H294161 ,H294223 ,H294286 ,H294350 ,H294412 ,H293655 ,H293719
,H293782 ,H293844 ,H293907 ,H293971 ,H294035 ,H294099 ,H294162 ,H294224
,H294287 ,H294351 ,H294413 ,H293656 ,H293720 ,H293783 ,H293845 ,H293908
,H293972 ,H294036 ,H294100 ,H294163 ,H294225 ,H294288 ,H294352 ,H294414
,H293657 ,H293721 ,H293784 ,H293846 ,H293909 ,H293974 ,H294037 ,H294101
,H294164 ,H294226 ,H294289 ,H294353 ,H294415 ,H293658 ,H293722 ,H293785
,H293847 ,H293910 ,H293975 ,H294038 ,H294102 ,H294165 ,H294227 ,H294290
,H294354 ,H294416 ,H293659 ,H293723 ,H293786 ,H293848 ,H293911 ,H293976
,H294039 ,H294103 ,H294166 ,H294228 ,H294291 ,H294355 ,H294417 ,H293660
,H293724 ,H293787 ,H293849 ,H293912 ,H293977 ,H294040 ,H294104 ,H294167
,H294229 ,H294292 ,H294356 ,H294418 ,H293661 ,H293725 ,H293788 ,H293850
,H293913 ,H293978 ,H294041 ,H294105 ,H294168 ,H294230 ,H294293 ,H294357
,H294419 ,H293662 ,H293726 ,H293789 ,H293851 ,H293914 ,H293979 ,H294042
,H294106 ,H294169 ,H294231 ,H294294 ,H294358 ,H294420 ,H293663 ,H293727
,H293790 ,H293852 ,H293915 ,H293980 ,H294043 ,H294107 ,H294170 ,H294232
,H294295 ,H294359 ,H294421 ,H293664 ,H293728 ,H293791 ,H293854 ,H293916
,H293981 ,H294044 ,H294108 ,H294171 ,H294233 ,H294296 ,H294360 ,H294422
,H293665 ,H293729 ,H293792 ,H293855 ,H293917 ,H293982 ,H294045 ,H294109
,H294172 ,H294234 ,H294297 ,H294361 ,H294423 ,H293666 ,H293730 ,H293793
,H293856 ,H293918 ,H293983 ,H294046 ,H294110 ,H294173 ,H294235 ,H294298
,H294362 ,H294424 ,H293667 ,H293731 ,H293794 ,H293857 ,H293919 ,H293984
,H294047 ,H294111 ,H294174 ,H294236 ,H294299 ,H294363 ,H294425 ,H293668
,H293732 ,H293795 ,H293858 ,H293920 ,H293985 ,H294048 ,H294112 ,H294175
,H294237 ,H294300 ,H294364 ,H294426 ,H293669 ,H293733 ,H293796 ,H293859
,H293921 ,H293986 ,H294049 ,H294113 ,H294176 ,H294238 ,H294301 ,H294365
,H294427 ,H293670 ,H293734 ,H293797 ,H293860 ,H293922 ,H293987 ,H294050
,H294114 ,H294177 ,H294239 ,H294302 ,H294366 ,H294428 ,H293671 ,H293735
,H293798 ,H293861 ,H293923 ,H293988 ,H294051 ,H294115 ,H294178 ,H294240
,H294303 ,H294367 ,H294429 ,H293672 ,H293736 ,H293799 ,H293862 ,H293924
,H293989 ,H294052 ,H294116 ,H294179 ,H294241 ,H294304 ,H294368 ,H294430
,H293673 ,H293737 ,H293800 ,H293863 ,H293925 ,H293990 ,H294053 ,H294117
,H294180 ,H294242 ,H294305 ,H294369 ,H294431 ,H293674 ,H293738 ,H293801
,H293864 ,H293926 ,H293991 ,H294054 ,H294118 ,H294181 ,H294243 ,H294306
,H294370 ,H294432 ,H293675 ,H293739 ,H293802 ,H293865 ,H293927 ,H293992
,H294055 ,H294119 ,H294182 ,H294244 ,H294307 ,H294371 ,H294433 ,H293676
,H293740 ,H293803 ,H293866 ,H293928 ,H293993 ,H294056 ,H294120 ,H294183
,H294245 ,H294308 ,H294372 ,H294434 ,H293677 ,H293741 ,H293804 ,H293867
,H293929 ,H293994 ,H294057 ,H294121 ,H294184 ,H294246 ,H294309 ,H294373
,H294435 ,H293678 ,H293742 ,H293805 ,H293868 ,H293930 ,H293995 ,H294058
,H294122 ,H294185 ,H294247 ,H294310 ,H294374 ,H294436 ,H293679 ,H293743
,H293806 ,H293869 ,H293931 ,H293996 ,H294059 ,H294123 ,H294186 ,H294248
,H294311 ,H294375 ,H294437 ,H293681 ,H293744 ,H293807 ,H293870 ,H293933
,H293997 ,H294060 ,H294124 ,H294187 ,H294249 ,H294312 ,H294376 ,H294438
,H293682 ,H293745 ,H293808 ,H293871 ,H293934 ,H293998 ,H294061 ,H294125
,H294188 ,H294250 ,H294313 ,H294377 ,H294439 ,H293683 ,H293746 ,H293809
,H293872 ,H293935 ,H293999 ,H294062 ,H294126 ,H294189 ,H294251 ,H294314
,H294378 ,H294440 ,H293684 ,H293747 ,H293810 ,H293873 ,H293936 ,H294000
,H294063 ,H294127 ,H294190 ,H294252 ,H294315 ,H294379 ,H294442 ,H293685
,H293748 ,H293811 ,H293874 ,H293937 ,H294001 ,H294064 ,H294128 ,H294191
,H294253 ,H294316 ,H294380 ,H294443 ,H294444 ,H294508 ,H294577 ,H294640
,H294702 ,H294768 ,H294832 ,H294897 ,H294963 ,H295030 ,H295097 ,H295170
,H295275 ,H294445 ,H294509 ,H294578 ,H294641 ,H294704 ,H294769 ,H294833
,H294898 ,H294965 ,H295031 ,H295099 ,H295171 ,H295276 ,H294446 ,H294510
,H294579 ,H294642 ,H294705 ,H294770 ,H294834 ,H294899 ,H294966 ,H295032
,H295100 ,H295172 ,H295277 ,H294447 ,H294511 ,H294580 ,H294643 ,H294706
,H294771 ,H294835 ,H294900 ,H294967 ,H295033 ,H295101 ,H295173 ,H295279
,H294448 ,H294512 ,H294581 ,H294644 ,H294707 ,H294772 ,H294836 ,H294901
,H294969 ,H295034 ,H295102 ,H295174 ,H295281 ,H294449 ,H294513 ,H294582
,H294645 ,H294708 ,H294773 ,H294837 ,H294902 ,H294970 ,H295035 ,H295103
,H295176 ,H295282 ,H294450 ,H294514 ,H294583 ,H294646 ,H294709 ,H294774
,H294838 ,H294903 ,H294971 ,H295036 ,H295105 ,H295177 ,H295284 ,H294451
,H294515 ,H294584 ,H294647 ,H294710 ,H294775 ,H294839 ,H294904 ,H294972
,H295037 ,H295106 ,H295178 ,H295287 ,H294452 ,H294516 ,H294585 ,H294648
,H294711 ,H294776 ,H294840 ,H294905 ,H294974 ,H295038 ,H295107 ,H295179
,H295290 ,H294454 ,H294517 ,H294586 ,H294649 ,H294712 ,H294777 ,H294841
,H294906 ,H294975 ,H295039 ,H295108 ,H295180 ,H295291 ,H294455 ,H294518
,H294587 ,H294650 ,H294713 ,H294778 ,H294842 ,H294907 ,H294976 ,H295040
,H295109 ,H295181 ,H295299 ,H294457 ,H294519 ,H294588 ,H294651 ,H294714
,H294779 ,H294843 ,H294908 ,H294977 ,H295041 ,H295110 ,H295182 ,H295300
,H294458 ,H294520 ,H294589 ,H294652 ,H294715 ,H294780 ,H294844 ,H294910
,H294978 ,H295042 ,H295111 ,H295183 ,H295301 ,H294459 ,H294521 ,H294590
,H294653 ,H294716 ,H294781 ,H294845 ,H294911 ,H294979 ,H295043 ,H295112
,H295184 ,H294460 ,H294522 ,H294591 ,H294654 ,H294717 ,H294782 ,H294846
,H294912 ,H294980 ,H295044 ,H295114 ,H295185 ,H294461 ,H294523 ,H294592
,H294655 ,H294718 ,H294783 ,H294847 ,H294914 ,H294981 ,H295045 ,H295116
,H295186 ,H294462 ,H294524 ,H294593 ,H294656 ,H294719 ,H294784 ,H294848
,H294915 ,H294982 ,H295046 ,H295117 ,H295188 ,H294463 ,H294525 ,H294594
,H294657 ,H294720 ,H294785 ,H294850 ,H294916 ,H294983 ,H295047 ,H295118
,H295193 ,H294464 ,H294526 ,H294595 ,H294658 ,H294723 ,H294786 ,H294851
,H294917 ,H294984 ,H295048 ,H295119 ,H295195 ,H294465 ,H294527 ,H294596
,H294659 ,H294724 ,H294787 ,H294853 ,H294918 ,H294986 ,H295049 ,H295120
,H295196 ,H294466 ,H294528 ,H294597 ,H294660 ,H294726 ,H294788 ,H294854
,H294919 ,H294987 ,H295050 ,H295121 ,H295199 ,H294467 ,H294529 ,H294598
,H294661 ,H294727 ,H294789 ,H294855 ,H294920 ,H294988 ,H295051 ,H295122
,H295201 ,H294468 ,H294530 ,H294599 ,H294662 ,H294728 ,H294790 ,H294856
,H294921 ,H294989 ,H295053 ,H295123 ,H295202 ,H294469 ,H294531 ,H294600
,H294663 ,H294729 ,H294791 ,H294857 ,H294922 ,H294990 ,H295054 ,H295124
,H295203 ,H294470 ,H294532 ,H294601 ,H294664 ,H294730 ,H294792 ,H294858
,H294923 ,H294991 ,H295056 ,H295125 ,H295206 ,H294471 ,H294533 ,H294602
,H294665 ,H294731 ,H294793 ,H294859 ,H294924 ,H294992 ,H295057 ,H295126
,H295207 ,H294472 ,H294534 ,H294603 ,H294666 ,H294732 ,H294794 ,H294860
,H294925 ,H294993 ,H295058 ,H295127 ,H295208 ,H294473 ,H294535 ,H294604
,H294667 ,H294733 ,H294795 ,H294861 ,H294926 ,H294995 ,H295059 ,H295128
,H295209 ,H294474 ,H294536 ,H294605 ,H294668 ,H294734 ,H294796 ,H294862
,H294927 ,H294996 ,H295060 ,H295129 ,H295210 ,H294475 ,H294537 ,H294606
,H294669 ,H294735 ,H294797 ,H294863 ,H294928 ,H294997 ,H295061 ,H295130
,H295211 ,H294476 ,H294538 ,H294607 ,H294670 ,H294736 ,H294798 ,H294864
,H294929 ,H294998 ,H295062 ,H295131 ,H295212 ,H294477 ,H294539 ,H294608
,H294671 ,H294737 ,H294799 ,H294865 ,H294930 ,H294999 ,H295063 ,H295132
,H295213 ,H294478 ,H294540 ,H294610 ,H294672 ,H294738 ,H294800 ,H294866
,H294931 ,H295000 ,H295064 ,H295133 ,H295214 ,H294479 ,H294541 ,H294611
,H294673 ,H294739 ,H294801 ,H294867 ,H294932 ,H295001 ,H295065 ,H295134
,H295216 ,H294480 ,H294542 ,H294612 ,H294674 ,H294740 ,H294802 ,H294868
,H294933 ,H295002 ,H295066 ,H295135 ,H295217 ,H294481 ,H294543 ,H294613
,H294675 ,H294741 ,H294803 ,H294869 ,H294934 ,H295003 ,H295067 ,H295136
,H295221 ,H294482 ,H294544 ,H294614 ,H294676 ,H294742 ,H294804 ,H294870
,H294935 ,H295004 ,H295068 ,H295138 ,H295223 ,H294483 ,H294545 ,H294615
,H294677 ,H294743 ,H294806 ,H294871 ,H294936 ,H295005 ,H295069 ,H295139
,H295224 ,H294484 ,H294546 ,H294616 ,H294678 ,H294744 ,H294807 ,H294872
,H294937 ,H295006 ,H295070 ,H295140 ,H295225 ,H294485 ,H294547 ,H294617
,H294679 ,H294745 ,H294808 ,H294874 ,H294938 ,H295007 ,H295071 ,H295141
,H295226 ,H294486 ,H294548 ,H294618 ,H294680 ,H294746 ,H294809 ,H294875
,H294940 ,H295008 ,H295072 ,H295142 ,H295227 ,H294487 ,H294549 ,H294619
,H294681 ,H294747 ,H294810 ,H294876 ,H294941 ,H295009 ,H295073 ,H295143
,H295228 ,H294488 ,H294550 ,H294620 ,H294682 ,H294748 ,H294811 ,H294877
,H294942 ,H295010 ,H295074 ,H295144 ,H295229 ,H294489 ,H294551 ,H294621
,H294683 ,H294749 ,H294812 ,H294878 ,H294943 ,H295011 ,H295075 ,H295145
,H295231 ,H294490 ,H294552 ,H294622 ,H294684 ,H294750 ,H294813 ,H294879
,H294945 ,H295012 ,H295076 ,H295146 ,H295235 ,H294491 ,H294553 ,H294623
,H294685 ,H294751 ,H294814 ,H294880 ,H294946 ,H295013 ,H295077 ,H295147
,H295236 ,H294492 ,H294554 ,H294624 ,H294686 ,H294752 ,H294815 ,H294881
,H294947 ,H295014 ,H295078 ,H295148 ,H295238 ,H294493 ,H294555 ,H294625
,H294687 ,H294753 ,H294816 ,H294882 ,H294948 ,H295015 ,H295080 ,H295149
,H295239 ,H294494 ,H294556 ,H294626 ,H294688 ,H294754 ,H294817 ,H294883
,H294949 ,H295016 ,H295081 ,H295150 ,H295240 ,H294495 ,H294557 ,H294627
,H294689 ,H294755 ,H294818 ,H294884 ,H294950 ,H295017 ,H295082 ,H295151
,H295241 ,H294496 ,H294564 ,H294628 ,H294690 ,H294756 ,H294820 ,H294885
,H294951 ,H295018 ,H295083 ,H295152 ,H295242 ,H294497 ,H294565 ,H294629
,H294691 ,H294757 ,H294821 ,H294886 ,H294952 ,H295019 ,H295084 ,H295154
,H295244 ,H294498 ,H294567 ,H294630 ,H294692 ,H294758 ,H294822 ,H294887
,H294953 ,H295020 ,H295085 ,H295155 ,H295245 ,H294499 ,H294568 ,H294631
,H294693 ,H294759 ,H294823 ,H294888 ,H294954 ,H295021 ,H295086 ,H295156
,H295246 ,H294500 ,H294569 ,H294632 ,H294694 ,H294760 ,H294824 ,H294889
,H294955 ,H295022 ,H295087 ,H295157 ,H295247 ,H294501 ,H294570 ,H294633
,H294695 ,H294761 ,H294825 ,H294890 ,H294956 ,H295023 ,H295089 ,H295158
,H295248 ,H294502 ,H294571 ,H294634 ,H294696 ,H294762 ,H294826 ,H294891
,H294957 ,H295024 ,H295090 ,H295159 ,H295249 ,H294503 ,H294572 ,H294635
,H294697 ,H294763 ,H294827 ,H294892 ,H294958 ,H295025 ,H295091 ,H295160
,H295250 ,H294504 ,H294573 ,H294636 ,H294698 ,H294764 ,H294828 ,H294893
,H294959 ,H295026 ,H295093 ,H295161 ,H295252 ,H294505 ,H294574 ,H294637
,H294699 ,H294765 ,H294829 ,H294894 ,H294960 ,H295027 ,H295094 ,H295162
,H295254 ,H294506 ,H294575 ,H294638 ,H294700 ,H294766 ,H294830 ,H294895
,H294961 ,H295028 ,H295095 ,H295163 ,H295255 ,H294507 ,H294576 ,H294639
,H294701 ,H294767 ,H294831 ,H294896 ,H294962 ,H295029 ,H295096 ,H295169
,H295265;
2) H140002 H140087 H180135 H180273
H180463 H180554 H180623 H180685 H180747 H180809 H180868 H181026 H181085
H140003 H140089 H180143 H180274 H180466 H180555 H180624 ,H180686 ,H180748
,H180810 ,H180869 H181027 H181086 H140004 H140091 ,H180144 ,H180275 ,H180467
,H180556 H180625 H180687 H180749 H180811 ,H180870 ,H181028 ,H181087 ,H140005
H140093 H180148 H180283 H180469 ,H180559 ,H180626 ,H180688 ,H180750 H180812
H180871 H181029 H181088 ,H140006 ,H140094 ,H180157 ,H180287 H180470 H180560
H180627 H180689 ,H180751 ,H180813 ,H180872 ,H181030 H181089 H140007 H140095
H180159 ,H180290 ,H180471 ,H180561 ,H180628 H180690 H180752 H180814 H180873
,H181031 ,H181090 ,H140008 ,H140099 H180161 H180291 H180472 H180562 ,H180629
,H180691 ,H180753 ,H180815 H180874 H181032 H181091 H140009 ,H140100 ,H180166
,H180292 ,H180473 H180564 H180630 H180692 H180754 ,H180816 ,H180875 ,H181033
,H181092 H140010 H140102 H180167 H180293 ,H180474 ,H180566 ,H180631 ,H180694
H180755 H180817 H180876 H181034 ,H181093 ,H140012 ,H140103 ,H180171 H180294
H180475 H180568 H180632 ,H180695 ,H180756 ,H180818 ,H180877 H181035 H181094
H140014 H140105 ,H180175 ,H180295 ,H180476 ,H180569 H180633 H180696 H180757
H180819 ,H180878 ,H181036 ,H181095 ,H140015 H140106 H180178 H180296 H180477
,H180570 ,H180634 ,H180697 ,H180758 H180820 H180879 H181037 H181096 ,H140016
,H140108 ,H180179 ,H180298 H180478 H180572 H180635 H180698 ,H180759 ,H180821
,H180880 ,H181038 H181097 H140017 H180328 H180493 ,H180584 ,H180645 ,H180708
,H180769 ,H180831 H180891 H181048 H181107 ,H140034 ,H180004 ,H180208 ,H180329
,H180494 H180585 H180646 H180709 ,H180770 ,H180832 ,H180892 ,H181049 ,H181108
H140035 H180007 H180209 ,H180404 ,H180498 ,H180587 ,H180647 ,H180710 H180771
H180833 H180894 ,H181050 ,H181109 ,H140036 ,H180008 ,H180210 H180405 H180499
H180588 ,H180649 ,H180711 ,H180772 ,H180834 ,H180895 H181051 H181110 H140037
,H180016 ,H180214 ,H180406 ,H180501 ,H180589 H180650 H180713 H180773 ,H180835
,H180896 ,H181052 ,H181111 ,H140040 H180019 H180215 H180407 ,H180507 ,H180590
,H180651 ,H180714 ,H180774 H180836 H180897 H181053 ,H181112 ,H140041 ,H180025
,H180217 ,H180409 H180509 H180591 H180652 ,H180715 ,H180775 ,H180837 ,H180898
,H181055 H181113 H140042 H180028 ,H180218 ,H180411 ,H180510 ,H180592 ,H180653
H180716 H180776 H180838 ,H180899 ,H181056 ,H181114 ,H140044 ,H180040 H180220
H180412 H180511 ,H180593 ,H180654 ,H180717 ,H180777 ,H180839 H180901 H181057
H181115 ,H140049 ,H180043 ,H180221 ,H180414 ,H180514 H180594 H180655 H180718
,H180778 ,H180840 ,H180907 ,H181058 ,H181116 H140050 H180044 H180222 ,H180417
,H180515 ,H180595 ,H180656 ,H180719 H180779 H180841 H181000 ,H181059 ,H181117
,H140051 ,H180048 ,H180225 H180420 H180517 H180596 ,H180657 ,H180720 ,H180780
,H180842 ,H181001 H181060 H181118 H140052 ,H180049 ,H180226 ,H180424 ,H180518
,H180597 H180658 H180721 H180781 ,H180843 ,H181002 ,H181061 ,H181119 ,H140053
H180050 H180228 H180425 ,H180519 ,H180598 ,H180659 ,H180722 ,H180782 H180844
H181003 H181062 ,H181120 ,H140054 ,H180052 ,H180229 ,H180426 H180520 H180599
H180660 ,H180723 ,H180783 ,H180845 ,H181004 ,H181063 H181121 H140055 H180056
,H180233 ,H180427 ,H180521 ,H180600 ,H180661 H180724 H180784 H180847 ,H181005
,H181064 ,H181123 ,H140056 ,H180058 H180235 H180433 H180522 ,H180601 ,H180662
,H180725 ,H180785 ,H180848 H181006 H181065 H181124 ,H140057 ,H180059 ,H180236
,H180434 ,H180527 H180602 H180663 H180726 ,H180786 ,H180849 ,H181007 ,H181066
,H181125 H140059 H180065 H180237 ,H180437 ,H180528 ,H180603 ,H180664 ,H180727
H180788 H180850 H181008 ,H181067 ,H181126 ,H140060 ,H180066 ,H180238 H180440
H180529 H180604 ,H180665 ,H180729 ,H180789 ,H180851 ,H181009 H181068 H181127
H140061 ,H180073 ,H180240 ,H180441 ,H180532 ,H180605 H180667 H180730 H180791
,H180852 ,H181010 ,H181069 ,H181128 ,H140063 H180080 H180242 H180442 ,H180533
,H180606 ,H180669 ,H180731 ,H180792 H180853 H181011 H181070 ,H181129 ,H140064
,H180083 ,H180243 ,H180444 H180534 H180607 H180670 ,H180733 ,H180793 ,H180854
,H181012 ,H181071 H181130 H140065 H180087 ,H180245 ,H180445 ,H180535 ,H180608
,H180671 H180734 H180794 H180855 ,H181013 ,H181072 ,H181132 ,H140066 ,H180098
H180249 H180446 H180536 ,H180610 ,H180672 ,H180735 ,H180796 ,H180856 H181014
H181073 H181133 ,H140068 ,H180099 ,H180250 ,H180447 ,H180538 H180611 H180673
H180736 ,H180798 ,H180857 ,H181015 ,H181074 ,H181134 H140069 H180102 H180254
,H180449 ,H180539 ,H180612 ,H180674 ,H180737 H180799 H180858 H181016 ,H181075
,H181135 ,H140072 ,H180103 ,H180255 H180450 H180541 H180614 ,H180675 ,H180738
,H180800 ,H180859 ,H181017 H181076 H181136 H140074 ,H180104 ,H180256 ,H180451
,H180542 ,H180615 H180676 H180739 H180801 ,H180860 ,H181018 ,H181077 ,H181137
,H140075 H180107 H180257 H180453 ,H180543 ,H180616 ,H180678 ,H180740 ,H180802
H180861 H181019 H181078 ,H181138 ,H140080 ,H180115 ,H180258 ,H180454 H180544
H180617 H180679 ,H180741 ,H180803 ,H180862 ,H181020 ,H181079 H181139 H140082
H180116 ,H180260 ,H180456 ,H180548 ,H180618 ,H180680 H180742 H180804 H180863
,H181021 ,H181080 ,H181140 ,H140083 ,H180117 H180262 H180457 H180549 ,H180619
,H180681 ,H180743 ,H180805 ,H180864 H181022 H181081 H181141 ,H140084 ,H180122
,H180268 ,H180459 ,H180550 H180620 H180682 H180744 ,H180806 ,H180865 ,H181023
,H181082 ,H181142 H140085 H180124 H180271 ,H180460 ,H180551 ,H180621 ,H180683
,H180745 H180807 H180866 H181024 ,H181083 ,H181143 ,H140086 ,H180129 ,H180272
H180462 H180553 H180622 ,H180684 ,H180746 ,H180808 ,H180867 ,H181025 H181084
H181145 H181146 ,H181210 ,H270260 ,H270447 ,H270634 ,H270783 H270894 H271101
H271225 ,H271320 ,H271409 ,H271499 ,H271575 ,H181147 H181211 H270261 H270451
,H270635 ,H270785 ,H270895 ,H271102 ,H271228 H271322 H271410 H271500 ,H271576
,H181148 ,H181212 ,H270267 ,H270452 H270639 H270787 H270897 ,H271103 ,H271229
,H271325 ,H271411 ,H271501 H271577 H181149 H181214 ,H270271 ,H270454 ,H270641
,H270789 ,H270898 H271104 H271231 H271327 ,H271412 ,H271503 ,H271578 ,H181150
,H181215 H270273 H270455 H270644 ,H270793 ,H270899 ,H271107 ,H271233 ,H271328
H271415 H271504 H271580 ,H181151 ,H181218 ,H270277 ,H270458 ,H270645 H270796
H271000 H271108 ,H271234 ,H271329 ,H271416 ,H271505 ,H271581 H181152 H181220
H270280 ,H270461 ,H270646 ,H270800 ,H271002 ,H271109 H271235 H271330 H271417
,H271506 ,H271583 ,H181153 ,H181222 ,H270281 H270462 H270647 H270801 ,H271004
,H271110 ,H271236 ,H271331 ,H271419 H271508 H271584 H181154 ,H181223 ,H270282
,H270464 ,H270649 ,H270803 H271005 H271111 H271238 ,H271332 ,H271421 ,H271509
,H271586 ,H181155 H181224 H270285 H270466 ,H270651 ,H270804 ,H271009 ,H271113
,H271239 H271333 H271422 H271510 ,H271589 ,H181156 ,H181225 ,H270290 ,H270470
H270652 H270805 H271013 ,H271114 ,H271241 ,H271336 ,H271423 ,H271511 H271590
H181157 H181228 ,H270297 ,H270473 ,H270653 ,H270806 ,H271015 H271115 H271245
H271337 ,H271424 ,H271512 ,H271591 ,H181158 ,H181230 H270302 H270474 H270654
,H270812 ,H271016 ,H271116 ,H271247 ,H271338 H271426 H271513 H271592 ,H181159
,H189999 ,H270303 ,H270476 ,H270656 H270813 H271017 H271120 ,H271248 ,H271339
,H271428 ,H271514 ,H271595 H181160 H260916 H270307 ,H270477 ,H270660 ,H270817
,H271019 ,H271121 H271250 H271340 H271430 ,H271516 ,H271596 ,H181161 ,H260917
,H270310 H270479 H270661 H270820 ,H271020 ,H271127 ,H271251 ,H271342 ,H271431
H271517 H271597 H181162 ,H260923 ,H270314 ,H270482 ,H270662 ,H270821 H271021
H271132 H271256 ,H271344 ,H271434 ,H271518 ,H271598 ,H181163 H260926 H270323
H270483 ,H270664 ,H270822 ,H271023 ,H271135 ,H271257 H271345 H271435 H271520
,H271599 ,H181164 ,H260932 ,H270324 ,H270486 H270665 H270823 H271024 ,H271137
,H271259 ,H271347 ,H271436 ,H271521 H271600 H181165 H260938 ,H270332 ,H270487
,H270667 ,H270825 ,H271025 H271141 H271260 H271349 ,H271437 ,H271522 ,H271601
,H181166 ,H260939 H270335 H270489 H270669 ,H270826 ,H271026 ,H271142 ,H271263
,H271350 H271441 H271525 H271603 ,H181167 ,H260944 ,H270338 ,H270490 ,H270670
H270827 H271029 H271143 ,H271264 ,H271351 ,H271442 ,H271527 ,H271604 H181168
H260947 H270341 ,H270491 ,H270675 ,H270828 ,H271031 ,H271144 H271266 H271355
H271444 ,H271528 ,H271605 ,H181170 ,H270010 ,H270344 H270495 H270681 H270829
,H271032 ,H271145 ,H271267 ,H271356 ,H271445 H271529 H271606 H181171 ,H270021
,H270345 ,H270498 ,H270682 ,H270833 H271034 H271149 H271269 ,H271357 ,H271447
,H271530 ,H271608 ,H181172 H270026 H270351 H270499 ,H270684 ,H270836 ,H271035
,H271151 ,H271270 H271358 H271448 H271532 ,H271609 ,H181173 ,H270027 ,H270352
,H270500 H270686 H270837 H271037 ,H271152 ,H271272 ,H271359 ,H271449 ,H271533
H271610 H181174 H270029 ,H270353 ,H270501 ,H270690 ,H270838 ,H271038 H271154
H271273 H271360 ,H271450 ,H271535 ,H271611 ,H181175 ,H270030 H270356 H270502
H270692 ,H270840 ,H271039 ,H271155 ,H271274 ,H271361 H271451 H271536 H271612
,H181176 ,H270038 ,H270360 ,H270503 ,H270693 H270842 H271042 H271161 ,H271276
,H271363 ,H271453 ,H271537 ,H271613 H181177 H270045 H270364 ,H270504 ,H270698
,H270844 ,H271043 ,H271166 H271280 H271364 H271454 ,H271538 ,H271614 ,H181178
,H270047 ,H270370 H270505 H270699 H270846 ,H271044 ,H271172 ,H271281 ,H271365
,H271456 H271539 H271615 H181179 ,H270073 ,H270371 ,H270511 ,H270700 ,H270852
H271051 H271178 H271282 ,H271366 ,H271457 ,H271540 ,H271616 ,H181180 H270079
H270373 H270512 ,H270705 ,H270854 ,H271052 ,H271182 ,H271283 H271367 H271458
H271541 ,H271617 ,H181181 ,H270088 ,H270378 ,H270513 H270708 H270857 H271055
,H271183 ,H271284 ,H271368 ,H271459 ,H271542 H271619 H181182 H270096 ,H270383
,H270516 ,H270710 ,H270858 ,H271056 H271184 H271285 H271369 ,H271461 ,H271543
,H271620 ,H181183 ,H270101 H270384 H270518 H270714 ,H270859 ,H271059 ,H271185
,H271287 ,H271371 H271462 H271544 H271621 ,H181184 ,H270106 ,H270385 ,H270520
,H270716 H270861 H271061 H271189 ,H271289 ,H271372 ,H271464 ,H271545 ,H271622
H181185 H270110 H270386 ,H270522 ,H270718 ,H270863 ,H271062 ,H271190 H271290
H271374 H271465 ,H271546 ,H271627 ,H181187 ,H270111 ,H270391 H270525 H270719
H270864 ,H271063 ,H271191 ,H271291 ,H271375 ,H271466 H271547 H271629 H181188
,H270118 ,H270392 ,H270526 ,H270725 ,H270865 H271067 H271192 H271292 ,H271376
,H271467 ,H271548 ,H271630 ,H181189 H270126 H270395 H270529 ,H270726 ,H270866
,H271068 ,H271193 ,H271294 H271377 H271468 H271550 ,H271631 ,H181190 ,H270134
,H270398 ,H270532 H270727 H270867 H271071 ,H271195 ,H271295 ,H271379 ,H271470
,H271551 H271632 H181191 H270138 ,H270399 ,H270535 ,H270728 ,H270868 ,H271072
H271197 H271296 H271380 ,H271472 ,H271552 ,H271633 ,H181192 ,H270201 H270401
H270539 H270732 ,H270869 ,H271073 ,H271198 ,H271297 ,H271381 H271473 H271553
H271634 ,H181193 ,H270209 ,H270410 ,H270540 ,H270733 H270871 H271075 H271199
,H271298 ,H271382 ,H271474 ,H271554 ,H271636 H181194 H270211 H270416 ,H270542
,H270735 ,H270872 ,H271076 ,H271200 H271299 H271383 H271475 ,H271555 ,H271637
,H181195 ,H270215 ,H270418 H270544 H270737 H270873 ,H271080 ,H271201 ,H271300
,H271384 ,H271476 H271557 H271638 H181196 ,H270216 ,H270419 ,H270550 ,H270739
,H270875 H271082 H271204 H271302 ,H271386 ,H271477 ,H271558 ,H271639 ,H181197
H270219 H270421 H270601 ,H270746 ,H270877 ,H271083 ,H271205 ,H271303 H271387
H271478 H271561 ,H271641 ,H181198 ,H270227 ,H270425 ,H270602 H270749 H270878
H271085 ,H271207 ,H271304 ,H271388 ,H271479 ,H271562 H271642 H181199 H270228
,H270426 ,H270606 ,H270751 ,H270879 ,H271086 H271208 H271305 H271389 ,H271481
,H271563 ,H271643 ,H181200 ,H270229 H270427 H270607 H270752 ,H270880 ,H271088
,H271209 ,H271306 ,H271390 H271483 H271564 H271645 ,H181201 ,H270230 ,H270428
,H270613 ,H270755 H270882 H271091 H271210 ,H271307 ,H271392 ,H271484 ,H271565
,H271646 H181202 H270232 H270430 ,H270615 ,H270757 ,H270883 ,H271092 ,H271211
H271309 H271395 H271485 ,H271566 ,H271647 ,H181203 ,H270235 ,H270431 H270622
H270761 H270884 ,H271093 ,H271212 ,H271312 ,H271398 ,H271487 H271567 H271648
H181204 ,H270244 ,H270435 ,H270623 ,H270762 ,H270885 H271094 H271216 H271314
,H271399 ,H271488 ,H271568 ,H271649 ,H181205 H270245 H270437 H270624 ,H270763
,H270886 ,H271095 ,H271217 ,H271315 H271401 H271492 H271570 ,H271650 ,H181206
,H270251 ,H270438 ,H270625 H270766 H270888 H271097 ,H271218 ,H271316 ,H271402
,H271493 ,H271571 H271651 H181207 H270252 ,H270439 ,H270627 ,H270777 ,H270889
,H271098 H271222 H271317 H271403 ,H271494 ,H271572 ,H271652 ,H181208 ,H270255
H270443 H270631 H270781 ,H270891 ,H271099 ,H271223 ,H271318 ,H271404 H271495
H271573 H271653 ,H181209 ,H270256 ,H270446 ,H270632 ,H270782 H270892 H271100
H271224 ,H271319 ,H271406 ,H271498 ,H271574 ,H271654 H271655 ,H271739 H271816
,H271901 ,H271983 ,H272071 ,H272149 ,H272251 H272348 H272447 H272865 ,H272957
,H273062 ,H271656 ,H271740 ,H271817 H271902 H271984 H272072 ,H272150 ,H272253
,H272350 ,H272448 ,H272867 H272958 H273067 H271657 ,H271741 ,H271818 ,H271904
,H271985 ,H272073 H272151 H272254 H272351 ,H272449 ,H272869 ,H272959 ,H273068
,H271658 H271743 H271819 H271907 ,H271986 ,H272074 ,H272152 ,H272255 ,H272352
H272450 H272870 H272962 ,H273069 ,H271659 ,H271744 ,H271820 ,H271908 H271987
H272075 H272154 ,H272256 ,H272353 ,H272451 ,H272871 ,H272965 H273070 H271660
H271746 ,H271821 ,H271909 ,H271988 ,H272076 ,H272160 H272257 H272356 H272452
,H272875 ,H272966 ,H273071 ,H271666 ,H271747 H271822 H271910 H271989 ,H272077
,H272161 ,H272258 ,H272357 ,H272453 H272877 H272967 H273074 ,H271667 ,H271748
,H271823 ,H271911 ,H271990 H272079 H272164 H272261 ,H272358 ,H272455 ,H272878
,H272969 ,H273078 H271670 H271749 H271824 ,H271912 ,H271991 ,H272080 ,H272165
,H272262 H272361 H272457 H272879 ,H272970 ,H273079 ,H271671 ,H271750 ,H271826
H271913 H271992 H272082 ,H272166 ,H272263 ,H272363 ,H272459 ,H272880 H272972
H273081 H271672 ,H271751 ,H271827 ,H271914 ,H271996 ,H272083 H272167 H272264
H272364 ,H272461 ,H272881 ,H272974 ,H273085 ,H271673 H271752 H271828 H271915
,H271997 ,H272084 ,H272170 ,H272265 ,H272366 H272463 H272882 H272975 ,H273087
,H271674 ,H271753 ,H271829 ,H271916 H271998 H272087 H272171 ,H272266 ,H272367
,H272465 ,H272883 ,H272976 H273090 H271675 H271754 ,H271830 ,H271917 ,H271999
,H272088 ,H272175 H272268 H272368 H272467 ,H272884 ,H272977 ,H273091 ,H271676
,H271756 H271832 H271919 H272002 ,H272090 ,H272176 ,H272269 ,H272369 ,H272468
H272886 H272979 H273094 ,H271677 ,H271757 ,H271834 ,H271920 ,H272004 H272091
H272177 H272270 ,H272371 ,H272471 ,H272887 ,H272980 ,H273095 H271679 H271758
H271836 ,H271922 ,H272005 ,H272092 ,H272181 ,H272272 H272373 H272472 H272888
,H272983 ,H273097 ,H271680 ,H271759 ,H271837 H271923 H272006 H272093 ,H272182
,H272273 ,H272374 ,H272475 ,H272891 H272984 H273098 H271682 ,H271761 ,H271839
,H271924 ,H272007 ,H272094 H272183 H272274 H272375 ,H272477 ,H272892 ,H272988
,H273099 ,H271683 H271764 H271840 H271925 ,H272010 ,H272095 ,H272185 ,H272276
,H272376 H272478 H272894 H272989 ,H290001 ,H271684 ,H271765 ,H271841 ,H271927
H272011 H272096 H272186 ,H272278 ,H272378 ,H272482 ,H272895 ,H272991 H290003
H271685 H271766 ,H271842 ,H271928 ,H272012 ,H272097 ,H272189 H272279 H272379
H272483 ,H272896 ,H272992 ,H290004 ,H271686 ,H271768 H271843 ,H271929 H272013
,H272098 ,H272190 ,H272280 ,H272382 ,H272484 H272897 H272994 H290014 ,H271687
,H271769 ,H271844 ,H271930 ,H272015 H272099 H272191 H272281 ,H272383 ,H272485
,H272898 ,H272995 ,H290026 H271688 H271770 H271846 ,H271932 ,H272017 ,H272100
,H272192 ,H272282 H272384 H272489 H272900 ,H272997 ,H290030 ,H271689 ,H271771
,H271847 H271933 H272018 H272101 ,H272193 ,H272284 ,H272385 ,H272490 ,H272901
H272998 H290031 H271692 ,H271772 ,H271848 ,H271934 ,H272019 ,H272102 H272194
H272285 H272386 ,H272493 ,H272902 ,H273000 ,H290032 ,H271695 H271774 H271849
H271936 ,H272020 ,H272103 ,H272196 ,H272286 ,H272387 H272495 H272903 H273001
,H290033 ,H271696 ,H271775 ,H271850 ,H271937 H272021 H272104 H272198 ,H272287
,H272389 ,H272496 ,H272904 ,H273002 H290034 H271697 H271776 ,H271851 ,H271938
,H272022 ,H272106 ,H272199 H272288 H272390 H272800 ,H272906 ,H273005 ,H290104
,H271698 ,H271777 H271852 H271939 H272024 ,H272107 ,H272200 ,H272289 ,H272391
,H272802 H272907 H273007 H290105 ,H271699 ,H271778 ,H271854 ,H271940 ,H272025
H272108 H272201 H272292 ,H272392 ,H272807 ,H272913 ,H273008 ,H290109 H271700
H271779 H271855 ,H271943 ,H272026 ,H272109 ,H272205 ,H272294 H272393 H272808
H272915 ,H273010 ,H290110 ,H271701 ,H271781 ,H271857 H271944 H272027 H272110
,H272206 ,H272295 ,H272395 ,H272809 ,H272916 H273012 H290111 H271702 ,H271782
,H271859 ,H271945 ,H272028 ,H272112 H272207 H272296 H272396 ,H272810 ,H272917
,H273013 ,H290113 H271703 H271784 H271861 H271946 ,H272030 ,H272113 ,H272208
,H272297 H272398 H272812 H272918 H273014 ,H290114 ,H271704 ,H271786 ,H271863
,H271947 H272032 H272114 H272210 ,H272298 ,H272399 ,H272813 ,H272919 ,H273017
H290116 H271705 H271787 ,H271864 ,H271948 ,H272034 ,H272116 ,H272211 H272300
H272400 H272814 ,H272920 ,H273018 ,H290118 ,H271706 ,H271788 H271865 H271949
H272036 ,H272117 ,H272212 ,H272303 ,H272402 ,H272815 H272921 H273019 H290121
,H271707 ,H271789 ,H271867 ,H271950 ,H272038 H272119 H272213 H272304 ,H272403
,H272817 ,H272923 ,H273020 ,H290122 H271708 H271790 H271868 ,H271952 ,H272039
,H272121 ,H272215 ,H272305 H272404 H272818 H272924 ,H273021 ,H290123 ,H271709
,H271791 ,H271869 H271953 H272040 H272122 ,H272217 ,H272308 ,H272406 ,H272819
,H272925 H273022 H290125 H271710 ,H271792 ,H271871 ,H271955 ,H272041 ,H272123
H272218 H272312 H272408 ,H272821 ,H272927 ,H273023 ,H290128 ,H271715 H271793
H271873 H271956 ,H272046 ,H272124 ,H272221 ,H272314 ,H272409 H272823 H272928
H273024 ,H290130 ,H271716 ,H271794 ,H271874 ,H271957 H272047 ,H272125 H272222
,H272315 ,H272411 ,H272829 ,H272929 ,H273025 H290132 ,H271717 H271795 ,H271876
,H271958 ,H272048 ,H272126 ,H272224 H272316 ,H272414 H272830 ,H272930 ,H273028
,H290135 ,H271718 ,H271796 H271878 ,H271959 H272051 ,H272127 ,H272227 ,H272317
,H272415 ,H272831 H272931 ,H273033 H290137 ,H271719 ,H271798 ,H271880 ,H271960
,H272053 H272128 ,H272228 H272319 ,H272417 ,H272833 ,H272934 ,H273034 ,H290138
H271722 ,H271799 H271882 ,H271962 ,H272054 ,H272130 ,H272229 ,H272320 H272419
,H272836 H272935 ,H273037 ,H290139 ,H271723 ,H271800 ,H271883 H271963 ,H272055
H272131 ,H272230 ,H272322 ,H272420 ,H272838 ,H272937 H273038 ,H290141 H271725
,H271803 ,H271884 ,H271965 ,H272056 ,H272132 H272232 ,H272323 H272424 ,H272839
,H272939 ,H273043 ,H290145 ,H271726 H271804 ,H271885 H271967 ,H272057 ,H272133
,H272233 ,H272330 ,H272429 H272840 ,H272940 H273044 ,H290151 ,H271727 ,H271805
,H271886 ,H271968 H272058 ,H272135 H272234 ,H272331 ,H272430 ,H272843 ,H272942
,H273045 H290152 ,H271728 H271806 ,H271887 ,H271969 ,H272059 ,H272139 ,H272238
H272332 ,H272431 H272844 ,H272943 ,H273046 ,H290154 ,H271731 ,H271807 H271888
,H271970 H272060 ,H272140 ,H272241 ,H272334 ,H272433 ,H272845 H272944 ,H273047
H290155 ,H271732 ,H271808 ,H271889 ,H271971 ,H272061 H272141 ,H272242 H272335
,H272435 ,H272850 ,H272947 ,H273050 ,H290156 H271733 ,H271809 H271891 ,H271977
,H272063 ,H272142 ,H272244 ,H272337 H272436 ,H272851 H272948 ,H273051 ,H290157
,H271734 ,H271810 ,H271892 H271978 ,H272065 H272144 ,H272245 ,H272338 ,H272437
,H272852 ,H272949 H273054 ,H290158 H271735 ,H271812 ,H271894 ,H271979 ,H272066
,H272145 H272246 ,H272340 H272439 ,H272853 ,H272950 ,H273058 ,H290159 ,H271736
H271813 ,H271897 H271980 ,H272067 ,H272146 ,H272247 ,H272342 ,H272441 H272855
,H272952 H273059 ,H290160 ,H271737 ,H271814 ,H271898 ,H271981 H272068 ,H272147
H272248 ,H272344 ,H272442 ,H272861 ,H272954 ,H273060 H290161 ,H271738 H271815
,H271899 ,H271982 ,H272070 ,H272148 ,H272249 H272347 ,H272443 H272863 ,H272955
,H273061 ,H290162 ,H290164 ,H290280 H290357 ,H290531 H290165 ,H290281 ,H290359
,H290532 ,H290166 ,H290283 H290360 ,H290534 H290168 ,H290284 ,H290363 ,H290535
,H290173 ,H290285 H290364 ,H290536 H290175 ,H290286 ,H290365 ,H290537 ,H290176
,H290288 H290366 ,H290539 H290181 ,H290289 ,H290367 ,H290542 ,H290182 ,H290290
H290368 ,H290543 H290185 ,H290291 ,H290369 ,H290544 ,H290187 ,H290292 H290370
,H290545 H290189 ,H290293 ,H290373 ,H290546 ,H290191 ,H290296 H290374 ,H292832
H290193 ,H290297 ,H290377 ,H290194 ,H290299 ,H290378 H290195 ,H290301 H290380
,H290196 ,H290302 ,H290382 ,H290197 ,H290305 H290383 ,H290198 H290306 ,H290384
,H290203 ,H290307 ,H290385 ,H290206 H290309 ,H290387 H290209 ,H290310 ,H290388
,H290210 ,H290311 ,H290389 H290211 ,H290312 H290390 ,H290212 ,H290314 ,H290393
,H290213 ,H290315 H290394 ,H290214 H290316 ,H290395 ,H290215 ,H290318 ,H290396
,H290216 H290319 ,H290398 H290217 ,H290320 ,H290404 ,H290219 ,H290322 ,H290406
H290220 ,H290323 H290408 ,H290223 ,H290324 ,H290409 ,H290225 ,H290325 H290410
,H290226 H290327 ,H290411 ,H290227 ,H290328 ,H290412 ,H290229 H290329 ,H290413
H290231 ,H290330 ,H290415 ,H290235 ,H290331 ,H290416 H290236 ,H290332 H290417
,H290237 ,H290333 ,H290424 ,H290239 ,H290334 H290425 ,H290240 H290335 ,H290426
,H290242 ,H290336 ,H290427 ,H290244 H290337 ,H290500 H290246 ,H290338 ,H290502
,H290247 ,H290339 ,H290506 H290250 ,H290340 H290507 ,H290255 ,H290342 ,H290513
,H290257 ,H290343 H290514 ,H290258 H290344 ,H290515 ,H290259 ,H290345 ,H290516
,H290260 H290346 ,H290518 H290261 ,H290347 ,H290521 ,H290264 ,H290348 ,H290522
H290265 H290349 H290523 ,H290266 ,H290350 ,H290524 ,H290268 ,H290351 H290525
,H290269 H290353 H290526 H290271 H290354 H290527 H290275 H290355 H290528
H290279 H290356 H290529
RECALLING FIRM/MANUFACTURER
Varian Medical Systems, Inc. Oncology
Systems, Palo Alto, CA,
REASON An event has been reported to Varian
which entails excessive connector resistance, which caused the actual jaw
positions to differ from the intended jaw positions without warning the
operator.
VOLUME OF PRODUCT IN COMMERCE
5714 units
DISTRIBUTION Nationwide and Internationally
PRODUCT
1) LigaSure V 5 mm Dolphin Tip
Laparoscopic 37 cm Sealer/Divider, LS1500, Sterile EO. Hand activated
sealer/dividers that can be used to seal vessels up to and including 7 mm.
lymphatics, and tissue bundles.
2) LigaSure V 5 mm Dolphin Tip Open 20
cm Sealer/Divider, LS1520, Sterile EO. Hand activated sealer/dividers that can
be used to seal vessels up to and including 7 mm. lymphatics, and tissue
bundles.
CODE
1) Lot numbers: S1CB001, S1CB002,
S1CB003, S1CB004, S1CB005, S1CB006, S1CB007R, S1CB008X, S1CB009X, S1CB012RX,
S1CB013PRX, S1CB014PRX, S1CB015PRX, S1CB016PRX, S1CB017PRX, S1CB018PRX,
S1CB019PRX, S1CB020PRX, S1CB021PRX, S1CB022PRX, S1CB023PRX, S1CB024PRX,
S1CB025PRX, S1DB006X, S1DB007X, S1DB008X, S1DB009X, S1DB010X, S1DB011X,
S1DB012X, S1DB013X, S1DB014X, S1DB015X, S1DB016X, S1DB017X, S1DB018X,
S1DB019X, S1DB020X, S1DB021X, S1DB022X, S1DB023X, S1EB001X, S1EB002X,
S1EB003X, S1EB004X, S1EB005X, S1EB006X, S1EB007RX, S1EB008RX, S1EB009X,
S1EB010RX, S1EB011RX, S1EB016P, S1EB017P, S1EB018P, S1EB019P, S1EB020P,
S1EB021P, S1EB022P, S1EB023P, S1EB024P, S1EB025P, S1EB026P, S1EB027P,
S1EB028P, S1EB029P, S1EB030P, S1EB031P, S1EB032P, S1EB033P, S1EB034P,
S1EB035P, S1EB036P, S1EB037P, S1EB038P, S1EB039P, S1EB040P, S1EB041P,
S1EB042P, S1FB001P, S1FB005PX, S1FB006PX, S1FB007PX, S1FB008PX, S1FB009PX,
S1FB010PX, S1FB011PX, S1FB012PX, S1FB013PX, S1FB014PX, S1FB015PX, S1FB016PX,
S1FB017PX, S1FB027PX, S1FB028PX, S1FB029PX, S1FB040X, S1FB041X, S1fb042X,
S1FB043X, S1FB044X, S1FB045X, S1FB046X, S1FB047X, S1FB048X, S1FB052X,
S1GB001X, S1GB002X, S1GB003x, S1GB004X, S1GB009X, S1GB010X, S1GB011X,
S1GB012X, S1GB013X, S1GB014X, S1GB015X, S1GB016X, S1GB017X, S1GB018X, S1HB002,
S1HB003, S1HN004, S1HB005, S1HB006, S1HB007, S1HB008, S1HB009P, S1HB014P,
S1HB015P, S1HB016P, S1HB017P, S1HB018P, S1HB019, S1HB021X, S1HB022X, S1HB023X,
S1HB024X, S1HB027PX, S1HB028PX, S1HB029PX, S1HB031X, S1HB032X, S1HB033X,
S1JB034X, S1JB035X, S1JB036X, S1JB038PX, S1JB041X, S1JB042X, S1JB043X,
S1JB044PX, S1JB045PX, S1JB046X, S1JB047X, S1JB048X;
2) Lot numbers: S1CB010X, S1CB011X,
S1DB001X, S1DB002X, S1EB012P, S1EB013P, S1EB014P, S1FB031PX, S1FB032PX,
S1FB033PX, S1FB035X, S1FB036X, S1FB037X, S1FB038X, S1FB039X, S1FB049X,
S1FB050X, S1FB051X, S1GB005X, S1GB006X, S1GB007X, S1HB010, S1HB011, S1HB012,
S1HB013P. S1CB001, S1CB002, S1CB003, S1CB004, S1CB005, S1CB006, S1CB007R,
S1CB008X, S1CB009X, S1CB012RX, S1CB013PRX, S1CB014PRX, S1CB015PRX, S1CB016PRX,
S1CB017PRX, S1CB018PRX, S1CB019PRX, S1CB020PRX, S1CB021PRX, S1CB022PRX,
S1CB023PRX, S1CB024PRX, S1CB025PRX, S1DB006X, S1DB007X, S1DB008X, S1DB009X,
S1DB010X, S1DB011X, S1DB012X, S1DB013X, S1DB014X, S1DB015X, S1DB016X,
S1DB017X, S1DB018X, S1DB019X, S1DB020X, S1DB021X, S1DB022X, S1DB023X,
S1EB001X, S1EB002X, S1EB003X, S1EB004X, S1EB005X, S1EB006X, S1EB007RX,
S1EB008RX, S1EB009X, S1EB010RX, S1EB011RX, S1EB016P, S1EB017P, S1EB018P,
S1EB019P, S1EB020P, S1EB021P, S1EB022P, S1EB023P, S1EB024P, S1EB025P,
S1EB026P, S1EB027P, S1EB028P, S1EB029P, S1EB030P, S1EB031P, S1EB032P,
S1EB033P, S1EB034P, S1EB035P, S1EB036P, S1EB037P, S1EB038P, S1EB039P,
S1EB040P, S1EB041P, S1EB042P, S1FB001P, S1FB005PX, S1FB006PX, S1FB007PX,
S1FB008PX, S1FB009PX, S1FB010PX, S1FB011PX, S1FB012PX, S1FB013PX, S1FB014PX,
S1FB015PX, S1FB016PX, S1FB017PX, S1FB027PX, S1FB028PX, S1FB029PX, S1FB040X,
S1FB041X, S1fb042X, S1FB043X, S1FB044X, S1FB045X, S1FB046X, S1FB047X,
S1FB048X, S1FB052X, S1GB001X, S1GB002X, S1GB003x, S1GB004X, S1GB009X,
S1GB010X, S1GB011X, S1GB012X, S1GB013X, S1GB014X, S1GB015X, S1GB016X,
S1GB017X, S1GB018X, S1HB002, S1HB003, S1HN004, S1HB005, S1HB006, S1HB007,
S1HB008, S1HB009P, S1HB014P, S1HB015P, S1HB016P, S1HB017P, S1HB018P, S1HB019,
S1HB021X, S1HB022X, S1HB023X, S1HB024X, S1HB027PX, S1HB028PX, S1HB029PX,
S1HB031X, S1HB032X, S1HB033X, S1JB034X, S1JB035X, S1JB036X, S1JB038PX,
S1JB041X, S1JB042X, S1JB043X, S1JB044PX, S1JB045PX, S1JB046X, S1JB047X,
S1JB048X
RECALLING FIRM/MANUFACTURER
Recalling Firm: Covidien, PLC, Boulder,
CO,
Manufacturer: Covidien Medical Products
(Shanghai) Manufacturing LLC, Shanghai, China. .
REASON Retaining pins in handle assembly may
disengage and fall into surgical field or compromise handle unlatching.
VOLUME OF PRODUCT IN COMMERCE
120,810 units
DISTRIBUTION Nationwide and Internationally
PRODUCT
1) Neulook Fancy FX is packaged in a
small glass bottle with a orange, white, and yellow affixed label with black
printing that reads, NeuLook COLOR CONTACT LENSES Distributed by EYEZONE
Contents : One Lens, 38% water and 62% Polymacon immersed in saline. Caution :
To ensure that your contacts stay comfortable and your vision stay clear, ask
your eye-care professional for product instruction. ***DIA. 14.0mm B.C
8.6mm***Rxonly STERILE***FDA 510K Cleared Made in Korea***Power : Plane Color
: Blue Angelic The color contact lens colors are: Aqua, blue, blue sapphire,
dark gray, dark violet, gray, green, hazel, honey, jade, magic angel, misty
blue, misty green, misty hazel, pearl gray, pure hazel, sky blue, sweet gray,
turquoise, violet, black out, vein, night, white cat, red cat devil, blue
angelic, white out, hot red spider web, psycho, cruller, solid yellow, and
blue wheel. prescription contact lens. Barcode on label: Bottle: No barcode,
Box: 8 806048 000468.
2) Neulook Deluxe is packaged in a
pink soft plastic box with white, black, and rainbow printing that reads,
NeuLook COLOR CONTACT LENSES Give yourself a whole new look***Contents : Two
sterile color contact lenses(62% Polymacon, 38% Water) Immersed in saline and
one lens case. Caution : To ensure that your contacts stay comfortable and
your Vision stay clear, a1) sk your eye-care professional for product
instruction. Made in Korea Distributed by EYEZONE***1023 FDA 510K
CLEARED***STERILE***B.C 8.6mm DIA. 14.0mm***8 806048 000468. The color contact
lens colors are: Black out, vein, night, white cat, red cat devil, blue
angelic, white out, and hot red. prescription contact lens. Barcode on label:
Bottle: No barcode, box: 8 806048 000468.
CODE
1) LOT: BD54 EXP 2014/09, include all
lots;
2) LOT. BD53, EXP. 2014/08, include all
lots
RECALLING FIRM/MANUFACTURER
Recalling Firm: Eyezone, Santa Fe
Springs, CA,
Manufacturer: Bescon Co., Ltd.,
Cheonan-City,
Republic Of South Korea.
REASON Pishon Trading dba Eyezone have
distributed soft color lens while on FDA hold and were later found to be
misbranded.
VOLUME OF PRODUCT IN COMMERCE
24,628 total pairs
PRODUCT Allen C-Flex Head Positioning System
non-invasive medical device used to support a patient's head during spine
surgery Catalog Numbers: A-70700 and A-70701-2.
CODE
Units: A128573, A157480, A164957,
A165624, A179587, A200505, A200508, A203121, A203123, A203124, A203126,
A205717, A205730, A205731, A207969, A207970, A207971, A207973, A208274,
A208275, A208277, A208278, A208932, A208933, A208934, A209578, A210915,
A210916, A210917, A210919, A210920, A214252, A214253, A214254, A214255,
A214256, A214582, A214583, A214584, A215498, A215489, A215499, A215500,
A216459, A216460, A216461, A217594, A217595, A217596, A217597, A218945,
A219496, A219623, A219624, A220031, A220032, A220033, A220664, A220665,
A222121, A222122, A222123, A222394, A223961, A223962, A223963, A223964,
A224084, A224085, A224144, A224145, A225106, A225107, A225553, A225554,
A22555, A225972, A225973, A226214, A226215, A226216, A226217, A226218, A226910
RECALLING FIRM/MANUFACTURER
Allen Medical Systems, Inc., Acton, MA,
REASON C-Flex Polar Head Positioner may have an
adjustment screw loosen and affect load capacity
VOLUME OF PRODUCT IN COMMERCE
83 units
DISTRIBUTION Nationwide and Internationally
PRODUCT Model 750 24' MRI Reusable/Disposable
Circuit Accessory for use with Model 750 Portable Ventilator.
CODE p/n 802-1750-05, no expiration;
802-17850-24
RECALLING FIRM/MANUFACTURER
Impact Instrumentation, Inc., West
Caldwell, NJ,
REASON Product does not have a 510(k).
REASON Pishon Trading dba Eyezone have
distributed soft color lens while on FDA hold and were later found to be
misbranded.
VOLUME OF PRODUCT IN COMMERCE
564 units
PRODUCT Dimension(R) Creatinine
Flex(R) Reagent
Cartridges. The CREA method is an in vitro diagnostic test for the
quantitative measurement of creatinine in human serum, plasma, and urine on
the Dimension Vista System. Catalog number DF33A.
CODE Lot numbers: GA2196, GA2203, DC2221,
DB2221, GA2229 and FA2237.
RECALLING FIRM/MANUFACTURER
Siemens Healthcare Diagnostics, Inc.,
Newark, DE,
REASON Siemens has confirmed the potential for
falsely elevated creatinine results when processing tests from a well that has
been punctured.
VOLUME OF PRODUCT IN COMMERCE
31,680 units
DISTRIBUTION Nationwide and Internationally
PRODUCT Spacelabs Medical Ultraview SL Command
Module, Model 91496. The Spacelabs Multi-parameter Module is intended for use
with the Patient Care Management System (PCMS) to acquire, monitor, and
process various clinical parameters from an adult or neonate/infant
populations in any type of clinical environment other than home use.
Physiological parameters that may be monitored include ECG with arrhythmia
detection, respiration, invasive and noninvasive blood pressure, temperature,
oxygen saturation (SpO2) and cardiac output. Acquired data may then be
communicated to an information network for display, recording, editing and
analysis.
CODE Serial number: 1496-029792.
RECALLING FIRM/MANUFACTURER
Spacelabs Healthcare, LLC, Issaquah, WA,
REASON Due to software error.
VOLUME OF PRODUCT IN COMMERCE
1 unit
PRODUCT Computed Tomography X-Ray Systems
(Brilliance 64, Model Number: 728231; Brilliance Big Bore, Model Number:
728243; Brilliance iCT, Model Number: 728306; Brilliance iCT SP, Model Number:
728311; and Ingenuity CT, Model Number: 728326). Intended to produce
cross-sectional images of the body by computer reconstruction of x-ray
transmission data taken at different angles and planes.
CODE None
RECALLING FIRM/MANUFACTURER
Philips Medical Systems (Cleveland)
Inc., Cleveland, OH,
REASON The Common Image Reconstruction System
(CIRS)
on the affected Brilliance Systems encounters an error and requires reboot of
the system. The unit failed to initialize axial scan due to timeout that
occurred waiting for "READY" state
VOLUME OF PRODUCT IN COMMERCE
548 units
DISTRIBUTION Nationwide and Internationally
PRODUCT
1) Natural-Knee II Durasul All-Poly
Patella, Size 0, 7mm The Zimmer Natural-Knee II Durasul Patella product is a
semi-constrained total knee prosthesis consisting of four anatomically
designed components: the femoral, tibial, articular surface, and patellar
prosthesis. The all-polyethylene patella is made from Durasul Highly
Crosslinked Ultra High Molecular Weight Polyethylene. The N-K II Durasul
polyethylene patella components are indicated for patient conditions of
non-inflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis,
osteoarthritis, and inflammatory degenerative joint disease (IJD), e.g.,
rheumatoid arthritis; correctable valgus-varus deformity and moderate flexion
contracture; those patients with failed previous surgery where pain,
deformity, or dysfunction persists; and revision of previously failed knee
arthroplasty.
2) Natural-Knee II Durasul All-Poly
Patella, Size 1, 7mm The Zimmer Natural-Knee II Durasul Patella product is a
semi-constrained total knee prosthesis consisting of four anatomically
designed components: the femoral, tibial, articular surface, and patellar
prosthesis. The all-polyethylene patella is made from Durasul Highly
Crosslinked Ultra High Molecular Weight Polyethylene. The N-K II Durasul
polyethylene patella components are indicated for patient conditions of
non-inflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis,
osteoarthritis, and inflammatory degenerative joint disease (IJD), e.g.,
rheumatoid arthritis; correctable valgus-varus deformity and moderate flexion
contracture; those patients with failed previous surgery where pain,
deformity, or dysfunction persists; and revision of previously failed knee
arthroplasty.
3) Natural-Knee II Durasul All-Poly
Patella, Size 2, 7mm The Zimmer Natural-Knee II Durasul Patella product is a
semi-constrained total knee prosthesis consisting of four anatomically
designed components: the femoral, tibial, articular surface, and patellar
prosthesis. The all-polyethylene patella is made from Durasul Highly
Crosslinked Ultra High Molecular Weight Polyethylene. The N-K II Durasul
polyethylene patella components are indicated for patient conditions of
non-inflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis,
osteoarthritis, and inflammatory degenerative joint disease (IJD), e.g.,
rheumatoid arthritis; correctable valgus-varus deformity and moderate flexion
contracture; those patients with failed previous surgery where pain,
deformity, or dysfunction persists; and revision of previously failed knee
arthroplasty.
4) Natural-Knee II Durasul All-Poly
Patella, Size 3, 7mm The Zimmer Natural-Knee II Durasul Patella product is a
semi-constrained total knee prosthesis consisting of four anatomically
designed components: the femoral, tibial, articular surface, and patellar
prosthesis. The all-polyethylene patella is made from Durasul Highly
Crosslinked Ultra High Molecular Weight Polyethylene. The N-K II Durasul
polyethylene patella components are indicated for patient conditions of
non-inflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis,
osteoarthritis, and inflammatory degenerative joint disease (IJD), e.g.,
rheumatoid arthritis; correctable valgus-varus deformity and moderate flexion
contracture; those patients with failed previous surgery where pain,
deformity, or dysfunction persists; and revision of previously failed knee
arthroplasty.
5) Natural-Knee II Durasul All-Poly
Patella, Size 0, 10mm The Zimmer Natural-Knee II Durasul Patella product is a
semi-constrained total knee prosthesis consisting of four anatomically
designed components: the femoral, tibial, articular surface, and patellar
prosthesis. The all-polyethylene patella is made from Durasul Highly
Crosslinked Ultra High Molecular Weight Polyethylene. The N-K II Durasul
polyethylene patella components are indicated for patient conditions of
non-inflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis,
osteoarthritis, and inflammatory degenerative joint disease (IJD), e.g.,
rheumatoid arthritis; correctable valgus-varus deformity and moderate flexion
contracture; those patients with failed previous surgery where pain,
deformity, or dysfunction persists; and revision of previously failed knee
arthroplasty.
6) Natural-Knee II Durasul All-Poly
Patella, Size 1, 10mm The Zimmer Natural-Knee II Durasul Patella product is a
semi-constrained total knee prosthesis consisting of four anatomically
designed components: the femoral, tibial, articular surface, and patellar
prosthesis. The all-polyethylene patella is made from Durasul Highly
Crosslinked Ultra High Molecular Weight Polyethylene. The N-K II Durasul
polyethylene patella components are indicated for patient conditions of
non-inflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis,
osteoarthritis, and inflammatory degenerative joint disease (IJD), e.g.,
rheumatoid arthritis; correctable valgus-varus deformity and moderate flexion
contracture; those patients with failed previous surgery where pain,
deformity, or dysfunction persists; and revision of previously failed knee
arthroplasty.
7) Natural-Knee II Durasul All-Poly
Patella, Size 2, 10mm The Zimmer Natural-Knee II Durasul Patella product is a
semi-constrained total knee prosthesis consisting of four anatomically
designed components: the femoral, tibial, articular surface, and patellar
prosthesis. The all-polyethylene patella is made from Durasul Highly
Crosslinked Ultra High Molecular Weight Polyethylene. The N-K II Durasul
polyethylene patella components are indicated for patient conditions of
non-inflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis,
osteoarthritis, and inflammatory degenerative joint disease (IJD), e.g.,
rheumatoid arthritis; correctable valgus-varus deformity and moderate flexion
contracture; those patients with failed previous surgery where pain,
deformity, or dysfunction persists; and revision of previously failed knee
arthroplasty.
8) Natural-Knee II Durasul All-Poly
Patella, Size 3, 10mm The Zimmer Natural-Knee II Durasul Patella product is a
semi-constrained total knee prosthesis consisting of four anatomically
designed components: the femoral, tibial, articular surface, and patellar
prosthesis. The all-polyethylene patella is made from Durasul Highly
Crosslinked Ultra High Molecular Weight Polyethylene. The N-K II Durasul
polyethylene patella components are indicated for patient conditions of
non-inflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis,
osteoarthritis, and inflammatory degenerative joint disease (IJD), e.g.,
rheumatoid arthritis; correctable valgus-varus deformity and moderate flexion
contracture; those patients with failed previous surgery where pain,
deformity, or dysfunction persists; and revision of previously failed knee
arthroplasty.
CODE
1) 6301-07-100; all lots;
2) 6301-07-101; all lots;
3) 6301-07-102; all lots;
4) 6301-07-103; all lots;
5) 6301-10-104; all lots;
6) 6301-10-105; all lots;
7) 6301-10-106; all lots;
8) 6301-10-107; all lots
RECALLING FIRM/MANUFACTURER
Zimmer Inc., Warsaw, IN,
REASON The firm has 16 reports of patella peg
shear out of 109,386 distributed units (0.0137%). The failures have been
reported where the pegs on the back side of the patella have sheared at the
peg/patella interface. Analysis shows that the highly cross-linked
polyethylene, Durasul, patella was shown to have a lower fracture fatigue
strength than the Sulene counterpart and that the cement technique used by the
surgeon can lead to insufficient fixation where it reduces the fatigue
strength by 50-60%. Design differences as compared to other all-poly patella
designs including a smaller cement groove contact area, the shape of the edge
of the cement grooves, and smaller pegs affect the fracture fatigue strength.
Also, it is hypothesized that a free-handed unlevel resection of the patella
bone could result in residual stresses that reduce the fixation integrity of
the patella implant.
VOLUME OF PRODUCT IN COMMERCE
111,454 units
DISTRIBUTION Nationwide and Internationally
PRODUCT Siemens AXIOM Aristos FX Plus solid
state x-ray imager. Image-intensified fluoroscopic x-ray system.
CODE Serial number 1273.
RECALLING FIRM/MANUFACTURER
Recalling Firm: Siemens Medical
Solutions USA, Inc., Malvern, PA,
Manufacturer: Siemens AG, Medical
Solutions, Business Unit XP, Erlangen, Germany. .
REASON Fixation bolts for the mounting of the
X-ray tube holder became mixed with an incorrect bolt type
VOLUME OF PRODUCT IN COMMERCE
1 unit
PRODUCT Stryker PCD Precision System 13G Needle
with Spineplex, sterile, single use, disposable , 4 units per box. units per
box. Part Number: 0505-585-000 Match Ground Four facet Tip Introduction Needle
- PIN 0306-130-011 13G Bevel Tip Stylet - PIN 0306-131-012 The intended use of
the Stryker PCD System is for mixing bone cement and delivering it
percutaneously. Bone cement is mixed in the chamber then transferred directly
into an extension nozzle which has been attached to the introduction needle in
the fracture site by a luer lock fitting. Cement is then dispensed directly
into the fracture site under fluoroscopy.
CODE Lot: 11223012 13G
RECALLING FIRM/MANUFACTURER
Stryker Puerto Rico, Arroyo, PR,
REASON Product was packaged with a four inch
needle in stead of the correct five inch needle. There is a potential that the
needle may not be able to reach the fracture site, which may require
additional medical intervention. In addition, if the stylet needs to be
changed and the additional stylet that comes in the package is used, this
additional stylet is one inch longer than the needle. which could lead to a
more serious injury and require additional medical intervention.
VOLUME OF PRODUCT IN COMMERCE
46 boxes (184units)
PRODUCT
1) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen --- Max CT Flow Rate: 4 mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12100805 --- Made in the USA --- indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
2) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen with REVERSE TAPER --- Max CT Flow Rate: 4mL/SEC ---
Injector Pressure Limit: 300 PSI --- Catalog No. [REF] 12102604 --- Made in
the USA --- indicated for short or long term peripheral access to the
central venous system for intravenous therapy and power injections of contrast
media.
3) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen with REVERSE TAPER --- Max CT Flow Rate: 4mL/SEC ---
Injector Pressure Limit: 300 PSI --- Catalog No. [REF] 12102605 --- Made in
the USA --- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
4) AngioDynamics Morpheus Smart PICC CT
65cm 6F Dual Lumen with REVERSE TAPER --- Max CT Flow Rate: 6mL/SEC ---
Injector Pressure Limit: 300 PSI --- Catalog No. [REF] 12102606 --- Made in
the USA --- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
5) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen with REVERSE TAPER --- Max CT Flow Rate: 4mL/SEC ---
Injector Pressure Limit: 300 PSI --- Catalog No. [REF] 12102611 --- Made in
the USA --- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
6) AngioDynamics Morpheus Smart PICC CT
65cm 6F Dual Lumen with REVERSE TAPER --- Max CT Flow Rate: 6mL/SEC ---
Injector Pressure Limit: 300 PSI --- Catalog No. [REF] 12102613 --- Made in
the USA--- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
7) AngioDynamics Morpheus Smart PICC CT
55cm 5F Dual Lumen with REVERSE TAPER --- Max CT Flow Rate: 5mL/SEC ---
Injector Pressure Limit: 300 PSI --- Catalog No. [REF] 12102616 --- Made in
the USA --- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
8) AngioDynamics Morpheus Smart PICC CT
55cm 6F Dual Lumen with REVERSE TAPER --- Max CT Flow Rate: 5mL/SEC ---
Injector Pressure Limit: 300 PSI --- Catalog No. [REF] 12102617 --- Made in
the USA --- indicated for short or long term peripheral access to the
central venous system for intravenous therapy and power injections of contrast
media.
9) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen --- Max CT Flow Rate: 4mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12102805 --- Made in the USA --- indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
10) AngioDynamics Morpheus Smart PICC CT
65cm 6F Dual Lumen --- Max CT Flow Rate: 6mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12102806 --- Made in the USA --- indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
11) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen --- Max CT Flow Rate: 4mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12102812 --- Made in the USA --- indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
12) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen --- Max CT Flow Rate: 4mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12102813 --- Made in the USA --- indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
13) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen --- Max CT Flow Rate: 4mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12102821 --- Made in the USA --- indicated
for short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
14) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen --- Max CT Flow Rate: 4mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12102829 --- Made in the USA --- indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
15) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen with REVERSE TAPER --- Max CT Flow Rate: 4mL/SEC ---
Injector Pressure Limit: 300 PSI --- Catalog No. [REF] 12104605 --- Made in
the USA --- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
16) AngioDynamics Morpheus Smart PICC CT
65cm 6F Dual Lumen with REVERSE TAPER --- Max CT Flow Rate: 6mL/SEC ---
Injector Pressure Limit: 300 PSI --- Catalog No. [REF] 12104606 --- Made in
the USA --- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
17) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen with REVERSE TAPER --- Max CT Flow Rate: 4mL/SEC ---
Injector Pressure Limit: 300 PSI --- Catalog No. [REF] 12104612 --- Made in
the USA --- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
18) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen --- Max CT Flow Rate: 4mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12104805 --- Made in the USA --- indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
19) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen --- Max CT Flow Rate: 4mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12104813 --- Made in the USA --- indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
20) AngioDynamics Morpheus Smart PICC CT
65cm 6F Dual Lumen --- Max CT Flow Rate: 6mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12104814 --- Made in the USA --- indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
21) AngioDynamics Morpheus Smart PICC CT
65cm 5F Dual Lumen --- Max CT Flow Rate: 4mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12104821 --- Made in the USA --- indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
22) AngioDynamics Morpheus Smart PICC CT
65cm 6F Dual Lumen --- Max CT Flow Rate: 6mL/SEC --- Injector Pressure Limit:
300 PSI --- Catalog No. [REF] 12104830 --- Made in the USA --- indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
23) AngioDynamics Morpheus Smart PICC CT
5F Dual Lumen 65cm --- Max CT Flow Rate: 4mL/SEC --- Catalog No.: 12110904 ---
Assembled and Packaged in the USA. indicated for short or long term peripheral
access to the central venous system for intravenous therapy and power
injections of contrast media.
24) AngioDynamics Morpheus Smart PICC CT
5F Dual Lumen w/ Reverse Taper 65cm --- Max CT Flow Rate: 4mL/sec --- Catalog
No.: 12110917 --- Assembled and Packaged in the USA. indicated for short or
long term peripheral access to the central venous system for intravenous
therapy and power injections of contrast media.
25) AngioDynamics Morpheus Smart PICC CT
6F Dual Lumen w/ Reverse Taper 65cm --- Max CT Flow Rate: 6mL/sec --- Catalog
No.: 12110918 --- Assembled and Packaged in the USA. indicated for short or
long term peripheral access to the central venous system for intravenous
therapy and power injections of contrast media.
26) AngioDynamics Morpheus Smart PICC CT
5F Dual Lumen w/ Reverse Taper 65cm --- Max CT Flow Rate: 4mL/sec --- Catalog
No.: 12110919 --- Assembled and Packaged in the USA. indicated for short or
long term peripheral access to the central venous system for intravenous
therapy and power injections of contrast media. ;
27) AngioDynamics Morpheus Smart PICC CT
with Stylet 5F Dual Lumen 65cm --- Max CT Flow Rate: 4mL/sec --- Catalog No.:
12110923 --- Assembled and Packaged in the USA. indicated for short or long
term peripheral access to the central venous system for intravenous therapy
and power injections of contrast media.
28) AngioDynamics Morpheus Smart PICC CT
with Stylet 5F Dual Lumen w/ Reverse Taper 65cm --- Max CT Flow Rate: 4mL/sec
--- Catalog No.: 12110929 --- Assembled and Packaged in the USA. indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
29) AngioDynamics Morpheus Smart PICC CT
5F Dual Lumen w/ Reverse Taper 65cm --- Max CT Flow Rate: 4mL/sec --- Catalog
No.: 12110934 --- Assembled and Packaged in the USA. indicated for short
or long term peripheral access to the central venous system for intravenous
therapy and power injections of contrast media. ;
30) AngioDynamics Morpheus Smart PICC CT
5F Dual Lumen w/ Reverse Taper 55cm --- Max CT Flow Rate: 5mL/sec --- Catalog
No.: 12110936 --- Assembled and Packaged in the USA. indicated for short or
long term peripheral access to the central venous system for intravenous
therapy and power injections of contrast media.
31) AngioDynamics Morpheus Smart PICC CT
with Stylet 5F Dual Lumen w/ Reverse Taper 55cm --- Max CT Flow Rate: 5mL/sec
--- Catalog No.: 12110937 --- Assembled and Packaged in the USA. indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
32) AngioDynamics Morpheus Smart PICC CT
6F Triple Lumen w/ Reverse Taper 55cm --- Max CT Flow Rate: 5mL/sec ---
Catalog No.: 12110941 --- Assembled and Packaged in the USA. indicated for
short or long term peripheral access to the central venous system for
intravenous therapy and power injections of contrast media.
33) AngioDynamics Special Order Morpheus
Smart PICC CT 65cm 5F Dual Lumen with Reverse Taper --- Max CT Flow Rate:
4mL/sec --- Injector Pressure Limit --- Catalog No.: 52100602 --- Made in the
USA --- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
34) AngioDynamics Special Order Morpheus
Smart PICC CT 65cm 5F Dual Lumen with Reverse Taper --- Max CT Flow Rate:
4mL/sec --- Injector Pressure Limit --- Catalog No.: 52102618 --- Made in the
USA --- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
35) AngioDynamics Special Order Morpheus
Smart PICC CT 65cm 6F Dual Lumen with Reverse Taper --- Max CT Flow Rate:
6mL/sec --- Injector Pressure Limit --- Catalog No.: 52102624 --- Made in the
USA--- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
36) AngioDynamics Special Order Morpheus
Smart PICC CT 65cm 5F Dual Lumen with Reverse Taper --- Max CT Flow Rate:
4mL/sec --- Injector Pressure Limit --- Catalog No.: 52102638--- Made in the
USA --- indicated for short or long term peripheral access to the central
venous system for intravenous therapy and power injections of contrast media.
CODE
1) FG Lot Numbers: 553481, 554552;
2) FG Lot Number: 552392, 554557;
3) FG Lot Number: 551897, 552393,
552394, 553196, 553197, 554085, 554558, 554559;
4) FG Lot Number: 550962, 551912,
552395, 555170;
5) FG Lot Number: 551899, 553211,
553482, 55408;
6) FG Lot Number: 554561;
7) FG Lot Number: 553483;
8) FG Lot Number: 555930;
9) FG Lot Number: 551900, 552398,
552399, 553200, 554077, 554563;
10) FG Lot Number: 553486;
11) FG Lot Number: 553480, 554088;
12) FG Lot Number: 552440;
13) FG Lot Number: 553193;
14) FG Lot Number: 553487;
15) FG Lot Number: 552444, 553205,
553488, 554079;
16) FG Lot Number: 552445;
17) FG Lot Number: 553948, 554090;
18) FG Lot Number: 55244;
19) FG Lot Number: 553213;
20) FG Lot Number: 552450, 553208,
554091, 557219;
21) FG Lot Number: 553949;
22) FG Lot Number: 554081;
23) FG Lot Number: 551915, 553495;
24) FG Lot Number: 552452, 553216;
25) FG Lot Number: 552454, 554541,
554576;
26) FG Lot Number: 551529, 553405,
554573;
27) FG Lot Number: 552455, 553217;
28) FG Lot Number: 551762, 552457;
29) FG Lot Number: 552459, 551536,
553499;
30) FG Lot Number: 553333, 553760,
558278;
31) FG Lot Number: 553761;
32) FG Lot Number: 553218, 554092;
33) FG Lot Number: 553407, 554413;
34) FG Lot Number: 554082, 554554;
35) FG Lot Number: 552387, 553194;
36) FG Lot Number: 552388, 553291,
554555, 555166
RECALLING FIRM/MANUFACTURER
Recalling Firm: Angiodynamics Worldwide
Headquarters, Latham, NY, .
Manufacturer: Angiodynamics, Inc.,
Queensbury, NY.
REASON The red luer on the Morpheus CT PICC
(Peripherally Inserted Central Catheter) has the potential to crack during use
of the product.
VOLUME OF PRODUCT IN COMMERCE
10,392 units
PRODUCT FloWire Doppler Guide wires. The Volcano
FloWire Doppler guide wire is a single use guide wire intended for use in all
blood vessels including both coronary and peripheral arteries to measure blood
flow velocities during diagnostics angiography and/or interventional
procedures. Part number 1400.
CODE
Serial numbers 011-0318-002;
011-01323-012; 011-01319-001; 011-01319-002, 011-01319-005, 011-01319-006;
011-01319-007; 011-01318-001, 011-01321-003, 011-01321-006; 011-01321-008;
011-01321-009; 011-01322-001, 011-01323-010
RECALLING FIRM/MANUFACTURER
Volcano Corp., Rancho Cordova, CA,
REASON Reports of no signal displayed on the
monitor when FloWire Doppler Guide wires were connected to the Combomap.
VOLUME OF PRODUCT IN COMMERCE
14 units
DISTRIBUTION Nationwide, Japan and Europe
PRODUCT GEMINI TF 16 PET/CT System & GEMINI
TF 64 PET/CT System. The device is a diagnostic imaging system for fixed or
mobile installations that combines Positron Emission Tomography (PET) and
X-ray Computed Tomography (CT) systems. The CT subsystem produces cross
sectional images of the body by computer reconstruction of x-ray transmission
data. The PET subsystem produces images of the distribution of PET
radiopharmaceuticals in the patient body (specific radiopharmaceuticals are
used for whole body, head, heart and other organ imaging). Attenuation
correction is accomplished by CTAC. The device also provides for list mode,
dynamic, and gated acquisitions. Image processing and display workstations
provide software applications to process, analyze, display, quantify and
interpret medical images/data. The PET and CT images may be registered and
displayed in a "fused" (overlaid in the same spatial orientation)
format to provide combined metabolic and anatomical data at different angles.
Trained professionals use the images in: -The evaluation, detection and
diagnosis of lesions, disease and organ function such as but not limited to
cancer, cardiovascular disease, and neurological disorders. -The detection,
localization, and staging of tumors and diagnosing cancer patients. -Treatment
planning and interventional radiology procedures. Cardiac imaging software
provides functionality for the quantification of cardiology images and
datasets including but not limited to myocardial perfusion for the display of
wall motion and quantification of left-ventricular function parameters from
gated myocardial perfusion studies and for the 3D alignment of coronary artery
images from CT coronary angiography onto the myocardium.
CODE Serial Number 7224; Serial Number 7217
RECALLING FIRM/MANUFACTURER
Philips Medical Systems (Cleveland)
Inc., Cleveland, OH,
REASON Philips became aware of an issue
relating to a UPS cabinet that contained a melted battery which caused damage
to other batteries lower in the cabinet, leading to battery acid leaking on
the floor.
VOLUME OF PRODUCT IN COMMERCE
2 units
PRODUCT Broselow/Hinkle Pediatric Emergency
System. The system supplies you with all the information and equipment needed
for a young patients' emergency resuscitation requirements, all contained in a
color coded, easy-to-use kit. Class 1 Exempt Device No Medical Device Listing
Number exp date - 13 months from the date received by the distributor.
Povidone Iodine Pads: The Povidone Iodine Pad is packaged in the Busse IV
Start Kit that is included in the Intra Venous Module, which is contained in
the Broselow products referenced below. Affected Armstrong product numbers:
Kits: AE-4700, AE 4701, AE-4712, and PBL-PC-9A Modules: 7700RIV, 7700PIV,
7700YIV, 7700WIV, 7700BIV, 7700OIV, 7700GIV, 7730RED, 7730PUR, 7730YEL,
7730WHI, 7730BLU, 7730ORG, and 7730GRN Affected Vital Signs product numbers:
Kits: 7730ALS, 7730IALS, 7730FLY, 7730MOD Modules: 7700RIV2, 7700PIV2,
7700YIV2, 7700WIV2, 7700BIV2, 7700OIV2, 7700GIV2, 7730RED5, 7730PUR5,
7730YEL5, 7730WHI5, 7730BLU5, 7730ORG5, and 7730GRN5 Sub-Assembly: 7700RIV,
7700PIV, 7700YIV, 7700WIV, 7700BIV, 7700OIV, and 7700GIV.
CODE Lot codes: 2008: 228F-366F 2009:
010K-258K and 010L-258L
RECALLING FIRM/MANUFACTURER
Vital Signs Devices, a GE Healthcare
Co., Totowa, NJ,
REASON Povidone Iodine Pads included in the
Vital Signs Broselow Pediatric Emergency Kit may be contaminated with
Elizabethkinga meningoseptica.
VOLUME OF PRODUCT IN COMMERCE
6675 units with 20166 pads in kits
PRODUCT Foot Rest Kit
(pN 6252700-003) for
Models 6251 and 6252 Stryker (EMS Stair -PRO), OE 2065939, 1739631, 2066066,
6066803 and 20666941 with optional. Intended to provide the patient greater
comfort and a sense of security.
CODE None
RECALLING FIRM/MANUFACTURER
Stryker Medical Division of Stryker
Corp., Portage, MI,
REASON The Foot Rest Kit may have the wrong
pivot spacers making the joint more susceptible to relaxation and prone to
fatigue. There is a moderate risk of injury to the patient should the handles
become detached during use.
VOLUME OF PRODUCT IN COMMERCE
3 kits
PRODUCT
1) Blanketrol II 48C Hyper-Hypothermia
System, Model Numbers: 222 and Model Number: 222R. The Blanketrol II 48C
System provides a flow of temperature-controlled water to a heat exchanger.
The device, in conjunction with a heat exchanger, is used to warm/cool the
fluid that enters and exits a heat exchanger. The conductive heat/cool
transfer is accomplished through any type of blood oxygenator/heat exchanger
unit. The operator of the unit regulates the desired fluid temperature.
Frequent observation of the actual blood/fluid temperature and body
temperature is required. The Blanketrol II 48 degree C does not contact the
patient or the fluids; the water being circulated by the Hemotherm is
separated from the fluid by the heat exchanger. To active temperatures above
42 Degrees C, the user must follow explicit instructions in to the owner's
manual in order to activate the higher setting. Without following the explicit
instructions, the device can only operate within the normal maximum
temperature range of: 42 degrees C. The Blanketrol II 48 Degree C System can
be used with adult and pediatric patients for extracorporeal circulation of
the patient's limb, entire body, or a chemo medication lavage of the abdominal
area. The 48C Blanketrol device is used to lower/raise the blood/fluid
temperature and/or maintain blood/fluid temperature, as required, through
conductive heat transfer. The conductive heat transfer is accomplished through
any type of Blood Oxygenator/Heat Exchanger unit. This device is intended to
hyperthermic treatment of AIDS and cancer patients and associated research.
The device can be set to reach an upper temperature of: 48 degrees C.
2) Hemotherm 48C Dual Reservoir
Heater/Cooler System, Model Number: 400 and Model Number: 400MR. The 48 o (
Degree) C Hemotherm Dual Reservoir Heater/Cooler I provides a flow of
temperature-controlled water to a heat exchanger. The device, in conjunction
with a heat exchanger, is used to warm/cool the fluid that enters and exits a
heat exchanger. The conductive heat/cool transfer is accomplished through any
type of blood oxygenator/heat exchanger unit. The operator of the unit
regulates the desired fluid temperature. Frequent observation of the actual
blood/fluid temperature and body temperature is required. The Hemotherm does
not contact the patient of the fluids; the water being circulated by the
Hemotherm is separated from the fluid by the heat exchanger. To active
temperatures above 42 Degrees C, the user must follow explicit instructions in
to the owner's manual in order to activate the higher setting. Without
following the explicit instructions, the device can only operate within the
normal maximum temperature range of: 42 degrees C. The Hemotherm can be used
with adult and pediatric patients for extracorporeal circulation of the
patient's limb, entire body, or a chemo medication lavage of the abdominal
area. The Hemotherm 48C device system is used to lower/raise the blood/fluid
temperature and/or maintain blood/fluid temperature, as required, through
conductive heat transfer. The conductive heat transfer is accomplished through
any type of Blood Oxygenator/Heat Exchanger unit. This device is intended to
hyperthermic treatment of AIDS and cancer patients and associated research.
The device can be set to reach an upper temperature of: 48C.
CODE
1) Serial Numbers: 902-8645, 902-8646,
911-9426, 914-10001, 914-10140, 914-10177, 943-12074, 992-14052, 994-14290,
994-14291, 002-14541, 002-14545, 012-15024, 014-15260, 022-15490, 074-18981,
962-12825, and 874-6277;
2) Serial Numbers: 862-8586M, 882-9143M,
882-9163M, 914-9837M, 952-10505M, 973-10918M, 992-11343M, 002-11540M,
003-11585M, 003-11586M, 003-11592M, 003-11600M, 004-11830M, 004-11631M,
014-11807M, 0311-12041M, 042-12250M, 071-12654M, 071-12655M, 073-12117M,
972-10872M, 062-12536M, and 963-10737M.
RECALLING FIRM/MANUFACTURER
Cincinnati Sub-Zero Products Inc.,
Cincinnati, OH,
REASON The suspect hyper-hypothermia device
systems sold to physicians for research and/or the treatment of AIDS and
Cancer patients, were not appropriately cleared by the FDA prior to
distribution.
VOLUME OF PRODUCT IN COMMERCE
20-Hemotherm 48C, 400M &400MR
systems; and 16 Blanketrol II Model 222 & 222R 48C System units were
distributed in U.S (rest shipped overseas).
PRODUCT Siemens CLINITEST hCG -Pregnancy Test
(qualitative), Radio immunoassay,Human Chorionic Gonadotropin (hCG) test
system. Package of 25. Product Code 1760/ Catalog# 06484105). For in vitro
diagnostic use as a qualitative method in the rapid detection of human
chorionic gonadotropin (hCG) in urine specimens.
CODE
Lot Expiration 028619, 6/8/2012; 028924,
7/5/2012; 029997, 10/7/2012; 030240, 11/6/2012; 030431, 11/24/2012; 030602,
12/8/2012; 030826, 12/24/2012; 030998, 1/11/2013; 031332, 2/4/2013; 031460,
2/19/2013; 031662, 3/6/2013; 031798, 3/20/2013; 031924, 3/30/2013; 032180,
4/28/2013; and 032106, 4/19/2013
RECALLING FIRM/MANUFACTURER
Recalling Firm: Siemens Healthcare
Diagnostics, Tarrytown, NY,
Manufacturer: Alfa Scientific Designs
Inc., Poway, CA. .
REASON A potential for false positive or
increase in borderline results with patients and commercially available
negative control material.
VOLUME OF PRODUCT IN COMMERCE
48,578 Kits (25 tests / kit)
DISTRIBUTION Nationwide and Internationally
PRODUCT
1) AU Clinical Chemistry Analyzers:
Model Number: AU400 Clinical Chemistry Analyzer, AU400(e) Clinical Chemistry
Analyzer, AU480 Clinical Chemistry Analyzer. The AU Clinical Chemistry
Analyzers are fully automated photometric analyzers intended for clinical
laboratory use. Applications include colorimetric, turbidimetric, latex
agglutination, and homogeneous enzyme immunoassay. Recall # Z-0974-2012;
2) AU Clinical Chemistry Analyzers:
Model Number: AU600 Clinical Chemistry Analyzer, AU640 Clinical Chemistry
Analyzer, AU640(e) Clinical Chemistry Analyzer, AU680 Clinical Chemistry
Analyzer. The AU Clinical Chemistry Analyzers are fully automated photometric
analyzers intended for clinical laboratory use. Applications include
colorimetric, turbidimetric, latex agglutination, and homogeneous enzyme
immunoassay.
3) AU Clinical Chemistry Analyzers:
Model Number: AU2700 Clinical Chemistry Analyzer . The AU Clinical Chemistry
Analyzers are fully automated photometric analyzers intended for clinical
laboratory use. Applications include colorimetric, turbidimetric, latex
agglutination, and homogeneous enzyme immunoassay. ;
4) AU Clinical Chemistry Analyzers:
Model Number: AU5400 Clinical Chemistry Analyzer. The AU Clinical Chemistry
Analyzers are fully automated photometric analyzers intended for clinical
laboratory use. Applications include colorimetric, turbidimetric, latex
agglutination, and homogeneous enzyme immunoassay.
CODE All analyzer serial numbers.
RECALLING FIRM/MANUFACTURER
Recalling Firm: Beckman Coulter Inc.,
Brea, CA,
Manufacturer: Beckman Coulter Kk,
Sunto-gun,
Japan.
REASON The AU Clinical Chemistry Analyzers may
experience cuvette overflow.
VOLUME OF PRODUCT IN COMMERCE
8,225 units
DISTRIBUTION Nationwide and Internationally
PRODUCT Terumo, Sarns Sternal Saw I System,
Catalog 6090. The Sarns Sternal Saw I is indicated for use in splitting the
sternum during thoracic surgery.
CODE Serial Numbers 0 - 10186
RECALLING FIRM/MANUFACTURER
Terumo Cardiovascular Systems Corp, Ann
Arbor, MI,
REASON Terumo Cardiovascular Systems has
received one report of minute metal shavings leaking from the Sarns Sternal
Saw I System at the point where the saw blade is inserted into the saw
handpiece.
VOLUME OF PRODUCT IN COMMERCE
10187
DISTRIBUTION Nationwide and Internationally
PRODUCT
1) EZ Steer ThermoCool Nav Part Number
D-1292-01-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ
STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter is a
multi-electrode luminal catheter with a deflectable tip designed to facilitate
electrophysiological mapping of the heart and to transmit radiofrequency (RF)
current to the catheter tip electrode for ablation purposes. The catheter
shaft measures 7.5 F with 8 F ring electrodes. For ablation, the catheter is
used in conjunction with an RF generator and a dispersive pad (indifferent
electrode) for the treatment of: - Type I atrial flutter in patients age 18 or
older. - Recurrent drug/device refractory sustained monomorphic ventricular
tachycardia (VT) due to prior myocardial infarction (Mil in adults. - Drug
refractory recurrent symptomatic paroxysmal atrial fibrillation, when used
with compatible three-dimensional electroanatomic mapping systems. The EZ
STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter provides
location information when used with CARTO~ EP Navigation Systems.
2) EZ Steer ThermoCool Nav Part Number
D-1292-02-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ
STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter is a
multi-electrode luminal catheter with a deflectable tip designed to facilitate
electrophysiological mapping of the heart and to transmit radiofrequency (RF)
current to the catheter tip electrode for ablation purposes. The catheter
shaft measures 7.5 F with 8 F ring electrodes. For ablation, the catheter is
used in conjunction with an RF generator and a dispersive pad (indifferent
electrode) for the treatment of: - Type I atrial flutter in patients age 18 or
older. - Recurrent drug/device refractory sustained monomorphic ventricular
tachycardia (VT) due to prior myocardial infarction (Mil in adults. - Drug
refractory recurrent symptomatic paroxysmal atrial fibrillation, when used
with compatible three-dimensional electroanatomic mapping systems. The EZ
STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter provides
location information when used with CARTO~ EP Navigation Systems.
3) EZ Steer ThermoCool Nav Part Number
D-1292-03-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ
STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter is a
multi-electrode luminal catheter with a deflectable tip designed to facilitate
electrophysiological mapping of the heart and to transmit radiofrequency (RF)
current to the catheter tip electrode for ablation purposes. The catheter
shaft measures 7.5 F with 8 F ring electrodes. For ablation, the catheter is
used in conjunction with an RF generator and a dispersive pad (indifferent
electrode) for the treatment of: - Type I atrial flutter in patients age 18 or
older. - Recurrent drug/device refractory sustained monomorphic ventricular
tachycardia (VT) due to prior myocardial infarction (Mil in adults. - Drug
refractory recurrent symptomatic paroxysmal atrial fibrillation, when used
with compatible three-dimensional electroanatomic mapping systems. The EZ
STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter provides
location information when used with CARTO~ EP Navigation Systems.
4) EZ Steer ThermoCool Nav Part Number
D-1292-04-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ
STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter is a
multi-electrode luminal catheter with a deflectable tip designed to facilitate
electrophysiological mapping of the heart and to transmit radiofrequency (RF)
current to the catheter tip electrode for ablation purposes. The catheter
shaft measures 7.5 F with 8 F ring electrodes. For ablation, the catheter is
used in conjunction with an RF generator and a dispersive pad (indifferent
electrode) for the treatment of: - Type I atrial flutter in patients age 18 or
older. - Recurrent drug/device refractory sustained monomorphic ventricular
tachycardia (VT) due to prior myocardial infarction (Mil in adults. - Drug
refractory recurrent symptomatic paroxysmal atrial fibrillation, when used
with compatible three-dimensional electroanatomic mapping systems. The EZ
STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter provides
location information when used with CARTO~ EP Navigation Systems.
5) EZ Steer ThermoCool Nav Part Number
D-1292-05-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ
STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter is a
multi-electrode luminal catheter with a deflectable tip designed to facilitate
electrophysiological mapping of the heart and to transmit radiofrequency (RF)
current to the catheter tip electrode for ablation purposes. The catheter
shaft measures 7.5 F with 8 F ring electrodes. For ablation, the catheter is
used in conjunction with an RF generator and a dispersive pad (indifferent
electrode) for the treatment of: - Type I atrial flutter in patients age 18 or
older. - Recurrent drug/device refractory sustained monomorphic ventricular
tachycardia (VT) due to prior myocardial infarction (Mil in adults. - Drug
refractory recurrent symptomatic paroxysmal atrial fibrillation, when used
with compatible three-dimensional electroanatomic mapping systems. The EZ
STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter provides
location information when used with CARTO~ EP Navigation Systems.
CODE
1) Lot Numbers: 14050236, 14069164,
14085850, 14085852, 14089636, 14098511, 14102112, 14138785, 14142301, and
15091243;
2) Lot Numbers: 15214996M, 14070102A,
15177599, 15156690, 15156649, 15123384, 15111287, 15079216, 15076033,
15070702, 14138791, 14117351, 14117345, 14117339, 14113236, 14109201,
14102119, 14102116 14093807, 14091431, 14087720, 14081176, 14077787, 14076743,
14075452, 14071539, 14071538, 14068519, 14066570, 14064255, 14060307,
14058853, 14058117, 14056270, 14056269, 14053165, 14052248, 14052247,
14049472, 14043262, 14040370, 14039651, and 14036714
3) Lot Numbers: 15076036, 15076035,
15059708, 14113238, 14105865, and 13439370;
4) Lot Numbers: 15221536M, 15215004M,
15215001M, 15215000M, 15208579M, 15208577M, 15208574M, 14069262A, 14069168A,
14069084A, 15164269, 15164264, 15159917, 15159912, 15156608, 15148849,
15144223, 15128115, 15128113, 15119884, 15119883, 15116509, 15116506,
15103250, 15098193, 15098191, 15098190, 15095199, 15083389, 15083386,
15080818, 15080815, 15070703, 15067809, 15064960, 15032155, 15014525,
15014522, 15014521, 15010028, 15010027, 15007251, 15007250, 14152525,
14149131, 14149130, 14149129, 14146082, 14142307, 14138855, 14138853,
14138846, 14138795, 14120030, 14117394, 14117388, 14117385, 14117376,
14117371, 14117370, 14117356, 14113251, 14113249, 14109218, 14109216,
14102071, 14102068, 14097083, 14095220, 14093166, 14092574, 14091435,
14089642, 14089641, 14088995, 14087721, 14080304, 14080303, 14080302,
14079405, 14078387, 14078385, 14077789, 14075460, 14075459, 14071639,
14071638, 14071542, 14070108, 14068526, 14067399, 14065778, 14065064,
14065059, 14064262, 14064259, 14064258, 14063230, 14062655, 14062654,
14062652, 14062651, 14062042, 14062040, 14062039, 14062038, 14061249,
14060315, 14060311, 14058856, 14058855, 14057214, 14056275, 14056273,
14056272, 14053738, 14053167, 14052547, 14052249, 14051584, 14043263,
14042518, 14042517, 14038851, 14036716, 14036583, and 14036582;
5) Lot Numbers: 15219359M, 15219356M,
15214930M, 15214908M, 15214902M, 15208638M, 15208634M, 15208631M, 15208628M,
15208626M, 15208619M, 15208612M, 15208607M, 15208593M, 15205050M, 15205047M,
15205031M, 15205018M, 15204883M, 15199820M, 15199819M, 15199811M, 15199801M,
15199794M, 15199792M, 14069263A, 15171668, 15171665, 15171663, 15171652,
15169965, 15169957, 15169946, 15164293, 15164271, 15159938, 15159935,
15159922, 15156563, 15151976, 15148785, 15148772, 15144234, 15144151,
15144134, 15144119, 15136209, 15132418, 15128133, 15128123, 15124171,
15124167, 15123408, 15120034, 15120022, 15116441, 15103256, 15103255,
15098206, 15098200, 15098198, 15098195, 15095224, 15095223, 15095220,
15091259, 15091254, 15083409, 15081335, 15081333, 15076058, 15076049,
15076046, 15071044, 15070717, 15070708, 15067823, 15067817, 15057475,
15040868, 15035767, 15035766, 15032160, 15028721, 15028712, 15028708,
15028701, 15025094, 15021180, 15021179, 15021175, 15021168, 15021164,
15021163, 15021148, 15017983, 15010043, 15010040, 15010033, 15010006,
15009982, 15009980, 15009978, 15009975, 15007267, 15007261, 15007255,
15007254, 14152530, 14149138, 14149136, 14146104, 14146099, 14146095,
14146090, 14146086, 14146084, 14142340, 14142339, 14142335, 14142333,
14142326, 14142325, 14142324, 14142323, 14142322, 14142317, 14142316,
14142314, 14142312, 14142311, 14138878, 14138878, 14138871, 14138866,
14138865, 14138864, 14138860, 14138859, 14120108, 14120105, 14120100,
14120092, 14120078, 14120076, 14120075, 14120067, 14120062, 14120055,
14120015, 14120014, 14120010, 14120006, 14120002, 14120000, 14119986,
14119985, 14119977, 14117438, 14117431, 14117430, 14117429, 14117409,
14117406, 14117404, 14117400, 14117398, 14113274, 14113268, 14113262,
14113257, 14109239, 14109234, 14109231, 14109230, 14109229, 14109228,
14102076, 14099138, 14099135, 14098716, 14098709, 14098706, 14098530,
14098525, 14097835, 14095223, 14090358, 14089643, 14088997, 14088996,
14087736, 14087731, 14085863, 14085143, 14085142, 14085141, 14081912,
14081911, 14081907, 14081809, 14081808, 14081805, 14081804, 14081185,
14080312, 14080310, 14079411, 14078390, 14076755, 14076256, 14076254,
14076253, 14075465, 14075464, 14071644, 14071547, 14068530, 14066590,
14066582, 14065792, 14065065, 14064778, 14064271, 14064266, 14064265,
14063238, 14063236, 14063234, 14062663, 14062660, 14062044, 14061259,
14061256, 14060323, 14060321, 14060320, 14058864, 14058857, 14058125,
14058124, 14058122, 14057221, 14057220, 14053745, 14053740, 14053739,
14053728, 14053724, 14053709, 14053168, 14052551, 14050933, 14049476,
14043896, 14042520, 14039654, 14038050, 14029870, 14029043, 14028071,
14027237, 14025878, 14025127, 14019684, 14016117, and 13435854
RECALLING FIRM/MANUFACTURER
Biosense Webster, Inc., Irwindale, CA,
REASON Biosense Webster, Inc. has recently
become aware of the omission of the following two Warning and Precaution
statements on the Instructions For Use (IFU) for the above catheter: 1. If
phrenic nerve location is a concern, precautionary measures are recommended to
evaluate the proximity of the nerve to the ablation electrode, such as pacing
maneuvers. 2. To minimize risk of atrio-esophageal fistula, precautionary
measures should be taken when ablating on posterior wall of the left atrium in
proximity to the esophagus. This issue affects only the IFU for the EZ Steer
ThermoCool NAV Diagnostic/Ablation Catheter, not the catheter itself. Please
note that these warnings and precautions are already included in the NaviStar
ThermoCool Catheter IFU. Both of these potential complications, and procedural
mitigations for avoiding them, are well documented in clinical and scientific
literature.
VOLUME OF PRODUCT IN COMMERCE
24,713 units
PRODUCT VITROS Valproic Acid
(VALP) Reagent,
REF/Catalog No. 6801710. An in vitro diagnostic. For in vitro diagnostic use
only. Used to quantitatively measure valproic acid (VALP) concentration in
human serum and plasma.
CODE
Reagent Lots from Generation (GEN) 14:
1511-14-9637 (exp. 6/30/11), 1511-14-9776 (exp. 9/30/11), 1511-14-9894 (exp.
11/30/11), 1511-14-1094 (exp. 1/31/12), and 1511-14-1230 (exp. 4/30/12)
RECALLING FIRM/MANUFACTURER
Ortho-Clinical Diagnostics, Rochester,
NY,
REASON Results using VITROS Chemistry Products
VALP Reagent were negatively biased compared to other manufacturers. OCD
confirmed that patient results were also affected by this issue.
VOLUME OF PRODUCT IN COMMERCE
Approximately 5836 units
DISTRIBUTION Nationwide and Internationally
PRODUCT Integra OS Osteoconductive Scaffold
Putty (a 2.5 cc Part numbers - IOS10125; b) 5 cc Part numbers - lOS10155).
Integra OS Putty, combined with bone marrow aspirate, is intended for use as a
bone void filler to fill voids or gaps of the skeletal system. Following
placement in the bony void or gap (defect), Integra OS Putty is resorbed and
replaced with bone during the healing process.
CODE a) Lot numbers - N199416 and 215944; b)
Lot numbers - 208500, 206504 and N216026
RECALLING FIRM/MANUFACTURER
Integra LifeSciences Corp, Plainsboro,
NJ,
REASON Integra OS Osteoconductive Putty is
labeled with incorrect storage conditions.
VOLUME OF PRODUCT IN COMMERCE
45 units
PRODUCT Gamma camera systems: AXIS SPECT Systems
& IRIX SPECT Systems, Model #: AXIS- 882130, IRIX- 882140. Philips
Healthcare. Used to do nuclear medicine diagnostic imaging of various body
organs.
CODE
System Code # 882130, Serial # 10, 97,
101, 103, 104, 107-112, 116, 117, 119-121, 123, 125, 127, 130, 132, 133, 135,
138, 142-144, 149-151, 153-158, 162-164, 166, 169-173, 180, 181, 185, 186,
189, 193-204, 214, 215, 218, 219, 221-223, 227, 229-231, 233-237, 240, 242,
244-246, 249-251, 260-262, 264, 267, 272, 273, 275, 277, 280, 281, 283, 285,
291-293, 295, 297-300, 305, 308-310, 312, 316, 319-322, 326-332, 334-336,
338-340, 346, 347, 349, 351, 352, 354, 355, 358-363, 368-370, 374, 377-379,
385, 387, 388, 390-396, 401-407, 411, 412, 416-419, 426-432, 434, 437, 438,
442, 443, 445-447, 449, 452, 457-462, 464, 465, 468, 471, 474, 480-485, 487,
490-493, 495, 496, 499, 502, 503, 505, 508, 509, 511, 513-518, 520, 521, 523,
529, 530, 534, 543-546, 548-560, 562, 563, 569-572, 574-576, 580, 582, 583,
585, 586, 590, 594, 595, 597, 602-607, 611, 612, 618-621, 624-628, 630, 635,
636, 640, 641, 647-650, 654, 656-658, 660-665, 668-673, 675, 679, 681-684,
686, 687, 690-695, 697-699, 701-708, 711-719, 723-729, 733-736, 738, 739,
743-746, 749-753, 755, 756, 759, 760, 762, 764-767, 771-773, 776-778, 780,
781, 786, 789, 790-795, 797, 799, 801, 803-806, 808, 809, 811-814, 818,
820-822, 825-827, 831-834, 836, 837, 839, 840, 841, 843-845, 848, 849,
851-855, 857, 858, 860-862, 865-869, 872-874, 876-881, 884-886, 889, 893, 894,
900-903, 907-922, 924, 927-930, 932, 934, 938-941, 943, 944, 946-948, 951,
953, 954, 958-963, 965-967, 970, 972-980, 982, 983, 985, 986, 991-994,
996-1000, 1003-1008, 1010-1013, 1016, 1017, 1019, 1022, 1023, 2007, 2008,
211037, 2008-0068, 555A, & 778A. System Code #882140, Serial # 102, 134,
141, 146, 147, 159, 160, 161, 167, 175, 176, 190, 209, 210, 216, 225, 226,
257, 259, 268-270, 274, 289, 290, 302, 304, 314, 315, 318, 323, 333, 341, 343,
344, 348, 350, 364, 366, 381, 386, 409, 414, 420, 421, 423, 436, 441, 455,
463, 469, 472, 473, 488, 489, 497, 498, 501, 531, 532, 536-539, 565, 566, 578,
588, 589, 591, 596, 598-600, 608, 610, 613, 615, 616, 622, 631, 632, 634, 637,
653, 666, 674, 676-678, 685, 689, 700, 709, 710, 731, 732, 741, 742, 748, 757,
763, 768, 774, 785, 802, 810, 816, 819, 824, 828, 829, 830, 838, 846, 856,
863, 864, 870, 871, 875, 884, 891, 892, 898, 904, 905, 931, 933, 936, 942,
945, 949, 950, 952, 955-957, 964, 968, 981, 989, 990, 995, 1001, 1002, 1009,
1015, 1018, & 1021. System Code #882130/40, Serial #184, 730, 122, 136,
254, 255, 284, 614, 696, 874, 906, & 984. System Code #882140/882130,
Serial #113, 148, 152, 178, 179, 187, 188, 217, 238, 243, 247, 248, 278, 288,
345, 367, 371, 373, 380, 384, 389, 397, 397, 399, 408, 413, 415, 435, 439,
439, 440, 453, 477-479, 486, 504, 512, 525, 526, 528, 533, 535, 540, 547, 564,
567, 568, 577, 579, 581, 638, 639, 644, 645, 651, 652, 659, 667, 680, 747,
758, 815, 850, 859, 882, 883, 890, 896, 897, 899, 925, & 969.
RECALLING FIRM/MANUFACTURER
Philips Medical Systems (Cleveland)
Inc., Cleveland, OH,
REASON Philips was informed that the detector
buckets may experience weakness over time. As a result, this could cause the
detector bucket to separate from the mounting plate of the gantry and
eventually result in the detector falling.
VOLUME OF PRODUCT IN COMMERCE
795 units
DISTRIBUTION Nationwide and Internationally
PRODUCT Zoll pedi.padz Reduced Energy
Multi-Function Electrodes (MFE) Part number: 8900-0401 Used in connection with
Zoll M Series and E series Defibrillators specifically for pediatric patients.
CODE Lot Numbers 0510 thru 4511
RECALLING FIRM/MANUFACTURER
Bio-Detek, Inc., Pawtucket, RI,
REASON Lack of an adequate silicone seal around
the periphery of the attenuator enclosure may allow fluid to enter.
VOLUME OF PRODUCT IN COMMERCE
3,080 units
DISTRIBUTION Nationwide and Canada
PRODUCT
1. Model #03001-02: Periophil Biphasic,
Synthetic Biphasic Dental Bone Graft Granules, 1.0 cc. Model # 03003-01
Periobase Biphasic: Periobase, Synthetic Biphasic Dental Bone Biphasic
Granules, 0.5 cc. Model #03002-01: Periobone Biphasic, Synthetic HA/TCP Dental
Bone Graft Biphasic Granules, 0.5 cc. Periophil Biphasic (synthetic grafting
material) is intended for use as a bone grafting material to fill, augment, or
reconstruct periodontal or oral/maxillofacial defects.
CODE
1. Lot # Z13-927A, expiration date
2/1/2014; 2. 03001-02 Periophil Biphasic, Lot # Z13-927B, expiration date
2/1/2014. 3. 03001-01 Periophil Biphasic, Lot # Z13-926A, expiration date
2/1/2014. 4. 03001-01 Periophil Biphasic, Lot # N13-925C, expiration date
1/31/2014. 5. 03003-01 Periobase Biphasic, Lot # N13-925B, expiration date
1/31/2014. 6. 03002-01 Periobone Biphasic, Lot # N13-925A, expiration date
1/31/2012.
RECALLING FIRM/MANUFACTURER
Recalling Firm: Cytophil, Inc., East
Troy, WI,
Manufacturers: Cytophil, Inc., East
Troy, WI;
Steris Isomedix Services, Libertyville,
IL.
REASON Devices were inadvertently shipped after
receiving an inadequate dose of radiation and this product may not be sterile.
VOLUME OF PRODUCT IN COMMERCE
547
PRODUCT Temporary Pacing Wire, BIPOLAR, Models
a) BM, b) V, c) VE and d) VF, Rx Only, STERILE/EO, Single use only -- Mce
(MEDICAL CONCEPTS EUROPE) --- Shipping or Unit Package: 20 individual units
packaged per box with 6 boxes per case. --- Device Classification Name:
Electrode, pacemaker, temporary. The device is supplied sterile and intended
for single use. Used to pace the heart during or after open heart surgery. The
electrodes are placed on the heart; the other end of the pacing wire is
inserted through the skin of the patient with a thoracic needle. The thoracic
needle can snap-off at the end where the pacing wire is attached, after
snap-off the remaining pins are connected to an external pacemaker. For
temporary atrial and ventrical pacing and sensing for contemplated implant
duration of 7 days or less.
CODE
a) Lot numbers 1101044-10, 1012035-10,
1012036-10, 1011128-10, 1101023-10, 1101134-10, 1102035-10, 1102061-10,
1009089-10; b) Lot number: 1103066-10; c) Lot numbers: 1101055-10, 1101008-10;
d) Lot numbers: 1011092-10, 1011148-20, 1012038-10, 1101021-10, 1101136-10
RECALLING FIRM/MANUFACTURER
European Custom Manufacturing B.V,
Gemert, Netherlands
REASON The crimping operation during product
production was not performed properly. As such, the pacing wire may separate
from the connection pin, which causes non-functionality.
VOLUME OF PRODUCT IN COMMERCE
2,120 individual units (106 boxes)
PRODUCT Leica M822 Surgical Microscope Device
Controller Software (MDC version 3.6, pack F, affecting the control unit
component, Article #10448446, in combination with a Leica M822 Optics Carrier-XY-Focus-Tilt-Unit,
article #10448587; Leica Microsystems (Schweiz) AG, Max Schmidheiny-Strasse
201. The intended use of the microscope is for improving the visibility of
objects through magnification and illumination. It can be applied for
observation and documentation and for human and veterinary medical treatment.
The corrected units will have controller software version MDC V3.6 pack G.
CODE
Serial numbers: 181011001, 181011009,
181011005, 181011003, 250811007, 28101108, 151111011, 211111007, 251111006 and
251111004
RECALLING FIRM/MANUFACTURER
Recalling Firm: Leica Microsystems,
Inc., Buffalo Grove, IL, .
Manufacturer: Leica Microsystems
(Schweiz) Ag, Heerbrugg, Switzerland.
REASON Leica Microsystems received complaints
stating that the zoom function of the Leica M822 surgical microscope locked up
during use.
VOLUME OF PRODUCT IN COMMERCE
29 units
DISTRIBUTION Nationwide,
Austraila, Japan, Korea,
Pakistan, Romania, South Africa, Switzerland and Turkey
PRODUCT SLIDEX Staph kit (250 Tests). The SLIDEX
Staph-Kit is a rapid latex and red blood cell agglutination test for the
identification of Staphylococcus aureus strains from culture media. Catalog
#73113.
CODE Lot number: 1105104545
RECALLING FIRM/MANUFACTURER
Recalling Firm: bioMerieux, Inc.,
Durham, NC,
Manufacturer: Biomerieux Brasil, Rio De
Janeiro, Brazil.
REASON Reports of false negative results in
organism identification.
VOLUME OF PRODUCT IN COMMERCE100 kits
PRODUCT Ingenuity CT (Computed
Tomography)
System, 510 (k) #K033326, Model #728326. Product Usage: The Ingenuity CT is a
Computed Tomography X-Ray System intended to produce cross-asectional images
of the body by computer reconstruction of X-Ray transmission data taken at
different angles and planes. This device may include signal analysis and
display equipment, patient, and equipment supports, components and
accessories.
CODE Serial Numbers: 300003, 300004, 300005,
and 300010
RECALLING FIRM/MANUFACTURER
Philips Medical Systems (Cleveland)
Inc., Cleveland, OH,
REASON A 'gap' is created in the side covers
when the gantry of the device is tilted to +30 degrees (forward tilt). A
hazard to the operator/user presents itself if the operator puts his or her
fingers or hand inside the gap and the gantry tilt is returned in position
while the fingers or hand of the user are still in the gap.
VOLUME OF PRODUCT IN COMMERCE
4 units
DISTRIBUTION Nationwide, Australia and Germany
PRODUCT VARIANT II Link Hemoglobin A1c Program
Reorder Pack, 1600 tests, IVD, For the determination of Hemoglobin A1c in
human whole blood. Contents: Elution Buffer A , 1 x 2500 mL Elution Buffer B,
1 x 2100 mL Wash/Diluent Solution, 1 x 2500 mL Cartridge Set CD ROM - 1
Calibrator/Diluent Set Whole Blood Primer, 6 x 1 mL Sample vials, 2 x 100
Instruction Manual. Intended for the percent determination of hemoglobin A1c
in human whole blood using ion-exchange high-performance liquid chromatography
(HLCP). Catalog number: 270-2716.
CODE Lot 70211322; 70211432, 70211440; exp
3/31/2012
RECALLING FIRM/MANUFACTURER
Bio-Rad Laboratories, Inc., Hercules, CA
REASON It was discovered that the Variant II
Turbo Link Calibrator Values card contains incorrect values for IFCC units.
The numbers are missing a decimal point.
VOLUME OF PRODUCT IN COMMERCE
63 units
PRODUCT MERZ AESTHETICS, 27G X 1.0 " (25MM)
Cannula, Sterile R, QTY 20, Model 9015M1X20. The intended use of the Cannula
is the injection of fluids into, or withdrawal of fluids from, parts of the
body below the surface of the skin.
CODE Lot #1029402
RECALLING FIRM/MANUFACTURER
Recalling Firm: Merz Aesthetics, Inc.,
Franksville, WI,
Manufacturer: Tsk Laboratory, Tochigi
City, Japan. .
REASON Recall due to a labeling error on the
secondary (outer) packaging. The outer package label states that the carton
contains Cannulas that are 27GA x 25 mm in size instead of 27G x 38 mm. Length
of the cannula is a matter of user preference and is unlikely to impact
patient safety.
VOLUME OF PRODUCT IN COMMERCE
5 boxes (20 units per box)
PRODUCT
1) Axium Detachable Coil System
QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the
endovascular embolization of intracranial aneurysms and other neurovascular
abnormalities, such as arteriovenous malformations and arteriovenous fistulae.
2) Axium Detachable Coil System
QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the
endovascular embolization of intracranial aneurysms and other neurovascular
abnormalities, such as arteriovenous malformations and arteriovenous fistulae.
CODE
1) Lot # 7964576;
2) Lot # 7964590
RECALLING FIR/MANUFACTURER
Micro Therapeutics Inc, dba ev3
Neurovascular, Irvine, CA,
REASON ev3 has initiated a voluntary removal of
two lots of Axium Detachable Coil System for product mislabeling as the lots
were swapped in production.
VOLUME OF PRODUCT IN COMMERCE
27 units
DISTRIBUTION Nationwide, the UK and France
WEEK ENDING FEBRUARY 18
PRODUCT Bassinet
Models: BSNT/01; BSNT/02, BSNT/03, BSNT/04 The Nemschoff Perinatal Bassinet
products are wheeled carts with a top tray intended for an infant tray with
mattress. The bassinets are intended for medical purposes that consist of a
bed designated for the use of a pediatric patient.
CODE Manufactured
between 11/21/2003 and 04/04/2008. Bassinet tubs manufactured between
11/21/2003 to present
RECALLING FIRM/MANUFACTURER
NEMSCHOFF, Sheboygan, WI,
REASON To
correct issues with casters, door hinges, drawer slides, and the bassinet
surround. Also, to add labeling to the product regarding specific instructions
for cleaning the bassinet tub and drawer weight limits.
VOLUME OF PRODUCT IN COMMERCE
1214 units
DISTRIBUTION Nationwide,
and Canada
PRODUCT Smith
& Nephew 5.5mm DYONICS BONECUTTER ELECTROBLADE Resector Catalog Number:
7205962.
CODE Lot
Numbers: 830567, 830568, 830569, 863004, 868509, 868519, 875226, 875227, and
897191
RECALLING FIRM/MANUFACTURER
Recalling Firm: Smith &
Nephew, Inc. Endoscopy Division, Andover, MA, .
Manufacturer: Smith & Nephew,
Inc., Endoscopy Div., Mansfield, MA.
REASON Manufacturing
issue: The distal tip of the outer sheath (return electrode) may detach and
become loose in the joint during use.
VOLUME OF PRODUCT IN COMMERCE
1743 units
DISTRIBUTION Nationwide
and Internationally
PRODUCT Pinnacle3
Radiation Therapy Planning System, software version 9.0. Model # Description
453560446041 Pinnacle3 version 9.0 English 453560446051 Pinnacle3 version 9.0
Simplified Chinese 453560446061 Pinnacle3 version 9.0 Traditional Chinese
453560446091 Pinnacle3 version 9.0 Dutch 453560446101 Pinnacle3 version 9.0
French 453560446111 Pinnacle3 version 9.0 German 453560446141 Pinnacle3
version 9.0 Italian 453560446161 Pinnacle3 version 9.0 Japanese 453560446181
Pinnacle3 version 9.0 Spanish 453560446201 Pinnacle3 version 9.0 Turkish
Intended to provide support for radiation therapy treatment planning for the
treatment of benign or malignant disease processes.
CODE
System Code # 453560446041 453560446111
453560446041 453560446101 453560446041 453560446101 453560446041 453560446051
453560446041 453560446101 453560446041 453560446101 453560446041 453560446111
453560446041 453560446111 453560446051 453560446041 453560446141 453560446041
453560446161 453560446041 453560446161 453560446041 453560446181 453560446101
453560446041 453560446061 453560446041 453560446061 453560446041 453560446201
453560446041
RECALLING FIRM/MANUFACTURER
Philips Medical Systems (Cleveland),
Inc., Fitchburg, WI,
REASON Phillips
Medical Systems have recently determined that a software (Pinnacle3 Software
version 9.0) nonconformance can cause incorrect beam geometry.
VOLUME OF PRODUCT IN COMMERCE
715 devices
DISTRIBUTION Nationwide
and Internationally
PRODUCT GEMINI
TF 16 PET/CT System (with software version 3.5.1 and 3.5.1.1), a) Model #
882470 & GEMINI TF 64 PET/CT System (with software version 3.5.1 and
3.5.1.1), b) Model #882471. The device is a diagnostic imaging system for
fixed or mobile installations that combines Positron Emission Tomography (PET)
and X-ray Computed Tomography (CT) systems.
CODE
a) Serial #'s: 7003-7006, 7008-7011,
7014, 7015, 7017-7026, 7031, 7034-7036, 7039, 7041, 7043, 7045, 7047, 7049,
7050-7054, 7058, 7060, 7062, 7064, 7066, 7068, 7069, 7073-7075, 7077, 7080,
7081, 7084-7087, 7089, 7093-7099, 7104, 7106-7108, 7112, 7114, 7115, 7118,
7120-7123, 7125, 7128, 7129, 7131, 7133, 7135, 7137-7140, 7142, 7143, 7145,
7147, 7148, 7156, 7159, 7161, 7163, 7165, 7166, 7168, 7170, 7173, 7174, 7176,
7177, 7181, 7183-187, 7193-7200, 7205, 7208, 7210, 7211, 7215, 7216, 7220,
7223 & 7224; b) Serial #'s: 7007, 7012, 7013, 7016, 7027-7030, 7032,
7042, 7044, 7055, 7057, 7059, 7061, 7063, 7065, 7067, 7071, 7072, 7076, 7078,
7079, 7082, 7083, 7088, 7091, 7092, 7102, 7103, 7105, 7109, 7110, 7116, 7117,
7119, 7126, 7132, 7136, 7141, 7144, 7146, 7149, 7152-7154, 7158, 7160, 7164,
7167, 7169, 7171, 7172, 7175, 7178-7180, 7182, 7188-7190, 7192, 7201-7204,
7206, 7209, 7212-7214, 7217-7219 & 7221.
RECALLING FIRM/MANUFACTURER
Philips Medical Systems (Cleveland)
Inc., Cleveland, OH,
REASON Philips
received notification by a customer of an issue relating to unwanted auto
delete occurrences resulting in raw data being deleted and, in some cases, raw
data being deleted immediately after scan acquisition but before images are
reconstructed.
VOLUME OF PRODUCT IN COMMERCE
197 units
DISTRIBUTION Nationwide
and Internationally
PRODUCT Multi-Analyte
Urine DAU Calibrator (15 mL); Multi-Analyte,Intermediate Calibrator; for
in-vitro diagnostic use only. Contains 300 ng/mL benzoylecgonine, 750 ng/mL d-methamphetamine,
4000 ng/m: morphine, 50 phencyclidine, 500 ng/m: oxazepam, 500 ng/mL
secobarbital, 700ng/mL propoxyphene and 600 ng.mL methadone in urine buffer
with dosium azide as preservitive. Drug abuse testing. Catalog 0804.
CODE Lot
number: 1101138K2, exp 1/28/2013
RECALLING FIRM/MANUFACTURER
Lin-Zhi International Inc., Sunnyvale,
CA,
REASON Catalog
# 0804 Multi-Analyte Urine DAU Intermediate Calibrators were incorrectly
labeled and mistakenly filled with Catalog # 0805 DAU High Calibrators.
VOLUME OF PRODUCT IN COMMERCE
49 items
DISTRIBUTION Nationwide
and Germany
PRODUCT R2
Adult R2 Non-Radiolucent Multifunction Electrodes, Catalog/REF No. 3200-1715,
NON-STERILE, MADE IN USA. This device, when connected to a
monitor/defibrillator has the ability to monitor the electro-cardio function (ECG)
of the patient. The device, when electrically activated by the
monitor/defibrillator can pace the heart rate of the patient and may also
transmit the electrical current to the patient for cardio-conversion. This is
a non-sterile, disposable device.
CODE Lot
1010194
RECALLING FIRM/MANUFACTURER
Recalling Firm: ConMed Corp.,
Utica, NY,
Manufacturers: ConMed Corp.,
Utica, NY;
Consolidated Medical Equipment Co,
Chihuahua, Mexico.
REASON The
specific lot was assembled without the rivet cover.
VOLUME OF PRODUCT IN COMMERCE
850 sets (2 pads per set in individual
pouch)
DISTRIBUTION Nationwide,
Canada, Belgium, Germany and Hong Kong
PRODUCT 5100-03-707
-- MOST Option STEM DISTRACTOR MATL: SS NONSTERILE QTY-1 Designed to
facilitate intraoperative decisions with a variety of options for femoral and
tibial components, implant connections and fixation of the implants.
CODE Lots
1519123, 1616672, 1627174, 60281712, 60773654, 60795066, 60841064, 61061148,
61348637, and 61770544
RECALLING FIRM/MANUFACTURER
Zimmer Inc., Warsaw, IN,
REASON Fractured
tips: Any use of force in a certain direction to the flat side of the
instrument may result in fracturing of the mating geometry, creating the
potential for metal debris at the surgical site.
VOLUME OF PRODUCT IN COMMERCE
109 total
DISTRIBUTION Nationwide
and Switzerland
PRODUCT Alaris
PC unit model 8015. The Alaris PC unit is the central programming, monitoring
and power supply component for the Alaris System.
CODE The
affected units have unique serial numbers for traceability.
RECALLING FIRM/MANUFACTURER
CareFusion 303, Inc., San Diego ,
CA,
REASON The
recall was initiated because Carefusion has identified a potential risk with a
PLD component on the Alaris PC unit model 8015 power supply board. The PLD
component used in production between June 21, 2011 and August 15, 2011 has a
higher standby current resulting in the battery depleting prematurely.
VOLUME OF PRODUCT IN COMMERCE
1,039 units
DISTRIBUTION Nationwide
and Australia
PRODUCT enGen(TM)
Laboratory Automation Systems configured with Hettich Rotanta 46 RSC Robotic
Centrifuges, Catalog # enGen -- Common/Usual Name: enGen(TM) Track System,
Hettich Centrifuge -- For in vitro diagnostic use only. The VITROS 5,1 FS
Chemistry System with enGen(TM) Laboratory Automation System is intended for
use in the in vitro quantitative measurement of a variety of analytes of
clinical interest, using both VITROS Chemistry Products Slides (colorimetric
endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS
Chemistry Products MicroTip liquid reagents (spectrophotometric and
spectrophotometric immunoassay methods).
CODE All
enGen(TM) Laboratory Automation Systems configured with Hettich Rotanta 46 RSC
Robotic Centrifuges.
RECALLING FIRM/MANUFACTURER
Recalling Firm: Ortho-Clinical
Diagnostics, Rochester, NY,
Manufacturers: Ortho-Clinical
Diagnostics, Rochester, NY;
Thermo Fisher Scientific OY, Vantaa,
Finland;
Andreas Hettich Gmbh & Co. Kg,
Tuttlingen, Germany.
REASON Complaints
of centrifuge buckets disconnecting from their supports during centrifugation
when using the Hettich Rotanta 46 RSC Robotic Centrifuges, Models 4815, 4816
and 4817, that may be configured with enGen(TM) Laboratory Automation Systems.
VOLUME OF PRODUCT IN COMMERCE
78 units worldwide (18 Domestic and 68
Foreign)
DISTRIBUTION Nationwide
and Internationally
PRODUCT The
SIGMA HP Fixed Bearing Tibial Tray Impactor is part of a set of tools designed
specifically for the installation of the PFC SIGMA Knee. SIGMA HP Fixed
Bearing Tibial Tray Impactors Packaged in poly bag with optional protection
from sharp or fragile points. Instruments are supplied non-sterile and do not
have specified expiration date or shelf life. Usage: The Tibial Tray (with
protective cover) and instrument are assembled outside of the patient in
theatre; the tray is introduced and impacted into the cement containing tibia
using the impactor. Once the tray is fully seated in the cement containing
tibia, the instrument is intended to be removed. Catalog number
9505-01-306.
CODE
All lots: 071089, 071812, 071989,
072017, 072219, 072220, 072221, 072222, 080026, 080027, 080028, 080029,
080319, 080336, 080337, 080338, 080403, 080404, 080405, 080406, 080407,
080632, 080633, 080634, 080635, 080841, 080842, 080843, 080844, 080845,
080846, 080847, 080848, 080849, 081138, 082151, 082152, 082153, 082175,
082176, 082177, 082178, 082260, 082261, 082262, 082263, 082264, 082265,
082266, 082267, 082268, 082269, 083106, 083107, 083108, 083109, 083110,
083111, 083400, 083401, 083402, 083403, 083404, 083405, 083406, 083407,
083408, 083409, 083495, 090237, 090238, 090239, 090240, 090241, 090242,
090933, 090934, 090935, 090936, 090937, 090938, 090939, 090940, 090941,
090942, 094951, 094952, 100059, 100060, 100172, 100173, 100174, 100455,
100456, 100457, 100458, 100459, 100460, 100469, 100470, 100787, 102110,
102111, 102616, 102617, 102618, 102619, 102620, 102621, 102622, 102623,
102624, 104779, 104780, 105278, 105279, 110046, 110047, 110343, 110344,
110345, 110725, 110726, 110727, 110728, 110729, 111011, 111012, 111013,
112537, 112538, and 112539
RECALLING FIRM/MANUFACTURER
Recalling Firm: DePuy
Orthopaedics, Inc., Warsaw, IN,
Manufacturer: Enztec Ltd,
Christchurch, New Zealand.
REASON The
impactor can become intentionally locked on the Tibial Tray implant during
surgery.
VOLUME OF PRODUCT IN COMMERCE
1794 (398 in USA)
DISTRIBUTION Nationwide,
Canada, Ireland and South Africa
PRODUCT Product
is a 15x16x27" (WxDxH) air purifier device with one of two different
brands next to the control panel: HealthWay or CleanStation. HealthWay brand
is sold under model number Deluxe 20600-3 Air Purification System. The
CleanStation brand is sold under model number CS20000. The devices are
intended to remove particles from the air for medical purposes.
CODE
Serial #’s: 08070001-08070425,
01080001-01080420, 13070001-13070415, 10080001-10080455, 33070001-33070458,
21080001-21080355, 34070001-34070455, 22080001-22080420, 35070001-35070455,
28080001-28080455, 36070001-36070455, 21080001-21080190, 38070001-38070395,
47080001-47080410, 39070001-39070455, 49080001-49080423, 45070001-45070195,
50080001-50080420, 42070001-42070455, 35090001-35090447, 43070001-43070455
RECALLING FIRM/MANUFACTURER
Recalling Firm: Healthway Home
Products, Inc., Pulaski, NY,
Manufacturer: Shenzhen Healthway
Elec Co. Ltd, Shenzhenshenzhen, China.
REASON A
potential risk of electrical arcing that could cause overheating or fire
within the units.
VOLUME OF PRODUCT IN COMMERCE
9,107 units
DISTRIBUTION Nationwide
and Canada
PRODUCT Siemens
MAGNETOM Aera/Skyra, model numbers 10432914 and 10432915. Nuclear magnetic
resonance imaging system.
CODE Serial
numbers 45115, 41187, 41157, 41116, 45072, 45081, 41089, 41215, 41144, 41201,
45148, 41164, 41088, 41245, 41168, 41152, 45140, 45152, 45143, 45195, 41222,
41150, 45129, 45118, 45119, 41134, 41138, 45182, 45213, 41167, 41100, 52024,
45196, 45197, 41180, 41130, 45088, 41189, 41104, 41246, 41176, 45138, 15084,
41071, 45120, 45107, 41271, 45135, 45141, 41264, 45110, 41279, 41231, 41284,
41235, 45221, 45223, 41276, 41278, 41137, 41225, 41131, 41141, 41233, 45185,
41236, 41238, 45149, 41191, 45074, 45106, 45006, 45201, 41133, 41265, 41188,
45220, 41140, 41079, 45160, 45167, 41194, 45155, 45130, 41241, 41170, 45147,
41006, 45219, 45085, 45150, 45031, 41162, 45121, 45134, 45099, 45038, 45079,
41109, 41156, 45116, 41249, 45128, 45078, 45214, 41103, 45229, 41234, 45181,
52025, 52021, 45105, 45132, 41149, 45228, 45060, 45080, 45208, 41083, 45230,
45205, 41097, 41181, 45093, 45199, 45184, 45071, and 45111
RECALLING FIRM/MANUFACTURER
Recalling Firm: Siemens Medical
Solutions USA, Inc., Malvern, PA,
Manufacturer: Siemens AG, Medical
Solutions, Business Unit XP. Erlangen, Germany.
REASON During
product monitoring firm has discovered that on a few systems the gradient
cable connections did not meet their specifications after installation.
Locally, the cable connections overheated, making servicing necessary. The
potential existed for a cable connection to become overheated, leading to the
emission of smoke and possible material damage.
VOLUME OF PRODUCT IN COMMERCE
128 units
PRODUCT Varian
brand Varian Clinac Accelerators with Mark, Millennium MLC 120 (Multileaf
Collimator) and BrainLAB m3 MLC, Reference/FSCA CP-05591, Model Number: H53.
Intended use: The Varian Millennium 120 Collimator (MLC) is an x-ray
collimator designed to be mounted on a C or EX Series Varian Clinac radiation
therapy linear accelerator and is intended to shape the x-ray field perimeter.
Field shape can either be static (fixed) or dynamic. Dynamic field shapes are
controlled as a function of either Clinac dose fraction or gantry angle. The
intended use is to assist the radiation oncologist in the delivery of
radiation to defined target volumes while sparing surrounding normal tissue
and critical organs from excess radiation.
CODE
H530001, H530075, H530439, H533043,
H535023, H535082, H535128, H535180, H535220, H530004, H530078, H531008,
H534004, H535024, H535084, H535131, H535181, H535867, H530008, H530082,
H531010, H534015, H535025, H535085, H535133, H535182, H536005, H530012,
H530084, H531020, H534020, H535030, H535086, H535135, H535183, H536452,
H530014, H530085, H531025, H534021, H535031, H535087, H535137, H535184,
H536465, H530018, H530087, H531029, H534022, H535033, H535088, H535141,
H535186, H536479, H530021, H530104, H531033, H534026, H535035, H535091,
H535142, H535188, H530022, H530204, H531035, H534032, H535036, H535094,
H535144, H535190, H530023, H530231, H531038, H534036, H535039, H535098,
H535147, H535191, H530024, H530250, H531041, H534040, H535043, H535099,
H535149, H535192, H530030, H530311, H531043, H534223, H535044, H535100,
H535151, H535193, H530034, H530323, H531044, H535003, H535045, H535104,
H535154, H535194, H530035, H530339, H532005, H535004, H535046, H535105,
H535157, H535195, H530036, H530340, H532006, H535005, H535047, H535106,
H535159, H535197, H530045, H530396, H532014, H535008, H535053, H535108,
H535161, H535199, H530046, H530401, H532021, H535009, H535054, H535109,
H535162, H535200, H530049, H530405, H532027, H535012, H535055, H535111,
H535164, H535201, H530050, H530417, H533004, H535013, H535056, H535115,
H535166, H535207, H530051, H530423, H533008, H535014, H535058, H535116,
H535167, H535209, H530055, H530425, H533014, H535015, H535068, H535117,
H535168, H535211, H530057, H530427, H533027, H535016, H535070, H535120,
H535169, H535213, H530063, H530428, H533028, H535017, H535072, H535121,
H535173, H535214, H530064, H530429, H533029, H535018, H535073, H535122,
H535176, H535215, H530066, H530430, H533036, H535020, H535076, H535123,
H535177, H535218, H530068, H530433, H533037, H535021, H535081, H535124,
H535178, H535219
RECALLING FIRM/MANUFACTURER
Varian Medical Systems, Inc. Oncology
Systems, Palo Alto, CA,
REASON Failure
to retract or park the Varian MLC leaves prior to use of the demountable
BrainLAB m3 could result in the Varian MLC leaves interfering with the
treatment field defined by the BrainLAB m3 without an interlock being
asserted.
VOLUME OF PRODUCT IN COMMERCE
205 units
DISTRIBUTION Nationwide
and Internationally
PRODUCT VITROS
Chemistry Products Calibrator Kit 24, Catalog No. 680 1708. Common/Usual Name:
VITROS Calibrator Kit 24. For in vitro diagnostics use only. VITROS Chemistry
Products Calibrator Kit 24 is used to calibrate the VITROS 5,1 FS Chemistry
System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System
for the quantitative measurement of albumin in urine. Ortho-Clinical
Diagnostics Inc.
CODE Kit
Lot 2430 (exp. 11/16/11), Kit Lot 2440 (exp. 01/08/12), Kit Lot 2450
(03/09/12), Kit Lot 2471 (exp. 7/14/12)
RECALLING FIRM/MANUFACTURER
Ortho-Clinical Diagnostics, Rochester,
NY
REASON Ortho
Clinical Diagnostics received numerous complaints of lower than expected
results for control fluids and patient samples upon initial calibration or
following a recalibration using VITROS Chemistry Products Calibrator Kit 24
and VITROS Chemistry Products mALB Reagent.
VOLUME OF PRODUCT IN COMMERCE
4436 units worldwide (nationwide - 2201;
internationally - 2235)
DISTRIBUTION Nationwide
and Internationally
PRODUCT Volcano
pcFM Software Kit. The volcano pcFM Software Kit is a research only off line
tool intended for review and printout of physiological measurements recorded
by ComboMap System. It is not intended for analysis or manipulation of the
original measurement data. Model 6830 ComboMap Software kit; Kit.
CODE P/N
#'s: 803259001; 806431005; S2000061; 806431002
RECALLING FIRM/MANUFACTURER
Volcano Corp., Rancho Cordova, CA,
REASON Internal
review of validation of pcMF Software kit found that instructions directs user
to remove the rear panel, causing the potential for electric shock.
VOLUME OF PRODUCT IN COMMERCE
8 units
DISTRIBUTION Nationwide,
The Netherlands, Sweden and United Kingdom
PRODUCT Custom
surgical kits: DeRoyal(R) TRAUMA CRANIOTOMY PACK, REF 89-7639.01, STERILE EO,
DeRoyal(R) TRAUMA CRANIOTOMY PACK PGYBK, REF 89-7639.01, STERILE EO, DeRoyal(R)
TRAUMA CRANIOTOMY PACK PGYBK, REF 89-7639.02, STERILE EO, surgical kits
CODE Lot
Number 24579045 REF 89-7639.01 - Lot Number 24664258 REF 89-7639.02 - Lot
Numbers: 24828048, 25598985, 25359212, 25399676
RECALLING FIRM/MANUFACTURER
DeRoyal Processing Center, New Tazewell,
TN,
REASON The
firm distributed surgical kits which contained DermaHook Neuro Elastic
Retractors which were subsequently recalled by Teleflex.
VOLUME OF PRODUCT IN COMMERCE
214 units
PRODUCT Surgical
lighting systems iLED and TruLight in use with the ALC+ function. The TRUMPF
surgical lights are intended to locally illuminate an operating or examination
area of the patients body with a high intensity light
CODE Material
number 1583966
RECALLING FIRM/MANUFACTURER
Recalling Firm: Trumpf Medical
Systems, Inc., Charleston, SC, .
Manufacturer: Kreuzer Gmbh & Co. KG, Puchheim, Germany.
REASON The
ALC+ function may not be properly locked during use resulting in the
possibility of the handle falling off into the surgical field causing injury
to the patient.
VOLUME OF PRODUCT IN COMMERCE
8733 units (4709 US)
DISTRIBUTION Nationwide
and Internationally
PRODUCT New
Label: EtchMaster ***Cordless Handpiece***P/N 797 HAZARD WARNING: Using CO2
cartridges from other suppliers may cause an operational hazard and will void
the warrantee. Old Label: EtchMaster***Cordless Handpiece*** The Cordless
Etcher air brush is powered by CO2 mini-cartridges to deliver pressurized gas
to the EtchMaster pre-filled disposable micro air abrasive tips. The Cordless
Etcher handpiece is quickly re-loaded with a CO2 mini-cartridge. Each CO2
mini-cartridge provides sufficient power to run 3 EtchMaster tips. The
EtchMaster tips are used to abrade tooth surfaces and prosthetic devices prior
to bonding procedures.
CODE
GRO-SK-EM-CE-R-RB-Sk
GRO-SK-EM-CE-K-RB-SK GRO-SK-EM-CE-K-CA-SK ROPI-EM
RECALLING FIRM/MANUFACTURER
Groman Inc., Margate, FL,
REASON When
non-conforming cartridges are inserted into the Cordless Etcher they can
induce excessive stresses on the device causing cracks in the plastic housing
of the device. These crack can lead to a hazardous situation due to the high
pressure of the gas contained in the CO2 cartridges.
VOLUME OF PRODUCT IN COMMERCE
395 devices
DISTRIBUTION Nationwide,
and Netherlands
PRODUCT Imalux
Niris OCT Imaging System Probe, Model Number 1300. The Niris Imaging System
probe is used with the Niris Imaging System. The Niris Imaging System employs
Optical Coherence Tomography (OCT), which can be used to construct high
spatial resolution. The probe, is attached to the Imaging Console, and is used
to direct light to and from the patient tissue. A small electromechanical
scanning mechanism in the probe moves the optical beam laterally across the
tissue surface while simultaneously acquiring in-depth backscattering profile
at each lateral position. The probe is not labeled or sold as sterile.
CODE Serial
Number F464019
RECALLING FIRM/MANUFACTURER
Imalux Corp., Cleveland, OH,
REASON During
an FDA inspection, the firm was notified that the wrong tubing was used during
the manufacture of the probe for the unit.
VOLUME OF PRODUCT IN COMMERCE
1 unit
PRODUCT Integra
PAC1 Mod 2 Cables used with the Integra MPM, CAM01 Monitors. MPM, CAMO1,
PAC1.
CODE
MEF1100103; MED1102303; MEB1101602;
MEF1100203; MEC1100602; MED1101003; MED1102203; MEE1102003; MED1102103;
MEC1102302; MEB1101402; MEE1101003; MEF1100703; MEE1101903; MEF1101103;
MEB1100803; MEC1103002; MEE1100903; MED1101703; MED1101203; MEC1102902;
MEB1101103; MEC1102002; MEB1101502; MED1102203; MEF1100603; MEE1101203;
MEC1100102; MEE1101503; MEE1100703; MEF1100403; MED1101303; MED1101403;
MEF1100303; MEC1100702; MED1101103; MEC1102402; MEC1101602; MEF1100903;
MEE1101303; MED1102003; MEC1103302; MEE1101803; MEE1100803; MEF1100503;
MEC1102502; MEC1101702; MEE1101403; MEC1103402; MEF1100803; MEC1102802;
MEC1101802; MEF1101003; MEC1103102; MEC1101902; MEC1102102
RECALLING FIRM/MANUFACTURER
Integra LifeSciences Corp., Plainsboro,
NJ,
REASON Some
of the connector cables associated with Integra's intracerebral pressure (ICP)
monitors may cause temporary interference with the ICP waveform on the monitor
display.
VOLUME OF PRODUCT IN COMMERCE
221 units
DISTRIBUTION Nationwide
and Internationally
Flight Medical
Innovations Ltd. Provides Update on Recall of Ventilator Remote Alarm Cable
Flight Medical Innovations Ltd. (Flight
Medical) issued the following update regarding its recall of remote alarm cable,
part number V24-00400-29, used to connect a Newport HT50 mechanical ventilator
to a nurse station. The recalled cable may negatively affect Newport HT50
ventilators manufactured prior to 2007. Affected units may shutdown
unexpectedly.
Flight Medical initiated the recall in
August 2011, which affected 363 cables. The affected devices were
distributed between April 25, 2007 and July 6, 2011. A total of 328 cables
were distributed within the US. An additional 22 cables were distributed
outside the US. 13 cables remained in the distributor’s warehouse.
Customers, including Newport Medical
are requested to dispose of the affected cables properly
Customer inquiries related to this
action should be addressed to Newport’s Technical Service Group at (800) 451-3111,
extension 500.
WEEK ENDING FEBRUARY 11
PRODUCT Phantom-brand
Black Latex Powder Free Examination Gloves, 100 gloves by weight, size small
and Medium, Product Code PHM912, Lot No. 002-E1LF302-1, Manufacture date JUL
2011, Size Small Product Code PHM915, Lot No. 002-E1LF305-1, Manufacture date
JUL 2011, Size Medium Examination Disposable Gloves.
CODE Lot
No. 002-E1LF302-1 and Lot No. 002-E1LF305-1
RECALLING FIRM/MANUFACTURER
Recalling Firm: Adenna Inc.,
Ontario, CA,
Manufacturer: Tekmedic Sdn Bhd,
Melaka, Malaysia.
REASON The
recall was initiated because Adenna Inc. distributed latex gloves that have
been FDA refused.
VOLUME OF PRODUCT IN COMMERCE
427 cartons
PRODUCT Fresenius
Optiflux F250NRe Hemodialyzer, , intended for single use in acute and chronic
hemodialysis Catalog No. 0500325E The Fresenius Optiflux F250NRe Hemodialyzers
are intended for patients with acute or chronic renal failure when
conservative therapy is judged to be inadequate.
CODE
Lot Codes: 8SU203 EXPIRATION DATE:
12/31/2011, 8SU204 EXPIRATION DATE: 12/31/2011, 8SU205 EXPIRATION DATE:
12/31/2011, 09DU02013 EXPIRATION DATE: 04/30/2012, 09DU02014 EXPIRATION DATE:
04/30/2012, 09DU02022 EXPIRATION DATE: 04/30/2012, 09NU02011 EXPIRATION DATE:
10/31/2012, 09PU02011 EXPIRATION DATE: 11/30/2012, 09SU02001 EXPIRATION DATE:
12/31/2012, 09SU02007 EXPIRATION DATE: 12/31/2012, 10AU02013 EXPIRATION DATE:
01/31/2013, 10CU02001 EXPIRATION DATE: 03/31/2013, 10EU02010 EXPIRATION DATE:
05/31/2013, 10EU02017 EXPIRATION DATE: 05/31/2013, 10HU02013 EXPIRATION DATE:
06/30/2013, 10HU02014 EXPIRATION DATE: 06/30/2013, 10LU02001 EXPIRATION DATE:
09/30/2013, 10LU02013 EXPIRATION DATE: 09/30/2013, 10LU02014 EXPIRATION DATE:
09/30/2013, 10LU02015 EXPIRATION DATE: 09/30/2013
RECALLING FIRM/MANUFACTURER
Recalling Firm: Fresenius Medical
Care Holdings, Inc., Waltham MA,
Manufacturer: Fresenius Medical
Care, North America, Ogden, UT.
REASON Optiflux
F250NRe Hemodialyzers have increased risk of an internal blood leak.
VOLUME OF PRODUCT IN COMMERCE
599,436 units
DISTRIBUTION Nationwide
and Canada
PRODUCT STERIS
SYSTEM 1E Liquid Chemical Sterilant Processing System. Model P6500, 510 (k)
Numbers: K090036 Cleared 4/05/2010; K101409 cleared 8/2/2010; K102462, cleared
9/2/2010; add to file 2/24/2011. The SYSTEM 1E Liquid Sterilant Processing
System is intended for the liquid chemical sterilization of manually cleaned
immersible and reusable critical and semi-critical heat sensitive medical
devices, including endoscopes and their accessories.
CODE Serial
Numbers: 400000 through 405499
RECALLING FIRM/MANUFACTURER
Steris Corporation Hopkins Facility,
Mentor, OH,
REASON Software
deficiency. STERIS identified that water specifications were not being met at
some of their customer sites. Critical parameters for initiating cycles are:
water quality, water pressure and water temperature. Failure to meet these
criteria may lead to cycle aborts by the SYSTEM 1E Liquid Chemical Sterilant
Processing System. The firm intends to perform a software field correction
which will enable the SYSTEM 1E to handle a wider range of user facility water
characteristics.
VOLUME OF PRODUCT IN COMMERCE
3,882 units
DISTRIBUTION Nationwide,
and Bahrain, Hong Kong, Japan and South Korea
PRODUCT
1) Presource Custom Sterile Surgical and
Procedure Kits; general surgery trays; The kit catalog numbers: PBVFLCAVJ,
PCW1VPMUW, PE41TP55A, PE73EALUO, PE73MNLUM, PEDKEASTH, PEOCMEBL1, PEOCMEBLD,
PEVFBHAVK, PEVFENAVH, PEVFHNAVJ, PEVNMJCHF, PG73HNLUL, PG73OTALUN, PGCGOTUI8,
PGCGOTUII, PGVFGMAVM, PGVFLPAVC, PGVFMNAVG, PGVFOTAVD, PGVFTTAVK, PGVNGSCHJ,
PLVFOTAVJ, PN21CYBIG, PN21CYBIH, PN41CY557, PN73CR11F, PN73SPLUT, PN73VPLUV,
PNV2ALCPA, PNV2CY10Z, PNV2LM10R, PNV2OT10H, PNV2OTA10Q, PNVFLNAVK, PO73AALU1,
PO73AALUR, PP69OHMLO, PUOCOT13J, PVV2CA10B, SBA13KDBSY, SBA23KTNCC,
SBA23OPNCC, SBACGSTAHA, SCV11CACPR, SCV19OHCHX, SCV23MJNCH, SEN41TYCMC,
SID21ACUHA, SIDCGISAUA, SNE11ALCPA, SNE11CR10F, SNE11LM10C, SNE11NMCPX,
SNE14ACHHK, SNE14CRHHK, SNE14LLHHN, SNE21SHEGS, and SNE32CRSA7 contain one or
more of the following listed components: a) Silastic Tubing, 0.062 ID x 0.032
Wall x 50 Ft, component C-602-285, vendor part number SFM3-2850; b) Silastic
Tubing, 0.040 ID x 0.023 Wall x 50 Ft, component C-SFM3-2050, vendor part
number SFM3-2050; c) Pathology Container with Lid, 8 oz., component 3701Z,
vendor part number 3701; d) Silicone Tubing, 0.187 ID x 0.313 OD x 50 Ft.,
component C-31010SG, vendor part number 31010;
2) Presource Custom Sterile Surgical and
Procedure Kits; surgical trays; The kit catalog numbers kit catalog numbers
ICU955327G, ICU955327H, and MAT330041C contain one or more of the following
listed components: a) Cool Peri-Pad Self-Adhesive Absorbent Pad with Built-In
Cold Pack, component C-CP23, vendor part number CP23; b) Giraffe Disposable
Skin Temperature Probe, vendor part number 6600-0873-700.
CODE 1)
and 2) All recent lots
RECALLING FIRM/MANUFACTURER
Cardinal Health, McGaw Park, IL,
REASON Some
of the components contained in the Presource kits are being recalled because
Cardinal Health has determined that the component manufacturer(s) does not
have all necessary FDA registrations or filings needed for this product
VOLUME OF PRODUCT IN COMMERCE
7,213 each
PRODUCT Product
is a lithium ion battery charger sold as both a stand-alone accessory, and as
part of endoscopy system sets. --- Stand alone labeling: Vision Sciences
Battery Charger, Lithium Ion [REF] 07-3051; [MFG Symbol. Accessory to
endoscopy kits used to charge/recharge lithium ion batteries. Products which
include this item are VISION SCIENCES CAT. No. ITEM Description; 07-3051
Battery Charger, Lithium-Ion; 07-3051X Battery Charger, Lithium-Ion*; 07-3053
BLS Kit; 07-3053X BLS Kit*; 07-3052 Rechargeable Batteries, Lithium-Ion;
07-3052X Rechargeable Batteries, Lithium-Ion*; 02-2602 ENT-3000 Scope
(Standard Set); 02-2602-IN ENT-3000 Scope (Standard Set)*; 02-2602X ENT-3000
Scope (Standard Set)*; 02-4201 ENT-4000 Portable Set; 02-4201X ENT-4000
Portable Set.
Product is a lithium ion battery sold as
both a stand-alone accessory, and as part of endoscopy system sets. Stand
alone labeling: Vision Sciences Rechargeable Lithium Ion Batteries,
Replacement (2) [REF] 07-3052; [MFG Symbol] Accessory to endoscopy kits used
to charge/recharge lithium ion batteries. Products which include this item are
VISION SCIENCES CAT. No. ITEM Description; 07-3051 Battery Charger,
Lithium-Ion; 07-3051X Battery Charger, Lithium-Ion*; 07-3053 BLS Kit; 07-3053X
BLS Kit*; 07-3052 Rechargeable Batteries, Lithium-Ion; 07-3052X Rechargeable
Batteries, Lithium-Ion*; 02-2602 ENT-3000 Scope (Standard Set); 02-2602-IN
ENT-3000 Scope (Standard Set)*; 02-2602X ENT-3000 Scope (Standard Set)*;
02-4201 ENT-4000 Portable Set; 02-4201X ENT-4000 Portable Set.
CODE All
units
RECALLING FIRM/MANUFACTURER
Vision-Sciences, Inc. Orangeburg, NY,
REASON Potential
for over-heating and possible burning of the lithium-ion batteries being
charged.
VOLUME OF PRODUCT IN COMMERCE
1618 chargers, 3628 batteries
DISTRIBUTION Nationwide
and Internationally
PRODUCT KIMBERLY
CLARK MIC GASTRO-ENTERIC FEEDING TUBE, 22F. Gastro-Enteric Feeding Tube.
Catalogue number: 0210-22.
CODE Lot
number: AA1129E02
RECALLING FIRM/MANUFACTURER
Recalling
Firm: Kimberly-Clark Corp., Roswell, GA,
Manufacturer:
Avent S.A. De C.V, Nogales Sonora, Mexico.
REASON The
product is misbranded, i.e. the Gastric (center) port may be mislabeled as the
Jejunal port, and the Jejunal (side) port mislabeled as the Gastric port.
VOLUME OF PRODUCT IN COMMERCE
50 devices
DISTRIBUTION Nationwide
and Canada
PRODUCT DRAD-3000E:
Radrex-i X-Ray is a general purpose X-ray System. Labeling reads in part
"TOSHIBA***DIGITAL RADIOGRAPHY SYSTEM RADREX-i DRAD-3000E Systems Data
No. MSDXR0144EA***"
CODE
Serial numbers of the following:
U2A0892009, U2A08X2010, U2A08X2011, U2A08Y2012, U2A08Z2013, U2A08Z2014,
U2A0912015, U2A0912016, U2A0922017, U2A0942019, U2A0942020, U2A0942021,
U2A0952022, U2A0952023, U2A0972024, U2A0972025, U2A0982026, U2B09X2027,
U2B09Y2028, U2B09Y2029, U2B09Z2030, U2B1042031, U2B1042032, U2C1052033,
U2C1062034, U4A0982001, U4A0982002, U4A0992003, U4A09X2004, U4A09Z2005,
U4A09Z2006, U4A09Z2007, U4A1022009, U4B1022010, U4B1032011, U4B1052012,
U4B1052013, U4B1062014, U6A0972001 and U4B1012008
RECALLING FIRM/MANUFACTURER
Recalling Firm: Toshiba American
Medical Systems Inc., Tustin, CA,
REASON It
has been found that some of the starter cables used for x-ray tube high speed
rotation are not sufficiently durable.
VOLUME OF PRODUCT IN COMMERCE
39 units
PRODUCT Alere
Cholestech LDX Analyzer, Model Number: 10-004, 14-874, 14-875, in vitro
diagnostic.
CODE ROM
versions prior to 3.40
RECALLING FIRM/MANUFACTURER
Alere San Diego, San Diego, CA,
REASON Alere
has identified that a bias on results for certain analytes tested on the Alere
Cholestech LDX System may exist due to humidity variability.
VOLUME OF PRODUCT IN COMMERCE
22,000 meters
DISTRIBUTION Nationwide
and Internationally
North American Rescue,
LLC Announces A Recall of Crickit® Convenience Kits
North American Rescue, LLC (NAR)
announces the recall of the NAR Surgical Crichothyroidotomy Kit or Crickit®,
due to potential contamination of the povidone iodine prep pads, manufactured by
Triad Group.
The Crickit® Kit component may be
found inside a larger kit, the NAR Medic Trauma Pack Kits, available in 4 colors
for which the codes are listed below. Only the Crickit® component is being
recalled. Please see the attached lot numbers for the affected Crickit®
Kits. The affected lot numbers for the Medic Trauma Pack Kits can be found on
the North American Rescue website www.narescue.com.
Please note that this product has been discontinued and therefore is no longer
included in the Medic Trauma Pack Kits.
| Product
Code |
Item
Description |
Lot
Numbers |
| 10-0017 |
Surgical Cricothyroidotomy Kit
or Crickit® |
|
| 80-0070 |
Medic Trauma Pack Kit – BLK
(Black) |
|
| 80-0071 |
Medic Trauma Pack Kit – ODG
(Green) |
|
| 80-0072 |
Medic Trauma Pack Kit – COY
(Tan) |
|
| 80-0096 |
Medic Trauma Pack Kit – DUC
(Camouflage) |
|
Customers are instructed to immediately
discontinue the use of the Crickit® convenience kits and to contact North
American Rescue Customer Service for instructions for returning the product. Only
the Crickit® should be returned from the Medic Trauma Pack Kits.
Consumers with questions may contact
North American Rescue at 1-888-689-6277, Monday through Friday, 8:00am to 5:00pm
(ET).
Part Number: 10-0017
Description: Crickit or Surgical Cricothyroidotomy Kit
| Lot
# |
| 21096 |
| 21420 |
| 21421 |
| 21422 |
| 21423 |
| 21424 |
| 21910 |
| 21911 |
| 21912 |
| 21913 |
| 22151 |
| 22152 |
| 22289 |
| 22290 |
| 22344 |
| 22345 |
| 22346 |
| 22347 |
| 22725 |
| 22726 |
| 22727 |
| 22728 |
| 22946 |
| 22947 |
| 22948 |
| 22949 |
WEEK ENDING FEBRUARY 4
PRODUCT Syngo
Imaging XS, all versions Radiological image processing system. Model number
10496279.
CODE None
RECALLING FIRM/MANUFACTURER
Recalling Firm: Siemens
Medical Solutions USA, Inc., Malvern, PA,
Manufacturer: Siemens AG,
Medical Solutions, Business Unit XP, Erlangen, Germany.
REASON The
"Rename" functionality can cause an unintended patient merge if the
renamed patient is sent to another system.
VOLUME OF PRODUCT IN COMMERCE
38
PRODUCT VITROS
Chemistry Products Calibrator Kit 16, REF 680 1700, 5x1 mL. VITROS Chemistry
Products Calibrator Kit 16 is used in conjunction with VITROS Chemistry
Products FS Calibrator 1 to calibrate the VITROS 5, 1 FS Chemistry System, the
VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the
quantitative measurement of rheumatoid factor (RF).
CODE Lot
#: 1630
RECALLING FIRM/MANUFACTURE
Ortho-Clinical Diagnostics,
Rochester, NY,
REASON Product
will not likely meet its specified 12 months of stability.
VOLUME OF PRODUCT IN COMMERCE
768 kits
DISTRIBUTION Nationwide,
Internationally
PRODUCT Tobramycin
Calibrators, REF 7F93-01; Tobramycin is a commonly used antibiotic to treat
bacterial infections, in particular gram negative infections.
CODE Lots:
59281991, Exp 11-30-11, 59359468, Exp 2-29-12, 59590335, Exp 6-30-12,
59511648, Exp 8-31-12
RECALLING FIRM/MANUFACTURER
Microgenics Corp., Fremont, CA,
REASON Reports
that third party control material for 4 lots of Tobramycin Calibrators that
are recovering low and outside of the published range.
VOLUME OF PRODUCT IN COMMERCE
1,366 kits
DISTRIBUTION Nationwide
and Europe
PRODUCT Leo
3-Wheel Scooter and Leo Scooter, Model # LEO-4B; Intended to provide
transportation for a disabled or elderly person.
CODE
Serial # 10MCT000110, 10MCT000111,
10MCT000112, 10MCT000113, 10MCT000114, 10MCT000115, 10MCT000116, 10MCT000117,
10MCT000118, 10MCT000119, 10MCT000120, 10MCT000121, 10MCT000122, 10MCT000123,
10MCT000124, 10MCT000125, 10MCT000126, 10MCT000127, 10MCT000128, 10MCT000129,
10MCT000130, 10MCT000131, 10MCT000132, 10MCT000133, 10MCT000134, 10MCT000135,
10MCT000136, 10MCT000137, 10MCT000138, 10MCT000139, 10MCT000140, 10MCT000141,
10MCT000142, 10MCT000143, 10MCT000144, 10MCT000145, 10MCT000146, 10MCT000147,
10MCT000148, 10MCT000149, 10MCT000150, 10MCT000151, 10MCT000152, 10MCT000153,
10MCT000154, 10MCT000155 & 10MCT000156. Model # LEO-4S; Serial #
10MCT000001, 10MCT000002, 10MCT000003, 10MCT000004, 10MCT000005, 10MCT000006,
10MCT000007, 10MCT000008, 10MCT000009, 10MCT000010, 10MCT000011, 10MCT000012,
10MCT000013, 10MCT000014, 10MCT000015, 10MCT000016, 10MCT000017, 10MCT000018,
10MCT000019, 10MCT000020, 10MCT000021, 10MCT000022, 10MCT000023, 10MCT000024,
10MCT000025, 10MCT000026, 10MCT000027, 10MCT000028, 10MCT000029, 10MCT000030,
10MCT000031, 10MCT000032, 10MCT000033, 10MCT000034, 10MCT000035, 10MCT000036,
10MCT000037, 10MCT000038, 10MCT000039, 10MCT000040, 10MCT000041, 10MCT000042,
10MCT000043, 10MCT000044, 10MCT000045, 10MCT000046, 10MCT000047, 10MCT000048,
10MCT000049, 10MCT000050, 10MCT000051, 10MCT000052, 10MCT000053, 10MCT000054,
10MCT000055, 10MCT000056, 10MCT000057, 10MCT000058, 10MCT000059, 10MCT000060,
10MCT000061, 10MCT000062, 10MCT000063, 10MCT000064, 10MCT000065, 10MCT000066,
10MCT000067, 10MCT000068, 10MCT000069, 10MCT000070, 10MCT000071, 10MCT000072,
10MCT000073, 10MCT000074, 10MCT000075, 10MCT000076, 10MCT000077, 10MCT000078,
10MCT000079, 10MCT000080, 10MCT000081, 10MCT000082, 10MCT000083, 10MCT000084,
10MCT000085, 10MCT000086, 10MCT000087, 10MCT000088, 10MCT000089, 10MCT000090,
10MCT000091, 10MCT000092, 10MCT000093, 10MCT000094, 10MCT000095, 10MCT000096,
10MCT000097, 10MCT000098, 10MCT000099, 10MCT000100, 10MCT000101, 10MCT000102,
10MCT000103, 10MCT000104, 10MCT000105, 10MCT000106, 10MCT000107, 10MCT000108,
10MCT000109, 10MDG0001, 10MDG0002, 10MDG0003, 10MDG0004, 10MDG0005, 10MDG0006,
10MDG0007, 10MDG0008, 10MDG0009, 10MDG0010, 10MDG0011, 10MDG0012, 10MDG0013,
10MDG0014, 10MDG0015, 10MDG0016, 10MDG0017, 10MDG0018, 10MDG0019, 10MDG0020,
10MDG0021, 10MDG0022, 10MDG0023, 10MDG0024, 10MDG0025, 10MDG0026, 10MDG0027,
10MDG0028, 10MDG0029, 10MDG0030, 10MDG0047, 10MDG0048, 10MDG0049, 10MDG0050,
10MDG0051, 10MDG0052, 10MDG0053, 10MDG0054, 10MDG0055, 10MDG0056, 10MDG0057,
10MDG0058, 10MDG0059, 10MDG0060, 10MDG0061, 10MDG0062, 10MDG0063, 10MDG0064,
10MDG0065, 10MDG0066, 10MDG0067, 10MDG0068, 10MDG0069, 10MDG0070, 10MDG0071,
10MDG0072, 10MDG0073, 10MDG0074, 10MDG0075, 10MDG0076, 10MDG0077, 10MDG0078,
10MDG0079, 10MDG0080, 10MDG0081, 10MDG0082, 10MDG0083, 10MDG0084, 10MDG0085,
10MDG0086, 10MDG0087, 10MDG0088, 10MDG0089, 10MDG0090, 10MDG0091, 10MDG0092,
10MDG0093, 10MDG0094, 10MDG0095, 10MDG0096, 10MDG0097, 10MDG0098, 10MDG0099,
10MDG0100, 10MDG0101, 10MDG0102, 10MDG0103, 10MDG0104, 10MDG0105, 10MDG0106,
10MDG0107, 10MDG0108, 10MDG0109, 10MDG0110, 10MDG0111, 10MDG0112, 10MDG0113,
10MDG0114, 10MDG0115, 10MDG0116, 10MDG0117, 10MDG0118, 10MDG0119, 10MDG0120,
10MDG0121, 10MDG0122, 10MDG0123, 10MDG0124, 10MDG0125, 10MDG0126, 10MDG0127,
10MDG0128, 10MDG0129, 10MDG0130, 10MDG0131, 10MDG0132, 10MDG0133, 10MDG0134,
10MDG0135, 10MDG0151, 10MDG0152, 10MDG0153, 10MDG0154, 10MDG0155, 10MDG0156,
10MDG0157, 10MDG0158, 10MDG0159, 10MDG0160, 10MDG0161, 10MDG0162, 10MDG0163,
10MDG0164, 10MDG0165, 10MDG0166, 10MDG0167, 10MDG0168, 10MDG0169, 10MDG0170,
10MDG0171, 11ADG0001, 11ADG0002, 11ADG0003, 11ADG0004, 11ADG0005, 11ADG0006,
11ADG0007, 11ADG0008, 11ADG0009, 11ADG0010, 11ADG0011, 11ADG0012, 11ADG0013,
11ADG0014, 11ADG0015, 11ADG0016, 11ADG0017, 11ADG0018, 11ADG0019, 11ADG0020,
11ADG0021, 11ADG0022, 11ADG0023, 11ADG0024, 11ADG0025, 11ADG0026, 11ADG0027,
11ADG0028, 11ADG0029, 11ADG0030, 11ADG0031, 11ADG0032, 11ADG0033, 11ADG0034,
11ADG0035, 11ADG0036, 11ADG0037, 11ADG0038, 11ADG0039, 11ADG0040, 11ADG0041,
11ADG0042, 11ADG0073, 11ADG0074, 11ADG0075, 11ADG0076, 11ADG0077, 11ADG0078,
11ADG0079, 11ADG0080, 11ADG0081, 11ADG0082, 11ADG0083, 11ADG0084, 11ADG0085,
11ADG0086, 11ADG0087, 11ADG0088, 11ADG0089, 11ADG0090, 11ADG0091, 11ADG0092,
11ADG0093, 11ADG0094, 11ADG0095, 11ADG0096, 11ADG0097, 11ADG0098, 11ADG0099,
11ADG0100, 11ADG0101, 11ADG0102, 11ADG0103, 11ADG0104, 11ADG0105, 11ADG0106,
11ADG0107, 11ADG0108, 11ADG0109, 11ADG0110, 11ADG0111, 11ADG0112, 11ADG0113,
11ADG0114, 11ADG0115, 11ADG0116, 11ADG0117, 11ADG0118, 11ADG0119, 11ADG0120,
11ADG0121, 11ADG0122, 11ADG0123, 11ADG0124, 11ADG0125, 11ADG0126, 11ADG0127,
11ADG0128, 11ADG0129, 11ADG0130, 11ADG0131, 11ADG0132, 11ADG0133, 11ADG0134,
11ADG0135, 11ADG0136, 11ADG0137, 11ADG0138, 11ADG0139, 11ADG0140, 11ADG0141
& 11ADG0142
RECALLING FIRM/MANUFACTURE
Recalling Firm: Invacare
Corp., Elyria, OH,
Manufacturer: Chien Ti
Enterprise Co., Ltd., Hsin Chuang, Taipei, Taiwan. .
REASON The
scooter backrest fails unexpectedly due to seat flexing, and this has the
potential to result in a consumer falling causing death or serious injury.
VOLUME OF PRODUCT IN COMMERCE
391 scooters
DISTRIBUTION Nationwide
and Internationally
PRODUCT CADD
High Volume Administration Set, REF 21-7081V-01, Sterile EO, Rx only. Designed
for use with CADD pumps to allow fluid delivery from an IV bag.
CODE Lot
#: 21X222
RECALLING FIRM/MANUFACTURER
Recalling Firm: Smiths
Medical ASD, Inc., Saint Paul, MN,
Manufacturer: Smiths
Healthcare Manufacturing S.A. de C.V., Tijuana, B.C., Mexico.
REASON Smith
Medical has become aware of a small number of customer reports regarding
occlusions/high pressure alarms with the use of one lot of CADD High Volume
Administration Sets.
VOLUME OF PRODUCT IN COMMERCE
5,385 units
PRODUCT Siemens
syngo.plaza VA20A Server Farm setup integrated to a RIS, model number
10592457.
CODE Serial
#’s: 100174, 100177, and 100181
RECALLING FIRM/MANUFACTURER
Recalling Firm: Siemens
Medical Solutions USA, Inc., Malvern, PA,.
Manufacturer: Siemens AG,
Medical Solutions, Business Unit XP, Erlangen, Germany. .
REASON The
firm has become aware of the potential for an unintended merge of multiple
patients when using syngo.plaza version VA20 with Server Farm setup integrated
to a RIS. Sporadically it can happen that certain Patient IDs in the database
are deleted and sent to NULL. This can cause incorrect merging with patients
that already had a NULL Patient ID.
VOLUME OF PRODUCT IN COMMERCE
3
PRODUCT Surgical
Headlight labeled in part * Model Numbers: GMF-H001, GMF-H100, GMF-H101,
GMF-H102, GMF-H103, GMF-H104, GMF-H105, and GMF-H106. The Surgical Headlight
is a passive luminaire. It is illuminated by a source of light by way of the
fiber optics and provides supplemental lighting for medical and surgical
procedures.
CODE None
RECALLING FIRM/MANUFACTURER
Gulf Medical Fiberoptics, Inc.,
Oldsmar, FL
REASON The
cleaning instructions for the exterior surfaces of the headlight, headgear and
fiber cable with 70% isopropyl alcohol had not been validated.
VOLUME OF PRODUCT IN COMMERCE
381
DISTRIBUTION Nationwide
and Internationally
PRODUCT GE
LOGIQ P6 Ultrasound System. The device is intended for use by a qualified
physician for the evaluation of soft tissue and blood flow in the clinical
applications of: Fetal; Abdominal; Pediatric; Small Organ (breast, testes,
thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric);
Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology
(including prostate); Transesophageal (TE); Transrectal (TR); Transvaginal
(TV); and intraoperative (abdominal, thoracic, vascular and neurological).
CODE
Serial Number 144435SU5 133455SU6
140253SU6 140402SU9 143461SU2 135592SU4 135919SU9 143853SU0 143854SU8
143855SU5 143856SU3 143857SU1 143858SU9 144424SU9 136031SU2 137467SU7
138561SU6 140683SU4 144432SU2 144436SU3 144444SU7 144425SU6 144427SU2
144428SU0 144429SU8 144921SU4 146150SU8 146151SU6 146152SU4 146153SU2
146155SU7 146156SU5 146157SU3 146158SU1 146159SU9 146160SU7 146161SU5
146165SU6 146166SU4 146167SU2 146168SU0 146169SU8 146171SU4 146174SU8
148045SU8 148046SU6 148047SU4 148048SU2 148049SU0 148233SU0 148234SU8
148235SU5 148238SU9 148964SU0 148965SU7 148966SU5 148967SU3 149150SU5
149152SU1 149153SU9 149720SU5 149721SU3 149722SU1 149723SU9 150499SU2
150500SU7 150501SU5 151265SU6 151270SU6 151271SU4 151272SU2 151273SU0
151274SU8 151275SU5 151276SU3 151972SU7 151973SU5 151974SU3 151975SU0
151976SU8 151977SU6 151978SU4 151979SU2 151980SU0 152623SU5 152624SU3
152625SU0 152626SU8 153218SU3 153219SU1 153220SU9 153221SU7 153222SU5
153223SU3 153224SU1 153483SU3 153484SU1 153485SU8 154653SU0 154654SU8
155237SU1 155238SU9 155435SU1 155436SU9 155437SU7 155438SU5 155439SU3
155440SU1 156617SU3 156618SU1 156619SU9 156620SU7 156621SU5 156622SU3
157863SU2 157864SU0 157865SU7 157866SU5 157867SU3 157868SU1 157869SU9
157870SU7 161523SU6 161524SU4 162223SU2 162226SU5 133453SU1 136030SU4
136034SU6 136459SU5 138560SU8 141363SU2 143459SU6 144437SU1 144440SU5
144446SU2 144453SU8 133456SU4 133457SU2 133849SU0 137466SU9 138562SU4
140252SU8 134660SU0 140680SU0 140684SU2 136853SU9 141366SU5 144431SU4
135918SU1 141365SU7 144439SU7 144454SU6 135914SU0 143462SU0 138994SU9
136036SU1 137462SU8 140251SU0 144434SU8 136849SU7 133450SU7 133451SU5
140681SU8 140682SU6 144438SU9 144447SU0 144445SU4 133449SU9 135593SU2
135920SU7 140401SU1 134662SU6 143458SU8 141364SU0 141367SU3 144433SU0
139852SU8 138993SU1 135916SU6 139850SU2 133448SU1 144448SU8 136032SU0
133850SU8 144442SU1 144443SU9 144450SU4 144452SU0 133452SU3 133454SU9
134661SU8 136033SU8 136035SU3 136460SU3 136852SU1 136854SU7 137465SU1
138995SU6 139851SU0 143457SU0 143460SU4 143463SU8 143464SU6 135591SU6
135917SU3 133851SU6 135915SU8 136851SU3 137463SU6 137464SU4 144455SU3
135452SU1 135453SU9 135454SU7 147467SU5 147468SU3 147469SU1 133629SU6
133641SU1 149984SU7 145596SU3 148723SU0 148728SU9 148729SU7 148777SU6
149981SU3 151517SU0 154817SU1 136512SU1 136654SU1 137197SU0 145601SU1
147197SU8 149701SU5 149702SU3 149705SU6 135906SU6 143023SU0 148060SU7
148232SU2 148520SU0 140577SU8 140578SU6 144764SU8 134268SU2 141311SU1
146610SU1 149536SU5 155189SU4 158295SU6 159676SU6 160437SU0 156793SU2
157450SU8 161531SU9 149243SU8 149244SU6 149245SU3 149246SU1 149250SU3
149256SU0 149992SU0 150212SU9 150215SU2 150218SU6 150219SU4 154829SU6
154830SU4 154831SU2 154832SU0 155243SU9 155244SU7 155661SU2 155662SU0
135894SU4 135908SU2 135909SU0 135910SU8 135912SU4 136966SU9 137537SU7
137975SU9 137977SU5 137981SU7 137992SU4 138660SU6 142277SU3 142613SU9
142617SU0 142620SU4 143028SU9 143596SU5 143815SU9 147210SU9 148228SU0
148526SU7 148529SU1 133075SU2 149248SU7 149249SU5 149268SU5 153791SU9
155240SU5 155249SU6 155250SU4 155670SU3 155895SU6 135656SU7 135903SU3
136641SU8 138661SU4 139785SU0 142140SU3 143030SU5 143811SU8 143813SU4
144358SU9 145593SU0 146094SU8 148065SU6 148525SU9 148715SU6 134459SU7
142136SU1 146220SU9 147200SU0 148071SU4 133646SU0 139777SU7 136018SU9
136019SU7 149694SU2 137531SU0 137552SU6 147474SU1 147476SU6 147477SU4
133087SU7 148719SU8 148720SU6 149703SU1 149704SU9 154837SU9 136497SU5
136517SU0 136518SU8 136519SU6 136521SU2 136961SU0 136963SU6 137514SU6
137518SU7 137520SU3 137521SU1 137525SU2 137527SU8 138662SU2 139316SU4
140239SU5 142608SU9 143821SU7 148515SU0 148516SU8 139308SU1 139325SU5
137516SU1 136637SU6 136638SU4 136639SU2 139185SU3 139186SU1 139187SU9
139188SU7 139189SU5 139190SU3 139191SU1 139192SU9 139193SU7 139194SU5
139195SU2 139196SU0 139197SU8 139198SU6 139199SU4 140572SU9 140573SU7
140574SU5 140575SU2 140576SU0 140655SU2 140656SU0 140657SU8 140658SU6
140659SU4 140660SU2 140661SU0 140662SU8 140663SU6 140664SU4 141636SU1
141637SU9 141638SU7 141639SU5 141640SU3 141641SU1 141642SU9 141643SU7
142662SU6 142910SU9 142911SU7 142912SU5 142913SU3 142914SU1 144922SU2
144923SU0 145224SU2 145225SU9 145226SU7 145227SU5 145228SU3 145229SU1
145230SU9 145231SU7 145232SU5 145233SU3 145234SU1 145235SU8 145236SU6
145237SU4 145238SU2 146423SU9 146424SU7 146425SU4 146426SU2 146427SU0
146428SU8 146429SU6 147490SU7 147491SU5 148950SU9 148951SU7 148952SU5
148953SU3 148954SU1 148955SU8 148956SU6 148957SU4 148958SU2 148959SU0
148960SU8 148961SU6 148962SU4 148963SU2 151279SU7 151280SU5 151281SU3
151282SU1 151283SU9 152617SU7 152946SU0 152947SU8 152948SU6 155323SU9
155324SU7 155325SU4 155326SU2 155327SU0 156800SU5 156801SU3 156802SU1
156803SU9 156804SU7 156805SU4 159236SU9 159237SU7 159239SU3 159818SU4
159819SU2 159820SU0 143595SU7 143806SU8 137143SU4 137144SU2 143851SU4
143852SU2 144469SU4 145059SU2 135654SU2 138687SU9 140230SU4 142274SU0
142917SU4 143589SU0 147023SU6 147029SU3 147034SU3 147035SU0 147195SU2
147483SU2 149261SU0 149270SU1 149980SU5 159090SU0 159091SU8 133639SU5
133642SU9 153790SU1 155668SU7 135898SU5 136515SU4 136516SU2 136520SU4
136652SU5 137189SU7 137547SU6 137549SU2 138690SU3 139779SU3 142144SU5
142604SU8 142929SU9 143809SU2 144361SU3 146221SU7 147187SU9 148059SU9
148510SU1 133790SU6 146544SU2 149201SU6 132634SU7 156650SU4 133637SU9
133638SU7 155252SU0 134442SU3 139305SU7 139771SU0 153498SU1 157113SU2
157114SU0 140308SU8 140678SU4 132820SU2 139853SU6 140307SU0 136956SU0
142924SU0 145594SU8 155868SU3 139767SU8 145602SU9 147201SU8 136957SU8
133643SU7 133644SU5 148061SU5 133237SU8 133238SU6 133241SU0 133242SU8
133243SU6 133245SU1 133247SU7 133248SU5 133982SU9 133983SU7 133985SU2
133986SU0 133987SU8 133988SU6 133989SU4 133990SU2 133992SU8 133993SU6
134722SU8 134723SU6 134724SU4 134725SU1 134726SU9 134896SU0 134624SU6
134625SU3 134626SU1 134627SU9 134628SU7 135622SU9 135623SU7 135624SU5
135625SU2 135626SU0 135627SU8 135628SU6 135629SU4 135630SU2 135631SU0
136020SU5 136021SU3 136022SU1 136023SU9 136024SU7 136025SU4 136026SU2
136027SU0 136028SU8 136029SU6 136461SU1 136462SU9 136463SU7 136464SU5
136465SU2 136466SU0 136467SU8 136468SU6 136469SU4 136470SU2 136941SU2
136942SU0 136943SU8 136944SU6 136945SU3 136946SU1 136947SU9 136948SU7
137257SU2 137258SU0 137259SU8 137260SU6 137261SU4 137262SU2 137263SU0
138444SU5 138445SU2 138446SU0 138447SU8 138448SU6 138799SU2 138800SU8
138801SU6 138802SU4 138803SU2 139876SU7 139877SU5 142148SU6 142149SU4
142150SU2 142151SU0 142152SU8 142153SU6 142154SU4 142155SU1 142156SU9
142157SU7 142158SU5 142159SU3 142160SU1 142529SU7 142530SU5 142531SU3
142532SU1 142533SU9 142534SU7 142535SU4 142536SU2 143014SU9 143015SU6
143016SU4 143017SU2 143018SU0 143019SU8 143020SU6 143021SU4 143156SU8
143157SU6 143158SU4 143159SU2 143160SU0 143161SU8 143162SU6 143163SU4
143411SU7 143412SU5 143413SU3 143414SU1 144073SU4 144074SU2 144075SU9
144076SU7 144077SU5 144078SU3 144079SU1 144080SU9 145045SU1 145046SU9
145047SU7 145048SU5 145049SU3 145307SU5 145308SU3 145821SU5 145822SU3
145823SU1 145824SU9 145825SU6 146418SU9 146419SU7 146420SU5 146421SU3
146422SU1 146875SU0 146876SU8 146877SU6 146878SU4 146879SU2 146880SU0
146881SU8 146882SU6 146883SU4 146884SU2 146885SU9 146886SU7 146887SU5
146888SU3 146889SU1 147212SU5 147213SU3 147214SU1 147215SU8 147216SU6
147217SU4 147218SU2 148050SU8 148051SU6 148052SU4 148053SU2 148054SU0
148055SU7 148056SU5 148057SU3 148337SU9 148338SU7 148339SU5 148340SU3
148341SU1 148342SU9 148343SU7 148344SU5 148706SU5 148707SU3 148708SU1
148709SU9 148710SU7 149155SU4 149156SU2 149157SU0 149158SU8 149159SU6
149202SU4 149203SU2 149204SU0 149205SU7 149206SU5 150323SU4 150324SU2
150325SU9 150326SU7 150327SU5 150932SU2 150933SU0 150934SU8 150935SU5
151806SU7 151807SU5 151808SU3 151809SU1 152628SU4 152629SU2 152630SU0
152631SU8 153486SU6 153487SU4 153488SU2 153489SU0 153490SU8 153491SU6
153492SU4 153493SU2 153494SU0 153495SU7 153496SU5 154656SU3 154657SU1
154658SU9 154659SU7 154660SU5 154661SU3 154662SU1 154663SU9 154839SU5
155097SU9 155098SU7 155099SU5 155100SU1 155101SU9 155102SU7 155103SU5
155104SU3 155580SU4 155581SU2 155582SU0 155583SU8 155584SU6 155585SU3
155586SU1 155587SU9 156107SU5 156108SU3 156109SU1 156110SU9 156111SU7
156112SU5 156113SU3 156806SU2 156807SU0 156808SU8 156809SU6 156810SU4
157405SU1 157406SU9 157407SU7 157408SU5 157409SU3 158009SU1 158010SU9
158011SU7 158012SU5 158013SU3 158014SU1 158015SU8 158016SU6 159272SU4
159273SU2 159274SU0 159275SU7 159276SU5 159277SU3 159278SU1 159279SU9
159280SU7 159281SU5 160025SU3 160027SU9 160028SU7 160029SU5 160032SU9
160033SU7 161504SU6 161505SU3 161506SU1 161507SU9 161508SU7 161509SU5
161510SU3 158279SU0 160738SU1 149264SU4 149265SU1 149266SU9 149688SU4
149691SU8 149695SU9 149989SU6 153866SU9 153868SU5 154846SU0 155263SU7
155673SU7 155883SU2 155884SU0 155894SU9 156640SU5 156643SU9 156644SU7
156645SU4 156646SU2 156795SU7 136647SU5 137186SU3 137192SU1 137544SU3
137990SU8 138002SU1 138004SU7 138006SU2 138007SU0 138681SU2 138682SU0
142281SU5 142607SU1 142622SU0 142623SU8 142624SU6 142625SU3 143027SU1
143597SU3 143598SU1 144366SU2 144367SU0 144373SU8 144896SU8 144897SU6
144899SU2 144900SU8 145590SU6 145592SU2 145595SU5 145603SU7 145604SU5
145605SU2 146095SU5 147188SU7 147196SU0 147198SU6 147202SU6 147203SU4
147205SU9 147471SU7 148716SU4 136636SU8 136821SU6 136822SU4 139208SU3
139872SU6 139873SU4 139874SU2 139875SU9 140653SU7 140654SU5 146890SU9
133381SU4 133668SU4 133791SU4 134685SU7 135097SU4 138559SU0 140491SU2
142261SU7 143378SU8 143410SU9 143859SU7 144111SU2 142660SU0 142661SU8
144110SU4 145303SU4 145304SU2 145305SU9 145390SU1 145769SU6 146458SU5
147012SU9 147827SU0 147830SU4 148129SU0 148130SU8 148705SU7 150211SU1
150804SU3 151259SU9 151260SU7 151261SU5 151489SU2 151490SU0 154650SU6
158278SU2 159085SU0 159087SU6 159635SU2 159997SU6 158367SU3 135897SU7
143200SU4 143820SU9 143973SU6 150802SU7 155186SU0 155188SU6 133779SU9
135412SU5 136820SU8 133072SU9 133074SU5 133636SU1 133640SU3 155876SU6
155879SU0 155880SU8 155881SU6 155886SU5 156623SU1 156629SU8 156641SU3
156642SU1 156647SU0 156648SU8 156649SU6 156651SU2 156790SU8 156791SU6
156792SU4 156794SU0 157446SU6 157449SU0 158371SU5 159089SU2 161526SU9
161528SU5 148725SU5 148726SU3 148727SU1 149247SU9 149251SU1 149252SU9
149253SU7 149255SU2 149263SU6 149271SU9 149272SU7 149689SU2 149690SU0
149692SU6 149693SU4 149699SU1 149985SU4 149986SU2 149987SU0 149988SU8
149993SU8 149994SU6 150214SU5 150216SU0 150220SU2 150221SU0 150862SU1
150863SU9 150864SU7 150865SU4 150866SU2 150867SU0 150868SU8 150869SU6
150870SU4 150871SU2 151494SU2 151495SU9 151497SU5 151498SU3 151499SU1
151500SU6 151502SU2 151507SU1 151508SU9 151510SU5 151511SU3 151512SU1
151513SU9 151514SU7 151515SU4 151516SU2 153385SU0 153386SU8 153780SU2
153782SU8 153784SU4 153786SU9 153788SU5 153795SU0 153796SU8 153867SU7
153869SU3 153870SU1 153871SU9 153872SU7 153875SU0 153876SU8 154818SU9
154819SU7 154820SU5 154828SU8 154833SU8 154842SU9 155241SU3 155242SU1
155246SU2 155247SU0 155253SU8 155256SU1 155257SU9 155258SU7 155259SU5
155265SU2 155647SU1 155648SU9 155649SU7 155651SU3 155653SU9 155663SU8
155664SU6 155665SU3 155666SU1 155667SU9 155669SU5 155671SU1 155672SU9
155674SU5 155675SU2 134440SU7 134441SU5 134444SU9 134448SU0 134449SU8
134450SU6 134451SU4 134455SU5 134457SU1 134458SU9 135648SU4 135649SU2
135650SU0 135658SU3 135659SU1 135660SU9 135899SU3 135900SU9 135904SU1
135905SU8 135907SU4 135911SU6 135913SU2 135970SU2 135972SU8 136498SU3
136499SU1 136500SU6 136501SU4 136502SU2 136505SU5 136506SU3 136507SU1
136508SU9 136509SU7 136510SU5 136511SU3 136646SU7 136651SU7 136949SU5
136951SU1 136952SU9 136958SU6 136962SU8 136967SU7 136968SU5 136969SU3
136971SU9 136972SU7 137188SU9 137190SU5 137191SU3 137193SU9 137194SU7
137195SU4 137196SU2 137198SU8 137199SU6 137515SU3 137519SU5 137523SU7
137524SU5 137528SU6 137529SU4 137530SU2 137532SU8 137533SU6 137534SU4
137535SU1 137536SU9 137538SU5 137539SU3 137542SU7 137543SU5 137545SU0
137546SU8 137548SU4 137550SU0 137551SU8 137553SU4 137974SU2 137976SU7
137979SU1 137980SU9 137984SU1 137986SU6 137995SU7 137996SU5 137997SU3
137998SU1 138657SU2 138658SU0 138659SU8 138663SU0 138664SU8 138665SU5
138666SU3 138667SU1 138668SU9 138669SU7 138670SU5 138672SU1 138674SU7
138675SU4 138676SU2 138677SU0 138678SU8 138679SU6 138680SU4 138683SU8
138684SU6 138685SU3 138686SU1 138689SU5 139274SU5 139276SU0 139277SU8
139307SU3 139309SU9 139311SU5 139312SU3 139313SU1 139314SU9 139315SU6
139322SU2 139323SU0 139324SU8 139327SU1 139328SU9 139329SU7 139763SU7
139764SU5 139765SU2 139768SU6 139769SU4 139772SU8 139782SU7 139783SU5
139784SU3 140227SU0 140231SU2 140232SU0 140233SU8 140234SU6 140235SU3
140236SU1 140237SU9 140238SU7 140240SU3 140953SU1 140954SU9 140955SU6
140956SU4 140957SU2 140958SU0 140959SU8 140960SU6 140962SU2 141952SU2
141953SU0 141954SU8 141955SU5 141956SU3 141957SU1 141958SU9 141959SU7
141960SU5 141961SU3 141962SU1 141966SU2 142129SU6 142131SU2 142134SU6
142137SU9 142141SU1 142142SU9 142143SU7 142145SU2 142146SU0 142147SU8
142272SU4 142273SU2 142275SU7 142276SU5 142601SU4 142602SU2 142603SU0
142610SU5 142612SU1 142614SU7 142616SU2 142618SU8 142619SU6 142621SU2
142920SU8 142921SU6 142922SU4 142923SU2 142925SU7 143022SU2 143024SU8
143025SU5 143026SU3 143029SU7 143031SU3 143196SU4 143197SU2 143594SU0
143599SU9 143600SU5 143601SU3 143602SU1 143603SU9 143605SU4 143810SU0
143814SU2 143816SU7 143817SU5 143818SU3 143823SU3 143824SU1 143829SU0
144359SU7 144360SU5 144362SU1 144363SU9 144368SU8 144370SU4 144375SU3
144376SU1 144377SU9 144898SU4 144901SU6 144902SU4 144903SU2 144904SU0
145452SU9 145454SU5 145455SU2 145456SU0 145584SU9 145585SU6 145586SU4
145587SU2 145588SU0 145589SU8 145591SU4 145597SU1 145598SU9 145599SU7
146097SU1 146215SU9 146217SU5 146218SU3 146219SU1 147016SU0 147017SU8
147024SU4 147030SU1 147031SU9 147033SU5 147189SU5 147191SU1 147192SU9
147193SU7 147194SU5 147199SU4 147204SU2 147206SU7 147207SU5 147470SU9
147480SU8 147481SU6 147482SU4 147488SU1 147489SU9 148058SU1 148068SU0
148070SU6 148077SU1 148230SU6 148511SU9 148512SU7 148513SU5 148514SU3
148518SU4 148519SU2 148521SU8 148522SU6 148523SU4 148527SU5 148711SU5
148712SU3 148718SU0 158355SU8 133086SU9 149258SU6 149698SU3 153778SU6
153779SU4 153873SU5 154840SU3 154841SU1 155864SU2 155865SU9 155872SU5
134443SU1 135657SU5 135971SU0 136513SU9 136514SU7 136950SU3 136953SU7
137541SU9 137994SU0 138688SU7 139773SU6 139774SU4 139778SU5 140226SU2
141963SU9 141964SU7 142926SU5 143586SU6 143587SU4 143807SU6 143808SU4
143825SU8 143827SU4 145451SU1 146093SU0 147019SU4 147025SU1 147026SU9
147027SU7 147472SU5 147478SU2 147484SU0 147485SU7 147486SU5 147487SU3
148073SU0 148074SU8 148075SU5 148724SU8 149254SU5 151503SU0 153793SU5
154834SU6 155871SU7 135661SU7 136644SU2 138003SU9 139318SU0 139319SU8
142918SU2 143592SU4 145600SU3 154814SU8 149697SU5 153797SU6 142606SU3
153389SU2 153794SU3 134453SU0 139320SU6 139321SU4 135895SU1 138001SU3
140961SU4 148063SU1 144762SU2 156011SU9 136504SU8 137983SU3 137988SU2
139275SU2 139317SU2 133446SU5 133447SU3 133777SU3 133852SU4 133917SU5
138782SU8 139180SU4 139181SU2 139368SU5 139369SU3 139370SU1 140098SU5
140224SU7 140225SU4 140305SU4 140306SU2 140679SU2 140877SU2 140878SU0
140945SU7 141003SU4 141240SU2 141314SU5 142854SU9 144109SU6 145302SU6
145391SU9 145392SU7 145633SU4 145634SU2 145635SU9 145770SU4 145771SU2
145772SU0 145846SU2 146893SU3 149240SU4 149241SU2 149242SU0 149535SU7
149793SU2 149794SU0 149795SU7 150504SU9 150505SU6 150803SU5 151084SU1
151448SU8 151449SU6 151492SU6 151493SU4 151811SU7 151812SU5 151813SU3
151889SU3 151890SU1 151891SU9 151892SU7 152013SU9 152014SU7 152619SU3
152620SU1 152621SU9 152622SU7 152643SU3 152907SU2 153953SU5 153954SU3
153955SU0 153977SU4 155579SU6 156739SU5 157119SU9 157120SU7 157121SU5
157122SU3 157123SU1 157196SU7 157197SU5 157445SU8 157712SU1 157713SU9
157872SU3 158102SU4 158103SU2 158810SU2 159637SU8 159816SU8 159817SU6
159887SU9 159888SU7 159987SU7 159988SU5 159998SU4 160438SU8 138992SU3
139211SU7 139871SU8 143978SU5 132637SU0 132638SU8 132639SU6 132640SU4
134267SU4 134900SU0 134901SU8 134902SU6 134903SU4 134904SU2 141362SU4
141688SU2 144449SU6 144451SU2 144456SU1 144460SU3 148623SU2 148704SU0
151085SU8 151810SU9 156180SU2 156751SU0 158716SU1 161800SU8 161801SU6
153874SU3 132738SU6 133232SU9 133233SU7 133234SU5 133235SU2 133236SU0
134215SU3 134897SU8 134898SU6 134899SU4 135098SU2 135099SU0 135100SU6
135101SU4 135102SU2 135125SU3 135126SU1 135127SU9 135128SU7 135129SU5
135130SU3 135131SU1 135132SU9 135133SU7 135136SU0 135138SU6 135139SU4
135140SU2 135142SU8 135143SU6 135144SU4 135405SU9 135406SU7 135407SU5
135408SU3 135409SU1 135410SU9 135411SU7 135891SU0 135892SU8 135893SU6
138208SU4 141309SU5 141310SU3 141312SU9 141694SU0 141700SU5 142528SU9
144112SU0 144114SU6 144115SU3 144118SU7 144119SU5 144120SU3 144121SU1
144122SU9 144123SU7 144124SU5 144125SU2 144126SU0 144766SU3 144767SU1
144769SU7 144770SU5 144771SU3 145223SU4 145280SU4 145281SU2 145282SU0
145283SU8 145284SU6 145287SU9 145289SU5 145290SU3 145291SU1 146561SU6
146575SU6 147902SU1 147903SU9 147904SU7 147905SU4 147908SU8 149342SU8
149346SU9 149867SU4 149868SU2 150928SU0 152641SU7 152642SU5 156330SU3
157125SU6 157127SU2 157200SU7 157201SU5 157202SU3 157203SU1 157204SU9
157205SU6 157206SU4 157207SU2 157208SU0 148721SU4 149257SU8 154838SU7
143828SU2 146576SU4 146577SU2 146579SU8 146580SU6 133073SU7 133080SU2
153387SU6 153781SU0 153785SU1 153787SU7 155260SU3 155891SU5 156630SU6
158368SU1 137185SU5 137978SU3 137993SU2 138656SU4 139326SU3 139766SU0
142138SU7 142609SU7 143822SU5 144365SU4 144369SU6 144371SU2 145458SU6
145607SU8 147473SU3 147475SU8 148067SU2 148229SU8 148517SU6 148714SU9
142915SU8 142919SU0 137517SU9 149260SU2 149892SU2 162047SU5 162049SU1
139781SU9 139761SU1 133360SU8 132636SU2 132737SU8 133628SU8 133632SU0
133634SU6 133645SU2 133778SU1 133792SU2 133853SU2 133854SU0 133855SU7
134269SU0 134284SU9 134623SU8 135134SU5 135135SU2 135137SU8 135141SU0
136850SU5 137181SU4 138206SU8 138207SU6 139182SU0 139207SU5 139209SU1
139210SU9 139454SU3 140099SU3 140685SU9 141002SU6 141239SU4 141241SU0
141313SU7 141689SU0 141690SU8 141691SU6 141692SU4 141693SU2 141695SU7
141696SU5 141697SU3 141698SU1 141699SU9 142088SU4 142224SU5 143860SU5
144113SU8 144116SU1 144117SU9 144127SU8 144441SU3 144768SU9 144909SU9
144910SU7 145001SU4 145002SU2 145222SU6 145285SU3 145286SU1 145288SU7
145632SU6 145820SU7 146020SU3 146021SU1 146023SU7 146024SU5 146891SU7
146892SU5 147828SU8 147829SU6 147898SU1 147899SU9 147900SU5 147901SU3
147906SU2 147907SU0 148336SU1 149312SU1 149341SU0 149343SU6 149344SU4
149345SU1 149347SU7 149348SU5 149590SU2 149796SU5 150002SU4 150003SU2
150004SU0 151887SU7 151888SU5 151971SU9 152012SU1 152119SU4 152618SU5
153497SU3 153777SU8 156009SU3 156785SU8 157115SU7 157116SU5 157117SU3
157118SU1 157198SU3 157444SU1 157627SU1 158809SU4 159235SU1 159889SU5
159891SU1 135647SU6 135651SU8 135652SU6 135653SU4 135655SU9 135896SU9
135901SU7 135902SU5 143819SU1 147021SU0 148717SU2 135973SU6 135974SU4
136503SU0 136640SU0 136642SU6 136643SU4 136645SU9 136648SU3 136649SU1
136650SU9 136653SU3 136954SU5 136955SU2 136959SU4 136960SU2 136964SU4
136965SU1 136973SU5 137184SU8 137187SU1 137200SU2 137201SU0 137202SU8
137203SU6 137522SU9 137526SU0 137540SU1 137973SU4 137982SU5 137985SU8
137987SU4 137989SU0 137991SU6 137999SU9 138000SU5 138005SU4 138209SU2
138671SU3 138673SU9 139278SU6 139306SU5 139310SU7 139762SU9 139770SU2
139775SU1 139776SU9 139780SU1 139786SU8 140228SU8 140229SU6 141965SU4
142128SU8 142130SU4 142132SU0 142133SU8 142135SU3 142139SU5 142278SU1
142279SU9 142280SU7 142605SU5 142611SU3 142615SU4 142916SU6 142927SU3
142928SU1 143198SU0 143199SU8 143588SU2 143590SU8 143591SU6 143593SU2
143604SU7 143812SU6 143826SU6 144364SU7 144372SU0 144374SU6 144426SU4
144430SU6 144763SU0 144765SU5 144905SU7 145306SU7 145450SU3 145453SU7
145457SU8 145459SU4 145606SU0 145636SU7 145637SU5 146022SU9 146096SU3
146149SU0 146154SU0 146162SU3 146163SU1 146164SU9 146172SU2 146173SU0
146216SU7 146222SU5 146223SU3 146224SU1 146578SU0 147011SU1 147013SU7
147015SU2 147018SU6 147020SU2 147022SU8 147028SU5 147032SU7 147186SU1
147190SU3 147208SU3 147209SU1 147211SU7 147479SU0 148062SU3 148064SU9
148066SU4 148069SU8 148072SU2 148076SU3 148231SU4 148236SU3 148524SU2
148528SU3 148713SU1 148722SU2 149151SU3 149259SU4 149262SU8 149267SU7
149269SU3 149687SU6 149696SU7 149700SU7 149706SU4 149982SU1 149983SU9
149990SU4 149991SU2 150213SU7 150217SU8 150502SU3 150929SU8 150930SU6
150931SU4 151277SU1 151278SU9 151351SU4 151491SU8 151496SU7 151501SU4
151504SU8 151505SU5 151506SU3 151509SU7 151518SU8 153388SU4 153783SU6
153789SU3 153792SU7 153865SU1 153877SU6 153878SU4 153879SU2 153880SU0
154651SU4 154652SU2 154664SU7 154813SU0 154815SU5 154816SU3 154821SU3
154822SU1 154823SU9 154824SU7 154825SU4 154826SU2 154827SU0 154835SU3
154836SU1 154843SU7 154844SU5 154845SU2 154847SU8 154889SU0 154890SU8
154891SU6 154892SU4 154893SU2 155105SU0 155106SU8 155107SU6 155140SU7
155187SU8 155236SU3 155239SU7 155245SU4 155248SU8 155251SU2 155254SU6
155255SU3 155261SU1 155262SU9 155266SU0 155267SU8 155268SU6 155646SU3
155650SU5 155652SU1 155654SU7 155655SU4 155656SU2 155657SU0 155658SU8
155659SU6 155660SU4 155861SU8 155862SU6 155863SU4 155866SU7 155867SU5
155869SU1 155870SU9 155873SU3 155874SU1 155875SU8 155877SU4 155878SU2
155882SU4 155885SU7 155887SU3 155888SU1 155889SU9 155890SU7 155892SU3
155893SU1 156010SU1 156114SU1 156115SU8 156116SU6 156117SU4 156118SU2
156331SU1 156625SU6 156626SU4 156627SU2 156628SU0 156631SU4 156632SU2
156633SU0 156634SU8 156635SU5 156636SU3 156637SU1 156638SU9 156639SU7
156652SU0 156796SU5 156797SU3 156798SU1 156799SU9 157124SU9 157177SU7
157199SU1 157209SU8 157210SU6 157403SU7 157404SU4 157447SU4 157448SU2
157451SU6 157704SU8 157705SU5 157706SU3 157707SU1 157708SU9 157709SU7
157710SU5 157711SU3 158098SU4 158099SU2 158100SU8 158101SU6 158277SU4
158356SU6 158357SU4 158358SU2 158359SU0 158360SU8 158361SU6 158362SU4
158363SU2 158364SU0 158365SU7 158366SU5 158369SU9 158370SU7 158668SU4
158669SU2 158670SU0 158671SU8 158672SU6 158673SU4 158674SU2 158675SU9
158676SU7 158677SU5 158678SU3 158679SU1 158680SU9 158686SU6 158687SU4
158688SU2 158689SU0 158690SU8 158691SU6 158692SU4 158693SU2 158694SU0
158695SU7 158696SU5 158697SU3 158698SU1 158699SU9 158700SU5 158701SU3
158702SU1 158703SU9 158704SU7 158705SU4 158706SU2 158707SU0 158708SU8
158709SU6 158710SU4 158711SU2 158712SU0 158713SU8 158714SU6 158715SU3
158717SU9 158718SU7 159075SU1 159076SU9 159077SU7 159078SU5 159079SU3
159080SU1 159081SU9 159082SU7 159083SU5 159084SU3 159086SU8 159088SU4
159092SU6 159238SU5 159267SU4 159268SU2 159269SU0 159270SU8 159271SU6
159626SU1 159814SU3 159815SU0 159996SU8 160526SU0 161521SU0 161522SU8
161527SU7 161537SU6 161539SU2 162044SU2 162048SU3 162516SU9 162529SU2
RECALLING FIRM/MANUFACTURE
Recalling Firm: GE
Healthcare, LLC, Waukesha, WI,
Manufacturer: GE Ultrasound
Korea Co., Ltd., Kyunggi-Do Korea, Republic Of (South).
REASON GE
Logic P6 probe head has a possibility to cause burn injuries to patients or
operators.
VOLUME OF PRODUCT IN COMMERCE
2,270
DISTRIBUTION Nationwide
and Internationally
PRODUCT
1) Roller Pump, 4 Inch Diameter
The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours
on the extracorporeal circulation of blood for arterial perfusion, regional
perfusion, and cardiopulmonary bypass procedures, when used by a qualified
medical professional who is experienced in the operation of this or similar
equipment. Catalog Number 816570.
2) Roller Pump, 6 Inch Diameter
The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours
on the extracorporeal circulation of blood for arterial perfusion, regional
perfusion, and cardiopulmonary bypass procedures, when used by a qualified
medical professional who is experienced in the operation of this or similar
equipment. Catalog Number 816571.
CODE
1) Serial Numbers: 5939 – 5962;
2) Serial Numbers 6621 - 6640
RECALLING FIRM/MANUFACTURE
Terumo Cardiovascular Systems
Corp., Ann Arbor, MI,
REASON During
incoming inspection a single digital encoder cable assembly was rejected
because it had a loose wire. The assembly is used in the 4" and the
6" roller pumps for the Terumo Advanced Perfusion System 1.
VOLUME OF PRODUCT IN COMMERCE
44 units
DISTRIBUTION Nationwide,
Canada, Mexico, and Vietnam
PRODUCT Multi-Analyte
Urine DAU Calibrator (15 mL HDPE bottle inside a cardboard box) For in vitro
diagnostic use only. Intended use: Drug abuse detection. Catalog number 0803.
CODE Lot
#: 1101136K3, exp 1/28/13
RECALLING FIRM/MANUFACTURER
Lin-Zhi International Inc.,
Sunnyvale, CA,
REASON Multi
Analyte Urine DAU Cutoff Calibrator Gives a decreased absorbance reading due
to a loss of the BZO (oxazepam) analyte from the solution and may cause false
positive readings.
VOLUME OF PRODUCT IN COMMERCE
45 items
DISTRIBUTION Nationwide
and Germany
PRODUCT Visum
LED Surgical Lights, Ceiling Mounted Surgical Light. The Visum LED surgical
light is intended to illuminate the operative site during surgical procedures
with high intensity light. Catalog Number: 0682001250, 0682001251, 0682001252,
and 0682001253. Part Number: P10235, P10234, P12875, P12874. Product Numbers:
PI0235 (Visum LED Surgical Light, Camera Ready), PI0234 (Visum LED Surgical
Light), P12875 (Visum LED Surgical Light Camera Ready - Low Ceiling), PI2874 (Visum
LED Surgical Light - Low Ceiling). Expiration date: There is no expiration
date with the scope of the action.
CODE Serial
Numbers: Product manufactured between October 20I0 to July 20II and in use by
customer.
RECALLING FIRM/MANUFACTURER
Stryker Communications Corp.,
Flower Mound, TX,
REASON Stryker
became aware that there was a possibility that incorrect assembly could result
in the light head beginning to separate from the suspension arm. In the event
the light separates and falls from the arm, it could lead to serious injury.
VOLUME OF PRODUCT IN COMMERCE
1,143 devices
DISTRIBUTION Nationwide
and Internationally
PRODUCT Brand
name: Maxcem Elite, Product: a dental cement, Part Number: 34418. The intended
use of this device is for cementation of all indirect restorations including
ceramic, resin and metal-based inlays, onlays, crowns, bridges, posts and
veneers. Additional Indications include core-buildup material, pit and fissure
sealant and cementation of crown restorations to implants
CODE Lot
#: 3360900
RECALLING FIRM/MANUFACTURER
Kerr Corp., Orange, CA,
REASON The
recall was initiated because Kerr Corporation has confirmed that the use of
the affected lot of Maxcem Elite may not provide users with a minimum of 2
minutes of worktime thereby causing the product to prematurely set by 15
seconds.
VOLUME OF PRODUCT IN COMMERCE
463 units
DISTRIBUTION Nationwide
Canada, Australia and New Zealand
PRODUCT VITROS
Chemistry Products DT Calibrator Kit, REF/Catalog No. 195 7927, IVD ---
Common/Usual Name: VITROS DT Calibrator Kit -- VITROS Chemistry Products DT
Calibrator Kit is an in vitro diagnostic specially formulated for use as
calibrators for ALB, ALKP, ALT, AMYL, AST, TBIL, NBIL, BUN/UREA, Ca, CHOL, CK,
Cl-, CO2, CREA, CRSC, Fe, GGT, GLU, HDLC, K+, LAC, LDH, LIPA, Mg, Na+, NH3,
PHOS, TP, TRIG, urCR, and URIC on VITROS DT Chemistry Systems.
CODE Kit
Lot 37 (exp. 3/31/12) and Kit Lot 38 (exp. 3/31/13)
RECALLING FIRM/MANUFACTURER
Ortho-Clinical Diagnostics,
Rochester, NY,
REASON Negatively
biased NBIL results could occur for quality control, proficiency and patient
samples. As a result, VITROS Chemistry Products DT Calibrator Kit Lots 37
& 38 calibrator values were adjusted to allow customers to continue using
their inventory of affected kits.
VOLUME OF PRODUCT IN COMMERCE
12,750 units
DISTRIBUTION Nationwide
and Internationally
PRODUCT NanoKnife
Irreversible Electroporation System. System consists of multiple components:
Generator, footswitch, disposable electrode. Catalog #20300101 Intended use:
the generator system with six outputs is indicated for surgical ablation of
soft tissue.
CODE Batch
codes: 00800709, 00880909, 00870909, 00630609, 00610609, 00470109, 00080508,
00180608, 00050508, 00460109, 00830709, 00240708, 00500209, 00930909,
00430109, 00480209, 00371008, 01210911, 00650709, 00670709, 00690709,
00850909, 00920909, 00820709, 00860909, 00890909, 00940909, 00780709,
00770709, 00760709, 00910909, 00790709, 00361008, 00750709, 00510209,
00450109, 00421208, 00490209, 00530209, 00600609
RECALLING FIRM/MANUFACTURER
Recalling Firm:
Angiodynamics Worldwide Headquartery S, Latham, NY, .
Manufacturer: Angiodynamics,
Inc., Queensbury, NY.
REASON Angiodynamics
was marketing the NanoKnife System with the ablation zone estimator feature in
the US without FDA clearance.
VOLUME OF PRODUCT IN COMMERCE
40
PRODUCT
1) The Artificial Bur Patch is a
Temporary Fascial Prothesis. For Single Patient Use only. Use with Negative
Pressure Wound Dressing. Contents: Large Bur Patch (Sterile: EO), 20cm x 40cm
Loop Sheet, 20cm x 40cm Hook Sheet. Caution: Federal Law restricts this device
to sale by or on the order of a physician. Sterile: Contents are sterile for
maximum two years unless enclosed package has been opened or damaged. The
Wittmann Patch is used for a fascia expander prosthesis to temporarily close
the abdominal cavity when regular abdominal closure is not possible; to
prevent perfusion impairment of abdominal organs from increases
intra-abdominal pressure; and for advanced intra-abdominal pathology requiring
more than one abdominal operation to control the disease process.
2) The Artificial Bur HypoPack is
a Temporary Fascial Prothesis with Hypobaric Wound Dressing. For Single
Patient Use only. Contents: 1. Large Bur Patch (Sterile: EO), 20cm x 40cm Loop
Sheet, 20cm x 40cm Hook Sheet. 2. 60cm x 85 cm Steri Drape 3. Drain with
Connector 4. 4 x Gauze Rolls Caution: Federal Law restricts this device to
sale by or on the order of a physician. Sterile: Contents are sterile for
maximum two years unless enclosed package has been opened or damaged. The
Wittmann Patch is used for a fascia expander prosthesis to temporarily close
the abdominal cavity when regular abdominal closure is not possible; to
prevent perfusion impairment of abdominal organs from increases
intra-abdominal pressure; and for advanced intra-abdominal pathology requiring
more than one abdominal operation to control the disease process.
CODE
1) Lot #: 54643.2;
2) Lot #s: 1010.0, 1102.1, and
1103.1
RECALLING FIRM/MANUFACTURE
Recalling Firm: Novomedicus
LLC, Nokomis, FL,
Manufacturer: Hidih
Surgical, Doerrebach, Germany.
REASON Cytotoxicity
testing revealed a biocompatibility issue. The test results did not represent
a health risk.
VOLUME OF PRODUCT IN COMMERCE
47 patches and kits
DISTRIBUTION Nationwide
and Internationally
PRODUCT MEDPOR
TITAN Cranial Temporal with Template - Right, Catalog Number 81038. MEDPOR
Biomaterial with embedded Titanium Mesh Implants are intended for non-weight
bearing applications of craniofacial reconstruction/cosmetic surgery and
repair of craniofacial trauma.
CODE Lot
#: F001357
RECALLING FIRM/MANUFACTURE
Howmedica Osteonics Corp dba
Stryker Craniomaxillofacial, Newnan, GA
REASON The
template that is packaged with this implant was not packaged according to
specification. The template was packaged in a single pouch instead of a double
pouch.
VOLUME OF PRODUCT IN COMMERCE
10 units
DISTRIBUTION Nationwide
Australia and Canada
PRODUCT
1) Package of cross-drive bone
screws labeled in part: "***REF #95-1504***"1.5MM" SYSTEM HIGH
TORQUE (HT) CROSS-DRIVE SCREW, 5/ PK***1.5 X4MM TITANIUM***CAUTION: FEDERAL
LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN***"
"***REF: 91-1504***"1.5MM" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE
SCREW***1.5 X4MM TITANIUM***CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE
BY OR ON THE ORDER OF A PHYSICIAN. ***" Products subject to recall: REF
#95-1504***"1.5MM" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE SCREW, 5/ PK.
REF: 91-1504***"1.5MM" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE SCREW, 1/PK.
Internal fixation screw intended to aid in the alignment and stabilization of
bone in the oral cranio-maxillofacial skeletal system.
2) Package of center-drive bone
screws labeled in part: "***REF: 01-7204***1.5MM" SYSTEM
CENTER-DRIVE SCREW, 5/ PKG***1.5 X4MM TITANIUM*** CAUTION: FEDERAL LAW
RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN.***" REF:
99-7204***"1.5MM" SYSTEM CENTER-DRIVE SCREW, 1/ PKG***1.5 X4MM
TITANIUM***CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE
ORDER OF A PHYSICIAN.***" REF: 01-7204 1.5MM" SYSTEM CENTER-DRIVE
SCREW, 5/ PKG. REF: 99-7204 "1.5MM" SYSTEM CENTER-DRIVE SCREW, 1/
PKG. Internal fixation screw intended to aid in the alignment and
stabilization of bone in the oral cranio-maxillofacial skeletal system.
CODE
1) Lot # 622120;
2) Lot # 202450
RECALLING FIRM/MANUFACTURER
Biomet Microfixation, Inc.,
Jacksonville, FL,
REASON A
possibility that these products are mixed.
VOLUME OF PRODUCT IN COMMERCE
Center drive screws-484, Cross
drive screws-295
DISTRIBUTION Nationwide,
S Africa and Japan
PRODUCT Terumo
Cardiovascular Systems CAPIOX (r) SX18 Hollow Fiber Oxygenator. The X-coated
Capiox SX18 Hollow Fiber Oxygenators with Detachable Hardshell Reservoir are
used to exchange gases between extracorporeal blood and a gaseous environment
during cardiopulmonary bypass procedures. The integral heat exchanger is used
to warm or cool the blood or perfusion fluid and the detachable hardshell
reservoir is used to store blood during extracorporeal circulation from both
the venous line and the cardiotomy line. The reservoir contains a venous
section that is comprised of a filter and defoamer to facilitate air bubble
removal. Product code 3CXSX18RX.
CODE Lot
# HF22
RECALLING FIRM/MANUFACTURER
Terumo Cardiovascular Corp.,
Elkton, MD,
REASON Complaints
were received of expired X-coated Capiox SX18 Hollow Fiber oxygenators.
VOLUME OF PRODUCT IN COMMERCE
16 units (4 cases of 4 units)
DISTRIBUTION Nationwide